Radiology PowerPoint VT120
Radiology Study Notes
VT120 Introduction to the Applied Sciences
Instructor: Pippa Oval
Part I: X-Rays
History of X-Rays
In 1895, Wilhelm Roentgen discovered radiation, originally termed “Roentgenography”.
The hazards associated with radiation were not fully understood until later.
Electromagnetic Radiation
Definition: Energy that travels in a straight path through space in waves.
X-rays are characterized as short wavelength, high energy radiation.
Context: X-rays form a part of the electromagnetic spectrum.
Physical Properties of X-Rays
Travel Path: X-rays travel in straight lines.
Transmission: X-rays can penetrate many materials, resulting in transmitted x-rays.
Impact on Matter:
As x-rays pass through materials, they collide with electrons.
This results in:
Scatter Radiation: Change in the path of the x-ray.
Excitation: Increase in electron energy.
Fluorescence: Certain substances release energy in the form of light.
Ionization: Ejection of an outer electron from an atom.
The X-Ray Machine
Basic Components
X-Ray Tube:
A glass vacuum tube that includes the cathode and anode which produce x-rays.
Metal Housing: Protects the glass tube.
Oil: Serves as a coolant.
Window and Filter: The exit point for x-rays.
Cathode (-)
Function: Acts as the electron source.
Components:
Tungsten filaments, similar to light bulbs but emit electrons instead of light.
Focusing cups direct electrons towards the anode.
Application: Thicker end of the patient should be positioned at the cathode end of the x-ray tube.
Anode (+)
Anode Composition: Made of tungsten metal target.
Reactions on Impact:
When electrons strike, x-rays are produced (only 1% of the energy) and heat (99% of the energy).
Characteristics:
Beveled surface for directing x-rays down through the housing window.
Incoming wave pattern allows x-rays to fan out from a single point (focal spot).
Types of Anodes
Stationary Anode:
Made of tungsten target mounted on a stationary copper base.
Limited to low energy x-rays and prone to overheating. Often used for dental and portable radiography.
Rotating Anode:
Target is a rotating disc with beveled edges.
Distributes heat across a larger area and can produce higher energy x-rays. Often used for small and large animal radiography.
Collimator
Function: Controls the size of the primary x-ray beam, reducing exposure to the patient and personnel.
Benefits: Minimizes scatter and improves image quality.
Types:
Adjustable lead shutters.
Lead cones.
Tube Stand
Description: Supports the x-ray tube and the collimator.
Adjustability: Allows for adjusting tube height and angle, capable of moving left/right along the table.
Cassette Tray and Storage Bin
Function: Holds x-ray film cassette under the table.
Features:
Tray lock aligns tray and x-ray tube.
Moves left/right with tube stand and can be unlocked for independent movement.
Lead-lined storage bin for loaded cassettes.
Control Panel
Features:
On/Off switch.
Kilovolt selectors.
Milliamp (mA) selector.
Time selector (not present on newer machines).
Exposure foot switch for operation.
Radiation Safety
Hazards of Ionizing Radiation
Ionizing radiation can cause damage to rapidly dividing or sensitive cells, including:
Hematopoietic and lymphatic tissues.
Intestinal epithelium.
Gonads.
Thyroid.
Bone.
Skin.
Developing embryos.
Recommendation: Pregnant technicians should not take radiographs due to risks.
Historical Case Studies
Dr. John Hall-Edwards (1897):
Exposure led to malignant cancer on his left hand, resulting in amputation.
Typical “X-ray Hand” (1904-1913):
A technician experienced severe inflammation and cancer progressing rapidly after 6 months of exposure.
Differences between X-rays and Radioactive Radiation
X-rays:
Produced by electric charges between the cathode and anode.
Released only during exposure.
Pass through tissues without residual activity.
Radioactive Radiation:
Produced by gradual breakdown of unstable atoms.
Continuously released.
X-rays do not induce radioactivity in objects.
Maximum Permissible Dose (MPD)
Established by the National Council on Radiation Protection & Measurements (NCRP).
Permissible dose for professionals exceeds that of the general public, reflecting regular training and monitoring.
Units of measurement: Sieverts or rems, adjusted for biological effects based on different radiation strengths.
ALARA Principle: Stands for "As Low As Reasonably Achievable"—strives to maintain exposure rates well below limits.
Exposure Allowances
Weekly MPD:
0.001 Sv/week (0.1 rems/week).
Quarterly MPD:
0.0125 Sv/quarter (1.25 rems/quarter).
Yearly MPD:
0.05 Sv/year (5.0 rems/year).
Lifetime Exposure Calculation:
Formula: 5(N - 18), where N = current age.
Example: For a 50-year-old: 5(50 - 18) = 5(32) = 160 rems maximum.
Conversion: 100 rems = 1 Sievert.
Dosimetry
Dosimeter: A device for measuring radiation exposure.
Film Badge: Most common dosimeter worn at all times during radiography (California law).
Reporting: Reports on exposure issued quarterly, including quarterly, annual, and lifetime exposure data.
Radiation Exposure and Safety
Primary Beam:
Intended to penetrate the patient and produce an image.
Safety measure: Filtration—aluminum filters eliminate weak x-rays to prevent tissue damage.
Collimation: Required by California law, restricts beam size to necessary dimensions for the radiograph.
Secondary Radiation:
Also known as scatter radiation, poses exposure risks to patients and technicians.
Safety protocol:
Limit essential personnel in the area during radiography.
Use chemical or mechanical restraints for patients.
Maintain distance from the patient when applying manual restraints and always wear protective apparel.
Protective Apparel
Lead Lined Apparel: Should have a minimum of 0.5 mm lead equivalent thickness and include:
Aprons.
