II
Image Intensifier (I.I)
Characteristics of Image Intensifier
Brightness Gain
- Definition: Product of minification gain and flux gain.
- Purpose: Produce a bright enough image to be visible in daylight.
- Brightness Gain Range: Usually between 5000 and 30,000.Multifield Image Intensification (magnification mode)
- Function: Allows choice between multiple field sizes to produce magnified images.
- Examples:
- Tri-field Intensifier (field sizes: 12 cm, 17 cm, 25 cm)
- Dual field Intensifier (field sizes: 12 cm and 25 cm)Automatic Brightness Control (ABC) System
- Function: Maintains constant image brightness during fluoroscopy examinations.Fluoroscopic Technique
- Consideration: Operation of the fluoroscopy system should optimize for lower patient radiation doses.
- Recommended kVp and conditions stated in the subsequent sections.Fluoroscopic Image Quality
- Multiple factors (contrast resolution, spatial resolution, distortion, quantum mottle) impact image quality.
Brightness Gain
Purpose
- Brightness gain means higher illumination of the image.
- Traditionally, images had to adapt to darkness; now, they are produced bright enough for daylight.
Total Brightness Gain
- Formula:
- Typical range for total brightness gain is 5000 to 30,000.
Minification Gain
Definition and Process
- Minification Gain: The process in image intensifiers that increases image brightness by concentrating electrons onto a smaller output area.
- Input phosphor size can typically range from 15-60 cm, with the output phosphor size being about 2.5 cm.
- Formula for Minification Gain:
Example Calculation
- Given: Input phosphor = 25 cm, Output phosphor = 2.5 cm
Conclusion: Minification gain is 100.
Flux Gain
Definition
- Flux Gain: Related to the rate of transfer of particles (electrons) resulting in increased number of light photons emitted from the output phosphor.
- Formula for Flux Gain:
- Typical range for flux gain is between 50 and 150.
Example Calculation
- If 50 light photons are produced for every x-ray photon:
- Calculation:
Total Brightness Gain Example
Example Calculation
- When minification gain is 100 and flux gain is 60:
Conversion Factor
- Definition: Another measure of brightness gain focused on the efficiency of conversion from x-ray exposure to visible light.
- Formula:
Aging and Performance of Image Intensifiers
- Average loss of brightness gain over time: 10% per year.
- Consistent checks are necessary to maintain effective dosages to the patient.
- Typical lifespan of I.I.: 3-5 years.
Multifield Image Intensification
Magnification Mode Description
- The diameter of input phosphor defines magnification.
- Smaller field sizes (e.g., switching from 25 cm to 17 cm) is executed electronically.
- Effects seen include:
- Increased voltage on electrostatic lenses.
- Focal points for electrons shift, impacting the minification gain accordingly.
Effects of Magnification Mode
- Minification Gain Reduction
- When moving from 25 cm to 12 cm, the minification gain decreases significantly. - Magnification Factor:
- Magnification factor for going from 25 cm to 12 cm:
- Patient Dose Increase: Estimation based on ratios of the input phosphor areas.
Technical Considerations with Magnification Mode
- Ensure effective collimation to maintain visibility of the anatomy under exposure.
Vignetting and Image Artifacts
Definition
- Vignetting: Darkening of image edges when using I.I.
- Only the central portion contributes to the image during I.I. operation.
Effects of Magnification Mode
- Contrast Resolution: Magnifying improves ability to discern tissue differences.
- Spatial Resolution: Improved resolution when using a smaller FOV; clearer details are depicted.
- Increased Patient Dose: Higher doses linked to trade-offs in improving image quality.
Automatic Brightness Control (ABC)
Definition and Methods
- ABC equalizes brightness across diverse anatomical densities and thicknesses.
- Methods Used:
1. Automatic Dose Rate Control (ADC): Adjusts procedure parameters to maintain brightness.
2. Automatic Gain Control (AGC): Adjusts signal amplification for brightness enhancement post-image formation.
Gain Control Explained
- Gain is likened to adjusting volume; higher gain results in bright images, but may increase electronic noise.
Key Differences between ADC and AGC
| Feature | ADC | AGC |
|---|---|---|
| Full Name | Automatic Dose Rate Control | Automatic Gain Control |
| Adjusts | Radiation output | Electronic signal |
| Affects patient dose? | Yes | No |
| Primary Role | Control exposure | Image signal amplification |
| Where applied? | X-ray tube settings | Imaging Intensifier / Camera settings |
Fluoroscopic Technique
Operation and Settings
- Usage of higher kVp leads to decreased dose with increased transmission through the patient.
- Recommended kVp ranges articulated per the imaging type (e.g., 110-120 kVp for Barium Enemas).
Fluoroscopic Image Quality Factors
Key Aspects
- Quality is influenced by:
- Contrast Resolution (affected by vignetting and systems age)
- Spatial Resolution (based on receptors used)
- Distortions (pin-cushion and vignetting).
Distortion Effects
- Pincushion Effect: Distortion from imperfect electron focusing causing the image edges to pull inward.
- Mitigation: Can be minimized in mag mode operations. - Vignetting: Results in unequal brightness across the image, countered by using mag modes efficiently.
Quantum Mottle
- Explanation of the grainy appearance correlates to insufficient radiation exposure; remedy involves increasing mA.
Conclusion on Multifield Image Intensification
- Recommendations to utilize the largest Field of View (FOV) for the body part of interest to manage patient dose efficiently without compromising quality.
Recommendations for kVp in Digital Fluoroscopy
- Different ranges specified for various procedures to maintain optimal image quality and minimize patient exposure:
- Barium Enemas (solid-column): 110-120 kVp
- Small Bowel: 110-120 kVp
- Upper GI: 110-120 kVp
- Air Contrast Barium Enema: 90-100 kVp
- Abdomen and Pelvis: 80-90 kVp
- Iodine Procedures in Abdomen: 75-85 kVp
- Myelogram: 75-85 kVp
- Chest and Mediastinum: 75-85 kVp
- Joint Girdles: 70-80 kVp
- Extremities: 65-75 kVp