Masking Lecture Notes
Well-being Check-in
Dedicate time for well-being activities.
Engage in hobbies and activities.
Commitment to self-care, even small steps are beneficial.
Introduction to Masking
Masking is crucial for accurate audiograms, especially in cases of complex hearing loss.
The lecture aims to explain: Why, When, What, and How of masking.
Why Masking is Necessary?
Sound travels via vibration affecting the tympanic membrane, ossicular chain, and cochlea.
Loud sounds can vibrate the entire skull, potentially reaching the non-test ear.
Example: Sound presented to a left ear with no hearing can cross over to the right ear if loud enough.
The question is: How loud does the sound have to be to cross over and be heard by the non-test ear?
This is addressed by understanding interaural attenuation.
Interaural Attenuation
Interaural attenuation refers to the amount of sound lost when crossing from one side of the head to the other.
The amount varies for each individual due to differences in head size and density.
Clinical protocols are based on studies estimating interaural attenuation.
Studies show a range of interaural attenuation values across different frequencies.
Clinically, the most extreme (minimum) value is used to account for the worst-case scenario of sound crossover.
Minimum interaural attenuation for headphones is 40 decibels.
Interaural Attenuation Example:
Presenting 90 dB to the test ear:
With 40 dB interaural attenuation, 50 dB reaches the non-test cochlea.
With 80 dB interaural attenuation, 10 dB reaches the non-test ear.
Masking Defined
Masking: Introducing a distracting tone (narrowband noise) to the non-test ear.
Narrowband noise is frequency-specific and masks the tone in the non-test ear.
Practical Illustration on Affinity
Example scenario is provided with the left ear having no measurable hearing, while the right ear has normal hearing at 0 dB.
Without masking, a threshold is obtained in the left ear at 40 dB.
This occurs because the 40 dB tone crosses over, losing 40 dB, and is heard by the right ear at 0 dB.
The masking procedure is demonstrated using Affinity software to show practical application.
Masking Table Setup
Masking tables are crucial for recording masking levels and responses.
Key elements include frequency, test ear (the ear receiving the tone), masking transducer (headphones).
The masking table has two columns: Masking (dB) and Tone (response).
Start with the unmasked threshold before any masking is applied.
Masking starts by adding 10 dB of masking to the non-test ear, gradually increasing noise level.
If the patient doesn't respond, the tone volume is increased.
Responses are recorded in the masking table, noting masking level and tone volume.
Detailed Walkthrough on Affinity
The masking side of Affinity is typically red, indicating it's for the right ear.
The narrowband noise option is automatically selected for masking.
Start with 10 dB of masking, increasing gradually until the non-test ear is effectively masked.
If the patient doesn't hear the tone, increase the tone level in 5 dB steps until a response is obtained.
The masking table reflects only the levels at which a response is obtained.
The presenter clarifies that in masking we do not consider 2 out of 3 responses.
Increasing masking noise essentially gives the non-test ear a temporary hearing loss.
Advanced Masking Concepts
The tone is increased by 10 dB, then the masking occurs.
Further increases in masking with no response lead to higher tone presentations.
In cases when you reach 105 dB and can go no further because the equipment wont allow , use the \"N\" function to make a note. Also record as NR105
Scenario: Left Ear Can Hear at 80 dB
The speaker adjusts the example to assume the left ear can hear at 80 dB.
The masking table is the same, starting with a tone heard at 40 dB without masking.
The masking process is reiterated, clarifying whether the response is from the right or left ear.
The concept of a \"plateau\" is introduced.
Plateau in Masking
Plateau: consistent response across three masking presentations demonstrating sufficient masking.
If the left ear can hear, the masking is increased until the left ear begins to respond.
Under-masking: Not enough masking, the right ear is still responding.
Sufficient masking: The left ear is responding, which is at the plateau.
Additional Masking Concepts
10-for-10 shift: Increase masking by 10 dB, tone moves by 10 dB.
It may not be obvious that a plateau has been reached.
Increase masking and look for that plateau.
How to Decide Whether Masking Is Required
Masking decisions depend on transducers and interaural attenuation.
For headphones, the minimum interaural attenuation is 40 dB.
For tube phones (insert earphones):
70 dB interaural attenuation for frequencies 1 kHz or less.
50 dB interaural attenuation for frequencies above 1 kHz.
Bone Conduction Masking
Minimum interaural attenuation for bone conduction is 0 dB.
Sound presented via bone conduction can be heard by both cochleas.
Masking for bone conduction is needed to differentiate between conductive and sensorineural hearing loss.
Bone conduction is masked if there is a >15 dB gap between air and bone conduction.
Bone Conduction Setup
Need to add headphone to test the other ear, if not already in place.
Occluded if bone conduction done with the headphone on already
Unoccluded if bone conduction done without the headphone on already
When you test it unmasked, yes, it will always be what the better cochlear is
Clinical Considerations with Masking
There is a risk of over-masking.
Bone conduction will always give you the better cochlear thresholds. So you could give the headphone to occlude the ear.
Going Through an Audiogram and Deciding if Masking Is Required
Review of how the minimum interaural attenuation will affect the need to apply masking.
You can base stuff off of air conductions. The same rule as air conductions will applies using 40 dB rule to better cochlea.
If greater than there are greater than a 15 dB difference in air bone gap then we mask for bone conduction.
Can interpolate between the air and bone to decide if masking is required.
Mask thresholds will always be correct.
Additional Practice and Application
Discussion and review of additional exercises on when to mask.