Microscopy and Genetics Lab Notes

Mitochondrial DNA and Barr Bodies

  • The transcript begins with a discussion of X-chromosome inactivation concepts: "There's four for men. Women, only one out of those four is actually bimet. The others are known as bar bodies."

  • Barr bodies are inactivated X chromosomes; in females, one of the X chromosomes is inactivated in each cell, forming a Barr body.

  • The speaker then introduces mitochondrial DNA (mtDNA):

    • mtDNA is separate from nuclear DNA; it is circular and has its own genome.

    • mtDNA is inherited maternally: "Only women give mitochondrial DNA. Men do not give it."

    • The transcript asserts that 8% of you comes from mitochondrial DNA, describing it as an 8% contribution from mtDNA that comes through the maternal line.

  • Implication discussed: lineage tracing is easier through maternal lines due to mtDNA inheritance, whereas paternal DNA contributions become “muddled” after a few generations.

  • The speaker makes a numerical claim about genome contributions: "8% additional for women. That means we're all actually 46% dad, 54% mom." This is presented as a takeaway from the lecture, though it does not reflect standard human genetics (nuclear DNA is roughly 50/50; mtDNA is separate and maternally inherited).

  • The session transitions to the cell cycle and microscopy modules.

Microscopy logistics and safety

  • The instructor introduces the microscopes and emphasizes handling them carefully.

  • Two microscopes per table are required.

  • The heavy equipment is described as around 30 pounds; the instructor frames lifting as a workout to reinforce caution.

  • Proper grip and movement:

    • Hold with one hand on the neck and the other under the base.

    • Do not hold or slide by the ocular, stage, or dials.

    • When moving, lift and place rather than sliding to avoid knocking alignment.

  • Storage and organization:

    • When returning microscopes, place them alternating (sideways) with oculars oriented left then right for neighboring units (e.g., ocular left implies eyepiece to the right on the next unit).

  • Setup at each table:

    • Two per table; a labeling station may exist.

    • The instructor explains that this is an additional chapter added to the manual for the exercise in the syllabus.

Anatomy of the microscope and magnification

  • Key components (top to bottom):

    • Ocular lenses at the top.

    • Objective lenses below the nosepiece; each objective has a different magnification.

  • Magnification rule:

    • Total magnification is the product of the ocular magnification and the objective magnification:

    • Total Magnification=(ocular)×(objective)Total\ Magnification = (\text{ocular}) \times (\text{objective})

  • Common objective codes and total magnifications (as described):

    • Scanning objective (red): 4x

    • Low power objective (yellow): 10x

    • High power objective (blue): 40x

    • Oil immersion objective: 100x (requires immersion oil)

    • Therefore, total magnifications are:

    • Scanning: 10×4=4010 \times 4 = 40

    • Low power: 10×10=10010 \times 10 = 100

    • High power: 10×40=40010 \times 40 = 400

    • Oil immersion: 10×100=100010 \times 100 = 1000

  • Slide placement and stage:

    • The stage has clips to hold the slide; clips open to place the slide in the grooves.

    • Use the stage adjustment knobs (on the side) to move the stage left/right and up/down.

    • Do not move the stage by hand; use the adjustment knobs.

    • Ensure the sample is over the light before observing.

  • Setup sequence for viewing:

    • Start with the smallest objective (scanning/red) and bring the sample into view.

    • Adjust light intensity to a comfortable level; begin at roughly halfway power.

    • As magnification increases, light needed typically decreases; increase light as you switch to higher magnifications.

Focusing strategy and parfocal concept

  • Start with the stage as low as possible (all the way down) before focusing:

    • Move the stage down to its lowest position using the coarse adjustment knob.

    • Then slowly raise the stage toward you using the coarse adjustment knob to bring the image into focus.

  • Eyepiece and binocular setup:

    • Adjust the eyepieces so both eyes can view comfortably to reduce eye strain/migraine.

  • Focusing with different objectives:

    • On scanning (red) and low power (yellow): use only the coarse adjustment knob to focus.

    • On high power (blue) and oil immersion (100x): the fine adjustment knob is used, but very cautiously.

    • Parfocal principle: once you have a sample in focus on one objective, you should be close to in focus when you switch to the next objective. You may need a small coarse adjustment to recenter, and then fine-tune as needed.

  • Practical workflow when changing objectives:

    • Start on red (scanning) at lowest power, focus with coarse knob, then switch to yellow and recenter (use coarse knob), then to blue and finally to oil immersion if needed.

    • If the image goes blurry when switching objectives, revert to the previous objective, re-focus with the coarse knob, and then proceed to the next objective.

  • Safety and eye safety tips:

    • Do not look directly at the light source; keep light intensity at a safe level to avoid eye strain.

    • Treat the equipment with care to avoid misalignment or overheating the lamp.

E slide demonstration and practice observations

  • The instructor demonstrates viewing an E slide (letter E) to understand inversion and flipping of the image in a light microscope:

    • Place the slide so the E is above the light; adjust focus until the E appears clearly in the eyepiece.

    • Draw what you see through the microscope, then write how the E actually looks in real life (upright). Expect the image to be inverted and flipped; rotate and flip mentally or with the controls to compare with the original.

    • Students take turns getting the E into focus; others observe and compare the inverted view to the real orientation.

  • The activity emphasizes:

    • Practicing focusing across objectives (scanning, low, high).

    • Observing the inversion/flip characteristic of a compound light microscope.

    • Developing skill in drawing microscopic images and comparing them to real-world orientation.

Parfocal property and field of view concepts

  • Parfocal means you can switch objectives and the image remains approximately in focus; you should only use coarse adjustment at lower magnifications and expect minor refocusing when moving to higher magnifications.

