5. Automation
Automation Overview
Clinical laboratories generate about 80% of the information physicians rely on to make crucial treatment decisions.
Patients, physicians, and healthcare workers expect test results that are accurate and done quickly.
Expectations of turnaround times (TAT) have been key to the development of lab equipment as well as Point of Care Testing (POCT) devices.
Major advances in clinical lab:
Miniaturization
Linking patient results across provinces (Electronic Medical Record shared results)
Automated instruments of all sizes (both in robotics and IT)
Benefits of Automation
Reduction in Medical Errors: Critical errors still occur in mislabeling or misidentification.
Reduced Specimen Sample Volume: 200 µL or less.
Increase in Accuracy and Precision: Automation enhances test precision.
Improved Safety for Lab Staff: Reduces risks from sample handling, including eliminating the need for cap removal.
Faster Turnaround Time: Automation allows for quicker processing of test results.
Alleviation of Shortages: Automation helps counteract the impending shortage of skilled laboratory staff.
Automation Processes
Automation processes include special design, compatibility, and integration of automated clinical systems.
Standardized analyzers include devices for sampling patient specimens or other samples to be tested (such as blanks, controls, and standard solutions).
Analyzers possess features to add specimens to reagents in the correct sequence and incubation modules for specific reactions.
Measurement devices: Often utilize photometric technology to quantify the extent of reactions.
Recording mechanism: Provides a final reading or permanent record of results and interfaces with a computer.
Most analyzers can process a variety of specimens, increasing efficiency and reducing human error related to repetition and manipulations (e.g., pipetting errors).
Automated Analyzers
Choice of instrument depends on:
Volume of determination in the lab
Type of data profile generated
Level of staffing
Initial cost and maintenance
Operational costs and time required for each analysis.
Designed to perform frequently ordered tests. In large volume hospitals and reference centers, completely automated lab systems may be employed.
Each automated instrument can function separately or be integrated with other lab instruments.
Instruments can link in continuous operations, including robotic specimen processing.
Highly automated systems are used predominantly in larger clinical chemistry and hematology labs, while semi-automation is present in urinalysis, blood banks, and microbiology laboratories.
Three Phases in Testing
Pre-analytical Phase:
Involves specimen labeling, centrifuging, and sorting.
Errors often occur, such as labeling the wrong tube.
Analytical Phase:
Entails chemical reactions and measurement resulting in data generation.
Post-analytical Phase:
Data storage, reflex testing, and electronic reporting occur here.
Most laboratory errors (up to 70%) occur in the pre-analytical phase.
Major Types of Analyzers
Continuous Flow:
Liquids are pumped through a system of tubing, separating samples with air bubbles.
Centrifugal Analysis:
Uses centrifugal force for mixing reagents and samples, allowing multiple samples to run simultaneously.
Discrete Analysis:
Each sample is contained in its own reaction container (cuvette), and this is the current industry standard.
Components of a Discrete Analyzer
Sample Loader:
Utilizes circular carousels or continuous racks.
Sampling Probe:
Aspires precise microlitre volumes and includes liquid level sensing.
Reagents Station:
Refrigerated compartments for open or closed systems.
Mixing Unit:
Incorporates stirring paddles or ultrasonic vibrations.
Incubation Chamber:
Maintains a constant temperature of 37°C.
Measurement Technologies
The “brain” of the analyzer:
Spectrophotometry: Measures light absorbance to determine concentration.
Ion Selective Electrodes (ISE): Specifically for electrolytes (Na, K, Cl).
Chemiluminescence: Utilizes light-emitting chemical reactions for immunoassay testing.
Turbidimetry/Nephelometry: Measures light scatter for protein analysis.
Total Laboratory Automation
The Track System: Conveyor belts that transport tubes between modules.
Decappers: Automated removal of tube stoppers.
Onboard Storage: Automated retrieval of samples for repeat or reflex testing.
Quality Control and Calibrations
Westgard Rules: Criteria used to determine the acceptance or rejection of analytical runs.
Quality Control (QC): Involves running stable materials at different levels (normal/abnormal) to ensure systems are in statistical control regarding accuracy and precision.
Calibration: The process of setting a ruler for the instrument using known standards.
