Week 4: Transformation Results and Spectrophotometry

Transformation Results

if you were doing all parts of the experiment yourself, you would pick a white colony from the insert plate to grow in liquid culture and then measure in the spent media

you will be creating a table of your personal colony counts and calculating transformation and ligation efficiency during lab

  • this is done for you behind the scenes

resulting isolate plasmid will be used for plasmid mapping

Transformation Results OLM

  • we performed a restriction digest and ligation in lab 1 to get insert into our plasmid

  • transformed the recombinant plasmid into E.coli JM101 so the bacteria could express our insert

  • the transformation was plated and incubated overnight at 37 degrees C

  • plates kept at 4 degrees C until lab

Technical vs. Experimental Controls

technical: controls that are specific to the technique (i.e. restriction digest, ligation)

  • ensure that the technique is working as it should

  • getting expected results; no interference giving us something we shouldn’t be getting

experimental: control that represents the “normal” condition (if trying something new for e.g.)

  • what you compare your experimental sample to

  • can be considered one of the experimental conditions

    • if certain concentration makes a technique better/improves disease outcome — different concentrations would be different concentrations & no substance would be control (and also one of the conditions)

  • not all experiments have an experimental control

    • which plasmid is better at transforming — no normal plasmid (no experimental control)

all of our plates were either a sample of interest or a technical control

  1. sample of interest: “insert” plate

    • plate that we’re actually going to work with

      • will do downstream applications with

    • not a control — one we’re interested in

    • we are not comparing this sample to anything else, but if we were, it would be the experimental sample

      • would be experimental sample if comparing the effectiveness of different transformation protocols for e.g.

    • white colonies contain the insert (looking for the white colonies)

      • operon disrupted by insert (contain the insert)

      • will see a mix of white and blue colonies

  2. ligation technical control: “H2O” plate

    • water added instead of insert

      • must keep volume equal

      • always want to replace substance with neutral one

      • did not just add “nothing”

    • expect no ligation of insert into MCS so all blue colonies

    • showing that the only wat to get white colonies is if there is an insert present

  3. transformation technical control: uncut plasmid added to E.coli

    • expect transformation to happen into competent E.coli

    • expect blue colonies — no insert

    • makes sure that the bacteria are able to be transformed by this particular plasmid

  4. transformation technical control (another): TE buffer tube (not diluted)

    • TE buffer = liquid that the plasmid is floating in when given to you

      • by adding just TE, no plasmid was added to the E.coli during transformation

      • adding TE but NO PLASMID (E.coli don’t get plasmid during transformation

      • plasmid to be digested was floating in TE buffer

    • plated on LB + ampicillin plate

    • expect no colonies — don’t have gene to be able to grow in presence of ampicillin

    • makes sure we don’t get growth if we aren’t supposed to get growth

      • not contaminated

      • not mixing up samples

  5. transformation technical control (and another!): TE buffer (no plasmid) added to E.coli but plated on LB only plates

    • no ampicillin on plate

    • expect growth

    • makes sure that the E.coli are alive and did not get killed during transformation protocol

    • dilution because we want to be able to see individual colonies

      • not too many bacteria

      • if did not dilute, would get a lawn of bacteria = confluent growth

    • diluted to get countable number of colonies

Controls MUST show Predicted Results

  • if ANY of your technical controls do not show what is expected, you cannot trust your samples

  • you cannot analyze data or draw any conclusions

  • you cannot use sample of interest in downstream applications

What if plasmid does not ligate to insert or to self?

  • plasmid not ligated = linear

  • when linear plasmid enters bacteria, exonuclease will degrade linear DNA from end (no longer in plasmid)

In lab:

  • you will be counting the colonies and creating a table of results

  • table should include:

    • sample name (i.e. insert, H2O, TE, etc.)

