QA/QC Notes
Quality Assurance (QA) and Quality Control (QC)
Quality Assurance (QA)
Deals with patients and services, focusing on the patient's overall experience.
Example: A patient’s experience from start to finish.
Continuous Quality Improvement (CQI)
Focuses on improving processes within which employees function as team members for patient care.
All employees must be involved in the improvement process.
Example: How a patient’s experience was from the beginning to the end of their hospital stay.
The Joint Commission (TJC)
Ensures high-quality services and care in healthcare organizations.
TJC requires QA/QC programs.
Quality Control (QC) Overview
Product-oriented, focusing on defect identification.
Reactive approach using corrective tools.
Specific team responsibility.
Monitors radiology instrumentation, equipment, and artifacts; ensures radiographic procedures are performed safely to provide the highest quality images.
Each new piece of equipment should be tested prior to use.
Required by law to maintain license for room/department
Process of a QC Program:
Acceptance testing
Performance monitoring
Error maintenance
Annual evaluations are required.
Acceptance Testing
Performed before machine installation or major repairs.
Determines if equipment is working properly.
Performance Maintenance
Tests performed daily, monthly, quarterly, and annually to ensure equipment functions properly and to avoid failure.
Error Maintenance
Occurs due to poor equipment performance.
Corrective action must be taken.
QC Documentation
Consistent and accurate records are essential.
Unacceptable results must be documented with a corrective action plan.
Physicist’s Annual Testing (Partial List)
Noise, Spatial Resolution, Contrast Resolution
Dose Response/Signal, Dynamic Range (Exposure Linearity)
Detector & Dose Calibration, AEC & kVp Testing
Plate Reader Calibration, Residual Signal Erase, Image Artifacts
Repeat Analysis, Uniformity Test
Beam Filtration, Collimation
Monitor Display Function, Luminance & Reflection Testing
Workstations
Require regular QC testing.
Radiologist's viewing stations have higher quality than technologist's.
Assessment for Display Performance for Medical Imaging Systems: document for daily, monthly/quarterly tests.
SMPTE Test
Used to determine if contrast/brightness settings on monitor are acceptable.
Observation of 5% and 95% luminance patches to identify deviations in luminance adjustments.
Workstation Calibration – AAPM TG 18 Test
Commonly called TG18-QC.
Pattern used for overall assessment of a monitor’s adjustment & calibration.
Photometer
Tool to measure light intensity (luminescence) on a workstation monitor.
Elements of Radiographic Systems QC Program
Filtration, Collimation, Focal spot size
Spatial Resolution
PSP Imaging Plate Uniformity & Erasure Thoroughness
kVp calibration, Exposure timer accuracy
Exposure linearity & reproducibility
Filtration
State requires minimum total filtration of mm Al.
Measured by checking HVL, annually or with tube/housing changes.
Collimation/Beam Restriction Test
Beam alignment check relative to the image receptor (IR).
PBL (positive beam limitation) automatically collimates to film size.
Focal Spot
Small focal spot: needed when spatial resolution is needed.
Large focal spot: short exposure time priority, more photons emitted.
Spatial Resolution
Ability to render small objects on an image.
Smaller pixel size = better spatial resolution.
PSP Uniformity Test
Assesses uniformity of signal from exposed imaging plate (IP).
Detects artifacts; requires regular cleaning of cassettes/IP.
PSP Imaging Plate Erasure Thoroughness
Ensures minimal residual signal (ghosting).
White light is used to delete trapped energy from IP.
AEC
Method for setting exposure factors to ensure a quality radiographic image.
Exposure Timer Accuracy
Exposure Linearity: Constant radiation output for various mA and exposure time combinations. Example: If and , sec. () produces , then and , sec () should produce .
Exposure Reproducibility: Tests kVp, mA, and time in successive exposures.
Protective Apparel Test
Aprons, gonadal shields, gloves checked for defects under fluoroscopy.
Recognition & Reporting Malfunctions
Technologist's responsibility to report malfunctions.
Equipment failure requires repair and physicist recalibration.
Fluoro QC
Fluoro exams = high pt. dose
Entrance skin exposure (ESE) averages fluoro exam is to rads/min
Exposure Rate
State law: ESE shall not exceed = normal fluoro.
ESE allowed up to = interventional procedures.