Troubleshooting and Maintenance
Daily Maintenance: Includes cleaning probes, checking temperatures, and replenishing reagents (e.g., replacing ERF daily).
Common Issues:
Fibrin clots in probes.
Reagent depletion.
Lamp failure.
Instructor’s Rule:
If QC is out, the sequence of checks should involve checking the reagent, then the calibrator, then the instrument, and lastly the patient’s sample.
Steps in Automated Analysis
Major steps designed by manufacturers to replicate manual techniques:
Specimen collection and processing
Measurement and delivery of specimen and reagents
Chemical reaction phase
Measurement phase
Signal processing and data handling
Specimen Collections and Processing
Proper specimen collection, labeling, and transport to the lab are critical for accurate analysis.
Automation of specimen preparation includes:
The use of barcoded labels to eliminate clerical errors related to patient data entry.
Automation allows for efficient storage and retrieval processes.
Ultimately, proper identification is the responsibility of the staff member collecting the sample.
Specimen and Reagent Measurement and Delivery
Automated instruments combine reagents with measured specimen amounts:
Reagents must be introduced correctly for accurate analysis.
Random-access analyzers test specimens sequentially based on selected tests.
A microprocessor controls the addition of diluents and reagents, initiating spectrophotometric measurements of varied samples.
Some analyzers employ a circular or parallel configuration for testing.
Chemical Reaction Phase
Reagents can be classified as liquid or dry chemistry, with varying handling based on instrument capabilities.
The chemical reaction phase entails mixing, separation, incubation, and reaction time:
Continuous flow analyzers utilize air bubbles between sample plugs.
Discrete analyzers maintain reactants in individual containers, functioning as cuvettes for optical analysis.
Incubation allows time for reaction at controlled temperatures set by the analyzer.
Measurement Phase
Traditionally relies on photometry and spectrophotometry to measure absorbance.
Alternate methods include nephelometry, chemiluminescence, enzyme immunoassays, and ion-selective electrodes (ISE).
Regular standardization and calibration are essential for accurate results, with QC specimens needing to be analyzed frequently.
Signal Processing and Data Handling
Results are visualized on a readout using LEDs or monitors, convertible to hard copy or electronic formats.
Data management systems interface with analyzers and host Laboratory Information Systems (LIS).
Automated QC data management with evaluation against acceptable limits.
The LIS can auto-verify results within predefined parameters, sending them directly to patient files, while technologists review all data that falls outside these limits.
Automation in Clinical Chemistry
Automated analyzers replicate standard manual reactions.
Unique methods, such as chemiluminescence, are specifically developed for immunoassays.
Analyzer Methods - Spectrophotometry & Photometry
Photometric Instruments: Measure light intensity; often use filters or prisms to isolate wavelength ranges.
Beer’s Law: States that the concentration of a substance is directly proportional to the extent of light absorbed or inversely proportional to the logarithm of transmitted light.
Spectrophotometric Instruments: Evaluate light transmitted by a solution to ascertain concentrations of light-absorbing substances.
Radiant Energy
Radiant energy passing through an object is partially reflected, absorbed, or transmitted.
Absorbance (A): Defined as the amount of light absorbed, derived from %T using the formula: Extra open brace or missing close braceExtra open brace or missing close brace.
Absorbance depends on molecular and ionic types present and can vary with concentration, pH, or temperature.
Deviations from linearity can occur at high absorbances (around 2.0 units).
Components of the Spectrophotometer
Key components include:
Light Source: Tungsten or deuterium/mercury arc lamp (UV).
Monochromators: Wavelength isolators such as filters and prisms.
Sample Cell: Typically plastic or quartz, sensitive to scratches and contaminants.
Photodetectors: Convert radiant energy to electrical energy (e.g., phototube, photocell, photomultiplier tube).
Wavelength Selectors
Various types:
Filters
Prisms
Grating monochromators
Holographic gratings
Quality described by nominal wavelength, effective bandwidths, and bandpass.
Wavelengths
Nominal Wavelength: Peak transmittance in nanometers.
Spectral Bandwidth: Range of wavelengths halfway between the baseline and peak.
Bandpass: Total range of wavelengths transmitted.
Photomultiplier Tubes
Used in low radiant power situations (low analyte concentrations).
Similar to phototubes but output is amplified significantly (up to one million-fold).
