(Chapter 5 CLM)Keratometry and Topography: Corneal Measurement

Historical Foundations and Principles of Corneal Measurement

  • Fundamental Requirement: A contact lens fitter must measure the radius of curvature of the cornea to properly fit a contact lens and maintain a healthy lens-cornea relationship.

  • Optical Mechanism: The first refractive surface of the human eye (the tear/air interface) acts as a high-quality reflecting surface, allowing precise measurement of corneal curvature.

  • Historical Pioneers: In the late 19th century, Herman von Helmholtz, Antonio Placido Costa, Allvar Gullstrand, and E. Javal recognized that the anterior corneal surface functions optical-wise as a small, spherical convex mirror and designed methods to determine its curvature based on this principle.

  • Terminology Evolution:

    • Ophthalmometry: Coined for early instruments designed to determine the principal meridians (flattest and steepest curvatures) of the cornea.

    • Keratometry: The modern term used to denote the measurement of corneal curvature.

    • Keratoscopy: A technique introduced by Antonio Placido utilizing a disk marked with concentric dark and light rings and a central viewing aperture. Reflecting this disk off the cornea allows visualization of multiple meridians and detection of corneal surface irregularities.

    • Corneal Topography: The modern term for computer-driven instruments utilizing advanced algorithms and digital imaging to build detailed, full-surface spatial maps of corneal curvature.

Physics and Measurement Principles of Convex Surfaces

  • Reflection Geometry: Keratometry treats the anterior surface of the cornea as a highly polished convex spherical mirror. To measure its radius of curvature (rr), an illuminated target or mire of known dimensions is positioned at a known distance from the corneal surface.

  • Fundamental Keratometry Equation: Rays reflected from the cornea form a finite virtual image behind the surface. The size of this virtual image is directly proportional to:

    1. The radius of curvature of the cornea (rr).

    2. The physical size of the object or mire.

    3. The distance of the mire object from the cornea.

  • Internal Image Formation: The virtual image formed behind the cornea serves as the object for the telescope portion of the keratometer, which forms a real, inverted image inside the instrument on the reticle plane.

  • Fixation Tremor and Doubling Devices:

    • Fine physiological fixation tremor of the human eye prevents direct linear measurement of the reflected corneal image using a reticle scale.

    • To overcome eye movements, quantitative doubling devices split the image. Because both doubled images move identically in response to eye tremor, motion interference is completely neutralized.

    • The magnitude of doubling equals the actual size of the real image formed within the keratometer.

  • Types of Doubling Techniques:

    • Fixed Doubling with Variable Object Size: The doubling distance remains constant while the size of the target mire is adjusted.

    • Constant Object Size with Variable Doubling: The mire target size remains fixed while the doubling distance is adjusted. This is the primary method utilized in standard manual keratometers.


Keratometer mire reflected from spherical cornea showing plus and minus signs

Refractive Power Calculations and Index of Refraction

  • Measuring Power vs. Curvature: Keratometry directly measures only the radius of curvature (rr) at the corneal apex. To convert this radius into dioptric refractive power (DD), a refractive index formula is applied.

  • Refractive Power Formulas:

    • When radius rr is measured in meters:     D=1.3375−1.000r=0.3375rD = \frac{1.3375 - 1.000}{r} = \frac{0.3375}{r}

    • When radius rr is converted to millimeters:     D=337.5r (mm)D = \frac{337.5}{r\,(\text{mm})}

  • Selection of Assumed Index of Refraction:

    • To determine total corneal refracting power rather than just anterior surface power, early researchers selected an assumed combined refractive index accounting for the weak negative refracting power of the posterior corneal surface.

    • Listing and Helmholtz established 1.33751.3375 as the standard combined index of refraction for the cornea and aqueous humor.

    • Some instrument manufacturers calibrated dioptric scales using alternative corneal indices, such as 1.3361.336 or 1.3321.332

    • Instrument Calibration Discrepancies: While instruments calibrated to different refractive indices measure the exact same physical radius of curvature (rr), their listed equivalent dioptric powers can vary by as much as 0.75 D0.75\,D (three quarters of a diopter).

