8. QUALITY CONTROL
INTRODUCTION TO QUALITY CONTROL
Definition: The process of monitoring the characteristics of an analytical system to detect error and prevent the release of inaccurate results.
The Goal:
- Ensure Precision (reproducibility) and Accuracy (closeness to the true value).QC Materials:
- Utilize control samples with known concentrations (Low/Normal/High) that mimic patient samples.
STATISTICAL FOUNDATIONS
Mean (x): The average of the data points.
Standard Deviation (SD): Measures the “spread” or imprecision of data.
Coefficient of Variation (CV): Expresses SD as a percentage of the mean.
- Formula:
- A lower CV indicates better precision.
QUALITY CONTROL STATISTICS
Reference range for a particular measurement is generally related to a normal bell-shaped curve.
MEAN, MEDIAN, AND MODE
Mean (Average):
- Commonly used in laboratory measurements.
- Calculation: Summing values and dividing by the number of data points.
- Essential for estimating precision and control levels.Data Collection:
- Obtain a minimum of 20 data points from separate testing runs.
- If 20 runs are not feasible, use at least seven runs (three replicates per run).Provisional Ranges: Set using mean and standard deviation (SD).
Updating Mean and Limits:
- Replace mean and limits derived from abbreviated data when 20 separate runs’ data become available.
- Regular reassessment is crucial for accuracy and reliability.
MEDIAN
Definition: Represents the middle value in a dataset.
Calculation: Arrange data in order of magnitude and locate the value halfway between the highest and lowest variables.
Example: For the list 2, 2, 3, 4, 5, 6, 7, the median is 4.
MODE
Definition: The most frequently occurring value in a dataset.
Example: In the list 2, 2, 3, 4, 5, 6, 7, the mode is 2.
STANDARD DEVIATION
Definition: Measures the spread or variability in a dataset.
Calculation:
- Calculated as the square root of the variance of values.QC Tasks:
- Establish QC targets and ranges (SD) for assay controls.
- Determine mean and SD for new QC lots.
- Revised standards allow calculating target mean using 10 control specimens over 10 days.Statistical Interpretation:
- In a normal population:
- 68% of values fall within ±1 SD from the average.
- 95% fall within ±2 SDs.
- 99.7% fall within ±3 SDs.Reference Values:
- Reference interval includes 95% of test results for a healthy population.
- Replaces terms like “normal values” or “normal range.”
CONFIDENCE INTERVALS
Definition: Confidence Intervals (CIs) represent 2 SDs around the mean and encompass 95% of values.
Considerations: Acknowledge day-to-day shifts in analytical procedures.
Purpose of 95% CI: Accounts for sampling variability and method imprecision.
Laboratory Reference Ranges:
- Manufacturer’s ranges provide initial guidance.
- Individual labs adjust mean and QC limits based on patient population.Control Material Analysis:
- New control lots should be analyzed alongside existing material to ensure accurate mean and limits for quality control.
COEFFICIENT OF VARIATION
Definition: CV, expressed as a percentage, is the ratio of SD to the mean; it quantifies variability relative to the mean in a data set.
Purpose: Normalizes variability, allowing comparison of SDs across different means.
- Useful for assessing precision differences in assays and methods.Comparison:
- When comparing SDs, consider the mean to avoid misleading impressions.Control Limits:
- Setting control limits at ±2 SDs is common; however, it may lead to high false rejection rates.
EXAMPLE OF COEFFICIENT OF VARIATION
Calculation:
- ,
-
- % CV:
ext{% CV} = rac{SD imes 100}{Mean} = rac{0.36 imes 100}{3.14} = 11.5 ext{%}
DETERMINATION OF CONTROL RANGE
Control Solution Assessment:
- Laboratories must determine an acceptable control range for a specific analysis after purchasing an unassayed control solution.Method for Establishing Range:
- Assay an aliquot of the control serum alongside regular batches of assays over 15 to 25 days, treating the control sample like an unknown specimen.Calculating Acceptable Limits:
- Repeated determinations yield a normal bell-shaped curve. Calculate mean (x) and standard deviation (SD).
- Most labs use ±2 SDs from the mean as the allowable range, while others use it as a warning limit.Control Implementation:
- Include control specimens in each batch once the acceptable range is established.
