Frequencies and Tolerances

Tolerances

What we’re checking

Tolerance

SSD on patient

5mm

Collimator Axis of Rotation

2mm

Field Size Indicator Accuracy

2mm

Tx Couch Isocenter

2mm

Lasers

2mm

ODI

2mm

Field Light

2mm

Linear scales (indicating table direction)

2mm

Stability of Isocenter under Gantry Axis of Rotation

2mm

Dosimetry Process

5%

Entire conventional Tx process

10%

Film Processor (Diagnostic)

5-10%

Film Processor (Therapy )

15%

Cobalt/ Linac Spot Check Dose Rate

3%

Cobalt/ Linac Spot Check Beam Penetration

2%

Cobalt/ Linac Calibration Dose Rate

2%

Cobalt/ Linac Calibration Beam Penetration

2%

Field Flatness and Symmetry

2% across the inner 80% of the field

Isodose Distributions/ Verify MU Clacs

2%

Collimator Rotation Digital Readout

1°

Gantry Rotation Digital Readout

1°

Frequencies

Daily

Weekly

Monthly

Yearly

Quarterly

Room Condition

Chart Checks

Emergency OFF

Isodose Distributions/ Verify MU calcs

Limit Switches (Gantry Rotation)

Communication devices

Tx Accessories

Field Light

Motor Reversal

Interlocks (Door, Wedge, Tray)

Over-Ride (Motion Enable Switch)

Cobalt Spot Check Dose Rate

Cobalt Calibration Dose Rate

Warning Lights

ODI

Cobalt Spot Check Penetration

Cobalt Calibration Penetration

Back Up Systems (Timers, MU2)

Linac Spot Check Dose Rate

Linac Calibration Dose Rate

Lasers

Linac Spot Check Penetration

Linac Calibration Penetration

Film Processor (Diagnostic)

Film Processor (Therapy )

Linear Scales (Indicating table direction)

Gantry Rotation Digital Readout

CT sim lasers

Detector Calibration

Other Frequencies

What we are checking

Frequency

Tx couch isocenter

At time of Acceptance Testing

Camera Calibration

Every Few Days

Field Flatness & Symmetry

Weekly - Monthly

Stability of Isocenter Under Gantry Access of Rotation

Weekly- Monthly