Radiation Protection
Radiation Protection Considerations
I. Ionizing Effects of X-Radiation
A. Electromagnetic Radiation
Electromagnetic radiation consists of wave-like fluctuations of electric and magnetic fields that travel through space at the speed of light (3 × 10⁸ m/s).
Examples of electromagnetic radiation include:
Radio waves
Microwaves
Infrared radiation
Visible light
Ultraviolet radiation
X-rays
Gamma rays
These radiations differ mainly in wavelength, frequency, and energy.
Wavelength
Wavelength is the distance between two consecutive wave crests.
Frequency
Frequency refers to the number of cycles per second, measured in Hertz (Hz).
Relationship of Wavelength, Frequency, and Energy
These three quantities are related as follows:
Wavelength and frequency are inversely related
Energy increases as frequency increases
Energy decreases as wavelength increases
Formula:
Speed of light = Frequency × Wavelength
c = fλ
Radiations with short wavelength and high frequency have higher energy.
Ionizing Radiation
Ionizing radiation has sufficient energy to remove electrons from atoms, creating positive and negative ions.
Examples of ionizing radiation include:
X-rays
Gamma rays
Ionization can cause chemical and biological damage to tissues.
Diagnostic X-rays have extremely short wavelengths ranging from:
10⁻⁸ to 10⁻¹² meters
The unit previously used for very small wavelengths is the Angstrom (Å).
1 Å = 10⁻¹⁰ meters
Sources of Radiation Exposure
Humans are continuously exposed to radiation from two major sources.
Natural Background Radiation
Natural radiation comes from:
Cosmic radiation from the sun
Radioactive materials in the Earth’s crust
Radon gas in the air
Radioactive materials in food and water
Natural background radiation accounts for the largest portion of human radiation exposure.
Artificial (Man-Made) Radiation
Man-made radiation sources include:
Medical and dental x-rays
Nuclear testing fallout
Nuclear power plants
Occupational radiation exposure
Among artificial sources, medical radiation contributes the largest exposure.
II. Production of X-Rays at the Tungsten Target
Diagnostic x-rays are produced inside the x-ray tube when high-speed electrons strike the tungsten anode target.
Process:
Electrons are produced at the heated cathode filament.
A high voltage (kilovoltage) accelerates electrons toward the anode focal spot.
When electrons strike the tungsten target, their kinetic energy is converted to X-ray photons.
Two types of X-rays are produced.
1. Bremsstrahlung Radiation
Bremsstrahlung means “braking radiation.”
This occurs when a high-speed electron is deflected by the positively charged nucleus of a tungsten atom.
The loss of energy from the electron is emitted as an x-ray photon.
Characteristics:
Produces a continuous spectrum of x-ray energies
Accounts for 70–90% of the x-ray beam
Produces a polyenergetic beam
2. Characteristic Radiation
Characteristic radiation occurs when a high-energy electron ejects an inner-shell electron (usually K-shell) from a tungsten atom.
Steps:
An incoming electron ejects a K-shell electron.
A vacancy is created.
An outer-shell electron drops down to fill the vacancy.
The energy difference is released as a characteristic x-ray photon.
Characteristics:
Energy is specific to the target material
Tungsten K-characteristic x-rays ≈ 69 keV
Represents 10–30% of the x-ray beam
III. Interaction of X-Ray Photons with Matter
As X-rays pass through tissue, their intensity decreases. This process is called attenuation.
Attenuation occurs mainly through two interactions:
1. Photoelectric Effect
In the photoelectric effect:
A low-energy X-ray photon interacts with tissue.
The photon completely transfers its energy to an inner-shell electron.
The electron is ejected from the atom.
This creates a vacancy, which is filled by an outer-shell electron, producing characteristic radiation.
Characteristics:
Occurs more often in high atomic number materials (e.g., bone, contrast media)
Increases patient radiation dose
Produces short-scale contrast
Probability ≈ Z³ / E³
2. Compton Scatter
In Compton scatter:
A high-energy X-ray photon interacts with an outer-shell electron.
The electron is ejected (recoil electron).
The photon changes direction and continues with reduced energy.
Characteristics:
Most common interaction in diagnostic radiology
Produces scattered radiation
Causes image fog
Represents a radiation hazard to personnel
IV. Dose–Response Relationships
A dose-response relationship describes the association between radiation dose and the resulting biological effect.
Dose-response curves show how the biological response changes with radiation dose.
Two important concepts:
Threshold
The minimum radiation dose required before a biological effect appears.
Nonthreshold
A situation where any radiation dose can cause a biological effect.
Linear Nonthreshold (LNT) Model
The Linear Non-Threshold Model is used in radiation protection.
Characteristics:
No safe radiation dose
Even very small doses can cause damage
Used to estimate risk for low-level radiation exposure
This model applies to stochastic effects, including:
Cancer
Leukemia
Genetic mutations
Nonlinear Threshold Model
Some radiation effects occur only after a certain threshold dose.