- If a control value falls outside the limits, repeat the procedure before reporting patient results.
- Accreditation bodies mandate written procedures for monitoring and resolving out-of-control situations.
SOURCES OF VARIANCE OR ERROR
Achieving identical results for specific specimens is generally impossible due to inherent variability arising from:
- Sampling Factors:
- Collection time, patient position, physical activity, fasting duration, and storage conditions may all contribute to variance.
- Procedural Factors:
- Aging of chemicals, personal bias, variations in standards, reagents, environment, methods or apparatus can also influence variance.
- Experimental errors may arise from method changes, instrument variations, or personnel shifts.
LEVEY-JENNINGS CHARTS
Definition: The primary visual tool for monitoring QC over time.
Components:
- X-axis: Time/Date
- Y-axis: Concentration, marked with the Mean, ±1SD, ±2SD, and ±3SD.
- Value: Enables technicians to see shifts and trends immediately before they lead to out-of-control situations.
DAILY CONTROL SPECIMEN VALUES
Laboratories are required to plot these values on a Quality Control (QC) chart.
Many modern instruments automatically generate QC charts daily, flagging out-of-control results.
Purpose of Control Charting:
- Ensures reliable and stable laboratory processes and identifies unacceptable runs and deviations effectively.
DRIFT DETECTION
If a drift is detected, checks should include reagent age, calibration status, and potential adjustments to the mean.
GLUCOSE LEVEL EXAMPLE
Levey-Jennings Graph for Glucose Level: Marks concentration against time, demonstrating the process for visualizing QC data.
CONTROL SAMPLES
QC involves using at least two different control samples for a specific analyte, including both normal and abnormal control specimens.
Control Limits:
- The mean value and acceptable error limits are typically set at ±2 SDs or ±3 SDs from the mean.
- 2-SD serves as a warning limit; 3-SD acts as an action limit.Daily Monitoring: Allow for easy identification of values that fall “out of control,” helping detect trends or drift over time.
VALIDATION OF NEW PROCEDURES
Laboratories validate new procedures before routine use, determining reproducibility and confidence limits while establishing acceptable variation limits for control specimens.
Quality Control (QC) Program:
- Calculates mean and SD for each procedure and generates control charts for performance monitoring.
- Regular assessments help detect issues promptly and correct them timely.
SHIFTS, TRENDS, AND DISPERSION IN LEVEY-JENNINGS CHARTS
Shifts:
- Definition: A sudden and sustained change in one direction in control sample values.
- Indication: May be caused by sudden instrument malfunction.Trends or Systematic Drift:
- Defined as gradual change in control sample results over time.
- Indication: If direction consistently shifts from the mean for at least 3 days, signaling potential issues like reagent deterioration.Dispersion:
- Refers to increased random error or lack of precision, which can undermine data reliability.
WESTGARD RULES (MULTI-RULES)
Multi-rules are employed before reporting patient data and designed to detect random and systematic errors, maximizing detection ability while minimizing false rejection rates.
WESTGARD RULES (“THE DECISION TREE”)
1₂S: One control exceeds 2SD, often used as a warning.
1₃S: One control exceeds 3SD, indicating random error; the run must be rejected.
2₂S: Two consecutive controls exceed the same 2SD limit, pointing to systematic error.
R₄S: The range between two controls in the same run exceeds 4SD (e.g., one QC is +2SD, the other is -2SD), indicating random error.
4₁S: Four consecutive results exceed the same 1SD limit, which signifies systematic error.
10x: Ten consecutive results fall on the same side of the mean, indicating a shift or systematic error.
TABLE OF WESTGARD QUALITY CONTROL RULES AND INTERPRETATION
22s Rule:
- One control observation exceeding the mean ±2s may be used as a warning rule to initiate additional testing.13s Rule:
- One control observation exceeding the mean ±3s recommends rejecting patient results, sensitive to random error.2s Rule:
- Two consecutive control observations exceeding the same mean plus or minus 2s recommends rejecting patient results, sensitive to systematic error.R4s Rule:
- One observation exceeding the mean plus 2s and another exceeding the mean minus 2s recommends rejecting patient results, sensitive to random error.4s Rule:
- Four consecutive observations exceeding mean plus or minus 1s recommend rejecting patient results, sensitive to systematic error.10x Rule:
- Ten consecutive control observations falling on one side of the mean recommend rejecting patient results, sensitive to systematic error.
