Diagnostic Pulmonary Function Testing: Lung Volumes, Airway Resistance, and Gas Distribution Analysis
Overview of Advanced Pulmonary Function and Lung Volume Testing
- Advanced lung volume tests evaluate volumes and capacities that cannot be measured using simple spirometry, vital capacity maneuvers, or Maximum Voluntary Ventilation (MVV).
- Diagnostics obtained from lung volume testing are paired with Slow Vital Capacity (SVC) testing performed within the same session to make necessary volumetric calculations.
- Primary testing modalities include:
- Open-Circuit Multiple-Breath Nitrogen (N2) Washout.
- Single-Breath Nitrogen (N2) Washout (used briefly for assessing gas distribution).
- Closed-Circuit Multiple-Breath Helium (He) Dilution.
- Body Plethysmography (commonly referred to as the "body box").
- Secondary Gas Distribution Assessment:
- Distribution of ventilation can be secondarily derived from the results of a multi-breath nitrogen washout or a helium dilution test, though these are not standalone certified gas distribution tests.
- Central Target Measurement:
- The primary objective of these testing techniques is to measure Functional Residual Capacity (FRC).
- Alternative Diagnostic Modalities:
- Chest X-Ray: Allows visual assessment of rib spacing and the position of the diaphragm relative to the ribs to form educated estimates regarding air trapping or hyperinflation.
- Computed Tomography (CT) / Magnetic Resonance Imaging (MRI): Advanced radiological imaging techniques that quantify volumetric data but involve high financial costs, radiation exposure (CT), injectable contrast media, or strong magnetic fields (MRI).
- Routine pulmonary breathing tests remain the preferred standard diagnostic option.
Indications for Lung Volume Measurements
- Differentiating between obstructive and restrictive lung diseases. Restrictive lung disease can be definitively identified when measured lung volumes and capacities are reduced below normal limits.
- Staging the severity of restrictive lung disease.
- Evaluating therapeutic responses to medical or surgical interventions (such as evaluating volume expansion following a lobectomy).
- Preoperative patient evaluation.
- Quantifying the exact degree of hyperinflation and air trapping.
- Measuring Specific Airway Conductance (sGaw):
- Conductance (Gaw) represents the ease with which gas flows through the airways.
- The lowercase prefix "s" stands for specific, indicating conductance normalized to a specific lung volume.
- Specific conductance evaluates how effectively the airways conduct airflow at a designated lung volume, specifically measured at the End-Expiratory Level (EEL) or the level of FRC.
Circuit Designs and Equipment Configurations
- Body Plethysmography ("Body Box"):
- The only testing modality that requires placing the patient inside a sealed enclosure resembling a telephone booth.
- If a laboratory lacks a body box, lung volume measurements can still be effectively performed using nitrogen washout or helium dilution techniques.
- Open-Circuit Systems:
- A breathing circuit setup that allows the patient to inhale a fresh supply of gas with every breath.
- Used in open-circuit multiple-breath nitrogen washout testing.
- Closed-Circuit Systems:
- A rebreathing circuit setup where exhaled gas is recirculated through the system.
- Water Absorber (Desiccant): Required in testing circuits to eliminate condensation buildup in tubing, which would otherwise increase circuit resistance.
- Carbon Dioxide (CO2) Scrubber: Contains chemical absorbers (such as soda lime) to continuously remove exhaled CO2 from the circuit, preventing rebreathing of CO2 and maintaining normocapnic / eucapnic conditions (eucapnia).
- Supplemental Oxygen (O2): Added continuously into closed rebreathing circuits to replace oxygen consumed by the patient, maintaining normoxic conditions and constant circuit gas volume.
Lung Volumes, Capacities, and Diagnostic Ratios
- Functional Residual Capacity (FRC):
- Directly measured during volume testing.
- Defined as the volume of gas remaining in the chest at the end of a quiet, normal exhalation (End-Expiratory Level, or EEL).
- Residual Volume (RV):
- A calculated parameter obtained by subtracting Expiratory Reserve Volume (ERV) measured during Slow Vital Capacity (SVC) testing from the measured FRC:
RV=FRC−ERV
- Total Lung Capacity (TLC):
- A calculated parameter obtained by adding the measured FRC to the Inspiratory Capacity (IC) measured during SVC testing:
TLC=FRC+IC
- Residual Volume to Total Lung Capacity Ratio (RV/TLC):
- Represents the fraction or percentage of total lung volume that cannot be exhaled.
- Elevated RV/TLC ratios indicate air trapping and hyperinflation.
- Thoracic Gas Volume (VTG):
- Defined as the absolute volume of gas present in the chest at a specific point in time.
- Obtained exclusively via Body Plethysmography.
- When measured at the End-Expiratory Level (EEL), VTG is equivalent to FRC (VTG=FRC).
Pathophysiology of Airway Resistance and Compliance
- Obstructive Pathophysiology:
- Air trapping causes volume and pressure buildup inside the chest cavity.
- Patients are forced to breathe at higher lung volumes ("breathing on top of trapped air"), operating on a less compliant portion of the pressure-volume curve.
- Overcoming built-up intra-thoracic pressure increases airway resistance and the total work of breathing.
- Restrictive Pathophysiology:
- Lungs become pathologically stiff with high elastic recoil forces ("rubber band lungs").
- Patients breathe at lower lung volumes and must overcome excessive elastic resistance to inhale gas.
- Compliance is reduced, significantly increasing work of breathing during inspiration.