Gloves.
Thyroid shields.
Maintenance:
Hang aprons and gloves to dry to prevent cracking.
Conduct periodic inspections for damage.
Exposure Factors and Radiographic Quality
mAs (Milliampere-Seconds)
Definition: Determines the quantity of x-rays produced per second.
Settings for small animal x-ray machines: Typically 100 mA and 300 mA.
Calculation:
Example: 300 mA x 1/60 second = 5 mAs.
kVp (Kilovolt Peak)
Definition: Represents the strength of the x-rays during exposure.
Application:
Thicker tissues require higher kVp settings.
Larger patients also necessitate higher kVp.
kVp vs mAs
kVp: Affects the power/strength/quality of the x-ray beam; impacts density and contrast.
mA: Refers to the quantity of x-rays produced; quantity affects image density.
Time: Longer exposure time results in more emitted x-rays, thereby affecting density.
kVp and mAs relationship:
mAs = mA x time (seconds).
Distance Considerations
Source-Image Distance (SID)
Definition: Distance between any x-ray source and the receptor (film, digital sensor, etc.).
Focal Film Distance (FFD): Distance between the anode focal spot and the x-ray film.
Standard SID in small animal radiology is typically 40 inches and is marked on the tube stand.
Inverse Square Law
The intensity of the x-ray beam is inversely related to the square of the SID:
Doubling the distance reduces intensity to 1/4.
Halving the distance increases intensity 4-fold.
Tripling the distance reduces intensity to 1/9.
Scatter Radiation
Definition: Released in random directions and reduces image clarity while exposing patients and technicians to radiation.
Factors Increasing Scatter:
Large patients.
Poor collimation.
High kVp settings.
Visual Effects:
An image with kVp too high may appear entirely black.
Grids
Function: Positioned between patient and film to absorb scatter radiation, enhancing image clarity.
Composition: Vertical strips of lead or aluminum and spacer material, often mounted under the table and above the cassette tray.
Patient Motion
Considerations:
Veterinary patients may struggle to stay still, highlighting the need for firm restraint and quick motion during exposure.
Motion is typically not detected with exposure times less than 1/30 second (exposure options include 1/120 sec and other fractions).
Distortion in Imaging
Concept: The image of an object changes as its position changes.
Types of Distortion:
Foreshortening: Objects at angles appear shorter.
End-on View: Long objects appear as spots.
Best Practices: Ensure the body part of interest is parallel to the film.
Tissue Densities
Classification:
Gas: Least dense (air in lungs, gases in GI tract).
Fat: Fat cells filled with lipids.
Water: Body fluids (blood, urine, etc.), cytosol, and internal organs.
Bone: Most dense.
Radiographic Color:
Gas: Black.
Fat: Dark Gray.
Water: Light Gray.
Bone: White.
Radiograph Quality
Density: Depth of lightness/darkness of the image; controlled by mAs and kVp.
Contrast: Degree of difference in shades; controlled solely by kVp.
Improvement Techniques:
Adjust mAs (increase or decrease) for brightness issues.
Adjust kVp (increase or decrease) for contrast issues.
X-ray Film and Film Processing
Cassette
Description: A light-proof, rigid case for film, with internal foam backing to flatten film and intensifying screens on each surface.
Intensifying Screens
Function: Fluoresce when struck by x-rays, emitting visible light that exposes the film; improves efficiency.
Types:
Standard screens emit blue light; rare-earth screens emit green light.
Film Anatomy
Layers:
Protective coat.
Emulsion containing silver halide crystals.
Base for support.
Film Characteristics
Film Speed:
Fast Film: Large silver halide crystals, less exposure from x-rays, grainy image.
Medium Film: Good balance of speed and quality, commonly used in veterinary practice.
Slow Film: Small crystals provide detailed images, requiring more radiation exposure.
Film Identification
Requirement: Permanent label including patient name, owner’s name, clinic’s name, and date must be applied during radiography or development.
Film Processing Steps
Developer: Converts exposed silver halide to black metallic silver, revealing the latent image.
Rinse: Removes developer chemicals from film with circulating water.
Fixer: Eliminates unexposed silver halide, hardening the gelatin in the emulsion.
Wash: Circulating water removes fixer chemicals from film.
Dry: Hanging film to air dry.
Film Handling Guidelines
Avoid exposing film to white light, store in a light-tight box, and ensure hands are clean and dry when handling film.
Load and process only one film at a time to prevent cross-exposure or contamination.
Digital Radiography
Definition: Traditional film replaced with digital x-ray image receptors; images are captured on electronic sensors and can be viewed, edited, and stored on a computer.
Advantages:
Image editing capabilities for brightness and contrast adjustments.
Rapid viewing of images with reduced exposure risks for patients and technicians.
Digital storage allows easy link to medical records and file transfer.
Limitations:
Initial equipment costs.
Adjustments do not correct issues like patient positioning or collimation errors.
Technique Chart
Purpose: Provides proper exposure factors based on body part thickness in cm, including kVp, mA, and seconds to reduce retakes.
Use Guidelines: Determines exposure factors vary with film-screen systems, tissue density, desired contrast, and the use of grids.
Technique Chart Development:
Use a mid-sized patient for trials.
Adjust kVp and mAs based on trial outcomes per patient size adjustments.
Technical Artifacts and Errors in Film X-rays
Common Errors:
Black cloudy edges due to film exposure to white light.
Fingerprints from handling with contaminated hands.
Double exposure and underexposure resulting in unclear images.
Overlapping images from incorrect processing leading to chemical trapping.
Conclusion
Careful handling and adherence to protocols is essential to ensure x-ray quality and safety for both patients and technicians.