Examples:
Skin erythema
Cataracts
These are deterministic (nonstochastic) effects.
V. Late Effects of Radiation
Late radiation effects occur months or years after exposure.
Examples include:
Leukemia
Cancer
Genetic mutations
Cataracts
Infertility
Occupational radiation workers are mainly concerned with the late effects of radiation exposure.
Historical examples include:
Early radiologists developing leukemia
Hiroshima and Nagasaki survivors
Radium dial painters developing bone cancer
VI. Biologic Effects of Ionizing Radiation
Law of Bergonié and Tribondeau
This law states that cells are more radiosensitive if they are:
Young
Undifferentiated
Rapidly dividing (high mitotic activity)
Examples of highly radiosensitive cells:
Lymphocytes
Bone marrow cells
Intestinal epithelial cells
Embryonic cells
Radiosensitive vs Radioresistant Tissues
Highly radiosensitive tissues:
Lymphocytes
Bone marrow
Intestinal epithelium
Reproductive cells
Radioresistant tissues:
Muscle tissue
Nerve tissue
Mature bone
VII. Radiation Weighting and Tissue Weighting
Different radiation types cause different levels of biological damage.
Radiation Weighting Factor (Wr)
Represents the biologic effectiveness of different radiation types.
Example:
X-rays: Wr = 1
Alpha particles: Wr = 20
Tissue Weighting Factor (Wt)
Represents relative radiosensitivity of specific tissues.
Examples:
Gonads: high weighting factor
Skin: lower weighting factor
Effective Dose Formula
Effective Dose (EfD) =
Wr × Wt × Absorbed Dose
VIII. Linear Energy Transfer (LET)
LET describes the rate at which radiation deposits energy in tissue.
Higher LET radiation causes greater biological damage.
Examples:
Alpha particles → high LET
X-rays → low LET
Relative Biological Effectiveness (RBE)
RBE measures how damaging radiation is compared to X-rays.
Relationship:
Higher LET → Higher RBE → Greater biologic damage
IX. Direct and Indirect Radiation Effects
Direct Effect
Occurs when radiation directly damages DNA molecules.
More common with high-LET radiation.
Indirect Effect
Occurs when radiation interacts with water molecules, producing free radicals that damage DNA.
This is the most common mechanism in diagnostic radiology.
Because the body is 65–80% water, indirect effects predominate.
X. DNA Damage from Radiation
Radiation may cause several types of DNA damage:
Single-strand break
Double-strand break
Cross-linking
Base damage or mutation
Approximately 90% of radiation-induced damage is repairable.
However, severe or repeated damage can cause:
Cell death
Cancer
Genetic mutation
XI. ALARA Principle
Radiation exposure must always be kept:
As Low As Reasonably Achievable (ALARA).
Radiographers must minimize exposure by:
Proper collimation
Correct exposure factors
Avoiding repeat examinations
Shielding sensitive organs
Maintaining equipment quality control
Radiation safety is a professional responsibility of all radiologic technologists.
IV. Genetic Effects
Definition
Genetic effects are radiation effects that occur in the future offspring of exposed individuals due to damage or mutation of reproductive cells (gametes).
These effects do not appear in the exposed person, but rather in their children or future generations.
Mechanism
Radiation can damage the DNA of sperm or ova, causing gene mutation.
If the mutated gene is transmitted during reproduction, it can lead to hereditary disorders.
Important Concept: Genetically Significant Dose (GSD)
Genetically Significant Dose (GSD) refers to the average radiation dose received by the reproductive organs of a population that may affect future generations.
Purpose:
Used to estimate the genetic impact of radiation exposure on a population
Helps establish radiation protection standards
Radiation Protection for Reproductive Organs
Females
The ovaries are highly radiosensitive.
Possible effects:
Infertility
Genetic mutation in ova
Potential hereditary abnormalities in offspring
10-Day Rule
The 10-Day Rule states that:
Radiologic examinations of the abdomen or pelvis in women of childbearing age should be performed within 10 days after the start of menstruation.
Reason:
Ovulation has not yet occurred
The probability of pregnancy is minimal
Patient Questionnaire
Patients are asked about:
Pregnancy status
Last menstrual period (LMP)
This helps prevent accidental fetal exposure.
Posting
Radiology departments often place signs reminding patients to inform staff if they are pregnant.
Males
The testes are very radiosensitive.
Possible effects:
Temporary or permanent sterility
Reduced sperm production
Possible genetic mutation in sperm cells
Because of this:
Gonadal shielding is used whenever possible.
Children
Children are more sensitive to radiation because:
Cells divide more rapidly
Tissues are still developing
They have longer life expectancy, allowing more time for radiation effects to appear
Therefore:
Radiation exposure must be kept as low as possible.
