EU.MDR.AnnexVIII.Riskclassification.pdf

ANNEX VIII CLASSIFICATION RULES

CHAPTER I DEFINITIONS SPECIFIC TO CLASSIFICATION RULES

1. DURATION OF USE
  • 1.1. ‘Transient’

    • Means normally intended for continuous use for less than 60 minutes.

  • 1.2. ‘Short term’

    • Means normally intended for continuous use for between 60 minutes and 30 days.

  • 1.3. ‘Long term’

    • Means normally intended for continuous use for more than 30 days.

2. INVASIVE AND ACTIVE DEVICES
  • 2.1. ‘Body orifice’

    • Refers to any natural opening in the body, the external surface of the eyeball, or any permanent artificial opening, such as a stoma.

  • 2.2. ‘Surgically invasive device’

    • Defined as:

    • (a) An invasive device that penetrates inside the body through the surface, including mucous membranes of body orifices during surgical procedures.

    • (b) A device producing penetration by means other than through a body orifice.

  • 2.3. ‘Reusable surgical instrument’

    • An instrument intended for surgical tasks such as cutting, drilling, etc., without a connection to an active device, designed by the manufacturer for reuse after cleaning, disinfection, and sterilisation.

  • 2.4. ‘Active therapeutic device’

    • Any active device used to support, modify, replace or restore biological functions or structures to treat or alleviate an illness or disability.

  • 2.5. ‘Active device intended for diagnosis and monitoring’

    • An active device that provides information for detecting, diagnosing, monitoring, or treating physiological conditions or congenital deformities.

  • 2.6. ‘Central circulatory system’

    • Comprises the following blood vessels:

    • Arteriae pulmonales

    • Aorta ascendens

    • Arcus aortae

    • Aorta descendens

    • Arteriae coronariae

    • Arteria carotis communis

    • Arteria carotis externa

    • Arteria carotis interna

    • Arteriae cerebrales

    • Truncus brachiocephalicus

    • Venae cordis

    • Venae pulmonales

    • Vena cava superior

    • Vena cava inferior

  • 2.7. ‘Central nervous system’

    • Refers to the brain, meninges, and spinal cord.

  • 2.8. ‘Injured skin or mucous membrane’

    • An area of skin or mucous membrane showing pathological changes due to disease or wounds.

CHAPTER II IMPLEMENTING RULES

3.1
  • The application of the classification rules is governed by the intended purpose of the devices.

3.2
  • If the device is intended for combination use with another device, classification rules shall apply separately to each device. Accessories are classified on their own aside from the primary device.

3.3
  • Software that drives or influences a device falls within the same class as the device; independent software is classified separately.

3.4
  • If not intended for specific body parts, devices are classified based on the most critical use.

3.5
  • When multiple rules apply to a device, the strictest rule leading to the higher classification takes precedence.

3.6
  • Continuous use defined as:

    • (a) Entire duration of device use disregarding temporary interruptions for procedures or cleaning.

    • (b) Accumulated use if the manufacturer intended immediate replacement with a similar device.

3.7
  • A device allows direct diagnosis by providing a diagnosis by itself or decisive information for diagnosis.

CHAPTER III CLASSIFICATION RULES

4. NON-INVASIVE DEVICES
  • 4.1. Rule 1

    • All non-invasive devices are classified as class I, unless stated otherwise.

  • 4.2. Rule 2

    • Non-invasive devices for channeling or storing blood, etc., are classified as class IIa if:

    • They connect to class IIa, IIb, or III active devices.

    • They store blood or body liquids except for blood bags (class IIb).

  • 4.3. Rule 3

    • Non-invasive devices modifying human tissues or cells are classified as class IIb, except when involving filtration, centrifugation, or heat exchanges (class IIa).

    • In vitro devices in contact with human cells are classified as class III.

  • 4.4. Rule 4

    • Non-invasive devices in contact with injured skin/(or) mucous membranes are classified as:

    • class I for mechanical barriers/compression/exudate absorption.

    • class IIb for serious injuries to the skin/mucous membrane.

