UV and IR Spectroscopy and GC-MS Study Notes
UV Spectroscopy Basics
Definition: UV spectroscopy involves the absorption of light in the wavelength range of . This energy absorption leads to electronic transitions.
Mechanism: When atoms or molecules absorb light energy, they transition from a state of low energy (ground state) to a state of high energy (excited state). These transitions occur specifically between electronic energy levels within the ultraviolet and visible spectrum.
Electronic Transitions
Transition Types: In compounds other than alkanes, electrons can undergo several possible transitions depending on their orbital types. The most important transitions, ordered by increasing energy, are:
: Occurs in carbonyl compounds consistently.
: Occurs in compounds containing oxygen, nitrogen, sulfur, and halogens (lone pair electrons).
: Occurs in alkenes, carbonyl compounds, alkynes, and azo compounds.
: Found in alkanes (saturated hydrocarbons).
Significance: The transition of the lowest energy is generally the most important for analysis.
Selection Rules: Not all transitions are possible; they are governed by selection rules.
Spin Selection Rule: Changes in the spin quantum number of an electron during transition are not allowed. Such transitions are termed "forbidden."
Transition Probabilities
: Allowed Transition.
: Partially forbidden.
: Very weak (forbidden).
Beer-Lambert Law
Principle: Explains the relationship between the light absorbed by a solution and its concentration/path length. It states that absorbance is directly proportional to the concentration of the absorbing species and the path length of the sample.
Equation:
Variables:
: Absorbance.
: Intensity of light incident upon the sample cell.
: Intensity of light leaving the sample cell.
: Molar concentration of the solute ( or Molar).
: Length of the sample cell ().
: Molar absorptivity ().
UV Spectroscopy Exercise
Scenario 1: A solution of benzonitrile in water () is examined at . The absorbance and path length .
Question: What is the molar absorptivity ()?
Calculation: .
Scenario 2: The same solution is examined at where .
Question: What is the absorbance reading and the intensity ratio ?
Absorbance Calculation: .
Ratio Calculation: Since , then .
Instrumentation of UV Spectrophotometer
Components:
Light Source:
Deuterium lamp: Emits radiation in the UV region.
Tungsten lamp: Used for radiation in the visible region.
Monochromator: Acts as a diffractor to spread the beam into its component wavelengths.
Detector: Records the intensity of transmitted light. Common types include photomultipliers or photodiodes.
Data System: A computer records the plot of absorbance versus wavelength.
Solvent Selection in UV Spectroscopy
Requirement 1 (Transparency): The solvent must be transparent at the measurement wavelength. Solvents have a "cutoff wavelength," below which they absorb strongly.
Rule: Choose a solvent with a cutoff lower than the analyte's .
Common Solvent Cutoffs:
Acetonitrile:
Water:
Cyclohexane:
Isooctane:
n-hexane:
95% ethanol:
Methanol:
Trimethyl phosphate:
1,4-dioxane:
Chloroform:
Requirement 2 (Solubility): Solid analytes must dissolve completely to form a clear, homogenous solution. Cloudy solutions scatter light and distort results.
Requirement 3 (Reactivity): The solvent must not react with the analyte.
Requirement 4 (Volatility): The solvent should not evaporate quickly.
General Usage Guide:
Water: For polar compounds.
Ethanol/Methanol: Widely used for various polarities.
Hexane: For non-polar compounds.
Chromophores and Auxochromes
Chromophore: The part of a molecule responsible for absorbing UV or visible light. They contain electrons excitable by UV light.
Chromophore Classes and Typical :
(alkene): ()
(alkyne): ()
: ()
: ()
: ()
Aromatic rings: ()
: ()
: ()
: ()
(nitro): ()
(azo): ()
: ()
Auxochrome: Groups attached to a chromophore that increase the intensity and/or wavelength of absorption. Examples: , halogens, amino groups.
Effects on Absorption Band
Bathochromic Shift (Red shift): Shift to a longer wavelength (lower energy).
Hypsochromic Shift (Blue shift): Shift to a shorter wavelength (higher energy).
Hyperchromic Effect: An increase in absorption intensity.
Hypochromic Effect: A decrease in absorption intensity.
Practical Guide for UV Interpretation
Low/Medium intensity band < 220 nm: Usually indicates no transitions. Possibilities include amines, alcohols, and thiols.
Two low-intensity bands (250 - 360 nm) with no major short-wave absorption: Usually indicates transitions (e.g., ).
Two medium-intensity bands ($\lambda_{\max} > 200\,nm$): Generally indicates an aromatic system.
High-intensity bands (> 210 nm): Generally represent an -unsaturated ketone, a diene, or a polyene.
Multi-system compounds (e.g., simple ketones): Show two absorptions: an transition at longer wavelengths (> 300\,nm, low intensity) and a transition at shorter wavelengths (< 250\,nm, high intensity).
Highly colored compounds: Likely contain a long-chain conjugated system or a polycyclic aromatic chromophore.
