QA/QC Notes

Quality Assurance (QA) and Quality Control (QC)

Quality Assurance (QA)

  • Deals with patients and services, focusing on the patient's overall experience.

  • Example: A patient’s experience from start to finish.

Continuous Quality Improvement (CQI)

  • Focuses on improving processes within which employees function as team members for patient care.

  • All employees must be involved in the improvement process.

  • Example: How a patient’s experience was from the beginning to the end of their hospital stay.

The Joint Commission (TJC)

  • Ensures high-quality services and care in healthcare organizations.

  • TJC requires QA/QC programs.

Quality Control (QC) Overview

  • Product-oriented, focusing on defect identification.

  • Reactive approach using corrective tools.

  • Specific team responsibility.

  • Monitors radiology instrumentation, equipment, and artifacts; ensures radiographic procedures are performed safely to provide the highest quality images.

  • Each new piece of equipment should be tested prior to use.

  • Required by law to maintain license for room/department

Process of a QC Program:
  1. Acceptance testing

  2. Performance monitoring

  3. Error maintenance

  • Annual evaluations are required.

Acceptance Testing
  • Performed before machine installation or major repairs.

  • Determines if equipment is working properly.

Performance Maintenance
  • Tests performed daily, monthly, quarterly, and annually to ensure equipment functions properly and to avoid failure.

Error Maintenance
  • Occurs due to poor equipment performance.

  • Corrective action must be taken.

QC Documentation
  • Consistent and accurate records are essential.

  • Unacceptable results must be documented with a corrective action plan.

Physicist’s Annual Testing (Partial List)

  • Noise, Spatial Resolution, Contrast Resolution

  • Dose Response/Signal, Dynamic Range (Exposure Linearity)

  • Detector & Dose Calibration, AEC & kVp Testing

  • Plate Reader Calibration, Residual Signal Erase, Image Artifacts

  • Repeat Analysis, Uniformity Test

  • Beam Filtration, Collimation

  • Monitor Display Function, Luminance & Reflection Testing

Workstations

  • Require regular QC testing.

  • Radiologist's viewing stations have higher quality than technologist's.

  • Assessment for Display Performance for Medical Imaging Systems: document for daily, monthly/quarterly tests.

SMPTE Test

  • Used to determine if contrast/brightness settings on monitor are acceptable.

  • Observation of 5% and 95% luminance patches to identify deviations in luminance adjustments.

Workstation Calibration – AAPM TG 18 Test

  • Commonly called TG18-QC.

  • Pattern used for overall assessment of a monitor’s adjustment & calibration.

Photometer

  • Tool to measure light intensity (luminescence) on a workstation monitor.

Elements of Radiographic Systems QC Program

  • Filtration, Collimation, Focal spot size

  • Spatial Resolution

  • PSP Imaging Plate Uniformity & Erasure Thoroughness

  • kVp calibration, Exposure timer accuracy

  • Exposure linearity & reproducibility

Filtration
  • State requires minimum total filtration of 2.52.5 mm Al.

  • Measured by checking HVL, annually or with tube/housing changes.

Collimation/Beam Restriction Test
  • Beam alignment check relative to the image receptor (IR).

  • PBL (positive beam limitation) automatically collimates to film size.

Focal Spot
  • Small focal spot: needed when spatial resolution is needed.

  • Large focal spot: short exposure time priority, more photons emitted.

Spatial Resolution
  • Ability to render small objects on an image.

  • Smaller pixel size = better spatial resolution.

PSP Uniformity Test
  • Assesses uniformity of signal from exposed imaging plate (IP).

  • Detects artifacts; requires regular cleaning of cassettes/IP.

PSP Imaging Plate Erasure Thoroughness
  • Ensures minimal residual signal (ghosting).

  • White light is used to delete trapped energy from IP.

AEC
  • Method for setting exposure factors to ensure a quality radiographic image.

Exposure Timer Accuracy
  • Exposure Linearity: Constant radiation output for various mA and exposure time combinations. Example: If 70kVp70kVp and 50mA50 mA, 11 sec. (50mAs50mAs) produces 50mR50mR, then 70kVp70kVp and 100mA100 mA, 0.50.5 sec (50mAs50 mAs) should produce 50mR50mR.

  • Exposure Reproducibility: Tests kVp, mA, and time in successive exposures.

Protective Apparel Test
  • Aprons, gonadal shields, gloves checked for defects under fluoroscopy.

Recognition & Reporting Malfunctions
  • Technologist's responsibility to report malfunctions.

  • Equipment failure requires repair and physicist recalibration.

Fluoro QC

  • Fluoro exams = high pt. dose

  • Entrance skin exposure (ESE) averages fluoro exam is 33 to 55 rads/min

Exposure Rate
  • State law: ESE shall not exceed 10R/min10R/min = normal fluoro.

  • ESE allowed up to 20R/min20R/min = interventional procedures.