    • Standardization: A refractive index value of 1.33751.3375 is the universally adopted standard for corneal conversion tables and modern keratometers.

Chord Length Dynamics and Measurement Boundaries

  • Manual Keratometer Target Dimensions:

    • Mire target size: Fixed at approximately 64 mm64\,mm

    • Telescope focal plane: Focuses from the mire plane at approximately 75 mm75\,mm

  • Reflection Points and Chord Length:

    • The keratometer derives its curvature measurement from two discrete reflection points located at the extreme outer edges of the mire reflection along a given principal meridian.

    • The linear distance between these two reflection points on the cornea is defined as the chord length (CLCL).

    • The chord length is directly proportional to the radius of curvature of the surface being measured.

  • Instrument Extremes and Minimum Chord Lengths:

    • Steepest curvature limit: 52 D52\,D (radius of 6.49 mm6.49\,mm), requiring a minimum chord length of 2.6 mm2.6\,mm

    • Flattest curvature limit: 36 D36\,D (radius of 9.38 mm9.38\,mm), requiring a minimum chord length of 3.8 mm3.8\,mm


Chord length extremes measured by manual keratometer at 52D and 36D
  • Corneal Surface Requirements and Limitations:

    • Unmeasured Zones: The corneal area between the two reflection points and all regions peripheral to them are not evaluated during a single measurement.

    • Spherical Symmetry Requirement: The keratometer can only measure true primary curvature if the cornea has sufficient chord length to reflect all object rays symmetrically. If reflection points fall on asymmetric high or low spots, the instrument records a non-spherical mathematical average lying somewhere between the true curvatures of each reflection point.

Optics of Concave Surface Measurement

  • Application: Manual keratometers can measure concave reflecting surfaces, such as the posterior curve (base curve) of a contact lens.

  • Focal Shift Effect:

    • The keratometer focuses on the reflected image rather than on the physical reflecting surface itself.

    • When properly focused for a concave surface, the target mire is located at a greater physical distance from the reflecting surface than it would be for a convex surface of identical radius.

    • This increased object distance produces a slightly smaller reflected image.

    • The instrument interprets this smaller image as a smaller radius of curvature, leading to an artificial overestimation of dioptric power.

  • Correction Factor:

    • The systematic error over the entire operating range averages 0.03 mm0.03\,mm in radius of curvature, which corresponds to approximately 0.15 D0.15\,D

    • Standard contact lens conversion tables incorporate this concave-surface correction to yield true physical radii from drum readings.

Manual Keratometer Hardware Components

  • Historical Design: The standard manual keratometer design was created and patented by Bausch & Lomb, Inc. (Rochester, NY) in the late 1920s and early 1930s (e.g., patent US 1750931 A).


Components of a manual Bausch & Lomb keratometer
  • 15 Major Hardware Components:

    1. Vertical meridian marker: Identifies the vertical axis alignment.

    2. Axis scale: Displays axis orientation in degrees facing the operator.

    3. Rotating grip for locating the axis: Allows full rotation of the body tube to align with astigmatic meridians.

    4. Eyepiece: Focusable optical tube containing the internal reticle crosshair.

    5. Vertical measuring drum: Right-hand drum used to measure dioptric power along the vertical principal meridian (closest to 90∘90^\circ).

    6. Horizontal measuring drum: Left-hand drum used to measure dioptric power along the horizontal principal meridian (closest to 180∘180^\circ or 0∘0^\circ).

    7. Focusing knob: Adjusts target distance to sharpen doubled central mire circles.

    8. Locking knob: Secures the instrument head position relative to the base.

    9. Chin rest elevation knob: Adjusts patient chin height.

    10. Elevating knob: Adjusts vertical height of the entire instrument body.

    11. Forehead Rest Adjustment Knob: Sets depth position of the forehead support.

    12. Chin rest: Supports patient chin.

    13. Leveling sight pin: Pin on the side of the lamp house used to align instrument height with patient eyes.

    14. Head rest: Stabilizes patient forehead.

    15. Occluder: Swiveling white shield used to block vision in the non-measured eye.

Step-by-Step Manual Keratometry Protocol

  1. Eyepiece Adjustment:

    • Swivel the white-backed occluder (15) into a vertical position blocking the objective lens.

