UV-Visible Spectroscopy (UV-Vis)
UV-Visible Spectroscopy (UV-Vis)
Overview of UV-Vis Spectrometer
- Components:
- Light source
- Chopper
- Shutter
- Monochromator
- Sample compartment
- Mirrors
- Semitransparent mirror
- Detector (PT)
- Signal processor
- Temperature control
Introduction
- UV-Vis Spectroscopy uses light to measure chemical concentration.
- Light is described as a photon because of its ability to carry energy.
- E = hv
- Where:
- E is energy.
- h is Planck's constant (6.626 x 10^{-34} J.s).
- v is the frequency of light in Hz.
Electromagnetic Spectrum
- Electromagnetic Spectrum ranges from Radio waves to Gamma rays.
- Wavelengths range from 10^3 meters (Radio) to 10^{-12} meters (Gamma Ray).
- Frequencies range from 10^4 Hz (Radio) to 10^{20} Hz (Gamma Ray).
- Visible light is a small portion of the electromagnetic spectrum (.5 x 10^{-6}).
Electromagnetic Spectrum (Continued)
- Radiant energy propagated as a transverse wave.
- Vibrates perpendicular to the direction of propagation.
Wave Properties
- Wavelength (\lambda): The distance of one complete cycle.
- Frequency: The number of cycles passing a fixed point per unit time.
Wavenumber
- Wavenumber is the reciprocal of the wavelength (cm⁻¹).
- Relationship between wavelength and frequency:
- \text{Wavelength (cm)} = \frac{\text{Velocity of light (}3 \times 10^{10} \text{ cm/s)}}{\text{Frequency (Hz or s}^{-1})} (corrected LaTeX)
UV-Vis Absorption
- Electromagnetic radiation for the UV-Vis region ranges in wavelength from approximately 200 to 700 nm.
- Ultraviolet (UV) range: 200 to 350 nm.
- Visible range: 350 to 700 nm.
UV-Vis Absorption (Continued)
- When white light passes through an object, certain wavelengths are absorbed, and the unabsorbed wavelengths are transmitted or reflected.
Measurement
- The amount of light absorbed is measured as a function of wavelength (\lambda).
- This provides qualitative and quantitative information about the sample.
- \lambda_{max} is a characteristic value.
- Absorbing compounds must be examined in diluted solution using solvents like water, ethanol, hexane, and cyclohexane.
- Solvents with double or triple bonds or heavy atoms (e.g., S, Br, I) should be avoided.
Chromophore and Auxochrome
- A chromophore refers to the absorbing groups of atoms in a molecule.
- A molecule containing a chromophore is called a chromogen.
- Absorption results from the transition from the ground state (lower energy) to excited states (higher energy).
Beer's Lambert Law
- States the relationship between the absorbance of a solution and the concentration of the absorbing species.
- Equation:
- A = abc
- A = \epsilon bc
- Where:
- A = absorbance (no unit)
- a = absorptivity (L/g.cm)
- \epsilon = molar absorptivity (L/mol.cm)
- b = optical path length (cm)
- c = concentration (g/L)
Attenuation
- Describes how the amount of attenuation depends on the concentration of absorbing molecules and the path length over which absorption occurred.
- Absorption may be presented as:
- Transmittance (T = \frac{P}{P_0})
- Absorbance (A = \log \frac{P_0}{P})
- Where:
- P_0 = power of incident radiation
- P = power of transmitted radiation
- c = concentration
- b = path length
Transmittance and Absorbance
- Transmittance is often expressed as a percentage (%T).
- Absorbance is a logarithmic function of transmittance.
Absorbance Values
- Absorbance is zero when transmittance is 100%, meaning no light is absorbed by the sample.
- Absorbance will be infinite when transmittance is zero (all lights were absorbed by the sample).
- When P = P_0, T = 1.0 or 100%, and A = 0.
Beer’s Lambert Law Implications
- Referring to the Beer’s Lambert Law Equation [ A = \epsilon bc ] → the absorbance depends on the total quantity of the absorbing compound in the light path through the cuvette.
Standard Calibration Curves
- A plot of absorbance vs. concentration will yield a straight line passing through the origin (0,0).
- The linear relationship between concentration and absorbance is simple and straightforward.
- The plot of %T vs. concentration will give an exponential-type graph (complicated relationship).
Limitations of Beer's Lambert Law
- The Beer's Lambert Law is not obeyed at high concentrations.
Errors
- Gravimetric errors (e.g. mood, skills etc.)
- Incomplete spectral resolution (due to wrong slit width selection).
- Turbidity (must filter the sample to avoid cloudy samples).
- Aggregation at high concentration (the particles stick together).
- Contamination of cuvettes (must clean the cuvettes immediately after use).
Example of Calculation 1
- A 7.50 x 10^{-5} M solution of potassium permanganate has a transmittance of 36.4% when measured in a 1.05 cm cell at a wavelength of 525 nm.
- Calculate:
- (a) absorbance of this solution
- (b) molar absorptivity of KMnO4
- Answer:
- a) A = - \log T = - \log 0.364 = 0.439
- b) Use: A = \epsilon b c
- \epsilon = \frac{A}{bc} = \frac{0.439}{1.05 \text{cm} \times 7.50 \times 10^{-5} \text{mol/L}} = 5.57 \times 10^3 \text{L/mol.cm}
Example of Calculation 2
- At 580 nm, the wavelength of maximum absorption of complex Fe(SCN)^{2+} has a molar absorptivity of 7.0 x 10^3 L/mol.cm.
- Calculate the absorbance of a 2.5 x 10^{-5} M solution of the complex at 580 nm in a 1.0 cm cell.
Example of Calculation 3
- Find the concentration of samples deer 1 & deer 2.
- Plot a standard calibration curve.
Instrumentation: UV-Vis Spectrophotometer
- Consists of:
- Light Source
- Wavelength selector (monochromator)
- Sample containers (quartz & glass)
- Detector
- Signal processor and readout
Single Beam Spectrometer
- The simplest type of spectrometer employs a single source to supply radiation to the sample.
- Well-suited for quantitative absorption measurement at single-wavelength type.
- Simple, low cost, and ease of maintenance.
- The sample & blank must be placed alternately in the light path (time-consuming)
- For measurements at multiple wavelengths, the blank must be run at each wavelength.
Double Beam Spectrometer
- Has two light paths: one for the sample & one for a blank or reference.
- Measures both the sample and solvent.
- The detection electronics manipulate the measurements to give the absorbance.
Double Beam Spectrometer Advantages
- Automatic scanning & continuous recording of spectrum (absorbance versus wavelength).
- The absorption in the sample is automatically corrected for the absorption occurring in the blank.
- Automatic correction for changes of the source intensity & changes in the detector response with time or wavelength because the two beams are compared & measured at the same time.
Double Beam Spectrometer Disadvantages
- What are the disadvantages of Double Beam Instrument?