Electron Microscopy - Physics and Theories
Introduction to Electron Microscopy
Electron Microscopes
Hitachi SEM (High-End):
Equipped with multiple detectors for various applications.
Add-ons include EDS for elemental analysis, cryo transfers, and sero face imaging.
Versatile for diverse applications.
TEM (200 Kilowatts):
Features a lab 6 filament.
Capable of TEM and STEM.
Example Images:
Extracellular vesicles.
Biological structures.
Grains of metal alloy.
EDS elemental analysis results (colorful images).
What is Electron Microscopy?
Electron microscopes use a beam of electrons to inspect samples.
Enables high magnification images and provides information about the structure and chemistry of the sample.
Example: Close-up view of a bee's side using a benchtop SEM.
History of Electron Microscopy
Late 1800s: Discovery of the electron as a particle with known mass and charge.
1924: Hypothesis of wave-particle duality.
1927: JP Thompson's discovery that electrons behave as waves.
The concept of electron microscopy was born from the limitation of optical light microscopes due to the wavelengths of visible light.
The first electron microscope consisted of an electron source, lenses, a sample chamber, and a screen or camera.
Resolution in Microscopy
Microscopy involves imaging samples smaller than what can be resolved with the naked eye.
Common units used: micrometre, nanometer, and picometer.
Non-SI unit: Angstrom (1/10 of a nanometer), often used due to typical atomic bond lengths being 2-3 Angstroms.
Electron microscopy allows the study and imaging of viruses, individual molecules, and atomic resolution samples.
Resolution Limits
Maximum resolution for green light in fluorescence microscopy is about 227 nanometers (approximately half the wavelength of light).
The resolution limit at half the wavelength applies to all forms of imaging with waves.
Electrons as Waves
Electrons, like light, exhibit both particle and wave behavior.
The wavelength of an electron is related to its speed, which is related to its energy (accelerating voltage).
Equation: , where:
is the wavelength of the electron.
is Planck's constant.
is the rest mass of an electron.
is the charge of an electron.
is the accelerating voltage.
Constants:
Example:
At 20 kV, the calculated wavelength is 0.0086 nanometers, suggesting resolutions 10,000 times higher than with green light are possible.
Relativistic effects occur at these accelerating voltages, further reducing the real wavelength.
Image Formation
Image formation in electron microscopes relies on the interactions between electrons and the sample, broadly described as scattering.
Electrons strongly interact with matter due to their charge.
Interactions can be elastic or inelastic.
Elastic Scattering: Change in electron direction only, with no energy transfer.
Inelastic Scattering: Change in electron direction and transfer of energy to the electron or sample.
Different interactions provide useful information about the sample and electrons.
Modes of Scattering
Incoming high voltage electron beam interacts with the sample, resulting in:
Backscatter electrons
Secondary electrons
Unscattered electrons (if the sample is thin enough)
Forward scattered electrons
Elastic Scattering
Interaction that changes the electron's direction without altering its energy.
An electron is repelled by the electron cloud and attracted to the nucleus of an atom.
In crystalline materials, regularly spaced nuclei cause constructive and deconstructive interference.
Diffraction patterns result from electron interference when electron beams pass through a crystalline material (e.g., double-slit experiment).
Inelastic Scattering
Involves energy transfer between the beam and the sample.
Can damage the sample through localized heating (leading to diffusion or structural changes) and ionization (breaking covalent bonds in polymers and biological materials).
Provides valuable information about the sample's chemistry.
Elemental Analysis
Incoming electrons interact with the sample, transferring energy to electrons in the sample.
Excited electrons return to lower energy levels, releasing characteristic X-rays used for elemental analysis.
Another signal introduced by inelastic scattering is called Bras strong X-rays (breaking radiation).
Generated by primary electrons in interactions that don't eject core electrons.
Can have any energy up to the original energy of the electron and don't carry specific information.
Secondary Electrons
Electrons emitted from the sample due to direct interaction with the electron beam.
