Digital Display Device QA/QC Notes
Quality Assurance (QA) Program
- Deals with patient services, focusing on the patient's experience from start to finish.
Continuous Quality Improvement (CQI)
- Improves processes within which employees function as a team for patient care.
- Requires involvement of all employees.
The Joint Commission (TJC)
- Ensures high-quality services and care.
- Requires QA/QC programs.
Quality Control (QC)
- Product-oriented, focusing on defect identification.
- Reactive approach using corrective tools.
- Specific team responsibility.
- Monitors radiology instrumentation, equipment, and artifacts.
- Ensures safe radiographic procedures and high-quality images.
- Equipment must be tested before use.
- Required by law to maintain room/department license.
Process of a QC Program
- Acceptance testing.
- Performance monitoring.
- Error maintenance.
- Annual evaluations are required.
Acceptance Testing
- Performed before machine installation or major repairs.
- Determines if equipment is working properly.
Performance Maintenance (PM)
- Tests performed daily, monthly, quarterly, and annually to assure proper equipment function.
Error Maintenance
- Occurs due to poor equipment performance.
- Requires corrective action.
QC Documentation
- Consistent and accurate records are important.
- Unacceptable results must be documented with an action plan.
Physicist’s Annual Testing
- Includes noise, spatial resolution, contrast resolution, dose response/signal dynamic range, detector calibration, dose calibration, AEC testing, kVp calibration, plate reader calibration, residual signal erase, image artifacts, repeat analysis, uniformity test, beam filtration, collimation, monitor display function, luminance & reflection testing.
Workstations
- Require regular QC testing.
- Radiologist’s viewing stations are higher quality.
- Assessment for Display Performance for Medical Imaging Systems is completed by physicists and technologists for tests.
SMPTE Test
- Used to determine if contrast/brightness settings on monitor are acceptable.
- Observes 5% and 95% luminance patches.
Workstation Calibration – AAPM TG 18 Test
- Commonly called TG18-QC.
- Provides overall assessment of a monitor’s adjustment & calibration.
Photometer
- Measures light intensity (luminescence) on a workstation monitor.
Elements of a QC Program for Radiographic Systems
- Filtration, collimation, focal spot size, spatial resolution, PSP Imaging Plate Uniformity Test, PSP Imaging Plate Erasure Thoroughness, kVp calibration, exposure timer accuracy, exposure linearity, exposure reproducibility.
Filtration
- State requires minimum total filtration of 2.5 mm Al.
- Measured by checking HVL.
- Checked annually or when tube/housing is changed.
Collimation/Beam Restriction Test
- Checks beam alignment.
- Most systems have PBL collimators.
Focal Spot
- Important for image quality.
- Small focal spot is used when spatial resolution is needed.
- Large focal spot is used when a short exposure time is priority plus more photons emitted.
Spatial Resolution
- Ability to render small objects on an image.
- Smaller pixel size = better spatial resolution.
- Assesses uniformity of signal from exposed IP.
- Assists in detecting artifacts.
- Cassettes/IP must be cleaned regularly.
PSP Imaging Plate Erasure Thoroughness
- Ensures minimal residual signal (ghosting).
- White light is used to delete trapped energy from IP.
AEC
- Sets exposure factors to ensure a quality radiographic image.
Exposure Timer Accuracy
Exposure Linearity
- Ability of x-ray unit to produce constant radiation output for various mA and time combinations.
Exposure Reproducibility
- Tests kVp, mA, and time in successive exposures.
Protective Apparel Test
- Aprons, gonadal shields, gloves checked for defects under fluoroscopy.
Recognition & Reporting Malfunctions
- Technologist's responsibility to report malfunctions.
- Equipment failure must be repaired and recalibrated by a physicist.
Fluoro QC
- Fluoro exams = high pt. dose.
- Entrance skin exposure (ESE) averages 3 to 5 rads/min.
Exposure Rate
- State law: ESE shall not exceed 10R/min for normal fluoro.
- ESE allowed up to 20R/min for interventional procedures.