V. Somatic Effects
Definition
Somatic effects are radiation effects that occur in the exposed individual, not in their offspring.
These effects result from damage to body cells (somatic cells).
Somatic effects may appear:
Immediately (early effects)
Years later (late effects)
Major Somatic Effects
1. Carcinogenesis
Carcinogenesis is the development of cancer due to radiation exposure.
Radiation damages DNA, which may cause cells to grow uncontrollably.
Common radiation-induced cancers include:
Leukemia
Thyroid cancer
Breast cancer
Lung cancer
Bone cancer
These usually appear years after exposure.
2. Cataractogenesis
Cataractogenesis is the formation of cataracts (lens opacity) caused by radiation damage to the lens of the eye.
Effects:
Clouding of the lens
Decreased vision
Possible blindness in severe cases
The lens is highly radiosensitive.
3. Life Span Shortening
Radiation exposure may reduce life expectancy due to:
Cellular damage
Increased cancer risk
Degeneration of body systems
4. Reproductive Risks
Radiation exposure to reproductive organs may cause:
Temporary sterility
Permanent infertility
Genetic mutations
Embryologic / Fetal Effects
Radiation Effects on the Embryo and Fetus
The developing embryo and fetus are extremely radiosensitive.
Radiation exposure during pregnancy can cause:
Growth retardation
Congenital malformations
Mental retardation
Childhood cancer
Fetal death (at high doses)
Critical Periods of Fetal Development
1. Preimplantation Stage (0–2 weeks)
Effect:
All-or-nothing response
Possible outcomes:
Embryo dies
OREmbryo survives with no damage
2. Organogenesis (2–8 weeks)
This is the most critical stage.
Organs are forming, so radiation may cause:
Congenital malformations
Structural abnormalities
3. Fetal Stage (8 weeks – birth)
Possible effects:
Growth retardation
Mental retardation
Functional abnormalities
Increased cancer risk later in life
Important Radiation Protection Principles for Pregnancy
Radiologic technologists must:
Verify pregnancy status
Avoid unnecessary examinations
Use lead shielding when possible
Use lowest radiation dose possible (ALARA principle)
Consult the radiologist if pregnancy is suspected
Ultra-High Yield Board Exam Points
These are frequently tested facts:
• Genetic effects affect future generations
• Somatic effects affect the exposed individual
• Gonads are highly radiosensitive
• 10-Day Rule protects possible early pregnancy
• Organogenesis is the most sensitive fetal stage
• Radiation may cause cancer, cataracts, and genetic mutation
• Children are more radiosensitive than adults
Patient Protection
Radiologic imaging professionals have the ethical and professional responsibility to minimize radiation exposure to patients, themselves, and the public.
Although diagnostic imaging provides significant medical benefits, exposure to ionizing radiation carries potential long-term biologic risks. Therefore, radiation exposure must always follow the ALARA principle (As Low As Reasonably Achievable).
Ways to achieve patient protection include:
Beam restriction
Proper exposure factors
Filtration
Shielding
Proper patient communication
Accurate positioning
Use of automatic exposure control (AEC)
Appropriate image receptors
Use of grids or air-gap techniques
Safe fluoroscopic practice
I. Beam Restriction
Beam restriction refers to limiting the size of the x-ray beam to the area of diagnostic interest.
It is considered the most important method of reducing patient radiation dose.
A. Purpose
Beam restriction serves several important purposes:
1. Reduce Patient Radiation Dose
Only tissues necessary for diagnosis are exposed.
2. Reduce Scatter Radiation
Smaller irradiated areas produce less scattered radiation.
3. Improve Image Quality
Scatter radiation produces radiographic fog, which decreases contrast and image clarity.
Therefore, beam restriction improves both patient safety and radiographic image quality.
B. Types of Beam Restrictors
There are three types of beam restriction devices:
Aperture diaphragms
Cones (cylinders)
Collimators
1. Aperture Diaphragm
The aperture diaphragm is the simplest beam restrictor.
Structure
Flat piece of lead (Pb)
Has a central opening
The opening determines the shape and size of the x-ray beam
Common Uses
Skull units
Some chest radiography units
Advantages
Simple design
Easy to use
Disadvantages
Opening size is fixed
Field size cannot be adjusted
To change field size, a different diaphragm must be used.
2. Cones (Cylinders)
Cones are lead-lined cylindrical devices attached to the x-ray tube or collimator.
Types
Straight cylinder cone
Flared cone
Some cones are extendable (telescopic).
Uses
Small anatomical areas such as:
Paranasal sinuses
L5–S1
Small localized structures
Advantages
More efficient than aperture diaphragms
Restrict the beam over a longer distance
Disadvantages
Fixed opening size
Limited to small field sizes
3. Collimators
The collimator is the most efficient beam restricting device.
