Centrifugation Principles and Laboratory Operation

Principles and Function of Centrifugation

  • Purpose: Centrifuges and microcentrifuges separate particles from liquid far more rapidly than gravity by utilizing centrifugal force generated through high-speed rotation.
  • Applications: Used across biological research and diagnostics, including isolating cells, measuring hematocrit (red blood cell volume in blood), and preparing samples for infectious disease testing.

Operation and Balancing Protocol

  • Tube Specifications: Only use tubes specifically rated to withstand high centrifugal forces.
  • Rotor Balancing:
    • An unbalanced rotor causes severe instrument damage.
    • Balance tubes by placing them directly across from each other or distributing them symmetrically around the rotor.
    • Opposing tubes must contain matched volumes; use a blank balance tube with a similar volume if running an uneven number of samples.
  • Execution: Secure internal lids (if equipped) before closing the external centrifuge lid. Set run duration and speed according to protocol.
  • Speed Metrics: Set in RPM (revolutions per minute) or RCF (related to g-force), where higher rotational speeds correspond to higher centrifugal force.

Sample Handling and Post-Centrifugation Fractions

  • Hinge Alignment: Position microcentrifuge tube hinges pointing outward when loading to easily predict and locate the position of small or translucent pellets after spinning.
  • Unloading: Remove tubes gently without jostling to prevent resuspending sedimented particles.
  • Separation Products:
    • Pellet: The tightly packed mass of separated particles concentrated at the bottom of the tube.
    • Supernatant: The remaining liquid layer above the pellet.