    • plate type (LB or LB + amp — include amp concentration 100 ug/mL)

    • number of white colonies

    • number of blue colonies

  • you will then calculate your transformation and ligation efficiency

    • you will present and discuss this in your lab report

  • include the transformation and ligation efficiency in Benchling

if transformation did not work, present that in lab report with explanation of what went wrong (no marks lost)

Transformation Efficiency (TE) Calculation

TE = (# white + blue colonies on insert plate) / (total colonies LB plate)

= 12 / 85 × 10 ^6

= 12 / 8.5 × 10^7

= 1.41 × 10^-7

transformation efficiency — no units! not percentage! just ratio!

  • # white + blue colonies on insert plate = number of bacteria/colonies that were transformed with plasmid

    • e.g. 5 white & 7 blue

  • total colonies on LB plate (TE diluted) = total number of bacteria we started with

    • e.g. 85 white & 0 blue

    • must take dilution into account

    • multiply by 10 ^ +6

  • only concerned with number of colonies for Insert plate and diluted TE plate

Ligation Efficiency Calculation

LE = (# white on insert) / (all colonies - white + blue)

= 5 / 12

= 0.42

ligation efficiency — no units! not percentage! just ratio!

  • only concerned with insert plate (only plate with insert — only one that underwent ligation)

  • proportion of plasmids with insert ligated / plasmids that were transformed into bacteria

  • # white colonies in insert = 5

  • # blue colonies in insert = 7

Why Do My Colonies Look Funny? Satellite Colonies

might see that the colonies look funny — small dots around colonies

satellite colonies: commonly seen with ampicillin resistance

  • only large colony is “real”

  • occur because of the way the ampicillin resistance is created

    • transformed bacteria with plasmid with amp resistance cassette

  • satellite colonies are not considered real colonies — do not include in counts!

  • occur because of the way the ampicillin resistance is created

how does ampicillin work

  • ampicillin is bactericidal: kills bacteria

  • inactivates transpeptidases - cross link peptoglycans (in cell wall)

  • disrupts cell wall synthesis in actively growing cells leading to lysis

    • any new bacteria that are being formed cannot create a structural cell wall

    • end up lysing

    • they die

Ampicillin Resistance Gene: codes for Beta-Lactamase

  • cleaves a structural ring in ampicillin molecule itself

    • ampicillin loses its function — cell walls will be able to be formed in newly formed bacteria

  • results in loss of function of ampicillin

  • secreted by bacteria

    • can destroy the ampicillin surrounding the colony

    • allows non-transformed colonies (satellite) to form

    • leaves the bacteria — destroys ampicillin surrounding colony of bacteria

      • non-transformed colonies (satellite) form

white fog: high number of satellite colonies

  • big juicy white colony in middle surrounded by fog

  • sometimes there are fewer satellite colonies — could count them but DONT

commonly see satellite colonies in research

Transformation Efficiency

  • number of colonies on insert plate / number of colonies on DILUTED TE plate

  • # on insert plate is ANY colour (blue and white)

  • # on diluted TE plate is colonies on LB plate

    • need to account for dilution factor (10^-6)

    • multiple # of colonies by 10^6

  • record this number in your lab book

even if we overcome barriers, transformation is still a rare event

Ligation Efficiency

  • use insert plate only

  • # white colonies / (# blue + white colonies)

  • record this number in your lab book

use personal numbers (from lab) in lab report

Spectrophotometry

1: Principles of Spectrophotometry

measure the concentration of something in a sample (e.g. concentration of insulin)

absorption spectroscopy: using light to probe matter

  • how a sample absorbs light

  • lambda = wavelength (correspond to different colours)

  • molecule will absorb wavelength strongly (others won’t)

  • linear?

  • P sub n = incoming power

    • high intensity, lots of photons

  • material will absorb a lot of light

    • small amount of power out

  • lower concentration: more outcoming power (absorbs less light)

    • ratio of output power to input power means something

    • % transmittance: output light/input light x 100%

  • spectrophotometer: measures power in and out (specifically for visible light spectrum)

  • absorbance: A (unitless quantity) = 2 - log(%T)

  • A = 2 - log(50) = 0.3

    • for 50% transmittance?? absorbance?