- Clinical Objective:
- Restoring a patient's FRC toward normal levels restores optimal lung compliance and reduces excessive work of breathing.
Open-Circuit Multiple-Breath Nitrogen (N2) Washout Test
- Test Principle:
- Baseline gas in human lungs contains approximately 78% nitrogen (75%–80%) in equilibrium with ambient room air.
- The patient breathes 100% oxygen (O2) to clear or "wash out" nitrogen residing in the lungs.
- Circuit Equipment:
- Open-circuit setup connected to a 100% O2 supply.
- Expiratory flow spirometer.
- Nitrogen Analyzer: Uses Emission Spectroscopy (or rapid gas chromatography) connected via a sampling line at the patient's mouth.
- Testing Protocol:
- Patient applies nose clips and forms a tight mouth seal around the mouthpiece.
- Patient breathes quiet tidal breaths on room air to establish a stable End-Expiratory Level (EEL).
- Software switches the inhalation valve to 100% O2 at the start of an inhalation at EEL.
- Patient continues relaxed, normal tidal breathing while inhaling 100% O2 continuously.
- Exhaled N2 concentration is monitored on a washout tracing graph plotting volume versus N2 concentration.
- Test ends when exhaled N2 concentration drops continuously down to 1.5%.
- Acceptability and Quality Control Criteria:
- Trial Requirement: Only 1 technically acceptable trial is required.
- Rest Interval: If repeated, a 15-minute rest period is required between trials to allow lung N2 concentration to return to normal baseline conditions.
- Reproducibility: If multiple acceptable trials are performed, measured FRC values must agree within 10% of each other. The reported FRC is the average of acceptable trials.
- Washout Duration:
- Healthy subjects wash out nitrogen in approximately 3 minutes.
- Obstructive disease prolongs washout time proportional to severity (>7minutes indicates severe obstruction or air trapping).
- Circuit Leaks:
- Graph must show a continuously falling N2 curve down to 1.5%.
- Spikes or sudden increases in N2 concentration (>1% increase) indicate a system leak or patient error (e.g., breaking the mouth seal or inhaling room air).
- If a leak occurs, the trial must be aborted, followed by a 15-minute rest period before retesting.
Closed-Circuit Multiple-Breath Helium (He) Dilution Test
- Test Principle:
- Patient rebreathes a known volume of gas containing 10% helium (He) in a closed spirometer system.
- Helium is an inert gas not naturally present in the lungs. Gas redistributes down its concentration gradient until equilibrium is established between the system box and the lungs.
- Circuit Equipment:
- Closed rebreathing circuit.
- Helium Analyzer: Uses a Thermal Conductivity Analyzer (or gas chromatography).
- CO2 absorber (soda lime) to remove exhaled carbon dioxide.
- Supplemental O2 line adding continuous oxygen to replace consumed oxygen.
- Water absorber to prevent condensation buildup.
- Testing Protocol:
- Equipment is prepared, and patient seal is verified with nose clip and mouthpiece.
- Patient breathes room air normally to establish a stable EEL baseline.
- Switch-in: At EEL, the valve opens to connect the patient to the closed circuit containing 10% He.
- Patient continues normal tidal rebreathing.
- Helium concentration is monitored as it falls to a stable plateau.
- End-of-Test Criteria:
- Equilibrium is achieved when He concentration changes by no more than 0.02% over a 30-second period.
- OR when the maximum test time limit of 10 minutes is reached.
- Acceptability and Quality Control Criteria:
- Trial Requirement: Only 1 technically acceptable trial is required.
- Rest Interval: If repeated, a 5-minute rest period is required (helium clears rapidly into ambient air).
- Reproducibility: Multiple acceptable trials must have FRC values agreeing within 10% of each other. The average FRC is reported.
- Equilibration Time:
- Healthy lungs equilibrate in approximately 3 minutes.
- Prolonged equilibration time up to 10minutes indicates obstruction.
- Switch-In Error:
- A correction factor accounting for deviations from the exact EEL during switch-in.
- A large switch-in error requires stopping and repeating the test.
- Sample baseline output parameters may show an FRC value of 3.81L with an equilibration time of 4.5minutes.
Diagnostic Limitations of Gas Dilution and Washout Methods
- Reliance on Well-Ventilated Lung Units:
- Both N2 washout and He dilution rely on all lung regions being well-ventilated and accessible to circuit gas distribution.
- Underestimation in Obstructive Lung Disease:
- In patients with severe obstructive lung disease (e.g., COPD) or significant gas trapping, poorly ventilated lung units or bullae do not participate efficiently in gas dilution or washout within standard timeframes.
- As a result, FRC, RV, and TLC values derived from gas dilution/washout tests can be significantly underestimated.
- Clinical Interpretations:
- Body Plethysmography evaluates total thoracic gas volume regardless of ventilation status, avoiding volume underestimation.
- When using gas dilution/washout systems, extending test duration (>7minutes for washout, 10minutes for helium dilution) allows clinicians to infer the presence and severity of poorly ventilated lung regions and gas distribution abnormalities.
Questions & Discussion
- Question: What does "jaw" stand for in sjaw?
- Answer: Conductance (Gaw). The lowercase "s" stands for specific, representing specific airway conductance (sGaw), typically measured at the end-expiratory level (FRC).
- Question: What is the rule regarding FRC agreement across multiple trials?
- Answer: Only one technically acceptable trial is required. However, if institutional policy or physician preference requires multiple trials, the calculated FRC values from acceptable trials must agree within 10% of one another. The final reported FRC value is the average of those trials.