    • class IIa for managing the microenvironment of injured skin.

5. INVASIVE DEVICES
  • 5.1. Rule 5

    • Invasive devices regarding body orifices not connected to an active device are classified based on intended use as follows:

    • Class I for transient use.

    • Class IIa for short-term use (oral cavity conditions excluded).

    • Class IIb for long-term use (oral cavity conditions excluded as long as they are not absorbed by mucous membranes).

  • 5.2. Rule 6

    • Surgically invasive devices intended for transient use are classified as class IIa, except for:

    • Specific use for heart/circulatory system (class III).

    • Reusable surgical instruments (class I).

    • Direct contact with heart/central nervous system (class III).

    • Energy delivery via ionising radiation (class IIb).

    • Biological effect or complete absorption classification (class IIb).

  • 5.3. Rule 7

    • Surgically invasive devices for short-term use are classified as class IIa, except when:

    • Intended for heart/circulatory system (class III).

    • Energy delivery through ionizing radiation (class IIb).

    • Biological effects or absorption (class III).

    • Chemical changes in the body (class IIb), with exceptions for dental placements.

  • 5.4. Rule 8

    • Long-term implantable/surgically invasive devices are generally class IIb, unless:

    • Used in teeth (class IIa).

    • Direct contact with heart/circulatory/central nervous systems (class III).

    • Effects leading to absorption or chemical change (class III).

    • Active implantable devices or their accessories (class III).

    • Breast implants/surgical meshes (class III).

    • Joint replacements (class III) but excluding ancillary components.

6. ACTIVE DEVICES
  • 6.1. Rule 9

    • All active therapeutic devices for energy administration/exchange are classified as class IIa unless potentially hazardous (class IIb).

  • 6.2. Rule 10

    • Active diagnosis/monitoring devices are class IIa, unless used in critical monitoring scenarios (class IIb).

  • 6.3. Rule 11

    • Software for decision-making is class IIa unless it leads to severe health impacts (class III) or serious deterioration (class IIb).

  • 6.4. Rule 12

    • Active devices for administering/removing substances are class IIa, unless potentially hazardous (class IIb).

  • 6.5. Rule 13

    • All remaining active devices are classified as class I.

7. SPECIAL RULES

  • 7.1. Rule 14

    • Devices integral with substances acting as medicinal products are class III.

  • 7.2. Rule 15

    • Devices for contraception or STD prevention are class IIb unless implantable/long-term (class III).

  • 7.3. Rule 16

    • Devices for disinfecting contact lenses classified as class IIb; cleaning devices for others except physical action, class IIa.

  • 7.4. Rule 17

    • Devices for generating diagnostic images are class IIa.

  • 7.5. Rule 18

    • Devices from viable/inviable human/animal origin classified as class III unless for intact skin contact only.

  • 7.6. Rule 19

    • Devices with nanomaterials classified based on internal exposure potentials.

  • 7.7. Rule 20

    • Invasive devices for medicinal inhalation classified as class IIa unless critical for efficacy/safety (class IIb).

  • 7.8. Rule 21

    • Substances administered by body orifice or applied to skin classified based on absorption/systemic effects.

  • 7.9. Rule 22

    • Active devices with incorporated diagnostics significantly influencing patient management classified as class III.


Classification and Rule Overview based on Annex VIII - Medical Device Regulation
  1. Mobile device application to monitor drug use as prescribed

    • Classification: Class IIb

    • Rule: Rule 10

  2. Orthopedic MRI

    • Classification: Class IIb

    • Rule: Rule 10

  3. Intrauterine device (IUD)

    • Classification: Class IIb

    • Rule: Rule 15

  4. Contact lens solution

    • Classification: Class IIa

    • Rule: Rule 14

  5. Surgical scalpel

    • Classification: Class I

    • Rule: Rule 11

  6. Hip Implant

    • Classification: Class III

    • Rule: Rule 8

  7. Coronary Stent

    • Classification: Class III

    • Rule: Rule 8

  8. Cardiac Guide Catheter

    • Classification: Class IIb

    • Rule: Rule 6