ASRT CT 2

Module Overview

  • Title: CT Basics 2.0 - Principles, Parameters, and Display.

  • Objective: After completing this module, you will be able to:

    • Explain major components of the computed tomography (CT) computer system.

    • Describe each step in the CT image formation sequence.

    • Differentiate between axial, helical, and volumetric methods of data acquisition.

    • Apply operator console parameters to optimize CT image data acquisition.

    • Describe the elements of digital image display.

Components of a CT Scanner

  • Modern CT scanners: Blend of practical design and advanced technology.

  • Key components:

    • Imaging System: Consists of a gantry and patient table.

    • Computer: Responsible for processing CT image data.

    • Operator's Console: Controls the imaging process and displays the final images.

  • Functionality:

    • Patient lies on the table which moves through the scanner to produce an image.

    • Equipment and instrumentation behind the scenes facilitate scanning and image reconstruction.

Patient Table and Gantry

  • Patient Table (Couch):

    • Constructed from carbon fiber for minimal radiation absorption.

    • Facilitates controlled movements to position anatomy for scanning.

    • Weight limit typically up to 450 pounds (204 kilograms).

  • Gantry:

    • Large circular apparatus housing essential equipment for image acquisition: X-ray tube, detector array, and data acquisition system.

    • Contains a cooling system for managing heat generated during X-ray production.

    • Allows tilting (12-30 degrees) to align anatomy of interest.

Key Parameters in Scanning

  • Pitch:

    • Controls table movement through the gantry and relates to image acquisition.

    • Defined as the ratio of the distance the table moves to the total collimated X-ray beam width.

    • For Single Detector CT (SDCT): Pitch of 1:1 is optimal for image quality; defined as detector pitch.

    • For Multi Detector CT (MDCT): Beam pitch is determined by table movement per rotation divided by beam width.

    • Increasing pitch reduces patient dose but affects image quality negatively.

  • Coordinates:

    • X-axis: Horizontal position (right to left).

    • Y-axis: Vertical position (anterior to posterior).

    • Z-axis: Corresponds to the axial plane (superior to inferior).

    • Patient alignment is crucial for accurate scanning.

Data Acquisition Techniques

  • Axial Scanning:

    • Step-by-step acquisition of data slice by slice with pauses.

    • Misregistration artifacts potentially occur due to patient movement.

  • Helical (Volumetric) Scanning:

    • Continuous movement of the table through the gantry while the X-ray tube rotates.

    • Rapid data collection and ability to cover larger volumes.

    • Reduces misregistration risks by minimizing breath-holding times.

CT Tube and Generator

  • X-ray Tube:

    • Larger than traditional diagnostic tubes for managing heat.

    • Mounts within gantry and rotates around the patient.

    • Cathode and Anode:

    • Cathode: Contains tungsten filaments; controls flow of electrons based on milliampere setting.

    • Anode: Composed of tungsten, molybdenum, rhenium; supports high-speed rotation.

  • Generator:

    • Produces high voltage (80-140 kV) required for X-ray production.

    • Converts incoming power supply (480 V, 60 Hz) to high voltage necessary for operation.

Detectors

  • Types of Detectors:

    • Gas Ionization Detectors: Convert X-rays directly to electric signals.

    • Scintillation Detectors: Convert X-rays to light, then to electric signals (industry standard).

  • Detector Efficiency:

    • Comprised of various factors including capture efficiency, absorption efficiency, stability, response time, dynamic range, and reproducibility.

Collimators

  • Purpose: Controls beam width and limits patient exposure to X-rays.

  • Types:

    • Pre-Patient Collimators: Shape beam before it reaches the patient.

    • Post-Patient Collimators: Shape beam after it passes through the patient, improving image quality.

Slip Ring Technology

  • Functionality: Allows continuous rotation of the gantry components without cables, enabling faster scan times and continuous data acquisition for advanced imaging techniques.

  • Types of Slip Rings:

    • Low voltage (uses AC) and high voltage configurations.

  • Applications: Critical for modern helical CT and computed tomography angiography (CTA).

Image Reconstruction and Display

  • Digital Image Data:

    • Comprised of pixels assigned a shade of gray based on CT numbers from scan data.

  • Matrix and Pixels:

    • Common image matrices: 512x512, 1024x1024, 2048x2048.

    • Pixel size calculated as extPixelSize=racextFieldofViewextMatrixSizeext{Pixel Size} = rac{ ext{Field of View}}{ ext{Matrix Size}}.

  • Hounsfield Units (HU):

    • Determined by the linear attenuation coefficient, with water serving as baseline with a CT number of 0.

    • Dense structures like bone have higher CT numbers (up to 1000), while air has a CT number of -1000.

    • Average values in voxels can cause misdiagnosis, a phenomenon known as Partial Volume Effect.

Windowing and Regions of Interest (ROI)

  • Window Width (WW): Controls contrast; wider WW yields less contrast and wider range of densities.

  • Window Level (WL): Adjusts brightness of image; should be set according to tissue type being imaged.

  • Region of Interest (ROI): Tool for selecting specific areas in the image for quantitative analysis.

    • Measurement aids in identifying tissue types based on density.

Summary of Learning Outcomes

  • The module concludes by ensuring understanding of:

    • Major components of the CT system.

    • Steps in the CT image formation sequence.

    • Differences between data acquisition methods: axial, helical, and volumetric.

    • The use of console parameters for optimizing CT image acquisition.

    • Elements of digital image display and reconstruction.