COMD 461- Notes 9/9/26
Hearing Evaluation: Pure Tone and Speech Audiometry
Case History (Hx)
Interview and/or questionnaires
Hearing Loss
Tinnitus
Dizziness
Otalgia
Aural fullness
Considerations:
Use open ended questions, share personal case history — establish relationship
Functional impact
Onset (When does it start?)/duration/frequency
Otoscopy
Examine ear canal
Purpose: Examine direction and size of EAC
Examine the shape of the ear: small, large, more curvy
Determine if testing can be completed
Visualize tympanic membrane
See ear drums
Pure Tone Testing (AC & BC(bone conduction))
Purpose:
Determine a patient’s threshold
Pure tones across the speech frequency rancge
Speech threshhold
Instruction:
“You are gonna hear some beeps. Press the button when you hear the beeps…”
Patient’s response to the beeps can vary
Press a button
Raise your hand
Verbal response “yes”
For Kids: Stand up or count how many times
Time Locked Resposne
The patient presses the button right after presentation
Diagnostic Audiometer
Air Conduction Testing (AC)
Specifies degree
When Testing a Patient
Pateient slightly angles or facing away from window
With pediatric patients, may want them facing the window
No visual cues
False Negative Responses: Ex. Patient misunderstood/forgot the task — hearing worse then they should
False Positive Responses: EX. Patient responds to eagerly — tries to predict the sound
AC Frequency Testing Order
1000 Hz
250 Hz
500 Hz
1000 Hz re-check (important to check make to give threshold response)
2000 Hz
3000 Hz
4000 Hz
6000 Hz
8000 Hz
Interoctaves
3000 Hz & 6000 Hz
Usually tested through depends on clinic protocol
750 Hz & 1500 Hz
≥20 dB at adjacent frequencies
125 Hz
Vibrotactile — more felt than heard
PT Testing Method
aka “bracketing”
most common method of manual audiometry

PTA (Pure Tone Average)

Bone Conduction Considerations- specifies type
Placed on mastoid process behind the ear
Placed on forehead on occasion
Both ears uncovered
Occulusion effect: when an object fills the outer portion of EAC, it causes an increase in the level of sound generated in the EAC by BC
BC Testing Frequencies:
500 Hz
1000 Hz
2000 Hz
4000 Hz
Air Conduction — degree of hearing lost
Speech Audiometry
Complains about difficulties understanding speech so logical next step is to test hearing function using speech stimuli
Degree of hearing loss (HL) for speech
Ability to recognize and discriminate sounds of speech
Setup
Set up as pure tone testing
Diagnostic audiometer
Recording or Monitored Live Voice (MLV)
Should use pre-recorded material for consistency
If running MLV, use VU to make sure own signal stays around 0 VU
Monaural or biaural (one or both ears)
Headphones, inserts, or sound field speaker
Uses diagnostic audiometer
Speech Cognition Threshold (SRT)
Lowest hearing level at which speech is barely understood
Use spondaic words called spondees
Similar threshold searching as pure tone threshhold
50% of list of spondaic words is correctly identified
Consider familiarizing list of words for patients before searching for SRT
Can be checked relability of pure tones (PT) (SRT vs (PTA)
Spondee Words:

SRT Testing Procedure Example:

Speech Detection Threshold (SDT)
May not be able to obtain SRT due to language barrier or too difficult of a task for patient to follow even after familiarization
SDT: lowest speech level a patient can detect speech
SRT Cross Check

± 7 for PTA range
Word Recognition Score (WRS)
Percentade of a set of words that a patient accurately reports back
Purpose: to obtain a percentage measure that quantifies how clearly speech is heard
Score is a percentage (%)
Gives us an idea of audibility of speech
Words used are monosyllabic and phonetically balanced
Commonly used NU-6 word lists or W-22 word lists
NU-6 listes preferrable with English speaking adults
Instructions: hear sentences
Performance Index (PI) Function

Speech in Noise Test
Sentence testing with an increase of of background noise
5 target words per sentence
Signal to Noise Ratio: Reflects the relative intensity of the signal (speech) and competing noise

Dynamic Range: “Floor to Ceiling”
HTL = softest sound they hear
MCL= Comfortable
UCL = Highest sound