Highly sensitive to UV and visible radiation with fast response times.
Signal Processors and Readout
Signal processing involves several steps:
Amplification of the signal.
Current rectification and direction.
Phase alteration of the signal.
Filtering of unwanted components and mathematical computations.
Quality Assurance in Spectrophotometry
Wavelength Accuracy: Measured at specific wavelengths, e.g., protein readings at 540 nm.
Assessed with special glass optical filters (e.g., didymium, holmium oxide).
Linearity: Checked with colored solutions that may be diluted; deviations can arise from issues pertaining to light source and monochromator performance.
Absorbance
Ideal: Linear relationship according to Beer’s Law.
Deviation: Caused by electronic limitations or instrument malfunctioning.
Reflectometry
Principle: A filter photometer measures the quantity of light reflected from a liquid sample on a non-polished surface.
Reflectance correlates nonlinearly with concentration:
Ideal reflectance for a pure white standard equals total light reflection.
Pure black material represents zero reflectance, absorbing all light.
Instrumentation involves tungsten-halide lamps, photodiodes, and filters for wavelength isolation, converted by computers into concentration units.
Fluorometry
Basic instrumentation involves filter fluorometers that assess concentrations of solutions with fluorescing molecules.
Design: Source emits short-wavelength excitation light, with mechanical attenuators to adjust intensity.
Advantages: Enhanced specificity and sensitivity.
Disadvantages: Susceptible to environmental changes (e.g., chemicals, UV light can lead to quenching).
Turbidimetry
Measures light transmission reduction due to particle formation within a sample.
Sensitivity is generally higher compared to nephelometry and is applied in microbiology and coagulation studies.
Chemiluminescence
Involves chemical energy producing excited intermediates that return to ground states with photon emission.
No requirement for excitation radiation.
Utilized in various oxidation reactions and measures turbidity or protein concentrations.
Ion-Selective Electrodes (ISE)
Designed for sensitivity towards specific ions.
Types:
pH electrodes for detecting hydrogen ion activity.
Gas-sensing electrodes detect specific gases; distinct in chemical mechanisms.
Enzyme electrodes incorporate immobilized enzymes catalyzing reactions.
Potentiometry
Measures potential voltage between two electrodes in a solution; requires a constant voltage reference.
Relates electrode potential to ionic concentration via the Nernst equation.
Reference Electrode
Common types:
Calomel electrode
Silver/silver chloride electrode.
Provide analytical qualities necessary for reliable measurements in conjunction with ISEs.
Advantages of ISE Over Wet Chemistry
Direct measurement without reagent prep or standard curves.
Cost-effective, swift, sensitive, easily maintained, and adapted to automation.
pH Electrode
Glass electrodes measure hydrogen ion activity, using a chloride buffer and silver/silver chloride reference.
Hydrogen ions replace sodium ions, affecting potential difference and thus pH reading.
pCO2 Electrode (Gas Analyzers)
Comprises a plastic jacket containing sodium bicarbonate buffer with a gas-permeable membrane.
When CO2 from blood contacts the membrane, reacts with the buffer, shifting pH levels indicative of concentration.
Reaction: extCO2+extH2extOightleftharpoonsextH2extCO3ightleftharpoonsextHCO3−+extH+
Amperometry
Measures current flow produced during oxidation-reduction reactions, applicable in various electrode designs.
pO2 Gas Sensing Electrode
The Clark PO2 electrode permits dissolved oxygen passage through a membrane, reacting with a platinum cathode, producing electron flow proportional to oxygen levels.
Osmometry
Measures osmolarity of solutions (serum, plasma, urine) using techniques like freezing-point depression, vapor pressure, and osmotic pressure.
Osmolarity: Concentration expressed per volume of solution (osmol/L), while Osmolality: Concentration per kilogram of water (mOsmol/kg) is superior as it remains temperature-independent.
Freezing Point Osmometer
Utilizes rapid supercooling and agitates samples to establish equilibrium at freezing point, measured by thermistors and converted to milliosmoles/kg.
Conclusion
Automation enhances laboratory processes but does not replace Medical Laboratory Technologists (MLTs).
It shifts the role of MLTs toward data management and problem-solving, underscoring the importance of validation of analyzer results.