UV Spectroscopy Exercises
Possible Transitions:
Cyclopentane (): Saturated, only bonds. Transition: .
Acetaldehyde (): Contains and lone pairs on Oxygen. Transitions: , , and .
Dimethyl ether (): Transition: .
Methylvinyl ether (): bond and lone pairs. Transitions: , , .
Trimethylamine: Transition: .
Cyclohexane (): Saturated. Transition: .
Acetone Spectrum Interpretation:
: transition.
: transition.
: transition.
Infrared Spectroscopy (IR)
Purpose: Used primarily to identify functional groups in organic compounds.
EM Spectrum Range: Mid-IR used by chemists ranges from approximately . Standard scale is to .
Mechanism: Measures stretching and bending vibrational frequencies of covalent bonds. Absorption increases the amplitude of vibrational motion.
Dipole Requirement: Only bonds with a dipole moment can absorb IR radiation. Symmetric bonds (e.g., in symmetrical alkanes/alkenes) are IR inactive.
Allowed Vibrations:
Symmetric stretching
Antisymmetric stretching (occurs at higher frequencies than symmetric)
In-plane bending (e.g., scissoring, rocking)
Out-of-plane bending (e.g., wagging, twisting)
IR Frequency Regions
:
: Around
:
:
:
Fingerprint Region: Below (unique to each molecule).
IR Instrumentation and Sample Preparation
Spectrophotometer Types:
Dispersive: Uses a hot wire beam divided into sample and reference paths. Passed through a monochromator to a thermocouple detector. Records in the frequency domain.
Fourier Transform (FT-IR): Produces an interferogram (time domain) containing all frequencies. A mathematical Fourier Transform converts it to a frequency domain spectrum.
Sample Holders: Glass and plastic absorb IR, so cells are made of ionic substances like Sodium Chloride () or Potassium Bromide ().
: Useful for .
: Useful for .
Methods:
Neat: A drop of liquid pressed between salt plates.
KBr Pellet: Mixing ground solid with KBr powder and pressing under high pressure. (Disadvantage: KBr absorbs water).
Nujol Mull: Grinding solid with mineral oil. (Disadvantage: Nujol bands at may overlap sample).
Solution: Dissolving in solvents like Carbon Tetrachloride (). (Disadvantage: stretches at may obscure bands).
Factors Influencing Carbonyl () Frequency
Typical Ranges:
Acid Chloride:
Anhydride: and
Ester:
Aldehyde:
Ketone:
Carboxylic Acid:
Amide:
Influence Factors:
Conjugation: Adjacent lowers the frequency.
Ring Size: Decreasing ring size increases strain and increases frequency (e.g., six-membered cyclic ketone at , five-membered at ).
-Substitution: Halogens () on the alpha-carbon shift the band to a higher frequency due to electron withdrawal.
Hydrogen Bonding: Reduces absorption frequency (e.g., Methyl salicylate at ).
Specific Functional Group Absorptions
Alkanes:
stretch:
bend:
bend:
Long chain band (n > 4\,CH_2):
Alkenes:
Vinyl stretch: > 3000\,cm^{-1} ().
stretch: . (Trans is weak, Cis is stronger; symmetrical is inactive).
Alkynes:
Terminal : .
: .
Aromatics:
stretch: > 3000\,cm^{-1} ().
ring stretch: and .
Substitution patterns (Bending ):
Monosubstituted: Strong band near .
Ortho: One band near .
Meta: Three bands at , and .
Para: One strong band at .
Gas Chromatography-Mass Spectrometry (GC-MS)
Definition: A hyphenated analytical technique combining separation (GC) and identification (MS) of components in a complex gas mixture.
Importance:
Forensic/Toxicology: Identifying illicit drugs.
Environmental: Detecting VOCs and pesticides in soil/water.
Food: Identifying flavors and contaminants.
Pharmaceutical: Monitoring drug purity and degradation.
Components of GC:
Carrier Gas: Mobile phase (, or ); must be inert.
Injector: Heated port to vaporize the sample.
Column: Houses the stationary phase (polar or non-polar).
Oven: Temperature-controlled chamber for the column.
Detector: Measures eluting components (e.g., FID or MS).
Data System: Computer that generates the chromatogram.
MS Process:
Ionization: Electron Impact (EI) bombards sample with electrons to create positive fragment ions.
Mass Analysis: Fragments separated by mass-to-charge () ratio via analyzers like Quadrupole or Time-of-Flight (TOF).
Detection: Registers ion quantity and produces a Total Ion Chromatogram (TIC) and Mass Spectrum.
Interpreting GC-MS Data
GC Chromatogram (TIC):
X-axis: Retention Time (RT). Fast-moving (small or less polar) molecules have lower RT.
Y-axis: Peak area/height, used to quantify components.
Mass Spectrum:
X-axis: ratio.
Y-axis: Relative abundance.
Base Peak: The tallest peak, representing the most stable/abundant fragment.
Molecular Ion Peak (): Represents the un-fragmented molecule; reveals the molecular weight.