Clinical Laboratory Instrumentation: Spectrophotometry

Electromagnetic Spectrum and Spectroscopy

  • Spectroscopy analyzes the absorption and emission of electromagnetic radiation by substances through the dispersion of light into its constituent wavelengths.
  • Light wave relationship:   Wavelength=light velocityfrequency\text{Wavelength} = \frac{\text{light velocity}}{\text{frequency}}
  • Key clinical chemistry wavelength regions:
    • Ultra violet (UV): 250−350 nm250-350\,\text{nm}
    • Visible: 350−750 nm350-750\,\text{nm} (peak human eye sensitivity occurs at 555 nm555\,\text{nm})
    • Near infra red (Near IR): 750−1400 nm750-1400\,\text{nm}
  • Visible spectrum color bands:
    • Violet: 400−420 nm400-420\,\text{nm}
    • Indigo: 420−440 nm420-440\,\text{nm}
    • Blue: 440−490 nm440-490\,\text{nm}
    • Green: 490−570 nm490-570\,\text{nm}
    • Yellow: 570−585 nm570-585\,\text{nm}
    • Orange: 585−620 nm585-620\,\text{nm}
    • Red: 620−780 nm620-780\,\text{nm}

Principles of Light Absorption and Beer-Lambert Law

  • Observed color corresponds to the transmitted or reflected light, whereas the absorbed light is its complementary color:
    • Red objects reflect red light and absorb green light (500−520 nm500-520\,\text{nm}, perceived complement: red/purple).
    • Yellow objects reflect yellow light and absorb violet-blue light (420−440 nm420-440\,\text{nm}).
  • Transmittance (TT): The fraction of incident light (I0I_0) transmitted through a sample (II):   %T=II0×100%\%T = \frac{I}{I_0} \times 100\%%T\%T decreases logarithmically (not linearly) with increasing concentration: T=10−abcT = 10^{-abc}.
  • Absorbance (AA): The capacity of a medium to absorb radiation, directly proportional to solute concentration according to the Beer-Lambert law:   A=C×L×εA = C \times L \times \varepsilonC=AL×εC = \frac{A}{L \times \varepsilon}
    • AA: Absorbance
    • CC: Solute concentration
    • LL: Optical path length
    • ε\varepsilon: Extinction coefficient (substance- and wavelength-specific constant)

Core Spectrophotometer Components

Spectrophotometer component diagram

  • Light Source: Must produce a stable, continuous radiant beam over the wavelengths of interest:
    • Tungsten filament lamp: 360−800 nm360-800\,\text{nm} (visible and near IR ranges).
    • Halogen lamp: 290−800 nm290-800\,\text{nm} (broader continuous spectrum).
    • Hydrogen or deuterium lamp: 200−375 nm200-375\,\text{nm} (lower UV range).
  • Monochromator: Isolates a narrow band of nominal wavelength from polychromatic light via colored glass absorption filters or dispersion prisms and diffraction gratings combined with entrance and exit slits.
  • Sample Holder (Cuvettes): Standard path length of 1 cm1\,\text{cm}; flat optical faces minimize stray light:
    • Plastic: Visible range only (absorbs UV).
    • Glass: 320−950 nm320-950\,\text{nm} (absorbs UV).
    • Quartz: UV wavelengths below 320 nm320\,\text{nm}.
  • Detectors: Convert transmitted radiant photons into measurable electrical energy:
    • Charged Couple Device (CCD): Multi-channel detector array measuring the entire spectrum simultaneously; allows rapid reference subtraction; exhibits higher dark/electronic noise and poorer limits of detection (LOD).
    • Photomultiplier Tube (PMT): Single-element detector per wavelength multiplying secondary electrons across a series of dynodes; provides high sensitivity, high signal-to-noise (S/N\text{S/N}) ratio, lower LOD, and high efficiency in the UV spectrum.

Spectrophotometer Optical Configurations

Double-beam-in-space and double-beam-in-time spectrophotometer optical paths

  • Single-Beam Spectrophotometer:
    • Directs a single light beam through the sample.
    • Requires manual calibration with a reference blank to establish 0 Absorbance0\,\text{Absorbance} (100% T100\%\,T) prior to reading samples.
    • Cannot compensate for fluctuations in light source intensity during operation.
  • Split (Double) Beam Spectrophotometer:
    • Splits light into two paths originating from a single source: one through the sample and one through a reference blank, canceling out light drift:     Absorbance=Sample signal−Reference (blank) signal\text{Absorbance} = \text{Sample signal} - \text{Reference (blank) signal}
    • Double-beam-in-space: Employs two separate detectors to measure sample and reference beams concurrently.
    • Double-beam-in-time: Uses a single detector with an oscillating chopper or rotating mirror to alternate light between sample and blank paths.
  • Dual Wavelength Spectrophotometer:
    • Employs two monochromators to illuminate a sample simultaneously with two discrete wavelengths.
    • Corrects for background absorbance when an interfering substance overlaps the analyte spectrum:     True Absorbance=Sample Wavelength Absorbance−Reference Wavelength Absorbance\text{True Absorbance} = \text{Sample Wavelength Absorbance} - \text{Reference Wavelength Absorbance}

Clinical and Analytical Applications

  • Enzyme Assays: Rate-based quantitation by tracking coenzyme absorbance shifts (e.g., lactate dehydrogenase / LD):   L-lactate+NAD+→LDPyruvate+NADH\text{L-lactate} + \text{NAD}^+ \xrightarrow{\text{LD}} \text{Pyruvate} + \text{NADH}   NADH product exhibits maximum absorption at 340 nm340\,\text{nm}.
  • Metabolite Detection: Chromogenic condensation reactions (e.g., urea determination):   Urea+Diacetyl→Acid & heatH2O+Diazine\text{Urea} + \text{Diacetyl} \xrightarrow{\text{Acid \& heat}} \text{H}_2\text{O} + \text{Diazine}   Diazine product absorbs maximally at 540 nm540\,\text{nm}.
  • Protein Assays:
    • Biuret reaction: Alkaline chelation of copper to peptide bonds forms a violet complex measured at 540 nm540\,\text{nm}.     Protein+copper→Alkaline pHprotein-copper complex\text{Protein} + \text{copper} \xrightarrow{\text{Alkaline pH}} \text{protein-copper complex}
    • Direct UV absorption: Diluted protein quantitation at 280 nm280\,\text{nm} driven by aromatic amino acids (tyrosine and tryptophan).
  • Nucleic Acid Quantitation: Direct UV absorption of diluted DNA/RNA measured at 260 nm260\,\text{nm}.
  • Micro-Volume Spectrophotometry (NanoDrop 2000/2000c):
    • Analyzes 0.5 μL−2.0 μL0.5\,\mu\text{L} - 2.0\,\mu\text{L} volumes directly pipetted onto a pedestal without cuvettes.
    • Liquid column forms via surface tension upon arm closure, automatically setting optical path lengths from 0.05 mm0.05\,\text{mm} to 1 mm1\,\text{mm}.
  • Enzyme-Linked Immunosorbent Assay (ELISA):
    • Solid-phase microplate assay measuring antigens, antibodies, or glycoproteins.
    • Enzyme-conjugated antibodies convert substrate (SS) to colored product (PP), followed by spectrophotometric microplate reading.