Clinical Chemistry I: Introduction to Beckman DXC Analyzers
Overview of the Beckman DXC 800
The Beckman DXC 800 is a console computer-controlled analyzer.
It is utilized for analyzing various sample types including:
Serum
Plasma
Urine
Cerebrospinal fluid (CSF)
Pre-treated whole blood
The analyzer performs general tests, therapeutic drug monitoring (TDM), and other tests.
Modular Chemistry (MC) analytes analyzed by the DXC 800 include:
Sodium
Potassium
Chloride
Total carbon dioxide
Calcium
Phosphate
Glucose
Urea
Creatinine
Total Protein
Albumin
Uric acid
The typical analysis time for tests is approximately 1 minute.
Main Components of the DXC
Analyzer Components:
Sample Handling
Modular Chemistry (MC)
Cartridge Chemistry (CC)
Hydropneumatic System
System Console
Water Supply
Waste Disposal
Sample Handling Procedures
Loading Samples:
Samples are loaded using sample racks or an autoloader tray.
Control buttons include:
RUN
Priority Load
STOP
The sample handling system includes:
Pushers
Gates
Shuttles
Barcode Reader
Analysis of Samples:
Samples are analyzed using:
Sample carousel
Sample probes for Modular Chemistry (MC) and Cartridge Chemistry (CC)
Storage of Completed Samples:
Completed samples are stored in an off-load tray.
Modular Chemistries
Reagent Storage:
Various reagent components stored for analysis including:
Electrolyte Injection Cup (EIC)
Flow cell assembly
Alkaline buffer reagent for CO2 measurement.
Reaction Mechanisms:
Modular chemistries utilize discrete reaction cup modules for the following components:
Urea
Phosphorous
Glucose
Creatinine
Total protein
Albumin
Hydropneumatics System
Provides:
Vacuum
Compressed air
Diluted wash solution
Deionized water
Caution: static-sensitive area.
Computer and Console
System Console:
Includes:
Monitor (touchscreen)
Keyboard
Mouse
Printer
Monitoring System:
Monitored Areas Include:
Temperatures
Power status
Hydro systems
Diagnostics on components
Analytical Principles of DXC
Ion-Sensitive Electrodes (ISE):
Measured ions include:
Na
K
Cl
CO2
Ca
The concentration of the ion is calculated from the difference in potential between a reference electrode and the measuring electrode.
Flow Cell Composition:
The flow cell is constructed from an acrylic block that houses the electrodes used for ISE measurements.
Specific ion-sensitive membranes include:
Sodium: aluminum silicate glass
Potassium: valinomycin
Chloride: Ag/AgCl
Calcium: ionophore membrane
CO2: measures pH rate changes.
ISE Reference:
Ensures a consistent measuring environment by using known concentrations of analytes to keep electrodes stable.
A reference reading is taken after each sample and subtracted from the buffer/sample dilution reading to assess variations in potential.
Buffer Solutions
ISE Buffer:
Used for patient sample dilution to prevent significant pH fluctuations.
It contains known concentrations of analytes to maintain consistency.
Advantages of Indirect Measurement Include:
Less sample volume required
Quicker analysis times
Reduced interference from sample components
Analytical Techniques for Specific Analytes
Urea: Urease reagent and conductivity increase measured with a gold conductivity electrode.
Creatinine: Measured using a colorimetric method (Jaffe reaction with alkaline picrate).
Glucose: Measured through the rate of oxygen consumption in a glucose oxidase reaction using a polarographic electrode.
Phosphorous: Colorimetric analysis with ammonium molybdate reagent.
Total Protein: Colorimetric method utilizing the Biuret reaction.
Albumin: Colorimetric measurement by dye binding with BCP.
Instrument Maintenance
Essential Routine Maintenance Schedule:
Daily, Monthly, Bi-monthly, 6-month, and Yearly checks are required to maintain efficient operation.
It's noted that 90% of problems happen after maintenance; however, caution is encouraged.
Daily Maintenance Tasks:
Check for any error codes on the screen.
Verify reagent levels and load appropriately.
Prime reagents and inspect for leaks.
Check calibration status and recalibrate if necessary.
Run quality control (QC) measures and clear the sample program after QC completion.
Monthly Maintenance Tasks:
Replace the cuvette wiper.
Clean the exterior of probes.
Clean the flow cell and modular chemistry cups.
Perform cleaning on cartridge chemistry probes.
Check calibration spans for chloride measurements.
Calibration Procedures
Calibration Timing Guidelines:
Required when:
Recommended by the manufacturer or upon calibration expiration
Quality control (QC) results indicate out-of-control status based on Westgard rules
New bottle/lots of reagent are loaded
After maintenance or repairs
New lot calibrator diskette is used
After rebooting.
Use of Calibrators:
AquaCal Calibrators Levels 1, 2, 3:
Used for analytes including Na, K, Cl, CO2, glucose, urea, creatinine, calcium, phosphate
Long-term storage occurs in a refrigerator; in-use bottles are kept at room temperature.
Calibration is daily for Na, K, Cl, CO2, calcium, and every 48 hours for glucose, urea, creatinine, and phosphate.
Protein calibrators are required monthly for total protein and albumin (TPm and ALBm).
Calibration Methodology:
Generally, 2 or 3 levels of calibrators are utilized for multiple chemistries, which are run in quadruplicate, discarding the highest and lowest readings, and using the middle values to establish a new calibration curve.
ADC (analog to digital conversion) process is employed for electrode potentials and absorbance measurements.
Accuracy, precision, and sensitivity checks are performed prior to finalizing the calibration curve.
Calibration Failure Analysis
Identifying Failure Parameters:
Possible failures in calibration might include:
Range (accuracy) issues caused by calibrator switching or malfunctioning instruments
Back-to-back (precision) errors which might arise from bubbles in the reagent lines
Span (sensitivity) problems from incorrect, contaminated, or improperly stored calibration reagents.
Systems Approach to Calibration Failure:
Consideration should include:
Sample-related factors (e.g., volume, hemolysis)
Reagent integrity and preparation protocols
Instrument calibration and maintenance history.
Quality Control Procedures
Running QC:
At least one set of quality controls with a minimum of two levels should be run per shift.
QC must be run following any calibration and upon suspected analytical issues.
Sample results must not be processed until QC has been verified as acceptable.
Troubleshooting QC Issues:
Common factors leading to QC failure include improper temperature storage of control vials.
Safety Precautions
Always ensure that power connections are grounded before operating any analyzer components.
Avoid contact near moving parts during operation.
Adhere to laboratory safety protocols regarding infectious or pathogenic material handling.
References
Beckman Coulter Instructions For Use: UniCel® DXC Synchron Clinical Systems. Volume 1. 2008
Beckman Coulter In-Lab Training Manual: UniCel® DXC 800. 2005
Clinical Chemistry: Principles, Techniques, and Correlations. 8th Edition by Bishop, M.L.; Fody, E.P.; Schoeff, L.E., Wolters Kluwer, 2018
© Nova Scotia Community College, MLTD 1207 Clinical Chemistry I, 2024. All Rights Reserved. Any unauthorized reproduction, storage, transmittal, or distribution, in whole or part, is strictly prohibited.