VIOLATIONS OF QUALITY CONTROL RULES
Figures visualize various violations of established control rules.
Guidelines regarding violations help maintain QC integrity and reliability in lab practices.
MULTIRULES FOR THREE CONTROLS
Purpose: Monitor lower and upper analytical range critical for clinical values.
- Level 1: Low (Abnormal) - monitors sub-normal concentrations.
- Level 2: Normal - validates routine sample accuracy.
- Level 3: High (Abnormal) - tests linearity and protects against toxic levels.
RESOLUTION FOR OUT-OF-RANGE CONTROLS
Laboratory policies dictate procedures for managing out-of-range controls, detailing steps to identify and correct issues before reporting patient results.
RATIONALE BEHIND THE RULES
Multirules integrate individual rules with lower false rejection rates, raising error detection rates.
12s Rule Discussion: 95% of results fall within 2 SDs; 5% exceed it, raising concern regarding true errors.
13s Rule Discussion: 99.7% of results fall within 3 SDs; a violation indicates a probable error, balancing false rejection and detection effectively.
Multirules are recommended for increased complex error detection capability.
EXTERNAL QUALITY ASSESSMENT (eQA)
Definition: A process where an external agency sends “blind” samples to the laboratory.
The Process: Labs treat eQA samples like patient samples and submit results for comparison against peer group means.
Purpose: Ensures long-term accuracy and inter-laboratory comparability.
INTERNAL QC vs EXTERNAL QA
Features:
- Internal QC: Conducted daily; aims for immediate actions regarding report decisions.
- External QA: Done periodically; focuses on long-term accuracy and bias detection through peer comparisons.
ANALYZING eQA RESULTS
SDI (Standard Deviation Index): A common metric for evaluating eQA performance.
- Formula:
- Interpretation:
- 0.0: Perfect agreement
- ±1.0 to 1.5: Acceptable performance
- > ±2.0: Marginal; requires investigation
- > ±3.0: Unacceptable; requires immediate corrective action
eQA PROVIDERS
IQMH: Common in Ontario for proficiency testing and accreditation services.
QMP-LS: Associated with provincial quality mandates.
One World Accuracy: Provides oversight for tests not available from IQMH.
CAP: An international “gold standard” for specialized testing.
CORRECTIVE ACTION FOR eQA FAILURES
A systematic investigation is necessary when a “red flag” (unacceptable result) is reported.
Steps involve:
- Clerical Check: Confirm accuracy in transcription and units.
- Specimen Integrity: Ensure the proper handling of eQA samples.
- Instrument/Reagent Review: Assess documented shifts or QC failures on testing day.
- Peer Group Analysis: Verify if others in the peer group experienced similar results.
- Technical Competency Check: Confirm adherence to SOPs by lab staff.
DOCUMENTATION AND RESOLUTION
The Paper Trail:
- Every eQA failure requires a Corrective Action Report (CAR).
- Root cause analysis must identify the reason for variance and outline preventative actions.
- Final sign-off is done by the lab manager or medical director prior to review.
CASE STUDIES
Case Study #1: The Chemistry Shift
Scenario: Glucose control level 2 above mean for 9 days, today at +2SD.
Question: Which Westgard rule is violated and what is the likely cause?
Answer: Possibly a 10x warning, indicating systematic error, likely due to reagent lot change.
Case Study #2: The Random “Hiccup”
Scenario: WBC Level 1 at -2SD, Level 2 at +3SD.
Question: What is the rule violation and immediate action?
Answer: This is a 1₃S violation indicating random error. QC should stop, check for bubbles or clots, and repeat the QC.
Case Study #3: The eQA Surprise
Scenario: Lab result of Sodium at 138 mmol/L, peer group mean 142 mmol/L, peer group SD 1.0.
Question: Calculate the SDI.
Answer: indicating unacceptable performance, requiring a thorough review.
SUMMARY
Levey-Jennings Chart: Visualizes control data, aids in detecting shifts and trends.
Westgard Rules: Key for identifying anomalies in control values.
Problem-Solving Approach: Utilize tools to troubleshoot and maintain analytical accuracy.
Patient Care: Quality control is ultimately about ensuring patient care and safety.