It is attached directly to the X-ray tube housing.
Components
Collimators contain two sets of adjustable lead shutters (blades):
Upper shutters
Lower shutters
The lower shutters control the length and width of the radiation field.
This allows the radiographer to precisely control field size and shape.
C. Light-Localization Apparatus
The collimator contains a light-localization system.
This system allows visualization of the X-ray field before exposure.
Components
Light bulb
Mirror
The mirror reflects the light to produce a visible light field corresponding to the X-ray beam.
Important Principle
For proper alignment:
The x-ray tube focal spot and light bulb must be the same distance from the mirror.
If alignment is incorrect:
The light field and x-ray field will not match
An incorrect exposure area may occur
Repeat examinations may be required
D. Accuracy
Collimator accuracy must be regularly checked through quality assurance (QA) programs.
NCRP Standard
Manual collimation must be accurate within:
±2% of the Source-to-Image Distance (SID)
Example:
If SID = 100 cm
Maximum error = 2 cm
Positive Beam Limitation (PBL)
Most modern radiographic systems use Positive Beam Limitation (PBL).
Definition
PBL automatically adjusts the x-ray beam size to match the image receptor size.
Sensors in the Bucky tray detect cassette size and adjust the collimator shutters.
Advantages
Prevents unnecessary exposure
Ensures correct beam restriction
Produces an unexposed border on the radiograph
NCRP Standard
Accuracy within 3% per side
Maximum 4% total misalignment
II. Exposure Factors
Exposure factors significantly influence patient radiation dose.
A. mAs and kV
mAs (milliampere-seconds)
mAs controls the quantity (number) of X-ray photons produced.
Increasing mAs:
Increases radiation output
Increases patient dose
kVp (kilovoltage peak)
kVp controls the energy and penetrability of the X-ray beam.
Increasing kVp:
Produces more high-energy photons
Increases beam penetration
Decreases radiation absorption by the patient
Optimal Technique
To minimize dose:
Use high kVp and low mAs whenever possible.
Benefits:
Lower patient dose
Reduced tube heat
Longer x-ray tube life
B. Generator Type
Modern radiographic systems often use:
Three-phase generators
High-frequency generators
These generators produce a more constant voltage.
Benefits:
More efficient x-ray production
Slightly reduced patient dose
Improved image quality
III. Filtration
The X-ray beam contains many low-energy photons.
Low-energy photons:
Cannot penetrate the patient
Are absorbed by tissues
Increase patient skin dose
Filtration removes these photons.
A. Inherent Filtration
Inherent filtration consists of materials permanently built into the X-ray tube housing.
Sources include:
Glass envelope of the tube
Insulating oil
Collimator mirror
Tube housing window
Typical value:
≈1.5 mm aluminum equivalent
B. Added Filtration
Added filtration refers to thin aluminum sheets placed in the beam path.
Purpose:
Remove low-energy photons
Increase average beam energy
Reduce patient skin dose
C. NCRP Filtration Guidelines
Minimum total filtration:
Tube Voltage | Minimum Filtration |
|---|---|
< 50 kVp | 0.5 mm Al |
50–70 kVp | 1.5 mm Al |
>70 kVp | 2.5 mm Al |
Total filtration = Inherent + Added filtration
IV. Shielding
Shielding protects radiosensitive organs from unnecessary radiation exposure.
Highly radiosensitive organs include:
Gonads
Lens of the eye
Bone marrow
Thyroid
Breast tissue
A. Rationale for Use
Shielding should be used whenever:
Gonads are within 5 cm of the primary beam
Patient has reproductive potential
The shield does not interfere with the diagnosis
B. Types of Shields
1. Flat Contact Shields
Flat lead-impregnated vinyl
Placed directly over the gonads
Limitation:
Only useful for AP or PA recumbent positions
2. Shadow Shields
Attached to the x-ray tube housing.
Advantages:
Can be used for multiple patient positions
Do not contaminate sterile fields
3. Contour (Shaped) Shields
Shaped to fit the male gonads.
Held in place by disposable briefs.
Very effective for many positions.
4. Breast Shields
Used during procedures such as scoliosis imaging.
Often used in:
Spinal stoles
Compensating filters
Performing scoliosis studies in PA projection reduces breast dose significantly.
C. Patient Position
Positioning can reduce radiation exposure.
Examples:
PA abdominal radiography reduces gonadal dose
PA skull imaging reduces lens exposure
PA scoliosis imaging reduces breast exposure
V. Reducing Patient Exposure
A. Patient Communication
Good communication:
Builds patient confidence
Improves cooperation
Reduces motion
Prevents repeat exposures
Repeat exposures increase:
Patient dose
Facility cost
B. Patient Positioning
Accurate positioning ensures:
Correct anatomy visualization
Avoidance of repeat examinations
Positioning is especially important when using AEC systems.
