kVp
Chapter 16: Kilovoltage Peak (kVp)
Important Terms
kVp (Kilovoltage Peak)
Defined as the highest amount of electrical "pressure" in a given exposure.
keV (Kiloelectron Volt)
A unit of energy utilized in certain charts.
It states that no photon produced can exceed the set kVp in energy.
Quality
Refers to the energy level of the beam and its penetrating capability.
Controlled exclusively by the kVp value.
Kilovoltage Peak (kVp) Overview
Affects both the quantity and quality of the x-ray beam.
Determines the maximum kinetic energy of electrons accelerating from the cathode to the anode, thus setting the maximum energy for resulting x-ray photons.
As kVp increases:
The penetrability (quality) of x-rays improves.
There is an exponential increase in the quantity of x-rays produced within the tube.
kVp and X-ray Energy
The energy (or penetrability) of the x-ray beam is not uniform.
For instance, setting the kVp to 80:
Maximum energy of produced x-ray photons is 80 keV.
However, most x-rays produced will have lower energies than this maximum.
The average energy is approximately one-third of the kVp set value.
Changes in kVp
Increasing the kVp results in:
Elevated overall energy of the beam (quality).
Increased production of x-rays (quantity).
Subject Contrast
Subject contrast should not be confused with image contrast.
Subject Contrast
Defined as the information transmitted by the remnant beam after passing through the patient.
It involves various exposure levels in the remnant beam due to differential absorption across different tissue types.
A specific example is illustrated using a cross-section of a forearm being radiographed in the antero-posterior (AP) position.
Subject Contrast and Image Contrast
Every anatomical structure has an inherent subject contrast based on tissue types.
kVp settings can enhance or override this inherent subject contrast.
Manipulating kVp affects the latent image contrast:
Example:
An AP foot can exhibit high contrast with low kVp settings.
Raising the kVp allows more photons to penetrate, producing more similar appearances among tissue types, thus decreasing image contrast and resulting in more shades of gray.
kVp and Subject Contrast
Higher-energy x-ray photons are capable of penetrating various tissue types more effectively.
It is crucial to find a balance between x-ray penetration and desired grayscale in imaging.
Minimum kVp:
Identified as the lowest kV necessary to ensure sufficient penetration of all relevant tissues.
Chest X-Ray (CXR) with Lower kVp
Lower kVp results in decreased x-ray beam penetration.
This results in:
Increased absorption of x-rays and reduced transmission through tissues.
A higher level of image contrast is produced as a result.
Chest X-Ray (CXR) with Higher kVp
Higher kilovoltages enhance the penetrating ability of the x-ray beam.
Resulting effects include:
Decreased absorption and increased transmission through anatomical tissues.
This translates into lower image contrast due to reduced variation in x-ray intensities exiting the patient.
Excessive or Insufficient kVp
Too Much kVp:
Leads to:
Over-penetration of tissues.
Loss of subject contrast; images of extremities (e.g., hand or foot) appearing too dark.
Increased scatter radiation reaching the image receptor, diminishing image quality.
Generally requires insufficient mAs, risking quantum noise in images.
Too Little kVp:
Results in:
Excessive patient radiation exposure due to increased absorption of x-rays by tissue.
Overall images may appear underexposed.
Minimum kVp Considerations
No amount of mAs can correct for inadequate penetration due to low kVp settings.
Reference materials (Table 16-1) indicate minimum kVp values for Computed Radiography (CR) and Digital Radiography (DR).
15 Percent Rule
Increasing kVp leads to an increase in the number of x-rays produced in the tube and enhances penetration through anatomical structures, allowing more x-rays to reach the image receptor (IR).
The 15 Percent Rule states:
A 15 percent change in kVp results in an exposure to the IR factor change of 2.
Specifically:
Increasing kVp by 15 percent doubles the exposure to the IR.
Decreasing kVp by 15 percent halves the exposure to the IR.
This compensates for exposure changes without adjusting mAs values.
Area Under the Curve (15 Percent Rule)
The area under the energy curve represents the total number of x-rays at different energy levels for a specific beam.
A 15 percent increase in kVp doubles the area under this curve, for example moving from an 82 kVp to a higher value.
Compensation for 15 Percent Rule
Adjustments can be made to kVp to modify beam penetrability or subject contrast while maintaining IR exposure.
Example case involves changing kVp along with mAs:
While radiographing a hand at 60 kVp and 2 mAs, one could increase kVp using the 15 percent rule while keeping exposure indices constant.
kVp Compensation Method
This method can work in both directions.
If kVp is decreased by 15 percent, mAs must be doubled.
Compounding adjustments requires recalibrating mAs with each kVp shift.
Example:
Start with: 55 kVp, 20 mAs.
Adjust to: 62 kVp, 10 mAs → 71 kVp, 5 mAs → 81 kVp, 2.5 mAs (for consistent exposures).
15 Percent Rule Uses
Adjust exposure towards the IR by changing kVp without adjusting mAs.
To double the exposure to IR:
Increase kVp by 15 percent.
To halve the exposure:
Decrease kVp by 15 percent.
To maintain EI when an mAs value is satisfactory yet lower mAs is desired:
Increasing kVp by 15 percent and halving the mAs still preserves original EI.
Optimum kVp Considerations
Digital radiography systems provide significant latitude with technical settings.
In contrast to earlier film systems, exact kVp settings are not as critical for achieving the correct contrast.
Higher kVp settings above the minimum are often acceptable, offering a bonus of reduced patient exposure.
15 Percent Rule Impact on Patient Dose
Adhering to the digital-era guidelines suggests utilizing high kVp with low mAs settings.
Following the 15 percent rule properly adjusts mAs, resulting in a potential one-third reduction in patient radiation exposure.
Caution must be practiced: mAs that is too low risks causing image mottle.
Minimizing Patient Exposure
Strategies to minimize patient radiation exposure include:
Employing higher kVp and lower mAs values whenever possible.
Restricting the x-ray beam size through collimation techniques.
Using grids judiciously rather than routinely for each examination.
Selecting appropriate digital exposure techniques tailored for patient age and body habitus, as pediatric patients require different exposure considerations due to their smaller size.