Introduction to Radiologic and Imaging Sciences and Patient Care

Image Receptor Characteristics

  • Digital image receptors have inherent characteristics that determine spatial resolution limitations.

  • These characteristics include:

    • Matrix size

    • DEL size

    • DEL pitch (physical distance between adjacent DELs, measured from center to center)

    • Sampling frequency (rate at which the signal is sampled)

  • Higher matrix sizes, smaller DELs with shorter distances between them, and higher sampling frequencies yield higher spatial resolution.

  • Radiologic technologists cannot alter these inherent characteristics.

Contrast Resolution

  • Contrast resolution is the ability to distinguish between subtle differences in image receptor exposures (signal values) and differentiate them from noise in the image.

  • It is affected by image receptor contrast and subject contrast.

Image Receptor Contrast

  • Image receptor contrast is the range of signal values acquired by the image receptor.

  • It is controlled by:

    • Detective quantum efficiency (DQE): measure of the sensitivity and accuracy of the image receptor

    • Grayscale bit depth: the maximum number of shades of gray each detector element can handle

    • Dynamic response: the ability of a detector material to sense the incoming radiation

Fluoroscopic Imaging

  • Fluoroscopic examinations involve a combination of imaging processes.

  • The fluoroscopic image is a dynamic (moving) image, unlike static radiographic images.

  • The fluoroscopic examination is divided into:

    • Viewing a physiologic event in real-time

    • Archiving images for later review

  • Modern fluoroscopic units have the x-ray tube located either over or under the x-ray table.

  • Opposite the tube is the image intensifier unit or flat-panel detector (FPD), which intercepts the attenuated beam as it exits the patient.

  • The image intensifier or FPD is the image receptor in this case.

  • X-ray photons reaching the image intensifier or FPD are transformed into an electronic image, which is displayed on a monitor.

  • Radiologists can view physiologic events and observe abnormalities in function.

Image Archiving Methods

  • Spot imaging: the fluoroscopic unit changes to radiographic mode for the exposure duration; images are processed, viewed, and stored.

  • Digital fluoroscopy: has replaced traditional filming and allows for easy manipulation of fluoroscopic images after conversion to a digital signal.

Summary of Key Concepts

  • The task of the radiologic technologist is to capture images produced by x-rays in a format allowing for storage and repeated viewing.

  • The basic mechanism of x-ray production has not changed significantly.

  • A beam of x-rays, produced mechanically by passing high voltage through a cathode ray tube, traverses a patient and is partially absorbed.

  • An image receptor intercepts the x-ray photons that exit the patient.

  • Various IR systems are used in radiography, including CR, DR (direct and indirect conversion), and fluoroscopic imaging systems.

  • A good-quality image must have a proper balance of:

    • Image signal properties (IR exposure and contrast) affecting the visibility of the image

    • Image resolution properties (spatial resolution and contrast resolution) affecting the sharpness and accuracy of the image

  • The radiographer must be able to manipulate technical factors such as mAs and kVp.

Central Ray Angulation and Body Part Rotation

  • Changing the orientation of the body part undergoing radiography also affects the relationship of the beam, objec t, and IR.

  • If the object of interest is superimposed on another object, the resulting image is difficult to evaluate.

  • By rotating or obliquing the body, the object of interest can be projected free from the interference of the overlying object.

  • Frequently, a combination of part rotation and central ray angulation is used to best demonstrate the anatomic details free from superimposition by overlying structures.

  • Ideally, the goal of a radiologic technologist is to place the anatomic part parallel to the IR and have the central ray aligned perpendicular to the IR.

Controlling Object-to-Image Receptor Distance (OID)

  • Control over OID depends on the radiographer's knowledge of anatomy and positioning.

Distortion

  • Distortion is the misrepresentation of the true size or shape of an object.

Size Distortion

  • Magnification is the only possible size distortion in radiography, due to the divergent property of x-ray photons.

  • Magnification can be controlled by proper SID and OID.

Source-to-Image Receptor Distance (SID)
  • The size of the recorded image varies with the SID.

  • If the SID decreases, magnification increases.

  • Magnification decreases as SID increases.

  • The use of standardized SIDs allows the radiologist to assume a specific magnification factor is present on all images.

  • Noting any deviation from the standard SID is extremely important.

Object-to-Image Receptor Distance (OID)
  • Varying the OID also influences the magnification.

  • If the object is moved closer to the receptor, magnification decreases.

  • If the object is moved farther from the receptor, magnification increases.

  • Magnification decreases as OID decreases.

  • The best image is produced with a small OID and a large SID.

Shape Distortion

  • The misrepresentation of the shape of an object on an image is called shape distortion.

  • It is controlled by the alignment of the beam, part, and IR.

  • Influencing factors include central ray angulation and body part rotation.

Central Ray Angulation
  • The beam of radiation diverges from the source in an approximately pyramidal shape.

  • The photons in the center of the beam travel along the straightest pathway (central ray).

  • The most accurate representation of an object results from the passage of photons in a straight line through the area of interest.

  • When the central ray is angled, the relationships among the beam, part being imaged, and IR are altered.

