Phase Variables
Breath Design and Delivery
- Ventilator TCT: Time frame from inspiration start to expiration end.
- 4 Phases of Breath:
- Trigger Variable: Initiates inspiratory phase.
- Limit Variable: Stops lung inflation.
- Cycle Variable: Allows exhalation.
- Baseline Variable: Determines baseline level.
Phase Variables
- Breath Definition: One cycle of positive flow (inhalation) and negative flow (exhalation).
- 4 Phase Variables:
- Trigger
- Limit
- Cycle
- Baseline
Trigger Variable
- Trigger Definition: Begins the inspiratory phase. Can be:
- Patient-Initiated Breath: Triggered by a change in pressure or flow by the patient.
- Mandatory Breath: Time-triggered by the ventilator.
- Neurally Adjusted Ventilatory Assist (NAVA): Uses diaphragm's electrical activity to trigger a breath.
Types of Breaths
- Mandatory (Controlled) Breaths: Ventilator controls the breath.
- Assisted Breaths: Patient controls timing; ventilator provides remaining support.
- Spontaneous Breaths: Patient controls most aspects with minimal support.
Trigger Variable Types
- Time Triggered: Ventilator triggers a breath after a set time, e.g., rate set at 12 breaths/min = 1 breath every 5 seconds.
- Patient Triggered: RT sets sensitivity; breath is initiated based on patient effort.
- Manual Triggering: RT can manually trigger a breath via a button.
Sensitivity Settings
- Patient Triggering: Should be set to minimize effort without causing auto-triggering (ventilator activating without effort).
- Trigger settings too sensitive can lead to unwanted breaths.
- Monitor for increased Work of Breathing (WOB).
Pressure Triggering
- Mechanism: Initiates a breath when patient’s inspiratory effort drops airway pressure.
- Typical Sensitivity Setting: -0.5 to -1.5 cmH2O; sensitivity at 2.0 cmH2O means patient must create a drop of ≥ 2 cmH2O.
Limit Variable
- Definition: Maintains a preset level before inspiration ends without terminating inspiration.
- Possible Types:
- Pressure: Limits pressure within a certain range.
- Volume: Limits maximum volume delivered.
- Flow: Limits maximum flow during breath.
Case Study Discussion
- Patient Parameters: VT = 467 ml, RR = 15, PEEP 8, Trigger = -1 LPM, FiO2 = 0.30
- Identify triggers and limits for breaths.
- Concerns: Low exhaled volume alarm, low pressure alarm; monitor waveform abnormalities for insights into ventilatory issues.
Examples of Triggering and Limiting
- Time Triggering Example: Breath is triggered after a set time, e.g., every 5 seconds.
- Patient Pressure Triggering: Pressure triggering adjusts slightly based on PEEP levels; auto-PEEP can complicate this.
- Flow Triggering: Triggered when flow decreases significantly, generating a breath based on set parameters.