ASRT CT 2
Module Overview
Title: CT Basics 2.0 - Principles, Parameters, and Display.
Objective: After completing this module, you will be able to:
Explain major components of the computed tomography (CT) computer system.
Describe each step in the CT image formation sequence.
Differentiate between axial, helical, and volumetric methods of data acquisition.
Apply operator console parameters to optimize CT image data acquisition.
Describe the elements of digital image display.
Components of a CT Scanner
Modern CT scanners: Blend of practical design and advanced technology.
Key components:
Imaging System: Consists of a gantry and patient table.
Computer: Responsible for processing CT image data.
Operator's Console: Controls the imaging process and displays the final images.
Functionality:
Patient lies on the table which moves through the scanner to produce an image.
Equipment and instrumentation behind the scenes facilitate scanning and image reconstruction.
Patient Table and Gantry
Patient Table (Couch):
Constructed from carbon fiber for minimal radiation absorption.
Facilitates controlled movements to position anatomy for scanning.
Weight limit typically up to 450 pounds (204 kilograms).
Gantry:
Large circular apparatus housing essential equipment for image acquisition: X-ray tube, detector array, and data acquisition system.
Contains a cooling system for managing heat generated during X-ray production.
Allows tilting (12-30 degrees) to align anatomy of interest.
Key Parameters in Scanning
Pitch:
Controls table movement through the gantry and relates to image acquisition.
Defined as the ratio of the distance the table moves to the total collimated X-ray beam width.
For Single Detector CT (SDCT): Pitch of 1:1 is optimal for image quality; defined as detector pitch.
For Multi Detector CT (MDCT): Beam pitch is determined by table movement per rotation divided by beam width.
Increasing pitch reduces patient dose but affects image quality negatively.
Coordinates:
X-axis: Horizontal position (right to left).
Y-axis: Vertical position (anterior to posterior).
Z-axis: Corresponds to the axial plane (superior to inferior).
Patient alignment is crucial for accurate scanning.
Data Acquisition Techniques
Axial Scanning:
Step-by-step acquisition of data slice by slice with pauses.
Misregistration artifacts potentially occur due to patient movement.
Helical (Volumetric) Scanning:
Continuous movement of the table through the gantry while the X-ray tube rotates.
Rapid data collection and ability to cover larger volumes.
Reduces misregistration risks by minimizing breath-holding times.
CT Tube and Generator
X-ray Tube:
Larger than traditional diagnostic tubes for managing heat.
Mounts within gantry and rotates around the patient.
Cathode and Anode:
Cathode: Contains tungsten filaments; controls flow of electrons based on milliampere setting.
Anode: Composed of tungsten, molybdenum, rhenium; supports high-speed rotation.
Generator:
Produces high voltage (80-140 kV) required for X-ray production.
Converts incoming power supply (480 V, 60 Hz) to high voltage necessary for operation.
Detectors
Types of Detectors:
Gas Ionization Detectors: Convert X-rays directly to electric signals.
Scintillation Detectors: Convert X-rays to light, then to electric signals (industry standard).
Detector Efficiency:
Comprised of various factors including capture efficiency, absorption efficiency, stability, response time, dynamic range, and reproducibility.
Collimators
Purpose: Controls beam width and limits patient exposure to X-rays.
Types:
Pre-Patient Collimators: Shape beam before it reaches the patient.
Post-Patient Collimators: Shape beam after it passes through the patient, improving image quality.
Slip Ring Technology
Functionality: Allows continuous rotation of the gantry components without cables, enabling faster scan times and continuous data acquisition for advanced imaging techniques.
Types of Slip Rings:
Low voltage (uses AC) and high voltage configurations.
Applications: Critical for modern helical CT and computed tomography angiography (CTA).
Image Reconstruction and Display
Digital Image Data:
Comprised of pixels assigned a shade of gray based on CT numbers from scan data.
Matrix and Pixels:
Common image matrices: 512x512, 1024x1024, 2048x2048.
Pixel size calculated as .
Hounsfield Units (HU):
Determined by the linear attenuation coefficient, with water serving as baseline with a CT number of 0.
Dense structures like bone have higher CT numbers (up to 1000), while air has a CT number of -1000.
Average values in voxels can cause misdiagnosis, a phenomenon known as Partial Volume Effect.
Windowing and Regions of Interest (ROI)
Window Width (WW): Controls contrast; wider WW yields less contrast and wider range of densities.
Window Level (WL): Adjusts brightness of image; should be set according to tissue type being imaged.
Region of Interest (ROI): Tool for selecting specific areas in the image for quantitative analysis.
Measurement aids in identifying tissue types based on density.
Summary of Learning Outcomes
The module concludes by ensuring understanding of:
Major components of the CT system.
Steps in the CT image formation sequence.
Differences between data acquisition methods: axial, helical, and volumetric.
The use of console parameters for optimizing CT image acquisition.
Elements of digital image display and reconstruction.