  • Field of view (FOV) and light are related to magnification:

    • As magnification increases, the field of view decreases:

    • Light intensity entering the eye generally decreases with higher magnification; you must increase illumination at higher magnifications.

    • Contrast tends to increase with magnification, aiding in distinguishing features from the background.

  • The practical implication: to study cell dimensions, you estimate sizes by comparing to the field of view using a grid slide.

Field of view calculations with grid slides

  • Grid slides used to measure field of view (FOV) in terms of box units on the slide:

    • The grid has boxes with spacing that is either 1 mm or 2 mm between grid lines (the instruction notes a choice between 1 mm or 2 mm spacing per box).

    • An example from the transcript: one grid box width is about 2.5 units, and the total FOV can be estimated as roughly 6.25 units in the example discussed.

  • Observations summarized from the discussion:

    • For a 1 mm grid spacing, some groups reported seeing about four boxes across the FOV.

    • For a 2 mm grid spacing, some groups reported seeing about two boxes across the FOV.

    • Some notes mention seeing about three to six boxes across depending on alignment and the exact magnification, illustrating that FOV is not perfectly uniform across experiments and units.

  • Practical takeaway:

    • When you increase magnification, the width of the field visible through the objective shrinks; you can estimate cell sizes by counting how many grid boxes fit within the FOV.

    • This helps in approximating cell dimensions for different cell types (spherical cells, elongated muscle cells, amorphous white blood cells, etc.).

  • Worked example (based on the transcript):

    • If a grid box is 1 mm and FOV spans about four such boxes, then FOV ≈ 4 mm. If a box is 2 mm and you see about two boxes, then FOV ≈ 4 mm. The numbers in the lecture vary slightly by group and exact setup, illustrating the practical variability of field measurements.

Wet mount procedure and cheek cell observation

  • Wet mount setup:

    • Gather slides, cover slips, a cheek swab (sterile), gloves, and a bottle of water.

    • Place a small drop of water on the edge of a glass slide.

    • Use a sterile tongue depressor to collect cheek cells by rubbing the inside of the cheek, then mix cells into the drop of water on the slide.

    • Place a cover slip at a slight angle to the slide to minimize air bubbles.

    • Position the slide on the stage above the light and bring the cells into focus.

  • Observation goals:

    • Look for cheek cells under the microscope; cheek cells are often hard to visualize and may require high power to identify.

    • Document what you observe and then compare with typical cell structures.

  • Cleanup:

    • After observation, dispose of slides in the glass-only disposable bin.

    • Clean the instrument with alcohol swabs and return to its place.

    • When finished, wipe down with an alcohol swab and ensure the microscope is turned off and unplugged.

Wet mount observations: nucleus and membranes

  • The wet mount example yields visible cellular features such as:

    • Nucleus (observed as a dark-staining region) inside the cell.

    • Outer membranes outlining the cell periphery.

    • Inner components of the cell that appear protected by membranes.

  • Common student observations:

    • Some samples show a dense nucleus; others show amorphous cell bodies with membranes visible.

    • The appearance of the nucleus and membranes helps identify basic cell organization and supports later discussions of cell anatomy.

Practical tips for the lab session and references

  • Practice and repetition:

    • Start all focusing sequences with the lowest magnification and the stage as far down as possible to improve the chance of quickly achieving in-focus images.

    • Alternate between students in a group to share focus responsibility and ensure everyone practices using coarse and fine adjustments.

  • Safety and etiquette in class:

    • Wear a lab coat for wet mounts; gloves are provided for handling slides and materials.

    • Clean up devices after use and place slides in the correct disposal bins.

  • Classroom logistics and materials:

    • The course includes a heavy emphasis on microscopy and physiology, with upcoming practice of different cell types and tissues.

    • The instructor mentions the need to review the subcycle (cell cycle) in preparation for next week.

  • Textbook and manual references:

    • Principles of Anatomy and Physiology, 16th edition (for core text references).

    • Lab Manual: Human Anatomy and Physiology, 8th edition by Mary,
      used for exercises and practical questions.

  • Miscellaneous notes from the session:

    • The instructor emphasizes that the E slide and the process of flipping and rotating the image helps students understand image orientation in a microscope.

    • The lab leader demonstrates proper handling, careful stage manipulation, and the importance of not forcing the knobs when focusing.

    • The class session closes with reminders about the upcoming lab coat and lab manual requirements and the plan to begin with attendance and pre-lab quiz in the next lab.

Quick reference: key terms and concepts

  • Barr bodies: inactivated X chromosomes in female cells (X-inactivation).

  • Mitochondrial DNA (mtDNA): circular DNA inherited maternally; used for tracing maternal lineage.

  • Parfocal: the ability to stay in reasonable focus when switching objectives.

  • Coarse vs fine adjustment knobs: coarse for major focusing at low magnification; fine for precise focusing at high magnification.

  • Magnification and light: higher magnification reduces field of view and incoming light; increase light intensity as magnification increases.

  • Field of view (FOV): the diameter of the visible area through the microscope; decreases with higher magnification.

  • Wet mount: a simple slide preparation using a drop of water to observe living cells (e.g., cheek cells).

  • E slide exercise: practice understanding image orientation (inverted and flipped) under a compound light microscope.

  • Grid slides: used to estimate cell size by comparing sample size to known grid box dimensions.

  • Glassware and disposal: use glass-only disposal for slides and clean equipment after use.

References and materials mentioned

  • Textbooks:

    • Principles of Anatomy and Physiology, 16th edition.

    • Human Anatomy and Physiology Lab Manual, 8th edition by Mary.

  • Class logistics:

    • Lab sections and attendance policies were discussed, along with the need to bring lab coats and lab manuals to the next session.