SPECT Imaging Principles, Data Acquisition, and Reconstruction

Overview and Principles of SPECT

  • SPECT Definition: Single Photon Emission Computerized Tomography.
  • Camera Configurations:
    • Rotates around the patient and uses a computer to generate cross-sectional 3D views.
    • Single-head camera must rotate 360∘360^\circ.
    • Two-head camera rotates 180∘180^\circ per head.
    • Triple-head camera rotates 120∘120^\circ per head.
  • Comparison with Other Modalities:
    • SPECT: Computerized emission tomography.
    • CT: Computerized transmission tomography.
    • PET: Emission tomography using annihilation photons from positrons.
  • Primary Advantages: Increased contrast resolution, added depth, size, and shape information, and 3D image representation.

SPECT Data Acquisition

  • Projections: Sequential 2D raw views used as intermediate steps to reconstruct slices (typically 64 stops, acquiring 128 2D raw images in a 128×128128 \times 128 matrix).
  • Timing and Orbits: Time per stop is constant for each stop. Acquisition uses circular or elliptical orbits across 180∘180^\circ (such as RAO to LPO) or 360∘360^\circ.
  • Acquisition Modes: Continuous acquisition, Step-and-Shoot acquisition, and Continuous Step-and-Shoot acquisition.
  • Gated SPECT: Synchronized acquisition based on the cardiac R wave.

Image Quality and Motion Detection

  • Motion Detection:
    • Identified when the same source is seen in two different places.
    • X-motion is easier to correct than Y-motion.
    • Detection techniques include Cine, Sinogram, and Summed images (see Powsner page 99).
  • Gating Errors: Irregular heart rates result in bad gating, identified using sinograms, count histograms, beat histograms, and accepted/rejected beat tallies.

Image Reconstruction Techniques

  • Reconstruction Types:
    • Filtered Back Projection (FBP): Most common reconstruction method during the 1980s and 1990s.
    • Iterative Reconstruction: Modern forefront technique used in SPECT and PET.
  • Back Projection: Smears counts along the radius/line of view into each pixel.
  • Star Artifact: Radial spokes/streaks extending from hot spots due to simple back projection, leading to image blurring.

Image Domains and Frequency Analysis

  • Spatial Domain: Data expressed as counts per pixel. Uses spatial filters like a 9-point smooth to redistribute counts to immediate neighbors via a filtering kernel (see Powsner pgs 154-157).
  • Frequency Domain: Data expressed as sine waves in \text{cycles/pixel}. Fourier Transformation converts data from the spatial domain to the frequency domain.
  • Nyquist Frequency: Set at 0.5 cycles/pixel0.5\,\text{cycles/pixel} (every other pixel is on), representing the highest useful imaging frequency. When expressed in cycles/cm\text{cycles/cm}, smaller pixels result in a larger Nyquist frequency.
  • Frequency Categories:
    • Low Frequency: Small pixel-to-pixel count differences (e.g., Gallium Scan).
    • High Frequency: Large pixel-to-pixel count differences with sharp edges (e.g., Bone Scan).

Image Components and Filtering

  • Raw SPECT Image Components:
    • Background: Low frequency; generated from back projection spokes.
    • Target Data: Desired image information; spans low to high frequencies depending on study type.
    • Noise: High frequency; unwanted statistical fluctuations inversely proportional to raw counts.
  • RAMP Filter: A high-pass filter that is always used in back projection to eliminate low-frequency background and reduce star artifacts.
  • User Filters: Low-pass filters applied to remove high-frequency noise without altering raw data (e.g., Hann, Hamming, Butterworth, Wiener, Parzen).