Ultrasound
A-scans
Stand for amplitude scan
One transducer held at one angle to skin
Pulse partly reflected and partly transmitted at boundary between two tissues.
Transducer receives reflected pulse
Allows a 1D, cross-sectional image to be seen
Scan results shown as peaks on a voltage against time graph on an oscilloscope screen
Peaks may decrease with time as reflected pulse will have less energy due to energy losses within the body and because energy from original pulse is transmitted through the boundary.
Time taken is the time for the pulse to travel to the boundary and be reflected, so distance must be halved after being calculated.
Example of use is for finding tissue depths in the eyes or determining the thickness of bone.
B-scans
Stand for brightness scan
Transducer at different angles to the skin/multiple transducers used
Transducer moved over patient’s skin.
Reflected ultrasound pulses from between boundaries are received by the transducer
Each reflected pulse received forms a bright spot on computer screen
Each dot corresponds to the boundary between two tissues.
Dot brightness is proportional to the intensity of the reflected ultrasound pulse.
Repeated pulses allow computer to form image
Image formed is 2D and a section through the patient.
Effectively multiple A scans (do not write this in the exam, just to help understand it!)
Example of use is for imaging unborn child.