CH 9 QAQC PP
Quality Assurance and Quality Control (QAQC) in Radiology
Introduction to QAQC
QAQC refers to Quality Assurance and Quality Control processes in radiology aimed at ensuring better patient care.
Importance of QAQC
Quality Assurance (QA): Documentation pertaining to policies and protocols related to patient care and safety.
Ensures safety and satisfactory performance of services offered to patients.
Quality Control (QC): Involves identifying and verifying actions necessary for equipment performance.
QC is an integral part of QA.
Activities of a Quality Assurance Program
Preventive maintenance
Quality control procedures
Equipment calibration
In-service education for personnel managing procedures involving:
Radiography
Laboratory equipment
Anesthesia machines
Evaluation of new products and technologies.
Purpose of a QAQC Radiology Program
Aim to minimize radiation exposure for:
Patients undergoing radiographic procedures.
Personnel assisting with the radiographs.
Ensures production of quality radiographs that enable accurate diagnoses.
Reduces the number of repeat films due to quality issues.
Most QAQC tests are conducted annually by:
Veterinary Technicians or designated service personnel.
All results from tests should be recorded in a QAQC log for future reference.
QAQC Testing for X-Ray Machines
Source-Image Distance (SID): Measured using a tape measure and carpenter's level.
Perpendicularity: Ensured with a carpenter's level.
Tube/Table/Crane Locks: Verifies that these components function properly.
X-ray Field Light: Tested with water and cloth.
Light Field Size: Measured with a tape measure.
Locks/Cables/Overhead Crane Movement: Assessed for functional integrity.
Angulation Indicator: Checked using a carpenter's level and protractor.
Light Field/X-ray Field Alignment: Verification done using nine pennies and a loaded cassette.
Nine Penny Test Detailed Procedures
Purpose: Verify the alignment of the x-ray field and collimator light.
Procedure:
Center the X-ray tube above the table at a set distance (SID) of 40 inches and ensure the collimator is level.
Place a loaded cassette in the Bucky tray, centered beneath the table.
Set the collimator field indicators to approximately 6” x 8”.
Turn the collimator light ON.
Place one penny at the middle of each edge of the light field (both inside and outside).
Pennies should be touching each other at the edges.
Take an exposure using settings suitable for imaging a carpus or stifle.
Develop the film to check the results.
The resultant x-ray should depict the pennies at their placed locations.
Measurement: The width of a penny is 0.75 inches, which is 0.2% of a 40-inch SID. If the alignment is off by the width of one penny, service personnel should be contacted for assistance.
Illustrative Figures
Figure 10-6: Light field/x-ray field alignment verification process.
Figure 10-7: Radiograph of the nine-penny test confirming satisfactory results within normal limits.