    • Turn the eyecap (4) counterclockwise as far as possible, then turn on the power.

    • Look through the eyepiece at the blurred reticle crosshair.

    • Slowly turn the eyecap clockwise until the crosshair comes into sharp focus.

    • Warning: Avoid turning the eyecap back and forth, as this stimulates observer accommodation and causes erroneous readings.

    • Note the reading on the outer edge of the eyecap scale and repeat to verify consistent eyepiece positioning.

  2. Patient Positioning:

    • Seat the patient comfortably with chin resting snugly in the chin rest (12) and forehead pressed firmly against the head rest (14).

  3. Instrument Height and Axis Alignment:

    • Grasp the rotating grip (3) and rotate the body so the axis scale reads exactly 90∘90^\circ and 180∘180^\circ

    • Release the locking knob (8) and swing the tube to align with the patient eye.

    • Adjust the elevating knob (10) until the leveling sight pin (13) on the lamp house lines up horizontally with the patient pupil center or outer canthus.

    • Swing to the contralateral eye to ensure both pupils line up evenly; adjust patient head position if tilted.

    • Cover the eye not being measured with the occluder (15).

  4. Patient Fixation:

    • Sighting from the side, align the illuminated circular mire reflection onto the center of the patient cornea.

    • Instruct the patient to look directly at the internal green fixation light inside the tube.

  5. Focusing the Instrument:

    • Look through the eyepiece to view the target mires.

    • Center the black reticle cross inside the double central focusing circle by swiveling the instrument and adjusting the elevating knob (10).

    • Lock the instrument position with the locking knob (8).

    • Rotate the focusing knob (7) until the doubled focusing circles merge into a single, sharp ring.

    • Astigmatism Consideration: In astigmatic eyes, the complete focusing circle will not sharpen simultaneously. Direct focus to sharpen the doubled horizontal plus sign first, disregarding the minus signs until later steps.


Reticle mire patterns during keratometer focusing and alignment
  1. Locating the Astigmatic Axis:

    • Turn the horizontal measuring drum (6) until the spur tips of the horizontal plus signs just touch.

    • Check the horizontal line passing through the plus signs:

      • If the horizontal white line is continuous and unbroken, the instrument is aligned with the corneal astigmatic axis.

      • If the horizontal line appears broken or stepped, rotate the entire barrel assembly using the rotating grip (3) until the line becomes continuous.

    • Axis Verification: Slightly defocus the image using the focusing knob (7) to double the plus sign. Verify that the single horizontal line of the left plus sign sits precisely at the midpoint between the doubled parallel lines of the right plus sign. Re-focus to single.

  2. Measuring the Horizontal Principal Meridian:

    • Turn the left-hand horizontal drum (6) to move the left plus sign until it superimposes perfectly onto the central focusing circle plus sign.

    • Read the dioptric power value from the left drum scale. This records the corneal curvature in the meridian closest to 180∘180^\circ (or 0∘0^\circ).

  3. Measuring the Vertical Principal Meridian:

    • Re-focus the central image using the main focusing knob (7) until the vertical minus signs (BB and B′B') above and below the center circle appear clear.

    • Turn the right-hand vertical drum (5) to superimpose the short minus sign (BB) onto the longer minus sign (B′B').

    • Read the dioptric power value from the right drum scale. This records the corneal curvature in the meridian closest to 90∘90^\circ

    • Operational Tip: Keep one hand constantly on the focusing knob (7) while adjusting drum controls to maintain sharp mire focus during minor physiological eye movements.

  4. Determining Astigmatism and Axis Orientation:

    • Corneal astigmatism equals the numerical dioptric difference between the horizontal drum reading (left) and vertical drum reading (right).