C. Automatic Exposure Control (AEC)
AEC automatically terminates exposure when adequate radiation reaches the detector.
Purpose:
Maintain consistent image density
Prevent overexposure
1. Ionization Chamber
Located between the patient and the image receptor.
Measures radiation passing through the patient.
Exposure stops when a preset level is reached.
2. Phototimer
Located behind the image receptor.
Uses a fluorescent screen and a photomultiplier tube to detect radiation.
3. Backup Timer
Prevents excessive exposure if the AEC fails.
4. Minimum Response Time
The shortest exposure time possible for the AEC.
If the exposure time required is shorter than this, the image may become overexposed.
VI. Image Receptors
Faster image receptor systems require less radiation.
Rare-earth screens are commonly used because they:
They are four times faster than calcium tungstate screens
Require a lower radiation dose
Digital imaging also reduces dose due to:
Increased detector sensitivity
Efficient image capture
VII. Grids and Air-Gap Technique
Grids
Grids remove scattered radiation before it reaches the image receptor.
Benefits:
Improved radiographic contrast
Disadvantage:
Requires increased exposure (mAs)
Air-Gap Technique
Creates distance between the patient and the image receptor.
Scatter radiation diverges and misses the receptor.
It can function similarly to a low-ratio grid.
Common example:
Lateral cervical spine imaging at 72-inch SID without a grid
VIII. Fluoroscopy
Fluoroscopy generally delivers higher radiation doses than standard radiography.
Reasons:
X-ray tube closer to the patient
Longer exposure times
Protective measures include:
High kVp technique
Proper collimation
Limiting exposure time
Using pulsed fluoroscopy
Fluoroscopy units include a 5-minute cumulative timer.
IX. NCRP Recommendations for Patient Protection
Important standards include:
• Minimum 2.5 mm Al filtration for >70 kVp
• Exposure reproducibility variation must not exceed 5%
• Linearity variation must not exceed 10%
• Tube leakage radiation must be <100 mR/hr at 1 meter
• Minimum 12-inch SSD for radiography
• Minimum 15-inch SSD for stationary fluoroscopy
• Fluoroscopy beam intensity <10 R/min
• 5-minute cumulative fluoroscopy timer required
• Collimation accuracy within 2% SID
• SID indicator accuracy within 2%
ULTRA HIGH-YIELD BOARD FACTS
These are very commonly tested:
• Beam restriction is the most effective method of reducing patient dose
• Collimator = most efficient beam restrictor
• High kVp and low mAs reduce patient dose
• Minimum filtration >70 kVp = 2.5 mm Al
• Gonadal shielding is used if within 5 cm of the beam
• AEC stops exposure automatically
• Grids improve contrast but increase patient dose
• Fluoroscopy requires a 5-minute timer
Personnel Protection
Personnel protection refers to methods used to minimize radiation exposure to radiologic workers such as radiographers, radiologists, and other healthcare personnel.
The most important principle in personnel protection is the ALARA principle, meaning radiation exposure must be kept As Low As Reasonably Achievable.
Radiation exposure to personnel mainly comes from secondary radiation, especially scatter radiation from the patient.
I. GENERAL CONSIDERATIONS
A. Occupational Exposure
Radiographers must avoid unnecessary radiation exposure and ensure that patients’ radiation dose is kept to the minimum necessary for diagnostic imaging.
Sources of Radiation Exposure to Radiographers
Radiation exposure can come from:
Primary Radiation
Secondary Radiation
Scatter radiation
Leakage radiation
Radiographers must never be exposed to the primary (useful) beam.
Most Important Source of Scatter
The patient is the principal source of scattered radiation in diagnostic radiology.
Other scattering objects include:
X-ray table
Bucky-slot cover
Control booth wall
Occupational Dose Monitoring
The National Council on Radiation Protection and Measurements (NCRP) recommends personal radiation monitoring for individuals who may receive 10% of the occupational dose limit.
Occupational Dose Limit
Maximum occupational dose limit:
5 rem per year (50 mSv/year)
Therefore, monitoring is required if exposure may reach:
0.5 rem/year (5 mSv/year)
B. ALARA (As Low As Reasonably Achievable)
ALARA is the fundamental radiation safety philosophy used in radiology.
It means radiation exposure must be kept as low as reasonably achievable, even if exposures are below the legal limits.
Methods Used to Maintain ALARA
Radiation monitoring devices (dosimeters)
Regular radiation safety training
Radiation safety orientation for new employees
Radiation surveys of radiologic facilities
Review of monthly dosimeter reports
Monthly radiation monitoring reports are considered legal documents and must be carefully reviewed.
Even very small exposures should be reduced whenever possible.
II. OCCUPATIONAL RADIATION SOURCES
A. Scattered Radiation
Scatter radiation occurs when primary X-ray photons interact with matter and change direction.
