Computed Tomography (CT) – Comprehensive Bullet-Point Study Notes
Terminology & Core Concept of Tomography
- Root of the word
- “Tomography” derives from Greek “tomo/tomos” meaning “to cut, section, or layer”.
- Philosophy of CT
- Uses a computer to transform x-ray data into cross-sectional “cuts”.
- Each slice = an independent plane through the patient, analogous to a single slice of bread in a loaf.
- Image Orientations generated
- Transverse/axial (perpendicular to long axis).
- From the transverse data set, sagittal and coronal planes can be reconstructed.
- Newer scanners allow direct acquisition in multiple planes.
Historical Milestones
- 1970 – Sir Godfrey Hounsfield demonstrates first CT technique.
- 1979 – Nobel Prize in Physiology/Medicine shared by Godfrey Hounsfield & Alan Cormack (Tufts University).
Conventional Imaging vs. CT
- Conventional Radiography
- Area x-ray beam, screen-film receptor.
- Low contrast due to Compton scatter, high-intensity beam, superimposition of anatomy → degraded images.
- Conventional (Axial) Tomography
- Selectively blurs anatomy above/below an “in-focus” plane.
- Enhanced contrast of chosen plane, but overall image still dull/blurred using film-screen.
- Digital Radiographic Tomosynthesis
- Area beam + multiple projections → 3-D data set; reconstruction of any plane, improved contrast.
- Computed Tomography (CT)
- Produces true cross-sectional (transverse) images by reconstructing hundreds of projections collected as the x-ray tube/detectors translate & rotate.
- Eliminates superimposition; delivers high spatial & low-contrast resolution.
Evaluation Criteria for CT Image Quality
- Spatial Resolution – Ability to distinguish small, closely spaced objects.
- Low-Contrast Resolution – Ability to differentiate objects with minimal density difference.
- Temporal Resolution – Speed of data acquisition; critical for moving organs (e.g., heart) to minimize motion artifacts.
CT Data Geometry
- Voxel & Pixel Relationships
- Each slice subdivided into volume elements (voxels). Width = X axis, height = Y axis, thickness = Z axis.
- In-plane grid = matrix (commonly 512×512). Each matrix cell = pixel.
- Projections
- One translate-rotate sweep → intensity profile (projection) dependent on attenuation pattern.
- Multiple projections (often hundreds) are stored digitally; superimposed mathematically to reconstruct the slice.
Beam Attenuation Principles
- Basic rule: More attenuation ⇒ lighter pixel; less attenuation ⇒ darker pixel.
- Determinants of Attenuation
- Atomic number Z
- Photoelectric absorption probability ∝Z3.
- Mass density (kg m−3) – more matter per volume → more interactions.
- Thickness traversed.
- Linear Attenuation Coefficient (μ)
- Expresses fraction of beam attenuated per unit path length.
- Example at 125 kVp for water: μ≈0.18cm−1.
- General trend: μ ↓ with ↑ photon energy, ↑ with ↑ Z or density.
- Representative Equation (conceptual): I=I0e−μx (not in transcript but underlies mathematics).
Hounsfield Units (HU)
- Quantifies relative attenuation on a scale where
- Water = 0HU
- Air = −1000HU
- Dense bone/metal ≈ +1000HU or higher.
- Purpose: Compare unknown tissue values to known ranges for diagnosis.
- Typical Ranges (approx.)
- White matter +20–+30HU
- Gray matter +30–+40HU
- Muscle +20–+40HU
- Hemorrhage +65–+95HU
- Fat −30–−70HU
Polychromatic X-ray Beams & Beam Hardening
- All diagnostic x-ray tubes emit a spectrum of photon energies.
- Detectors treat all arrived photons equally; cannot adjust for low-energy photons that are preferentially absorbed.
- Preferential absorption → beam hardening artifact (streaks, cupping), especially adjacent to dense bones (e.g., petrous ridges).
- Filtration with materials (Teflon, Al) removes low-energy photons, narrowing energy spread, mitigating artifact & patient dose.
Volume Averaging
- Voxels with mixed tissues display averaged HU → potential to obscure small or low-contrast lesions.
