Review of Masking Rules
Bone Conductor:
Zero decibel minimum interaural attenuation (IA).
Headphones:
40 dB minimum IA.
Tube Phones:
50 or 70 dB minimum IA, depending on frequency.
Unmasked vs. Masked Audiograms
Given a masked audiogram, you should be able to:
Calculate backwards to determine the unmasked audiogram, considering minimum IAs.
Calculate backwards using different IAs provided.
Example Audiogram Review
Best hearing cochlea is the right cochlea (hearing at 0-5 dB range).
Unmasked thresholds would be at the same level as masked thresholds for the best hearing cochlea.
Left bone conduction thresholds would also be at the same level as the right due to zero dB IA for bone conduction.
Right ear air conduction thresholds are already unmasked.
Left ear air conduction thresholds require calculation to determine unmasked thresholds.
Calculating Unmasked Thresholds
Example: 500 Hz left ear air conduction.
Right cochlea hears at 0 dB.
Sound played into the left ear crosses over to the right ear, losing 40 dB (IA for headphones).
Unmasked left air conduction threshold would be at 40 dB.
Playing a 40 dB tone into the left ear results in 0 dB at the right cochlea, which it can hear.
Lowering to 30 dB results in -10 dB at the right cochlea, which it cannot hear.
Increasing to 35 dB results in -5 dB at the right cochlea, which it cannot hear.
Increasing to 40 dB results in 0 dB at the right cochlea, which it can now hear and respond to.
Masking shifts the threshold to reveal the left ear's true hearing.
4000 Hz Exception
Right ear bone conduction hears at 0 dB.
Left ear truly hears at 35 dB.
Playing a 40 dB tone into the left ear technically both ears should be able to hear it.
At 30 dB, neither ear can hear (left ear can't hear 30 dB, right ear can't hear -10 dB).
At 35 dB, the left ear can hear it.
The better threshold will be the unmasked threshold.
Creating Masking Tables
Goal: Write masking tables to get from unmasked to masked audiogram as you would clinically.
Key Questions for Masking Tables
Without masking, where would I predict the left air conduction threshold to be?
Assume minimum interaural attenuation (IA).
At what level are we going to start presenting the masking in the right side?
Base masking prediction on the right air conduction threshold.
Add 10 dB to the right ear air conduction threshold to take that ear out of the equation.
Where is the true threshold of the ear that I'm masking there?
This is where the plateau will be.
Masking Table Example: 1 kHz Left Air Conduction
Unmasked threshold: 45 dB (right bone conduction threshold + IA).
Masking level: Start masking at 40 dB (right ear air conduction threshold + 10 dB).
True threshold: 70 dB (from masked audiogram).
Masking table will show under-masking, with a 10-for-10 shift until reaching 70 dB.
Plateau
Plateau between 50 dB, 60 dB, and 70 dB of masking.
Increase masking by 20 dB, but threshold shifts by no more than 5 dB.
This is acceptable as a plateau.
Masking Table Example: Bone Conduction Threshold
Without masking, bone conduction should be equal to the best ear bone conduction (5 dB).
Masking level starts at 40 dB (same as air conduction example).
The ear receiving masking is based on the air conduction threshold.
True threshold of the left bone conduction: 20 dB.
Masking with 10 dB effective masking or 40 dB of masking overall causes the right ear to shift by 10 dB.
Increasing to 50 dB masking elicits a 20 dB response truly coming from the left ear.
Central Masking
Central masking can cause a 5 dB shift in plateau even without true hearing loss.
The sound in the other ear is causing the response to be 5 dB poorer.
With air conduction, a 5 dB shift is still accepted as a threshold.
Air Conduction Example
If masking air conduction at 35 dB, put 60 dB of masking into the other ear to start the plateau.
Increase masking from 60 to 70 to 80, a 20 dB increase.
If the response shifts only 5 dB, it's a plateau.
Special Air Conduction Case:
If all 3 responses are only 5 dB away from the unmasked, assume the 5 dB shift has been because of central masking.
Example: If the best bone condiction (unmasked) is at 20dB, and the AC responses are at 25 then record the final number as 20dB
Bone Conduction
Because bone conduction is calibrated for central masking, take the poorer threshold.
Questions
What if the responses are 20, 25, and 30dB?
Central pattern.
Repeat the test as this is most often a confusion.
Legit cases are rare.
Order of Testing
Standard Practice
Best Ear Air Conduction, Poorest Ear Air Conduction, Poorest Ear Bone Conduction, Best Ear Bone Conduction
PTA starts with the better hearing ear.
Then move to the poorer hearing ear with air conduction.
Reason is due to masking being administered as the test occurs
If a difference appears, mask them immediately
*You Mask Thresholds Immediately for Efficiency!
*This also allows you to not reset equipment, as you only need to turn on masking.
Always remember that equipment is not voodoo.
Question to Patient?
Do you feel like one ear hears better than the other?
Used so that you can mask the better ear.
How AC Mask with Bone?*
Mask the poorer ear with air. You compare to AC to BC, but if the BC doesn't exist you assume that it will be better or the same AC.
If it 40db or more, mask immediately.
*There is the last final step!
Final Step- check to see if there has been missed AC masking. Compare previous AC test with current scores. If there has been 40db missed there, do a test.
There may be a huge gap between 3 tested. If you missed any, make sure they are properly included.
Order of Testing Example
Right Ear tested first, then left.
If there is 40db difference. mask the poorer ear.
15 DB is a key number!
As if there is 15db or more between BC and AC, then test bone immediately.
If they are off, repeat the test!
There must be a real reason as to why any of the equipment scores are off.
General Testing Overview!
Air tested on the richer ear, poorer ear. Next BC test on the poorer then richer ear.*
Only test BC where HL exists in the AC
Last step is to recheck past work! Make sure that with current information you have no gaps in tests.
Testing Occluded
Testing occluded occurs when the BC test is applied under the head set.
This only occurs when the professional thinks that the patient may need masking.
Can be identified as it is more effient to apply it during the initial test than later.*
*BC Interaural Attenuation- 0 *
*Criterion- 15
*The combination of these two principles gives professionals the knowledge to mask patients BCHL*
Testing when is there BC Potential?
Potential of 10-15, professionals expect it with tests.
They will make sure that they will initially do testing to try to identify if there is BC HL early in the test.
Tests always change. Never assume everything is good and ready.
BC must be checked during all stages of tests!
Does the BC have HL?- If yes then setup occluded! If not and you are wrong, you will need to retest. No biggie.
What if HL occurs?*
If BC tests as 0 or is fine, we test that in all parts.
However, BC tests is where HL occurs and AC is still fine you would need to test that immediately!*
Do patients know what they are talking about?
Patients might just say things. These might be false or true, just continue to test them
As professionals are trying to give them proper care!*