This usually occurs through the Compton interaction.
Importance in Radiology
Scatter radiation is the most significant occupational radiation hazard in diagnostic radiology.
This is particularly important during:
Fluoroscopy
Mobile radiography
During fluoroscopy, high kilovoltage techniques produce energetic Compton scatter from the patient.
Scatter Radiation Intensity
The intensity of scatter radiation at 1 meter from the patient is approximately:
0.1% of the primary beam intensity
Because of this, the patient is considered the most important source of scatter radiation.
Other Scatter Sources
Additional sources include:
X-ray table
Bucky-slot cover
Control booth wall
B. Leakage Radiation
Leakage radiation is radiation that escapes from the X-ray tube housing in directions other than the primary beam.
NCRP Leakage Radiation Limit
According to NCRP regulations:
Leakage radiation must not exceed 100 mR per hour at a distance of 1 meter from the X-ray tube housing.
This limit ensures that radiation leakage from the tube housing remains minimal and safe.
C. NCRP Guidelines
The National Council on Radiation Protection and Measurements (NCRP) establishes guidelines to ensure radiation safety.
Important Equipment Safety Requirements
The control panel must indicate when X-rays are being produced.
Usually by a visible or audible signal
The exposure switch must be a dead-man switch.
A dead-man switch means the operator must continuously press the switch for the exposure to occur. If released, exposure stops immediately.
The exposure switch must be located so that it cannot be operated outside the shielded control booth.
Rules for Selecting Someone to Assist a Patient
If a patient cannot hold the required position (e.g., children or weak patients), assistance may be required.
Guidelines include:
Preferably a male older than 18 years
A female older than 18 years who is not pregnant may also assist.
Must wear protective apparel
Must stand as far as possible from the beam
Must never stand in the path of the useful beam
Radiology personnel must never hold the patient during exposures.
Instead, use:
Mechanical restraining devices
Patient relatives (if necessary)
Fluoroscopy Protection Guidelines
Several NCRP recommendations apply during fluoroscopic procedures:
• Image intensifier must provide 2.0 mm lead equivalent protection
• The exposure switch must be a dead-man type
• Undertable fluoroscopic tubes must have a Bucky-slot cover with 0.25 mm Pb equivalent
• A cumulative timer must alert the operator after 5 minutes of fluoroscopy
• Table-side lead drapes must have 0.25 mm Pb equivalent
• Protective lead aprons must be 0.50 mm Pb equivalent
• Lead gloves must be 0.25 mm Pb equivalent
• Hands must never be placed in the useful beam
• Maximum fluoroscopic exposure rate: 10 R/min
III. FUNDAMENTAL METHODS OF PROTECTION
A. Cardinal Rules of Radiation Protection
The three cardinal principles of radiation protection are:
Time
Distance
Shielding
These are the most effective ways to reduce radiation exposure.
Time
Reducing the time of radiation exposure reduces radiation dose.
Example:
If a worker receives 10 mrem in 1 hour, then 30 minutes exposure results in 5 mrem.
Distance
Increasing the distance from the radiation source greatly reduces exposure.
This is described by the Inverse Square Law.
Shielding
Placing a barrier between the radiation source and the worker reduces exposure.
Examples:
Lead walls
Lead aprons
Lead shields
B. Inverse Square Law
The Inverse Square Law states:
Radiation intensity is inversely proportional to the square of the distance from the source.
Formula concept:
If distance doubles → exposure becomes 1/4
Example:
If 40 mrem is received at 40 inches, then at 80 inches the exposure becomes 10 mrem.
Increasing distance is therefore one of the most effective protection methods.
IV. PRIMARY AND SECONDARY BARRIERS
A. Primary Barriers
Primary barriers protect against the primary (useful) x-ray beam.
Examples include:
Radiographic room walls
Lead doors
Typical primary barrier thickness:
1/16 inch (1.5 mm) lead
Primary barriers are usually 7 feet high.
B. Secondary Barriers
Secondary barriers protect against:
Scatter radiation
Leakage radiation
Examples include:
Upper portions of walls
Control booth wall
Lead aprons
Secondary barriers typically require:
1/32 inch lead thickness
Important:
Secondary barriers do not protect against the primary beam.
Control Booth
The control booth protects the radiographer from radiation exposure.
Requirements include:
Shielded barrier
Leaded glass window (about 1.5 mm Pb equivalent)
Exposure switch inside the booth
The primary beam must never be directed toward the control booth.
B. Protective Apparel and Care
Protective apparel used by radiologic personnel includes:
• Lead aprons
• Lead gloves
• Thyroid shields
• Leaded eyewear
Required Lead Equivalents
Lead aprons: 0.50 mm Pb equivalent
Lead gloves: 0.25 mm Pb equivalent
Thyroid shields: 0.50 mm Pb equivalent
These are secondary protective devices, meaning they protect against scatter radiation but not the primary beam.
