SPECT Imaging Principles, Data Acquisition, and Reconstruction
Overview and Principles of SPECT
- SPECT Definition: Single Photon Emission Computerized Tomography.
- Camera Configurations:
- Rotates around the patient and uses a computer to generate cross-sectional 3D views.
- Single-head camera must rotate 360∘.
- Two-head camera rotates 180∘ per head.
- Triple-head camera rotates 120∘ per head.
- Comparison with Other Modalities:
- SPECT: Computerized emission tomography.
- CT: Computerized transmission tomography.
- PET: Emission tomography using annihilation photons from positrons.
- Primary Advantages: Increased contrast resolution, added depth, size, and shape information, and 3D image representation.
SPECT Data Acquisition
- Projections: Sequential 2D raw views used as intermediate steps to reconstruct slices (typically 64 stops, acquiring 128 2D raw images in a 128×128 matrix).
- Timing and Orbits: Time per stop is constant for each stop. Acquisition uses circular or elliptical orbits across 180∘ (such as RAO to LPO) or 360∘.
- Acquisition Modes: Continuous acquisition, Step-and-Shoot acquisition, and Continuous Step-and-Shoot acquisition.
- Gated SPECT: Synchronized acquisition based on the cardiac R wave.
Image Quality and Motion Detection
- Motion Detection:
- Identified when the same source is seen in two different places.
- X-motion is easier to correct than Y-motion.
- Detection techniques include Cine, Sinogram, and Summed images (see Powsner page 99).
- Gating Errors: Irregular heart rates result in bad gating, identified using sinograms, count histograms, beat histograms, and accepted/rejected beat tallies.
Image Reconstruction Techniques
- Reconstruction Types:
- Filtered Back Projection (FBP): Most common reconstruction method during the 1980s and 1990s.
- Iterative Reconstruction: Modern forefront technique used in SPECT and PET.
- Back Projection: Smears counts along the radius/line of view into each pixel.
- Star Artifact: Radial spokes/streaks extending from hot spots due to simple back projection, leading to image blurring.
Image Domains and Frequency Analysis
- Spatial Domain: Data expressed as counts per pixel. Uses spatial filters like a 9-point smooth to redistribute counts to immediate neighbors via a filtering kernel (see Powsner pgs 154-157).
- Frequency Domain: Data expressed as sine waves in \text{cycles/pixel}. Fourier Transformation converts data from the spatial domain to the frequency domain.
- Nyquist Frequency: Set at 0.5cycles/pixel (every other pixel is on), representing the highest useful imaging frequency. When expressed in cycles/cm, smaller pixels result in a larger Nyquist frequency.
- Frequency Categories:
- Low Frequency: Small pixel-to-pixel count differences (e.g., Gallium Scan).
- High Frequency: Large pixel-to-pixel count differences with sharp edges (e.g., Bone Scan).
Image Components and Filtering
- Raw SPECT Image Components:
- Background: Low frequency; generated from back projection spokes.
- Target Data: Desired image information; spans low to high frequencies depending on study type.
- Noise: High frequency; unwanted statistical fluctuations inversely proportional to raw counts.
- RAMP Filter: A high-pass filter that is always used in back projection to eliminate low-frequency background and reduce star artifacts.
- User Filters: Low-pass filters applied to remove high-frequency noise without altering raw data (e.g., Hann, Hamming, Butterworth, Wiener, Parzen).