  • Objects may appear elongated or foreshortened.

  • Central ray angulation is used in many radiographic examinations to help demonstrate specific anatomic details.

Focal Spot Size

  • The width of the beam is controlled by the selection of the small or large focal spot.

  • The small focal spot is used when fine detail is required, as in the radiography of small bones.

  • The large focal spot is used for most general radiographic examinations.

Source-to-Image Receptor Distance (SID)

  • The distance from the source of x-ray radiation to the IR is a major influence on the spatial resolution of the image.

  • Penumbra (image unsharpness) obscures the true edge (umbra).

  • If the flashlight is positioned closer to the object, then as distance decreases, the penumbra around the true shadow increases.

  • In radiography, the greater the SID, the better the spatial resolution will be.

  • SIDs are standardized so that the degree of penumbra is at least a known factor.

  • The most common SID is 40 inches, but chest radiography uses a 72-inch SID.

  • The SID of an examination should always be indicated to allow for the calculation of image unsharpness.

Object-to-Image Receptor Distance (OID)

  • When the object is moved closer to the IR, penumbra decreases and image sharpness increases.

  • As the object is moved farther from the receptor, penumbra increases and sharpness decreases.

  • The smaller the OID is, the better the spatial resolution will be.

  • Many of the objects that must be radiographed are structures located deep within the body.

  • Getting them close to the IR is often impossible.

Beam Modification

  • One of the primary purposes of beam modification is scatter control.

  • Scattered radiation reaching the IR produces nondiagnostic exposures referred to as fog.

  • Removal of fog results in the loss of some specific IR exposures.

  • Anything that decreases scatter increases subject contrast.

Digital Image Receptor Systems

  • With digital image receptor systems, the traditional relationship between displayed image contrast and exposure variables does not exist.

  • Displayed image contrast is determined by the look-up table (LUT), which is applied through digital processing.

  • The proper LUT provides the proper grayscale for the displayed image, regardless of variations in kVp and mAs, resulting in consistent images.

  • The information is displayed in a range of brightness, which can then be adjusted through the manipulation of the window width.

  • Sufficient differences in exposure to the receptor are still necessary.

  • The radiographer must set proper exposure factors to create an acceptable image quality with minimum exposure of the patient.

Grids

  • Grids absorb scatter radiation, improving subject contrast.

  • Grids are available in a diverse range of ratios and frequencies.

  • These choices allow the radiographer to select how much scatter radiation is to be eliminated, and thereby impact subject contrast.

  • Due to the high sensitivity of digital image receptors to scatter radiation, the use of a grid is much more critical than in film-screen radiography.

  • When possible, the highest grid frequency available should be used with digital imaging to avoid certain grid errors.

Image Resolution Factors

Spatial Resolution

  • The sharpness with which an object's structural edges are represented on an image is referred to as spatial resolution.

  • It is also described as sharpness of detail, definition, and recorded detail.

  • Sharpness of detail is complemented by visibility of detail.

  • Good image quality requires a proper balance of the two.

  • The chief factors affecting spatial resolution include temporal resolution, beam geometry, and image receptor characteristics.

Temporal Resolution
  • Temporal resolution is the relationship between the duration of signal acquisition (exposure time) and dynamic motion of the anatomy.

  • The most common cause of image unsharpness is motion.

  • Voluntary motion can be controlled by careful instructions, suspension of respiration, short exposure times, and immobilization devices.

  • Involuntary motion is best controlled by the shortest exposure time possible.

  • Equipment motion can be decreased by the use of short exposure times.

Beam Geometry
  • Beam geometry is affected by the focal spot size, distance, and distortion.

  • The fundamental problem is attempting to represent a three-dimensional object on a two-dimensional image.

  • Lessening the effect of this inherent loss of detail is possible by adjusting the factors over which the technologist has control: focal spot size, SID, object-to-image receptor distance (OID), alignment, and angulation.

Kilovoltage Peak (kVp)

  • kVp still controls subject contrast and should be selected for the desired level of latent image contrast (pre-processing).

  • When the kVp is too low, most of the photons do not reach the image receptor because they are absorbed in the patient.

  • With high kVp, subject contrast is decreased because both thick and thin body parts are penetrated, which reduces the degree of differential attenuation.

Amount of Irradiated Material

  • The amount of irradiated material is determined by the thickness of the body part being imaged and the size of the irradiated field.

  • As body part thickness increases, more x-ray photons will be attenuated.

  • Conversely, as thickness decreases, so does attenuation.

  • Subject contrast is reduced with an increase in scatter production.

Type of Irradiated Material

  • The atomic number of the irradiated material and its tissue density influence subject contrast.

  • Materials with a higher atomic number absorb a greater percentage of the x-ray photons.

  • Tissue density is determined by how tightly the atoms of a given tissue are packed together.

  • When the difference in the densities of adjacent tissues is great, subject contrast is increased.

Contrast Media

  • Contrast media are substances that attenuate the beam to a different degree than the surrounding tissue.

  • Contrast media increase subject contrast by introducing greater differences in atomic number variations than those that exist naturally.

