Any visible structure on an image that does not accurately demonstrate the true presence, location, or reflectivity of that structure.
Image detail resolution artifacts
Locational artifacts
Attenuation artifacts
Doppler artifacts
Ultrasound system assumptions
Equipment failure/faults
Inappropriate equipment settings
Electrical/magnetic interference
Assumption 1: The transmitted wave travels along a straight-line path from the transducer to the object and back.
Assumption 2: All detected echoes originate from the axis of the main beam only, and the depth of objects is proportional to the round-trip time.
Assumption 3: The ultrasound wave travels in all soft tissues at a constant rate of 1540 m/s.
Assumption 4: The attenuation of sound in tissue does not vary.
Assumption 5: All received echoes are derived from the most recently transmitted pulse.
Assumption 6: The amplitude of the echo is related to the reflective characteristics of each object.
If any of the assumptions are violated, an ultrasound artifact will occur.
Artifacts can be seen in B-mode and Doppler imaging.
Pro: Some artifacts are desirable and can indicate tissue characteristics, aiding in diagnosis.
Example: Shadowing may suggest calcification, and enhancement may indicate cystic change.
Con: Some artifacts are undesirable and may obscure or distort anatomy.
They must be recognized to prevent misinterpretation of the image and potential misdiagnosis.
Shadowing
Enhancement
Edge shadowing
Propagation speed artifact
Reverberation artifact
Slice thickness artifact
Beam width artifact
Side lobe/grating lobe artifact
Refraction artifact
Mirror image artifact
Attenuation artifacts
Depth artifacts
Beam dimension artifacts
Beam path artifacts
Equipment and electrical artifacts
Posterior acoustic shadowing
Posterior acoustic enhancement
Edge refraction shadowing
Assumption 4: The attenuation of sound in tissue does not vary.
Assumption 6: The amplitude of the echo is related to the reflective characteristics of each object.
Attenuation (a) = α x f (MHz) x L (cm)
Table of Medium-Acoustic Values:
Various materials, densities, velocities, impedance, and attenuation coefficients are provided in the table.
Definition: Increased attenuation causes transmitted intensity deep to a region to be reduced compared to the intensity of tissues around it at the same depth.
Classifications:
Clean Shadowing: Posterior to calcification or bone.
Due to high percentage of absorption & reflection with no transmission.
Dirty Shadowing: Posterior to air-filled structures.
Due to a high percentage of reflection and a small percentage of transmission with secondary reflections.
Reduction: Can be reduced with compound imaging.
Occurs at the edges of rounded structures.
The sound beam is reflected or refracted, and the sound beam is deflected or broadened.
Can be reduced with compound imaging.
Definition: Decreased attenuation causes transmitted intensity deep to a region to be strengthened compared to the intensity of tissues around it at the same depth.
Minimized by compound imaging.
Often a beneficial artifact for tissue characterization.