    • Read axis angles off the meridian indicators facing the operator (2):

      • For Minus Cylinder correction: Use the axis mark corresponding to the drum with the lower dioptric reading.

      • For Plus Cylinder correction: Use the axis mark corresponding to the drum with the higher dioptric reading.

Calibration, Precision Eyepiece Tuning, and Error Correction

  • Accommodation Mechanics: Focusing the crosshair by rotating the dial clockwise from maximum counterclockwise position places the reticle in focus with maximum plus (or least minus) vergence power, locking observer accommodation in abeyance.

  • Calibration Verification:

    • Instruments must be calibrated daily against a known steel ball bearing or convex test plate with a radius of 7.94 mm7.94\,mm (equivalent to 42.50 D42.50\,D corneal refractive power).

    • Uncalibrated instrument errors typically range from 0.25 D0.25\,D to 0.75 D0.75\,D

    • Because calibration errors are non-linear across the scale, subtracting a fixed correction factor from all clinical readings is inaccurate.

  • Practical Calibration Adjustment Procedure:

    1. Set both measuring drums to the exact dioptric power of the calibrated test sphere (e.g., 42.50 D42.50\,D).

    2. Adjust the eyepiece dial counterclockwise until the doubled mire images in the reticle achieve perfect superimposition.

    3. This procedure recalibrates the instrument scale to measure true curves accurately across its entire range, even if the reticle crosshair appears slightly out of focus.

Extended Range, Topogometry, and Automated Systems

  • Range Extension Techniques:

    • Standard measurement range: 36 D36\,D to 52 D52\,D

    • Range extension lenses attached over the eyepiece:

    • Auxiliary +1.25 D+1.25\,D sphere: Extends upper range limit up to approximately 61 D61\,D

    • Auxiliary −1.00 D-1.00\,D sphere: Extends lower range limit down to approximately 30 D30\,D

  • Topogometer:

    • An auxiliary fixation device attached to the front of a manual keratometer.

    • Utilizes a moveable, calibrated point source of light to measure corneal curvature at decentered locations across the surface.

    • Largely replaced by digital corneal topographers.

  • Automated Keratometers:

    • Electro-optical instruments that perform automated mire analysis faster than manual techniques.

    • Available in stationary tabletop models and handheld portable units for high-volume clinical settings.

Computerized Corneal Topography and Mapping Methods

  • Limitations of Manual Keratometry:

    • Measures only a small central zone (2.5 mm2.5\,mm to 3.0 mm3.0\,mm chord length) along two principal meridians.

    • Fails to assess the full corneal surface (11 mm11\,mm to 12 mm12\,mm average diameter).

    • Provides no information regarding peripheral corneal flattening rates, eccentricity (ee-value), or structural contour changes.

  • Corneal Topography Concepts: Uses digital video capture and algorithms to measure thousands of discrete spatial points across the entire corneal surface.

  • Topography Map Formats:

    • Axial Map (Power Map):

    • Calculates sagittal curvature in diopters.

    • Uses color-coded scales (red represents steeper curvature; blue represents flatter curvature).

    • Primary clinical tool for assessing overall corneal shape, symmetry, and limbal extension of astigmatism.

    • Tangential Map (Instantaneous Radius Map):

    • Calculates local curvature based on a tangent-to-normal mathematical approach.

    • Provides superior sensitivity for detecting localized changes in curvature and corneal pathologies (e.g., keratoconus).

    • Refractive Map:

    • Displays actual optical refraction and vision quality across the pupillary zone rather than simple geometric curvature.

  • Measurement Technologies:

    • Reflection Systems: Evaluate light reflected from the tear film; restricted to mapping the anterior corneal surface.

    • Projection Systems: Project light through corneal tissues; measure both anterior and posterior corneal surfaces, allowing precise pachymetry (corneal thickness measurement) required for LASIK surgical evaluations.

  • Contact Lens Fitting Modules: Topography software contains simulation modules that display virtual fluorescein patterns to evaluate contact lens design fit prior to physical lens application.

Clinical Keratometry and Irregular Corneal Fixation

  • **Fitting on