RADT 1010- Filters
Filtration in Radiologic Technology (RADT 1010)
Introduction to Filtration
Definition: Filtration is the process of increasing the mean energy of a heterogeneous beam of radiation by passing it through an absorber.
Purpose: It is used to selectively remove low-energy x-rays from the beam.
Effects of Filtration
Impact on Beam Characteristics:
Changes both the quantity and quality of the radiation beam.
Reduces the patient dose by filtering out less effective radiation.
“Hardens” the beam, which refers to the increase in the mean energy of the x-ray photons in the beam.
Common Beam Filtering Materials
Materials Used in Filtration:
Brass
Copper
Aluminum
Lead
Tin
Molybdenum
Tungsten
Titanium
Zirconium
Additional other materials may be included.
Types of Filtration
Total Filtration:
Calculated as the sum of inherent filtration and added filtration.
Inherent Filtration
Definition: Refers to the filtration that results from the absorption of x-rays as they pass through the tube and housing; this is a constant that the technologist cannot modify.
Components Contributing to Inherent Filtration:
Envelope of the tube
Oil surrounding the tube
Port or exit window of the tube
Aging Impact on Filtration:
Inherent filtration increases as the tube ages because vaporized tungsten from the anode and cathode deposits on the inside of the envelope, which further attenuates the beam.
Measurement of Inherent Filtration:
Generally falls between 0.5 – 1 mm Al equivalent.
Aluminum Equivalent
Definition: This describes the thickness of aluminum that has the same radiation attenuation characteristics under specified conditions as that of the material being considered.
Added Filtration
Definition: Refers to the additional layers of metal placed in the beam, located outside the true tube housing, primarily made of aluminum, and sometimes copper.
Types of Filtration Devices
Compound Filters
Composition:
Contains three layers: tin, copper, aluminum.
Functionality:
Material with a higher atomic number (Z) is positioned closer to the tube.
Each layer absorbs the characteristic photons created in the previous layer.
Atomic Numbers:
Tin (Z = 50)
Copper (Z = 29)
Aluminum (Z = 13)
Compensating Filters
Purpose: These filters are utilized to compensate for parts of non-uniform thickness or density in radiographic imaging.
Functionality:
They rectify variations in anatomy and promote a more consistent image brightness across the radiographic image.
Materials Used:
Opaque plastics
Aluminum
Various other metals
Shapes Available:
Wedge
Step wedge
Bilateral wedge
Trough
Boomerang shape
Conclusion
Future Topics: Information on other devices related to filtration techniques will be discussed in upcoming sessions (TBL next week).
Data Representation
Photon Numbers and Energy Spectrum: Illustrates characteristic L-shell x-rays, Bremsstrahlung x-rays, and K-shell x-rays across energy levels. The x-ray spectrum reveals significant differences before and after inherent and additional filtration.
Energy Levels (keV): Illustrates the distribution and maximum energy like 20, 40, 60, 80, and 100 keV depending on filtration steps.