Ch 21 Artifacts
Artifacts
An artifact is an error in imaging, characterized as:
Not real
Not seen on the image
Incorrect shape or size
Incorrect position
Incorrect brightness
Reflections are a type of artifact.
Causes of Artifacts
Violation of assumptions
Equipment malfunction or poor design
The physics of ultrasound
Operator error
Most artifacts are explained by identifying violated assumptions during image creation.
Image Characteristics
Terms used to describe ultrasonic image features:
Hyperechoic
Hypoechoic
Anechoic
Isoechoic
Homogeneous
Heterogeneous
Hyperechoic
Portions of an image that are brighter than surrounding tissues or appear brighter than normal.
Hypoechoic
Portions of an image that are not as bright as surrounding tissues or appear less bright than normal.
Anechoic
An extreme form of hypoechoic, meaning entirely without echoes.
Echo-free.
Isoechoic
Describes structures with equal echo brightness.
Homogeneous
A portion of tissue or an image that has similar echo characteristics throughout.
Heterogeneous
A portion of tissue or an image that has different echo characteristics throughout.
Basic Assumptions of Imaging Systems
Sound travels in a straight line.
Sound travels directly to a reflector and back.
Sound travels in soft tissue at exactly 1,540 m/s.
Reflections arise only from structures positioned in the beam’s axis.
The image plane is very thin.
The strength of the reflector is related to the characteristics of the tissue creating the reflection.
General Artifact Behavior
Artifacts appear in some views but not others.
Actual anatomy remains visible.
Corrective measures can eliminate artifacts.
Instrument malfunction should be considered when obvious artifacts remain visible even after corrective measures are taken.
Reverberation Artifact
Multiple, equally spaced reflections appear on the display.
Caused by the bouncing of the sound wave between two strong reflectors positioned parallel to the beam.
Relates to the assumption that sound travels directly to a reflector and back, which is invalid.
Characteristics
Appear in multiples.
Appear equally spaced.
Located parallel to the sound beam’s main axis.
Located at ever-increasing depths.
Reverberation places too many relations on the image.
The 1st and 2nd reflections are real, but the remaining echoes that appear deeper do not correctly correspond to true anatomy.
Resemble a ladder or Venetian blind.
Comet Tail Artifact
Described as “reverberation with spaces squeezed out.”
Appears as a solid hyperechoic line directed downward.
Also known as a ring-down artifact.
Relates to the assumption that sound travels directly to a reflector and back, which is invalid.
Characteristics
Created when closely spaced reverberations merge together.
More likely to appear when reflecting surfaces are located in a medium with a very high propagation speed, such as a mechanical heart valve.
Can arise from the resonance or vibration of small structures, such as gas bubbles, after they have been bombarded by a sound pulse.
Appears as a single long hyperechoic echo.
Located parallel to the sound beam's main axis.
Shadow Artifact
Appears as a hypoechoic or anechoic region extending downward from a very strong attenuating medium.
Shadowing can be used to help with a diagnosis.
Shadowing is entirely unrelated to the speed of sound in a medium.
Relates to the assumption that the strength of a reflection is related to the characteristics of the tissue creating the reflection, which is invalid.
Characteristics
Hypo- or anechoic (background color).
The result of too much attenuation.
Located beneath a structure with abnormally high attenuation.
Prevents visualization of true anatomy on the scan.
Edge Shadow Artifact
A special form of shadowing extending along the edge of a curved reflector.
Created as sound beams refract and diverge along the edge of a curved structure.
Hypoechoic region extending down from the edge of a curved reflector.
Prevents the display of true anatomic structures that are positioned within the extended hypoechoic region.
The hypoechoic region is the same color as the background.
Also known as shadowing by refraction.
Relates to the assumption that the intensity of a reflection is related to the characteristics of the tissue creating the reflection, which is invalid.
Characteristics
Hypo- or anechoic (background color).
Results when the beam spreads after striking a curved reflector.
Extends downward from the curved reflector’s edge, parallel to the beam.
Prevents visualization of true anatomy on the scan.
Difference Between Shadow and Edge Shadow
Edge shadowing is caused by the shape, not attenuation.
Enhancement Artifact
Appears as a hyperechoic region beneath tissue with abnormally low attenuation.
The opposite of shadowing.
Can help with clinical diagnosis and provide valuable diagnostic information to help characterize tissue.
Entirely unrelated to the speed at which sound travels through a medium.
Relates to the assumption that the intensity of a reflection is related to the characteristics of the tissue, which is invalid.
Characteristics
Hyperechoic (appears the same as the foreground tissue).
