LS_CHT_BIA_L36_Maintenance in Thermal Analysis
Lesson 36: Maintenance in Thermal Analysis
Lesson Objectives
Cleaning of detector
Maintenance of DSC system
Standard calibration procedure for DSC system
Cleaning of Detector
Purpose: Remove substances formed during analysis.
Inspection: Check sample/reference pans, lids, and crucibles for contaminants before and after each use.
Cleaning function in TA-60W software:
Increases temperature to 600°C immediately and holds it constant for the set time.
Cleaning Procedure
Ensure no sample runs remain on the detector.
Click {Detector}-{Cleaning Run…}
Enter {Hold Time(min)}, select gas type, and press {Start}.
Cleaning gas example: Air at 200-300 ml/min.
Maintenance
Hardware Maintenance:
Clean instrument’s exterior with a cloth soaked in ethanol.
Long Storage:
Disconnect gas pipelines and close respective ports.
Close furnace lid and cover; ensure no items on detector plate.
Store away from sunlight, dust, and vibration.
Vacuum System:
Ensure functioning properly; check for leaks and vacuum pump condition.
Performance Maintenance
Check performance by measuring melting temperature and heat of fusion of standard material (e.g., Indium).
Heat indium sample from ambient to 180°C at 10°C/min:
Accepted range: Tm = 156.5 ± 1.0°C, Heat of fusion = 27 ~ 30 J/g.
Perform calibration if results are outside the accepted range.
Recommended checks: once a month or preferably weekly.
Calibration Procedure
Enter TA-60 collection program.
Ensure DSC-60 is connected.
Click DETECTOR menu and select TEMPERATURE CALIBRATION.
Click CALCULATE.
Set Expected & Measured values for:
Indium: 156.63°C
Zinc: 419.58°C
Click OK.
Additional Calibration Steps
Set Coefficient A & B to 0.00000.
Click DOWNLOAD, then YES, and EXIT.
Click DETECTOR menu and select HEAT CALIBRATION.
Set Expected & Measured values to:
28.45 J/g
Click OK, DOWNLOAD, then YES, then EXIT.
Sample Run Preparation
Place Indium sample in DSC-60, fill in:
Sample Name: Indium
Sample Weight: 5.5 mg
Pan: Aluminium
Atmosphere: Nitrogen
Rate flow: 50 ml/min
Program for Indium:
Hold Temp: 200.0°C, Rate: 10.00°C/min.
Analyze using DSC Peak analysis after the run.
Repeat for Zinc sample:
Sample Weight: 3.56 mg
Hold Temp: 500.0°C, Rate: 10.00°C/min.
Inspect and Clean Gas Flow Path
Check gas flow path and connections; clean/replace as necessary.
Inspect electrical connections for damage or loose fittings.
Check seals and gaskets for wear; replace as needed.
Replace gas filters as per manufacturer recommendations.
Troubleshooting Symptoms
Abnormal Temperature Indication:
Replace defective cold junction element.
Ensure correct connections for detector thermocouples.
Temperature Doesn’t Rise:
Check for blown/missing fuse.
Ensure proper connection at heater terminal.
Temperature Deviates:
Change control constants for stability.
Address issues with wiring and power supply.
References
Principles of Instrumental Analysis (5th Edition) Skoog, Holmer, Nieman.