PV Delay and Pacing Modes
The PV Delay
- The PV Delay interval is a timing concept important for understanding certain exam questions.
- Also known as P Tracking Interval and PV Interval.
- PV Delay can be a fixed or variable value.
- Modes that use the PV Interval: DDD(R), VDD(R), DDT(R), and VAT.
- DDT and VAT modes are rare and won't be discussed.
- The PV Delay interval begins when the atrial output escape interval (AEI) is interrupted by a sensed event on the atrial channel.
- A sensed event, not just a P wave, triggers the PV Delay. Far-field ventricular events and noise can be sensed and start the PV Delay.
- When an event is detected, the atrial channel becomes refractory, and after a specific duration (PV Delay), the ventricular output is generated.
- An R wave or any perceived ventricular event will stop the PV interval and inhibit the ventricular output pulse.
- On the exam, focus on the relationship between the P wave and the ventricular output spike.
- The P to V interval should not lengthen as the rate increases.
- Progressive lengthening of the PV Interval may indicate pacemaker Wenkebach.
- Pacemaker Wenkebach is a NORMAL function in dual-chamber pacing. It happens when the sensed P rate is faster than the programmed maximum tracking rate (or upper rate interval).
- In DDD pacing, P tracking occurs from rates slightly above the programmed base rate to the upper tracking rate limit.
- Rates at or below the base rate should appear as AV or AR pacing.
- VDD Timing Consideration
- Unlike DDD timing, VDD mode has NO atrial output pulse.
- If a rhythm strip shows P tracking above the base rate but also has episodes resembling VVI (non-P Synchronous) at low rates, suspect VDD mode.
- In DDD mode, a ventricular output cannot occur unless an atrial event (sensed or paced) precedes it.
- In VDD pacing, if the atrial rate drops below the base rate, the mode will appear to be VVI.
- VDD pacing idiosyncrasy: below base rate ventricular pacing behavior when the P rate is very close to the programmed base rate.
- Example: Device programmed to 60 ppm (1000 ms) with an AV delay of 200 ms, and the sinus rate is 61 bpm (984 ms).
- The V - V interval will be 1184 ms (approximately 51 ppm).
- This occurs because the AV Interval starts when the P wave is sensed, delaying the ventricular output.
- V−V interval=AV Delay+Sinus Rate where AV Delay and Sinus Rate are expressed in milliseconds.
- V−V interval=200 ms+984 ms=1184 ms
- Below base rate ventricular pacing is NORMAL in VDD pacing.
- Ventricular pacing may occur at or above the programmed rate.