DMS125-Chapter 4- Describing Pulsed Waves(1)
Objectives
Identify Pulsed Sound and Duration
Learn about Spatial Pulse Length, Pulse Repetition Period, and Pulse Repetition Frequency
Understand the relationship of PRP and PRF
Recognize Duty Factor
Pulsed Sound
Defined as short bursts of acoustic energy.
Contains 5 parameters, must have a beginning and an end.
Comprised of cycles that travel together.
Two main components:
Transmit: "On" time (active signal transmission)
Receive: "Off" time (listening period)
Parameters of Pulsed Sound
Pulse Duration: The time from start to end of a pulse, typically measured in microseconds (๐๐บ).
Value ranges from 0.3 to 2.0 ๐๐บ, determined by the source, adjustable: No.
Directly proportional to the number of cycles and the period:
Pulse Duration (๐๐บ) = #Cycles x Period (๐๐บ)
Inversely proportional to frequency:
Pulse Duration (๐๐บ) = #Cycles / Frequency (MHz)
Spatial Pulse Length (SPL): The distance the pulse occupies, measured in mm.
Value range: 0.1 to 1.0 mm, determined by both source and medium, adjustable: No.
Relation: SPL (mm) = #Cycles x Wavelength (mm), inversely proportional to frequency.
Pulse Repetition Period (PRP): The time from the start of one pulse to the start of the next pulse, measured in ms.
Value range: 100 ๐๐บ to 1 ms, determined by the source/depth, adjustable: Yes.
Directly related to Depth of View (increases with depth).
Pulse Repetition Frequency (PRF): The number of ultrasound pulses transmitted per second, measured in Hz (Hertz).
Value range: 1,000 to 10,000 Hz, determined by source/depth, adjustable: Yes.
Duty Factor: The percentage of time the system transmits a pulse.
Value ranges from 0.2% to 0.5%, determined by source, adjustable: Yes.
Formula: Duty Factor (%) = PRP x 100.
Continuous Wave vs Pulsed Wave
Both waves are described by:
Period
Frequency
Wavelength
Propagation Speed
Summary of Parameters for Pulsed Waves
Pulse Duration: Not adjustable, typical value: 0.5 to 3.0 ๐๐บ
Pulse Repetition Period: Adjustable, typical value: 0.1 to 1.0 ms
Pulse Repetition Frequency: Adjustable, typical value: 1 to 10 kHz
Spatial Pulse Length: Not adjustable, typical value: 0.1 to 1.0 mm
Duty Factor: Adjustable, typical value: 0.2% to 0.5%
Review Questions
Pulse Repetition Period is the time from the start of a pulse to the start of the next pulse: C. Pulse Repetition Period
Pulse Duration is from the start of a pulse to the end of that pulse: D. Pulse Duration
Pulse Repetition Period is the reciprocal of Pulse Repetition Frequency: B. Pulse Repetition Period
Which describes line C? (requires figure context)
Which describes line D? (requires figure context)
Which describes line A? (requires figure context)
Which describes line B? (requires figure context)
Changing image depth affects: (More than one) A. PRF, B. Duty Factor, D. PRP
Duty Factor calculation based on provided values.
True/False: Pulse Repetition Frequency and Pulse Repetition Period are determined only by the imaging depth.