QAQC
FINALS
I. ACCESSORIES IN FILM-SCREEN RADIOGRAPHY
Although digital radiography has become the dominant technology today, these accessories remain foundational. Understanding them clarifies how imaging evolved and why certain QC standards exist.
1. Viewbox (Negatoscope)
Definition
An illuminated surface used for viewing radiographic films.
Purpose
Provide uniform brightness to evaluate details, contrast, and density.
Essential for detecting fractures, lesions, and soft-tissue differences.
Key Features
Bright, even white light (minimum 1500 cd/m²).
Diffuser plate to eliminate hotspots.
Film clips or magnets.
Considerations / Relevance
Ambient light affects perception; too much brightness reduces contrast sensitivity.
Dirty or yellowed acrylic reduces diagnostic accuracy.
Evenness of illumination is critical for recognizing subtle pathology.
QC / Maintenance
Monthly cleaning.
Annual luminance uniformity test.
Replace dim or flickering bulbs.

2. Cassette
Definition
A light-tight container that holds film and pressed intensifying screens against it.
Purpose
Protect film from light.
Ensure good film-screen contact.
Provide rigid support during exposures.
Types
Flat cassettes
Curved cassettes (mammography/tomography)
Gridded cassettes (scatter reduction)
Important Factors
Must maintain perfect film-screen contact → affects sharpness.
Hinges/latches must prevent light leaks.
QC / Issues
Cassette warping → causes image blurriness.
Dirty screens → white spots.
Light leaks → fogged films.
3. Intensifying Screens
Definition
Phosphor-coated screens that convert X-rays into visible light, exposing the film.
Purpose
Reduce patient dose by 90–95%.
Provide fast imaging.
Types
Calcium tungstate (old)
Rare-earth screens (gadolinium, lanthanum phosphors – faster, more efficient)
Screen Speed
Slow (high detail)
Medium
Fast (low dose, less sharp)
Relevance to Exposure Factors
Faster screens → lower mAs → lower dose, but ↓ detail.
Screen-film combinations must be properly matched (blue or green sensitive).
QC Requirements
Screen cleaning every month.
Screen contact test with wire mesh.
Replace cracked or dead phosphor areas.
4. Grids (Accessory but essential)
Purpose
Reduce scatter radiation reaching film → improves contrast.
Types
Linear
Focused
Parallel
Crossed
Moving (Bucky)
Considerations
Grid ratio affects exposure: higher ratio → more mAs needed.
Misalignment causes a grid cut-off.
5. Lead Markers
Purpose
Legal requirement to indicate anatomical side (L/R) and technologist ID.
Relevance
Essential for medico-legal documentation.
6. Compression Devices / Sponges
Purpose
Stabilization
Reduce OID → improve detail
Reduce motion
II. RADIATION QUALITY CONTROL (QC)
Performed to ensure consistent image quality, patient safety, and equipment reliability.
Regulatory Bodies
FDA (sets radiation performance standards)
DOH / local health regulators (PH)
Hospital Medical Physicist → performs annual QC
Radiologic Technologist → daily checks
1. Peak kVp Accuracy Test
Purpose
Ensure generator delivers correct kVp → affects contrast, penetration, and dose.
Procedure
Using a digital kVp meter:
Expose at set kVp (60, 80, 100)
Meter compares actual vs displayed
Acceptance Criteria
±5 kVp from the indicated value.
If Out of Control
Recalibrate the generator.
Unsafe → equipment locked out by physicist.
2. Exposure Time Accuracy Test
Purpose
Verify timer accuracy → motion blur and density depend on correct timing.
Procedure
Use oscilloscope or spinning top test (for single-phase units).
Compare actual pulses vs expected.
Acceptance Criteria
±5% for >10 ms
±20% for <10 ms
3. mA Linearity Test / mAs Reciprocity Test
Purpose
To ensure that different mA and time combinations produce the same mAs and the same density.
Example
200 mA × 0.1 s = 20 mAs
400 mA × 0.05 s = 20 mAs
Both should produce equal film density.
Acceptance Criteria
Variation must be ≤10%.
If Failed
Indicates tube or generator malfunction.
4. Focal Spot Size Test
Purpose
Ensure the focal spot has not enlarged (excess heat causes blooming).
Affects sharpness and spatial resolution.
Methods
Pin-hole camera
Star-test pattern
Slit camera
Acceptance Criteria
Small FS: ≤50% variance
Large FS: ≤40% variance
If Failed
Loss of detail
The tube may need replacement
5. Beam Alignment Test (Central Ray Alignment)
Purpose
Confirm CR is centered and perpendicular to IR to prevent shape distortion.
Method
Use the beam alignment tool
CR and light field must match
Tolerance
±2% of SID.
6. AEC (Automatic Exposure Control) Test – One Pulse Test
Purpose
Ensure AEC terminates exposure consistently.
Method
Use acrylic blocks of varying thickness
Measure exposure time or mAs
Repeatability test (should be consistent)
If Out of Control
Overexposure or underexposure
AEC recalibration required by a physicist
7. Step Wedge Test (Sensitometry / Densitometry)
Purpose
Monitor film processor consistency.
Procedure
Expose film using an aluminum step wedge
Measure optical density
Compare with the control chart
Evaluates
Developer temperature
Chemical strength
Processor performance
III. STANDARD RADIOGRAPHIC TECHNIQUE CHART
A technique chart is a standardized guide indicating kVp, mAs, SID, and projection based on part thickness and patient habitus.
Its purpose:
Ensure consistent image quality
Reduce repeat exposures
Reduce dose
Provide a reference for technologists
STRUCTURE OF A TECHNIQUE CHART
1. Body Part
Chest
Abdomen
Skull
Extremities
Spine
2. Thickness Measurement
Measured using calipers (cm).
Thicker → more kVp or mAs
Thinner → less kVp or mAs
3. Body Habitus
Hypersthenic → ↑ exposure
Sthenic → standard
Hyposthenic → slightly ↓
Asthenic → ↓ kVp and mAs
4. Projection Type
AP
PA
Lateral
Oblique
Relevance
Different projections change OID, magnification, scatter contribution, and positioning.
RELATIONSHIPS TO OTHER FACTORS
kVp vs Contrast
High kVp → low contrast (chest)
Low kVp → high contrast (extremities)
mAs vs Density
Directly proportional.
SID (distance) effects
↑ SID → ↓ intensity → need ↑ mAs (Inverse Square Law)
Grids
Use of grid → requires ↑ mAs
Immobilization & OID
Large OID → magnification → need compensation by compression or higher SID.
PURPOSES OF A TECHNIQUE CHART
Standardize exposures
Reduce repeat rates
Ensure diagnostic quality
Control patient dose
Assist inexperienced technologists
EVALUATION AND ACTION (for technologists)
After every exam, evaluate:
1. Positioning Accuracy
If incorrect → repeat with explanation and corrective technique.
2. Density and Contrast
If too dark or light → adjust mAs or kVp accordingly.
3. Visibility of Detail
If blurred → assess motion / focal spot/exposure time.
4. Presence of Artifacts
Remove jewelry, clothing, and lines next exam.
5. Compliance With Technique Chart
If consistently off → chart needs revision.
ROUTINE OF CARE RATIONALE
Before Exam
Correct ID → safety and medico-legal
History taking → guides technique
Remove metal → prevents artifacts
During Exam
Proper immobilization → prevents motion blur
Shielding → reduces dose
Centering → avoids distortion
After Exam
Confirm image quality
Document exposure parameters
Prepare the room for the next patient