EKG Ch 7
EKG Adaptations and Troubleshooting
1. Discuss artifact and identify situations that require adaptations during EKG testing
EKG Technician's Role: To generate a correct and readable EKG tracing, adjustments may be needed under certain conditions of the patient or environment.
Definition of Artifact: Artifact refers to any interference with the recording of the EKG tracing. It can arise from:
Environmental factors
Equipment issues
Patient conditions
Consequences of Artifact: Artifact distorts the EKG tracing, potentially making it impossible for healthcare providers to interpret the rhythm, which could delay diagnosis and treatment.
Situational Awareness: EKG technicians must adapt procedures based on patient's health status and conditions noted in patient records, such as:
Pregnancy
Amputation
Presence of scar tissue
Example Visuals:
Fig. 7-1: Compares an artifact-free normal heart rhythm with an EKG tracing disrupted by artifact.
2. Infection Prevention and EKG Testing
Standard Precautions: EKG technicians must use Standard Precautions with all patients.
Transmission-Based Precautions: Additional considerations for patients under these precautions:
Droplet Precautions: Wear an N95 mask before entering the room, encourage patients to wear a mask if relocated.
Airborne Precautions: Utilize negative pressure rooms or portable air filters; wear N95 respirators and limit opening doors.
Contact Precautions: Wear gloves, gowns, masks, and eye protection before entering the room; follow cleaning protocols when exiting.
Common Pathogen: Methicillin-resistant Staphylococcus aureus (MRSA) requires adherence to Contact Precautions.
Reporting Suspected Illness: Technicians should inform supervisors immediately upon noticing unrecorded symptoms of communicable illnesses.
3. Demonstrate solutions to different types of EKG artifact
Types of Artefacts:
Somatic Tremor Artifact: Produced by patient movement, resulting in a choppy distortion.
Causes: Tremors, shivering.
Solutions:
Positioning hands under buttocks.
Adjusting room temperature and providing blankets for shivering patients.
Reassuring anxious patients about painlessness.
Wandering Baseline Artifact: The isoelectric line shifts unpredictably up and down.
Causes: Muscle movement from breathing or poor electrode contact.
Solutions:
Remind patients to breathe normally and remain still.
Placing limb leads on wrists and ankles may help.
Improve electrode contact through proper skin preparation:
Remove sweat with towels.
Clean skin with soap and water to remove oils.
Abrade skin gently for better electrode gel contact.
Prepare skin appropriately for electrode placement to avoid wanders.
Example Visuals:
Fig. 7-2: Illustrates somatic tremor artifact.
Fig. 7-3: Displays wandering baseline artifact.
Fig. 7-4: Shows the use of abrading tape for improving skin-electrode contact.
4. Sources of Electrical Interference and Solutions
Electrical Interference: Can reduce signal quality, also referred to as AC or 60-cycle interference.
Causes: Other appliances in operation nearby.
Solutions:
Turn off appliances that may be causing interference.
Avoid using extension cords.
Ensure the machine is plugged into a grounded outlet.
Turn off fluorescent lights if they cause interference.
Broken Recording (Interrupted Baseline): EKG machine fails to locate a signal, resulting in a zig-zag pattern.
Common Cause: Frayed or broken lead wires.
Solutions: Check and replace damaged lead wires before use.
Example Visuals:
Fig. 7-5: Illustrates appropriate use of medical razors or clippers to remove skin hair, ensuring proper electrode contact.
Fig. 7-6: Shows an EKG tracing affected by electrical interference.
Fig. 7-7: Depicts a case of broken recording due to faulty lead wires.
5. Adaptations for EKG Procedures
General Adaptations: Changes to electrode placement or patient positioning may be needed based on specific conditions that deviate from the standard procedures.
Documentation of Changes: Any changes must be documented in the patient's record and noted on the EKG tracing to avoid misinterpretation by the provider.
Specific Situations Requiring Changes:
Dextrocardia: Heart points to the right, requiring mirrored electrode placement on the right side.
Limb Amputation/Injury: Electrodes are placed on the torso or closer to the torso if positioned incorrectly.
Presence of Scar Tissue: Avoid placing electrodes on scar tissue from surgeries like mastectomy.
Medical Implants: Electrode placement must be adjusted around pacemakers or breast implants.
Large Breasts: The breast tissue is positioned to allow for electrode placement on the chest wall.
Pediatric Patients:
Under 2 years old: Use right-sided EKG placement (dextrocardia adaptions).
Age 2-12 years: Use standard placements with minor adjustments, such as moving V4 to the right.
Posterior EKG: Used for diagnosing inferior wall myocardial infarctions, requiring electrodes for V4, V5, and V6 to be placed on the back.
Example Visuals:
Fig. 7-8: Shows right-sided electrode placement for dextrocardia.
Fig. 7-9: Illustrates electrode adaptation for patients with limb amputations.
Fig. 7-10: Depicts alternate placements around surgical scars.
Fig. 7-11: Provides layout for pediatric age-based electrode placement.
Fig. 7-12 & 7-13: Indicate posterior EKG lead placements and labeling necessity.
Fig. 7-14: Displays modified positioning for pregnant patients to ensure comfort and safety.
6. Troubleshooting and Maintenance of EKG Machines
User Manuals: Essential for understanding safety, operation, troubleshooting, and maintenance. Manuals should be accessible both in print and online.
Basic Troubleshooting: The EKG technician should check:
Battery levels
Paper supply status
Expiry of consumables
Verify that manufacturer instructions are followed.
Reporting and Handling Equipment Problems:
Problems need to be reported per facility procedures.
Place defective or out-of-service tags on faulty equipment and avoid usage until repaired.
Proper Maintenance: Regular cleaning, storage practices, and maintenance can minimize equipment issues and prolong device lifespan.
Chapter Review
Artifact and EKG Tracings: Artifact can interfere with readability and interpretation of EKGs. Several types must be recognized, including somatic tremor, wandering baseline, electrical interference, and broken recording.
Adaptations for Specific Conditions: Changes in patient conditions must be accommodated, including altered electrode placements for amputations, scar tissue, and pediatric patients, among others.
Equipment Handling: Follow manufacturers' guidelines and institutional protocols to ensure optimal operation and readiness of EKG recorders.