Care of Protective Apparel
Protective equipment must be handled properly.
Guidelines include:
• Hang aprons on proper racks
• Never fold lead aprons
• Do not drop them on the floor
Improper handling can cause cracks in the lead material, reducing protection.
Protective apparel must be radiographically inspected annually to detect cracks.
C. Protective Accessories
Additional protection devices include:
Mobile Lead Barriers
Mobile lead shields provide full body protection from scattered radiation.
They are used when personnel must remain in the room during exposure.
Lead Apron Under Pillow Technique
During some fluoroscopic procedures (e.g., GI or barium enema), a lead apron may be placed under the patient’s pillow to protect the radiographer assisting at the head of the table.
V. SPECIAL CONSIDERATIONS
A. Pregnancy
Pregnant radiographers require special radiation protection considerations.
Declaration of Pregnancy
The radiographer should declare pregnancy in writing to her supervisor.
Once declared:
• Occupational exposure history is reviewed
• A fetal monitor (baby badge) is issued
Fetal Dosimeter Placement
The fetal monitor is worn:
Under the lead apron at waist level
Fetal Dose Limit
Maximum fetal radiation dose during pregnancy:
500 mrem (5 mSv)
Under normal conditions, typical occupational exposures are far below this limit.
B. Mobile Radiography
Mobile radiography requires additional protection measures.
Guidelines include:
• Each mobile unit must have a lead apron
• Radiographer must stand as far away as possible
• Exposure cord must allow the radiographer to stand at least 6 feet from the patient
• Lead apron must be worn during exposure
C. Fluoroscopic Units and Procedures
Fluoroscopy produces higher radiation doses than routine radiography.
Required Source-to-Skin Distance
Minimum:
12 inches (30 cm)
Preferred:
15 inches (38 cm)
Fluoroscopic Exposure Limits
Maximum tabletop exposure rate:
10 R/min
Typical fluoroscopic current:
1–3 mA
Maximum:
5 mA
Fluoroscopic Protection Requirements
• Image intensifier must provide 2.0 mm Pb equivalent protection
• Bucky-slot cover must provide 0.25 mm Pb equivalent
• Protective curtain/drape must provide 0.25 mm Pb equivalent
• Total filtration must be 2.5 mm Al equivalent
• Cumulative timer must alert after 5 minutes
• Beam collimation must be visible on monitor
HIGH-YIELD EXAM POINTS
Most occupational exposure occurs during:
Fluoroscopy and mobile radiography
Most important scatter source:
The patient
Cardinal radiation protection rules:
Time – Distance – Shielding
Maximum occupational dose:
5 rem/year (50 mSv/year)
Maximum fetal dose:
500 mrem (5 mSv)
Leakage radiation limit:
100 mR/hr at 1 meter
Fluoroscopy exposure rate limit:
10 R/min
Radiation Exposure and Monitoring
The Three Things We Measure in Radiation
Radiation measurement focuses on three different concepts.
Students often confuse them, so remember this structure:
What is measured | Unit |
|---|---|
Ionization in air | Exposure |
Energy absorbed in tissue | Absorbed Dose |
Biological effect | Dose Equivalent |
We will learn each one.
Lesson 3 — Exposure (Roentgen)
Exposure tells us:
How much ionization radiation produces in air.
The unit used is:
Roentgen (R)
Key idea:
Roentgen measures charge produced in air by x-rays or gamma rays.
Important points:
• applies only to photons (x-rays and gamma rays)
• measures ionization in air
• not used for tissue dose
SI Equivalent
The SI unit replacing Roentgen is:
Coulomb per kilogram (C/kg)
Conversion:
1 R = 2.58 × 10⁻⁴ C/kg
Example
If x-rays pass through air and produce a lot of ionization, the Roentgen value increases.
But remember:
Roentgen does not tell us how much dose the patient absorbs.
For that we need another unit.
Board Exam Tip
If the question asks:
“Unit of ionization in air”
The answer is:
Roentgen
Absorbed Dose (Rad / Gray)
Exposure tells us what happens in the air, but doctors need to know:
How much radiation energy is absorbed by tissue.
This is called:
Absorbed Dose
Definition:
Energy deposited in 1 kilogram of material.
Traditional Unit
Rad
Rad stands for:
Radiation Absorbed Dose
SI Unit
Gray (Gy)
Conversion:
100 rad = 1 Gray
Example
If tissue absorbs radiation energy:
• molecules break
• chemical reactions occur
• cells may be damaged
That’s why absorbed dose is important in biology and medicine.
Board Exam Tip
Question:
Unit measuring energy absorbed by tissue?
Answer:
Rad or Gray
Biological Damage (Rem / Sievert)
Different types of radiation cause different levels of biological damage.
Example:
Alpha particles cause much more damage than x-rays.