  • use wavelength that is absorbed by molecules but not others

  • measure absorbance spectrum (A vs. wavelength)

    • when absorbance is high, a lot of light is absorbed

  • peak of graph is the lambda max (short wavelength for this example); means a lot of light is being absorbed by the molecules in the sample

    • use this for spectroscopy experiment

    • typically choose this one; most bang for buck

  • a high concentration of a substance that absorbs a given wavelength will also have a high transmittance

2: Spectronic 20

A) How does a Spectronic 20 work?

spectrophotometer: measures the amount of light that gets transmitted

source → monochromator → sample → detector

  • source = tungsten filament light bulb — heats up & gives off white light

  • white light goes through slit (narrow band) and pass through prism or other devices

    • separate all different colours of light

  • can rotate prism to sweep different colours of light back and forth over exit slip — select specific wavelength

  • shines that wavelength through sample

    • some light get’s absorbed, some gets transmitted

  • detector converts the amount of light intensity into electrical signal that we can then amplify and measure

nanometers (nm) for visible light

B) How to Operate a Spectronic 20

how to generate an absorption spectra using a dye called DCP

this is a basic spectrophotometer with analog dials and manual setting of 0% transmittance and blanking

  • Spec 20s accept glass test tubes in the sample chamber

    • requires a high volume of sample

  • more advanced spectrophotometers: use same principles

    • automatically set 0% transmittance

    • automatically blanks

    • uses cuvettes (much smaller sample volume)

      • 100 uL — 2 mL

    • a bit faster

  • light that emits white light and bounces off diffracting gradient then deflected back at sample

    • wavelength camera measures wavelength at sample

  • can pass different energies of light through sample and measure absorbance

    • make absorption curve

  • use left on to turn on (clockwise) — 15 minutes on so lamp can warm up

  • use left knob to bring red meter to 0 percent transmittance

    • 2 scales: top is percent transmittances, bottom is absorbance

    • top: zero

    • bottom: measurement

  • 0% — mirror: can’t see the reflection in the mirror = correct level

  • blank should contain everything except the substance you are trying to measure

    • we are measuring 2 dyes dissolved in DI (distilled) water so the blank is DI water with no dye

  • right knob to 100% transmittance (no mirroring of dial)

  • told the spectrometer what nothing is and what everything is

  • filter at bottom: gives range

    • make sure it’s in the proper range (flips back and forth)

    • wouldn’t be able to zero spectrophotometer

  • take out DI water blank and create absorption spectrum

  • put in red dye: will flip to wherever (0.084)

    • if you can see the meter in the mirror, you are NOT perfectly lined up to read the value

  • three decimal places

  • next measurement: 0.053

  • repeat by changing the wavelength,

  • must re-blank every time you change the wavelength (must zero and do 100% transmittance)

  • greatest reading for absorbance

    • taking measurements at 20 nm intervals (between 360 and 600)

    • depending on where absorption is highest, is the lambda max

  • use lambda max for standard curve

3: Microplate Reader

spectronic 20: glass tubes and requires large sample volume of 3 mL — limited to 1 sample at a time

digital spectrophotometer: use glass/disposable plastic cuvettes

  • sample volume of 1 mL - 100 uL volume, 1 sample at a time

both read in a horizontal manner; light path is horizontal

microplate reader:

  • put samples on microplate

  • microplate is loaded into drawer

  • drawer will close and load the plate into the machine

  • each well is positioned under/over a lamp one-by-one automatically by the machine to acquire reading

    • can be read from top or bottom (read vertically)

  • 96 well plates are common → 96 samples!