The result of too little attenuation.
Located beneath a structure with abnormally low attenuation.
Focal Enhancement Artifact
A special form of enhancement in which a side-to-side region of an image appears brighter than tissue at other depths.
Also known as focal banding.
Has the appearance of an incorrect TGC.
Most prominent at the region of the focus.
Structures at the focus appear brighter than those at other depths.
An entire horizontal band of tissues appears hyperechoic.
Relates to the assumption that the intensity of a reflection is related to the characteristics of the tissue, which is invalid.
Characteristics
A hyperechoic side-to-side region (appears the same as the foreground color).
Results from increased intensity at the focus.
Mirror Image Artifact
Created when sound reflects off of a strong reflector (mirror) and is redirected towards a second structure.
Redirection causes a replica or second copy of the structure to incorrectly appear on the image.
The artifact is placed deeper than the real structure.
Mirror artifact may be seen with color Doppler; the shallower vessel is real, and the deeper one is an artifact.
Relates to the assumptions that sound travels in a straight line and sound travels directly to a reflector and back, which are invalid.
Characteristics
A second copy of a true reflector.
The artifact appears deeper than the true reflector.
A bright reflector, the mirror, lies on a straight line between the artifact and the transducer.
The true reflector and artifact are equal distances from the mirror.
Cross-talk artifact: A type of mirror image artifact that appears on spectral Doppler display.
Speed Error Artifact
Created when a sound wave propagates through a medium at a speed other than that of soft tissue ( or ).
The correct number of reflectors are displayed, but they appear at incorrect depths.
Also known as range error artifact or range ambiguity artifact.
Relates to the assumption that sound travels through soft tissue at exactly , which is invalid.
When the Medium’s Speed is Slower than Soft Tissue
Sound travels slower than the ultrasound system expects.
Pulses return slowly.
Go-return time is too long.
The system assumes reflectors are far from the transducer.
Reflectors are placed too deep on the image.
Distances are overestimated (reported number is too large).
When the Medium’s Speed is Faster than Soft Tissue
Sound travels faster than the ultrasound system expects.
Pulses return quickly.
Go-return time is too short.
The system assumes the reflectors are close to the transducer.
Reflectors are located too shallow on the image.
Distances are underestimated (reported number is too small).
Characteristics
Propagation speed errors are called range error artifacts.
Correct number of reflectors.
Improper depth.
Appears as a step-off.
Example: If two pins are apart but show up apart on the display, then if they were apart in the test object, they would appear apart due to the 10% underestimation. This means sound is traveling faster than soft tissue's speed.
Lobe Artifacts
Appear when sound energy is transmitted in a direction other than along the beam’s main axis.
Lobes degrade lateral resolution.
Side lobes: Created by a single crystal transducer.
Grating lobes: Created by an array transducer.
Characteristics
Generally weaker than the main beam and typically do not generate reflections.
A reflection may be created when a strong reflector is in the path of the lobe.
Impossible to distinguish the true reflector and artifact from a single image.
Real anatomy will appear in all views, whereas an artifact will not.
Related to the assumption that reflections arise from structures located along the beam’s main axis, which is invalid.
A second copy of the true reflector.
The artifact and the true reflector are located side-by-side at the same depth.
Reducing Grating Lobes
Subdicing: Dividing each PZT element into smaller pieces.
Apodization: Exciting the subdiced elements with different voltages (higher voltages for elements closer to the center and lower voltages for outermost elements).
Refraction Artifact
Occurs when a sound pulse changes direction during transmission (transmission with a bend).
Occurs only when a sound beam strikes a boundary obliquely, and the media on either side of the boundary have different propagation speeds.
Degrades lateral resolution.
Relates to the assumption that sound travels in a straight line, which is invalid.
Characteristics
A second copy of the reflector is seen.
The copy is side-by-side, or at the same depth as the true reflector.
Slice Thickness Artifact
The 2-D imaging plane is not thin and uniform; in reality, it is 3-D.
Reflections from structures above or below the assumed imaging plane may appear in the image.
The beam flares out like a bell of a trumpet.
Slice thickness is related to the dimension of the beam perpendicular to the imaging plane.
Elevation resolution is determined by the thickness of the imaging plane.
Appears when the beam dimension is greater than the reflector size.
Also called section thickness artifact or partial volume artifact.
The thickness of the imaging plane varies; thicker portions have inferior elevation resolution compared to thinner portions.
The true reflector lies either above or below the assumed imaging plane but is displayed within the image.
Usually fills in hollow structures such as a cyst.