Even if the absorbed dose is the same.
To account for this we use:
Dose Equivalent
Dose equivalent measures:
Biological effect of radiation.
Unit
Rem
Rem means:
Radiation Equivalent Man
Formula:
Dose Equivalent = rad × Quality Factor
SI Unit
Sievert (Sv)
Conversion:
100 rem = 1 Sv
Example
If a patient receives:
1 rad of x-rays → 1 rem
But:
1 rad of alpha particles → 20 rem
Because alpha radiation is much more damaging.
Quality Factor (QF)
Quality Factor represents:
How biologically damaging a radiation type is.
Typical values:
Radiation | Quality Factor |
|---|---|
X-ray | 1 |
Gamma | 1 |
Beta | 1 |
Alpha | 20 |
Fast neutrons | 20 |
Radiographers mostly deal with:
Low QF radiation (x-rays).
Linear Energy Transfer (LET)
Another concept related to biological damage is:
Linear Energy Transfer
LET describes:
How much energy radiation deposits as it travels through tissue.
High LET Radiation
Characteristics:
• deposits energy rapidly
• causes severe damage
Examples:
• alpha particles
• neutrons
Low LET Radiation
Characteristics:
• spreads energy over a distance
• causes less damage
Examples:
• x-rays
• gamma rays
• beta particles
Radiology uses low LET radiation.
Why Radiation Monitoring Is Needed
Radiographers work near radiation sources every day.
To ensure safety we must measure occupational exposure.
Monitoring devices help:
• measure worker dose
• ensure legal limits are not exceeded
• track lifetime exposure
Monitoring is required for workers who might receive:
More than 1/10 of the annual dose limit.
Dosimeters (Radiation Monitors)
A dosimeter measures how much radiation a worker receives.
The most common types are:
OSL dosimeter
TLD dosimeter
Film badge
Pocket dosimeter
We’ll study them one by one.
OSL Dosimeter (Most Common Today)
OSL means:
Optically Stimulated Luminescence.
Material used:
Aluminum oxide (Al₂O₃).
How It Works
Radiation hits the crystal.
Energy becomes trapped inside the crystal.
A laser stimulates the crystal.
The crystal releases blue light.
Light intensity = radiation dose.
Advantages
• very sensitive
• detects ~1 mrem
• reusable readings
• resistant to heat and moisture
Because of this, OSL is the most widely used dosimeter today.
TLD Dosimeter
TLD means:
Thermoluminescent Dosimeter.
Material:
Lithium fluoride (LiF).
How It Works
Radiation stores energy in the crystal.
Crystal is heated in a reader.
Heat releases light.
Light intensity indicates radiation dose.
Advantages
• very accurate
• not affected by heat or humidity
• reusable
Minimum detectable dose:
~5 mrem
Film Badge
Film badges use:
photographic film.
Radiation darkens the film.
The darker the film, the higher the dose.
Filters Used
Film badges contain metal filters such as:
• aluminum
• copper
• open window
These help determine:
• radiation energy
• radiation type
Disadvantages
• sensitive to heat
• sensitive to humidity
• must be processed monthly
Because of these issues, film badges are being replaced by OSL dosimeters.
Pocket Dosimeter
Pocket dosimeters provide immediate readings.
They look like small penlights.
Inside is a tiny ionization chamber.
When radiation enters:
• air becomes ionized
• electrical charge changes
• a fiber moves inside the device
You read exposure through the eyepiece.
Measurement Range
Usually:
0–200 mR
Limitation
Pocket dosimeters do not create permanent records, so they are usually used in high-exposure situations.
Where to Wear Dosimeters
Standard placement:
Collar level outside the lead apron
This measures exposure to:
• head
• neck
• thyroid
• eyes
Pregnant Workers
Two badges are used:
Collar badge (outside apron)
Fetal badge (under apron at waist)
This estimates radiation to the fetus.
Ring Dosimeter
Used when hands are near radiation.
Common in:
• fluoroscopy
• interventional radiology
• nuclear medicine
Measures dose to fingers.
Radiation Dose Limits
Dose limits are recommended by the
National Council on Radiation Protection and Measurements (NCRP).
Occupational Worker Limit
Annual limit:
5 rem (50 mSv)
Applies to workers 18 years and older.
Public Limit
0.5 rem (5 mSv) per year.
Pregnant Worker
Total fetal dose limit:
0.5 rem (5 mSv).
Monthly limit:
0.05 rem (0.5 mSv).
ALARA Principle
ALARA stands for:
As Low As Reasonably Achievable.
Goal:
Minimize radiation exposure.
Three protection methods:
Time – reduce exposure time
Distance – increase distance from source
Shielding – use lead barriers
Key Concepts to Remember
Exposure → Roentgen
Absorbed Dose → Rad / Gray
Biological Effect → Rem / Sievert