  • clear, white, or black

  • small volume of 100-200uL

Modes of Operation (most microplate readers have 3)

Absorbance: spectrophotometer

  • microplate reader will work in the same way as a spectrophotometer

  • for multiple samples at the same time

  • light from lamp → filter → collect wavelength of light → through each sample → transmittance measured and converted to absorbance reading

Fluorescence:

  • light from lamp → filter (select wavelength) → hit each well & sites fluorescence tag that is in each sample

  • fluorescent tag will give off different wavelength of light → filter & select for detection of that particular wavelength

    • detected by P & T photomultiplier tube

    • converted into numeric reading

Luminescence:

  • light that is naturally given off by sample (bioluminescence) passes through filter and is detected by P & T (photomultiplier) tube and converted into numeric reading

Applications: Quantification

  • create a standard curve just as with a spectrophotometer (serial dilution)

  • all standards and unknowns (samples) are loaded on the same plate

  • need to include each standard and sample should be loaded in triplicate

    • technical replicate

      • measure of precision

    • different from biological replicate

  • use the average of the triplicates to create standard curve and interpolate sample concentration

  • difference: rather than reading each standard & sample one at a time — all of the standards and unknowns on one plate

  • when using instrument, must be loaded in triplicates

  • triplicates = technical replicate (3 different readings of same sample)

    • measure of precision of instrument

  • very different from biological replicate: measuring distinct samples

  • when analyzes, use average of triplicates and determine sample concentration of average from each sample

Applications: Colorimetric Assays

  • similar to standard curve but is a relative quantification

    • do not have standards

  • measure things like: cell viability, cell proliferation, cytotoxicity

  • example is MTT assay

  • quantification application can also often be colorimetric

  • colorimetric = change in colour

4: Standard Curve

overview of the methodology needed to create a standard curve

  • will be creating the standard curve and measuring the same standards and samples with 2 methods: Spec 20 & Plate Reader

  • you will be measuring the amount of insulin in the spent media of our transformed E.coli JM101

  • Does having the INSR insert change the amount of insulin found in the media?

using both the spectrophotometer and plate reader to make the standard curve

standard curves: very commonly used technique to quantify (concentration etc.) an unknown sample

  • start with sample of unknown concentration + standard with known concentration

  • start by making a dilution series of your standard — serial dilution method

    • serial dilution of known standard

  • transfer the dilution series to a cuvette or test tube and measure the absorbance of each standard

    • transfer to spec tube (or cuvette) or the microplate

    • measure absorbance of each standard

  • take standards and absorbances and graph them

    • absorbance vs. concentration

  • create line of best fit — give an equation of line — determine where sample falls in the standard curve

  • if the sample is outside the range of the standard curve

    • if unknown is below or above — can’t use standard curve

    • if in the x range, can interpolate

    • cannot extrapolate, can only interpolate

  • R² value will give indication of how accurate line is

    • R² is 0.93 — not the best but not bad

  • evaluating whether you can trust

Our Experiment

  • we are examining the concentration of a substance of interest found in spent media

  • spent media: media that E.coli have been grown in for 24 hours

    • can measure amount of substance depleted from media or amount of substance secreted by bacteria into the media

    • which one you are measuring depends on class experiment (OWL)

transformed E.coli JM101 with plasmid with INSR insert put into it

  • white colony will be picked (has insert) and grown in liquid culture

  • liquid = media

colorimetric assay to measure concentration of substance of interest

colorimetric assay: just means there is a colour change

  • lambda max cannot be directly measured by spectrophotometer

  • increase of colour means increase in substance of interest

  • specific assay we’re using (not really the colorimetric)

    • Labrador retriever is a dog (dog = colorimetric)

details of the assay (name, manufacturer, lambda max)

standard curve was created with the standard provided in the kit

  • create serial dilution with knowns

  • have unknown samples here

  • one reagent in kit is a colour reagent — add it to everything (unknown and standards)

    • solutions will turn a colour depending on concentration that’s in there

    • we are indirectly measuring the substance of interest

  • will be creating a standard curve which will be measured using both the Spec 20 and the plate reader

    • one set of serial dilutions, one set of unknown samples

    • use them both in the Spec 20 and plate reader

  • use the standard curves to calculate the concentration of our substance of interest in the spent media

  • info will be included in your lab report

USE THE EXCEL SPREADSHEET TO GENERATE AN EQUATION OF THE LINE

  • check with answer spreadsheet