EU.MDR.AnnexVIII.Riskclassification.pdf
ANNEX VIII CLASSIFICATION RULES
CHAPTER I DEFINITIONS SPECIFIC TO CLASSIFICATION RULES
1. DURATION OF USE
1.1. ‘Transient’
Means normally intended for continuous use for less than 60 minutes.
1.2. ‘Short term’
Means normally intended for continuous use for between 60 minutes and 30 days.
1.3. ‘Long term’
Means normally intended for continuous use for more than 30 days.
2. INVASIVE AND ACTIVE DEVICES
2.1. ‘Body orifice’
Refers to any natural opening in the body, the external surface of the eyeball, or any permanent artificial opening, such as a stoma.
2.2. ‘Surgically invasive device’
Defined as:
(a) An invasive device that penetrates inside the body through the surface, including mucous membranes of body orifices during surgical procedures.
(b) A device producing penetration by means other than through a body orifice.
2.3. ‘Reusable surgical instrument’
An instrument intended for surgical tasks such as cutting, drilling, etc., without a connection to an active device, designed by the manufacturer for reuse after cleaning, disinfection, and sterilisation.
2.4. ‘Active therapeutic device’
Any active device used to support, modify, replace or restore biological functions or structures to treat or alleviate an illness or disability.
2.5. ‘Active device intended for diagnosis and monitoring’
An active device that provides information for detecting, diagnosing, monitoring, or treating physiological conditions or congenital deformities.
2.6. ‘Central circulatory system’
Comprises the following blood vessels:
Arteriae pulmonales
Aorta ascendens
Arcus aortae
Aorta descendens
Arteriae coronariae
Arteria carotis communis
Arteria carotis externa
Arteria carotis interna
Arteriae cerebrales
Truncus brachiocephalicus
Venae cordis
Venae pulmonales
Vena cava superior
Vena cava inferior
2.7. ‘Central nervous system’
Refers to the brain, meninges, and spinal cord.
2.8. ‘Injured skin or mucous membrane’
An area of skin or mucous membrane showing pathological changes due to disease or wounds.
CHAPTER II IMPLEMENTING RULES
3.1
The application of the classification rules is governed by the intended purpose of the devices.
3.2
If the device is intended for combination use with another device, classification rules shall apply separately to each device. Accessories are classified on their own aside from the primary device.
3.3
Software that drives or influences a device falls within the same class as the device; independent software is classified separately.
3.4
If not intended for specific body parts, devices are classified based on the most critical use.
3.5
When multiple rules apply to a device, the strictest rule leading to the higher classification takes precedence.
3.6
Continuous use defined as:
(a) Entire duration of device use disregarding temporary interruptions for procedures or cleaning.
(b) Accumulated use if the manufacturer intended immediate replacement with a similar device.
3.7
A device allows direct diagnosis by providing a diagnosis by itself or decisive information for diagnosis.
CHAPTER III CLASSIFICATION RULES
4. NON-INVASIVE DEVICES
4.1. Rule 1
All non-invasive devices are classified as class I, unless stated otherwise.
4.2. Rule 2
Non-invasive devices for channeling or storing blood, etc., are classified as class IIa if:
They connect to class IIa, IIb, or III active devices.
They store blood or body liquids except for blood bags (class IIb).
4.3. Rule 3
Non-invasive devices modifying human tissues or cells are classified as class IIb, except when involving filtration, centrifugation, or heat exchanges (class IIa).
In vitro devices in contact with human cells are classified as class III.
4.4. Rule 4
Non-invasive devices in contact with injured skin/(or) mucous membranes are classified as:
class I for mechanical barriers/compression/exudate absorption.
class IIb for serious injuries to the skin/mucous membrane.
class IIa for managing the microenvironment of injured skin.
5. INVASIVE DEVICES
5.1. Rule 5
Invasive devices regarding body orifices not connected to an active device are classified based on intended use as follows:
Class I for transient use.
Class IIa for short-term use (oral cavity conditions excluded).
Class IIb for long-term use (oral cavity conditions excluded as long as they are not absorbed by mucous membranes).
5.2. Rule 6
Surgically invasive devices intended for transient use are classified as class IIa, except for:
Specific use for heart/circulatory system (class III).
Reusable surgical instruments (class I).
Direct contact with heart/central nervous system (class III).
Energy delivery via ionising radiation (class IIb).
Biological effect or complete absorption classification (class IIb).
5.3. Rule 7
Surgically invasive devices for short-term use are classified as class IIa, except when:
Intended for heart/circulatory system (class III).
Energy delivery through ionizing radiation (class IIb).
Biological effects or absorption (class III).
Chemical changes in the body (class IIb), with exceptions for dental placements.
5.4. Rule 8
Long-term implantable/surgically invasive devices are generally class IIb, unless:
Used in teeth (class IIa).
Direct contact with heart/circulatory/central nervous systems (class III).
Effects leading to absorption or chemical change (class III).
Active implantable devices or their accessories (class III).
Breast implants/surgical meshes (class III).
Joint replacements (class III) but excluding ancillary components.
6. ACTIVE DEVICES
6.1. Rule 9
All active therapeutic devices for energy administration/exchange are classified as class IIa unless potentially hazardous (class IIb).
6.2. Rule 10
Active diagnosis/monitoring devices are class IIa, unless used in critical monitoring scenarios (class IIb).
6.3. Rule 11
Software for decision-making is class IIa unless it leads to severe health impacts (class III) or serious deterioration (class IIb).
6.4. Rule 12
Active devices for administering/removing substances are class IIa, unless potentially hazardous (class IIb).
6.5. Rule 13
All remaining active devices are classified as class I.
7. SPECIAL RULES
7.1. Rule 14
Devices integral with substances acting as medicinal products are class III.
7.2. Rule 15
Devices for contraception or STD prevention are class IIb unless implantable/long-term (class III).
7.3. Rule 16
Devices for disinfecting contact lenses classified as class IIb; cleaning devices for others except physical action, class IIa.
7.4. Rule 17
Devices for generating diagnostic images are class IIa.
7.5. Rule 18
Devices from viable/inviable human/animal origin classified as class III unless for intact skin contact only.
7.6. Rule 19
Devices with nanomaterials classified based on internal exposure potentials.
7.7. Rule 20
Invasive devices for medicinal inhalation classified as class IIa unless critical for efficacy/safety (class IIb).
7.8. Rule 21
Substances administered by body orifice or applied to skin classified based on absorption/systemic effects.
7.9. Rule 22
Active devices with incorporated diagnostics significantly influencing patient management classified as class III.
Classification and Rule Overview based on Annex VIII - Medical Device Regulation
Mobile device application to monitor drug use as prescribed
Classification: Class IIb
Rule: Rule 10
Orthopedic MRI
Classification: Class IIb
Rule: Rule 10
Intrauterine device (IUD)
Classification: Class IIb
Rule: Rule 15
Contact lens solution
Classification: Class IIa
Rule: Rule 14
Surgical scalpel
Classification: Class I
Rule: Rule 11
Hip Implant
Classification: Class III
Rule: Rule 8
Coronary Stent
Classification: Class III
Rule: Rule 8
Cardiac Guide Catheter
Classification: Class IIb
Rule: Rule 6