Low energy and can only escape from the surface of the sample.
Provide high-resolution images of the sample surface; very surface-sensitive.
Auger Electrons
Generated similarly to characteristic X-rays, but instead of photon emission, energy is transferred to another electron in a nearby band.
This electron is then emitted from the sample.
Collective Interactions
Electron-hole pairs and cathode luminescence (CL):
In direct gap semiconductors, electron-hole pairs are generated.
Recombination of electrons and holes gives off light (cathode luminescence).
Plasmons and phonons:
Plasmons: Collective oscillations of free electrons when a beam electron interacts with the free electron gas.
Phonons: Collective oscillations of atoms in a solid when the atom lattice is struck by a beam electron.
Beam Damage
Can cause alterations in structure, ionization, heating, phase changes, or diffusion.
Correct interpretation of data becomes more difficult after these events.
Polymers and biological samples are particularly prone to damage.
Electron Microscope Design
Electron microscopes are generally cylindrical.
Comparison of optical microscopes, TEM, and SEM:
Optical Microscope: Light bulb, condenser lenses, objective lens, projector lenses, eyepieces/camera.
TEM: Electron source, condenser lenses, objective lens, intermediate and projective lenses, camera/phosphor screen.
SEM: Similar to the top half of a TEM, with the objective lens focusing the beam into a small spot and raster scanning across the sample to generate pixel data based on signal intensity.
Typically requires more lenses (2-4 condenser lenses, 3-4 intermediate/project lenses) along with smaller alignment lenses.
Electronic lenses cannot be diverging.
Electromagnetic Lenses
Electrons cannot be focused with normal optics due to strong scattering.
Electromagnetic fields are used to focus the beam by taking several wraps of copper like a typical electromagnet, then they are wrapped in the soft iron case.
Electrons rotate around the central axis as they pass through the lens, causing image rotation.
Electron Gun
Materials contain electrons; the energy required to remove them is the work function.
Methods to extract electrons:
Heating (Thermionic Emission):
Tungsten: High melting point (3660K) allows it to be heated to extract electrons.
Lanthanum Hexaboride (LaB6): Lower work function, requiring heating to about 1700 degrees to extract electrons.
Field Emission:
The strength of an electric field on a spherical surface is proportional to the applied voltage and the radius of the sphere.
If the radius is less than 100 nanometers, the electric field gets so strong that electrons can tunnel out of the material
Requires a very small tip (radius < 100 nm) to create a high electric field, allowing electrons to tunnel out of the material.
Field emission guns are more expensive and require high vacuum for maintenance.
Electron Beam Formation
Components of the electron gun:
Filament (electron source)
Wehnelt: Negatively charged, focusing electrons inward.
Anode: Positively charged, attracting electrons downward with a small hole for beam formation.
Vacuum Requirements
Electron microscopes must operate under high vacuum to prevent electrons from scattering off gas molecules.
Higher resolution microscopes require higher performance vacuum systems.
Vacuum pumps: rough and a high vacuum pump. Also, Field emission guns and other high performance instruments usually add ion pumps or iron-getter pumps to achieve the highest possible vacuum.
Comparison with Other Radiation Sources
X-rays and neutrons can achieve similarly small wavelengths, dependent on energy.
Demonstrated resolution limits:
Electrons outperform X-rays and neutrons in terms of resolution due to ease of focusing.
Neutrons are difficult to focus due to their lack of charge, requiring crude focusing methods.
Advantages of Electron Microscopy
High resolution to observe nanoscale objects and individual atoms.
Strong scattering allows collection of useful information from small sample areas.
Summary of Part 1
Electrons exhibit wave-particle duality and have wavelengths 10,000 times smaller than visible light.
High-resolution imaging is possible due to the low diffraction limit.
Strong interaction with matter necessitates high vacuum to prevent scattering.
Inelastic scattering causes significant sample damage due to energy transfer.
Electron microscopy provides high spatial resolution and useful information from small areas and volumes.