- Influencing factors:
- Slice thickness (Z-axis) – thicker slices ↑ averaging.
- In-plane pixel size (X & Y) – larger pixels ↑ averaging.
- Trade-off: Thin slices (< 1mm) ↓ averaging but ↑ patient dose; protocol balances diagnostic need vs. ALARA.
Data Types
- Raw (Scan) Data – Detector samples prior to image formation; unsectioned, no HU assigned.
- Image Data – After reconstruction; each pixel has HU; can be displayed or archived.
- Retrospective Reconstruction – Reusing raw data with altered parameters (e.g., slice thickness, kernel) to generate new images without re-scanning.
Scan Modes
- Step-and-Shoot (Axial/Sequential)
- 360° tube rotation per slice → table increments → repeat; dominated 1980s.
- Helical/Spiral
- Gantry rotates continuously while table moves; path of tube traces a helix.
- Enabled by slip-ring technology (cable-free continuous rotation).
- Multidetector Row CT (MDCT)
- Multiple parallel detector rows (2 → 4 → 16 → 64 → 128+).
- Captures multiple slices per rotation; facilitates sub-second volume coverage & thin collimation.
Imaging Planes & Patient Orientation
- Standard Planes
- Axial (transverse), Sagittal, Coronal.
- Oblique Planes – Angled relative to standard planes.
- Indications for Non-Axial Scans
- Anatomy oriented vertically (e.g., spine, sinuses).
- To minimize artifacts from adjacent structures.
- Plane adjustment via patient positioning, gantry tilt, or multiplanar reformatting (MPR).
Basic Phases of Contrast-Enhanced Scans
- Arterial phase: 20–40s post-injection.
- Portal venous phase: ≈ 80s.
- Delayed phase: 6–10min.
- Choice depends on organ/lesion characterization requirements.
CT System Components
- Gantry – Houses rotating hardware: x-ray tube, collimators, detectors, DAS.
- X-ray Tube
- Filament heats → electrons emitted → accelerated across high kV → strike rotating anode.
- Focal spot – Where electrons hit anode; size affects spatial resolution.
- Tube current (mA) – Controls photon quantity.
- >99% electron energy → heat; tube’s heat capacity & dissipation are design limits.
- Generator – Supplies high voltage (kV) to tube.
- Detectors
- Solid-state scintillator (modern) or xenon gas (older).
- Convert x-ray energy → light/electrical signal; high efficiency, low after-glow.
- Data Acquisition System (DAS) – Samples detector signals, digitizes, forwards to CPU.
- Central Processing Unit (CPU) – Executes reconstruction algorithms (e.g., filtered back-projection, iterative methods).
Three Segments of the CT Process
- Data Acquisition – Photons generated, modulated by patient, captured by detectors.
- Image Reconstruction – Sorting & mathematical processing assign one HU per pixel.
- Image Display – HU matrix converted to greyscale; output to monitors/film/PACS.
- Matrix size: 512×512 common.
- Typical HU scale: −1000(air)→0(water)→+1000(dense bone/metal).
- Linear attenuation coefficient example: μwater≈0.18cm−1 at 125kVp.
- Photoelectric dependence: PPE∝Z3.
- Intensity law: I=I0e−μx.
- Contrast-enhancement timings: 20–40s (arterial), 80s (portal), 6–10min (delayed).
Practical & Ethical Considerations
- Radiation Dose Management
- Thin slices & multiphase protocols increase dose; protocols must balance diagnostic benefit vs. ALARA principle.
- Artifact Recognition & Mitigation
- Techs must recognize beam hardening, motion, partial-volume artifacts and use filtration, faster acquisitions, or alternative planes to reduce them.
- Manufacturer Variability
- Similar features often branded differently (e.g., “spiral CT”, “Helical CT”, “Volume CT”); understanding core principles prevents confusion.
Connecting Concepts to Clinical Relevance
- HU measurement allows identification of pathology (e.g., differentiating hemorrhage from calcification, fat characterization of adrenal lesions).
- Multiplanar & 3-D reconstructions critical for surgical planning, trauma assessment, and interventional guidance.
- Temporal resolution advances enable coronary CT angiography, reducing motion artifacts.