  • Examples include barium and iodine compounds and air.

  • Technical factors, particularly kVp, must be adjusted for adequate penetration.

Milliampere-Seconds (mAs)

  • mAs alters IR exposure (signal value) and therefore influences subject contrast.

  • If an image is grossly overexposed or underexposed, then contrast is affected.

  • For digital systems, the exposure indicator should be in the acceptable range to assure that the image receptor received the correct exposure.

  • No increase in mAs can compensate for inadequate penetration.

Digital Image Contrast

  • Digital image contrast must be considered in terms of subject contrast, raw image contrast, and displayed image contrast.

  • Displayed image contrast is determined by computer processing and can be altered by the adjustment of window width, independent of the technical factors used during image acquisition.

  • Displayed digital images with relatively few gray tones possess high contrast, short-scale contrast, and narrow window width.

  • Images with greater numbers of gray tones possess low contrast, long-scale contrast, and wide window width.

Exposure Indicator Deviation Index Control Limits for Clinical Images

Table 7.3

  • Details the exposure limits

Table 7.2

  • Details the Exposure indicators

Scatter Radiation

  • If scatter radiation reaches the IR, it is not carrying useful information.

  • Scattered photons that strike the IR degrade the quality of the image by contributing unwanted exposure known as fog.

  • By restricting the size and shape of the primary beam to the area of interest, we are decreasing the probability of the production of scatter radiation.

  • A decrease in scatter causes a decrease in IR exposure (signal value).

Grids

  • A grid is a device that is designed to remove scattered photons exiting the patient before they reach the IR.

  • A grid consists of thin radiopaque lead strips interspersed with radiolucent spacing material.

  • The grid is placed between the patient and the IR to intercept scattered photons.

  • Increasing the lead in a grid increases its ability to remove scatter from the remnant beam.

  • Grid ratios commonly range from 5:1 to 16:1, with the higher ratio grid able to prevent more scattered photons from reaching the IR.

  • Grid frequencies are generally 85 to 200 lines per inch.

  • Decreasing the amount of scatter enhances the radiographic contrast.

  • All other factors being equal, if the same exposure factors are used with a 5:1 grid and a 10:1 grid, the 10:1 grid produces an image with less IR exposure (signal value).

Higher Fill Factor

  • A higher fill factor percentage results in increased detector sensitivity.

  • Sensitivity is determined by how much charge a receptor produces per incident x-ray photon before the collected signal is amplified.

Evaluating IR Exposure (Signal Value)

  • With digital IRs, the key factor that must be considered when evaluating proper IR exposure (signal value) is the exposure indicator.

  • Additional considerations include exposure latitude, automatic rescaling, and window leveling.

Exposure Indicator

  • Exposure indicator is a numeric representation of the quantity of exposure received by a digital IR.

  • The target exposure indicator for all systems designates the middle of the image receptor's operating range as the optimal exposure.

  • The deviation index (DI) would be used in conjunction to determine if the appropriate radiographic technique factors were used during an examination.

  • The DI can be defined as a comparison of the target value with the actual exposure recorded by the IR.

Inverse Square Law

  • I=1d2I = \frac{1}{d^2}

mA and time

  • mA and time are all directly proportional to beam intensity

Beam Modification

  • Anything that changes the nature of the radiation beam, apart from the factors already discussed, is referred to as beam modification.

  • The beam may be modified before it enters the patient (primary beam modification) or after it exits the patient (scatter control).

  • The amount of attenuating material required to reduce the intensity of a beam to half the original value is referred to as the half-value layer.

  • Beam restriction involves the use of devices, such as a collimator, to confine the x-ray beam to the area of interest, thereby reducing exposure to body parts other than those under examination.

Patient Factors

  • Various patient factors affect IR exposure (signal value).

  • Patient size and thickness, the predominant atomic numbers of the materials, pathologic conditions, anomalies, temporarily compressed tissues, and a number of other techniques all change the subject density of the tissues being examined.

  • As subject density increases, IR exposure (signal value) decreases, and vice versa.

Kilovoltage Peak

  • In addition to the number of x-ray photons produced, the relative strength of the photons must be considered.

  • The kVp setting determines the highest energy level, or the peak, possible for the photons within that beam.

  • The relationship between kVp and IR exposure (signal value) is not as simple as that of mAs.

  • Increasing kVp 15% will approximately double IR exposure (signal value).

  • Decreasing kVp 15% will approximately halve IR exposure (signal value).

  • kVp can be changed while maintaining the same IR exposure (signal value) as follows

Image Quality Factors

Exposure Maintenance Formula

  • mAs<em>1mAs</em>2=D<em>12D</em>22\frac{mAs<em>1}{mAs</em>2} = \frac{D<em>1^2}{D</em>2^2}

Inverse Square law

  • I<em>1I</em>2=D<em>22D</em>12\frac{I<em>1}{I</em>2} = \frac{D<em>2^2}{D</em>1^2}

Four factors for x-ray production

  • Vacuum

  • Source of electrons

  • Method to accelerate electrons rapidly

  • Method to stop the electrons