Can be reduced with thinner imaging planes, such as those created with 1.5-dimensional array transducers.
Relates to the assumption that the imaging plane is extremely thin, which is invalid.
Lateral Resolution Artifact
Also known as point spread artifact.
Occurs when a beam is wider than the distance between two reflectors that are side to side.
Two objects appear as one.
Displays a small reflector as a wide line.
Focusing improves lateral resolution.
When two structures are closer together than the beam's diameter, only one reflection is seen on the image.
Axial Resolution Artifact
Created when a long pulse strikes two closely spaced structures, where one is in front of the other.
Only one reflector will appear on the image if the structures are closer together than one-half spatial pulse length.
Higher frequency transducers have better axial resolution (shorter SPL).
Other Artifacts
Multipath Artifact
Created when sound pulses glance off a second structure on the way to or from the primary reflector.
Results in the transmit path differing from the return path.
Results in subtle, nonspecific changes that can be identified.
Proves invalid the assumption that a pulse travels directly to a structure and back.
Curved & Oblique Reflectors Artifact
Occurs when a sound beam strikes a curved or oblique reflector; some of the sound may be directed away from the transducer.
Reflection is weaker, causing the reflector to appear absent, weak, or different from similar reflectors.
Temporal Resolution Artifact
Temporal Resolution (T.R.) is the ability to precisely position a moving structure.
Determined by frame rate (higher frame rate = better temporal resolution = better ultrasound movie).
Poor T.R. caused by low frame rates results in less accurate positioning of reflectors in motion.
Spatial Resolution Artifact
Related to the overall detail in an image.
Determined by sound beam’s spacing or line density.
Higher line density (closely packed pulses) on the image shows great detail and superior spatial resolution.
Lower line density (wider gaps between pulses) on the image shows less detail and the spatial resolution is inferior.
Digital Format
When the display information is in digital format, spatial resolution is related to pixel density.
Improves with a large number of pixels in the image of a fixed size.
Each pixel is small, resulting in more detail.
Degrades with low pixel density.
Each pixel is large.
Spatial Resolution Artifact occurs when the pixel size is greater than the size of the reflector.
Range Ambiguity Artifact
Assumes that all reflections are received by the transducer before the next pulse is transmitted.
Occurs when a reflection from a deep structure is received after the next pulse has been transmitted.
This reflector will be interpreted as coming from the most recent pulse and will appear as a very shallow reflector.
Occurs when a reflecting structure is located deeper than the imaging depth of the image, and this reflector is placed at a shallow level.
Can be eliminated by increasing the PRP (longer PRP means deeper imaging and decreased PRF).
Noise Artifact
Appears as small amplitude echoes and results from many sources:
Electrical interference
Signal processing
Spurious reflections
More likely to affect low-level hypoechoic regions rather than a bright echogenic area.
Other Forms of Noise
Speckle
Clutter
Speckle Artifact
Grainy area not directly related to the actual biologic tissue.
Noise resulting from the constructive and destructive interference of small sound wavelets.
Constructive: in-phase
Deconstructive: out-of-phase
The image will contain fake detail, such as the grainy tissue-like texture.
Can be reduced by Spatial Compounding (pg. 254).
Clutter Artifact
The presence of false echo signals arising from locations outside of the main sound beam.
Side lobes, grating lobes, and thickness artifacts are sources of clutter.
With Doppler, it is the contamination of valid Doppler signals with high-amplitude reflections from vessel walls or heart muscle.
These high-amplitude reflections overpower the true Doppler signals created by moving blood.
Harmonic Imaging
Using Harmonics reduces the image’s noise content.
Goal is to selectively distinguish meaningful reflections from noise, resulting in an increased signal-to-noise ratio (Refer to Ch 17).
Artifact Summary
Anatomic reflectors absent on image
Shadowing
Shadowing by refraction
Lateral resolution
Axial resolution
Anatomic reflector appears multiple times on image. Artifact positioned deeper than the true anatomy.
Comet tail
Ring down
Reverberations
Mirror image
Anatomic reflector appear multiple times on image. Artifact displaced to the side of the true anatomy
Refraction
Side lobe
Grating lobe
Anatomic reflectors appear with abnormal brightness
Enhancement (hyperechoic)
Banding (hyperechoic)
Shadowing (hypoechoic)
Shadowing by refraction (hypoechoic)
Anatomic structures appear at incorrect depth
Speed errors
Range ambiguity artifact
Anatomic structures appear in the incorrect imaging plane
Slice or section thickness
Anatomic structures do not correspond to echoes on the image
Acoustic speckle
Multipath