Lecture 2
NELSON MANDELA UNIVERSITY - HMSV392 Evaluation Methods II - Lecture 02: Equipment and the TEM
Anthropometry Equipment
Specific equipment is essential for taking measurements. The following tools are vital for an anthropometrist:
Stadiometer: Measures stature and sitting height.
Weighing Scales: Used for measuring body mass.
Anthropometric Tape: For measuring girths and locating skinfold sites.
Skinfold Caliper: For assessing body fat through skinfold thickness.
Anthropometer: Measures heights and lengths.
Large Sliding Caliper: Used for larger distances.
Segmometer: For segment length measurement.
Wide Spreading Caliper: Designed for specific body measurements.
Small Sliding Caliper: For small bone breadth measurements.
Anthropometric Box: Facilitates height and length measurements.
Stadiometer
Used to measure stature and sitting height; typically mounted to a wall for accuracy.
Features:
Should have a headpiece with a locking device.
Measurement range: 60 cm to 220 cm, accuracy to 0.1 cm.
Requires a level and hard floor for accurate measurement.
Alternative Methods when Stadiometer is Unavailable
Carpenters Tape: Fixed to the wall with a 90° head board.
Girth Tape: Similar setup as carpenters tape.
Laminated Graph Paper: Taped to a wall for height estimation using a head board (not acceptable in lab settings).
Weighing Scales
Beam Balance: Traditional choice, accuracy to nearest 100g.
Spring Balance Scales: Accuracy of 500g, must be regularly calibrated.
Electronic Scales: More convenient, accuracy to within 50g, should be calibrated using certified weights totaling at least 150kg.
Anthropometric Tape
Essential for measuring girths and locating skinfold sites accurately.
Specifications:
Must be non-extensible and flexible.
Should not exceed 7mm in width.
Must have a stub area of at least 4cm before the zero line.
Recommended length: at least 1.5m, calibrated in cm with mm gradations.
Regular calibration required, particularly for fibreglass tapes.
Skinfold Calipers
Maintain a constant closing pressure of 10g/mm².
Harpenden Calipers: Widely used, calibration up to 40mm in 0.2mm divisions.
Slim Guide Calipers: Alternative, calibrates to 80mm in 0.5mm divisions.
Caution advised when applying skinfold data from different calipers due to variations.
Anthropometer
Measures heights, lengths, and segment lengths directly.
Primarily for vertical heights, can also measure segment lengths, bone breadths, and stature.
Large Sliding Calipers
Upper segment of the anthropometer; measures large bone breadths.
Features two straight branches attached to a rigid scale.
Caution against incorrect assembly of calipers.
Segmometer
Comprises a steel tape with two straight branches (approx. 7cm in length) for direct segment length measurements.
Wide Spreading Calipers
Features recurved branches, ideal for measuring dimensions not suitable for straight branches due to anatomical constraints.
Small Sliding Calipers
Mitutoyo Adapted Calipers: Suitable for small bone breadths, dimensions include a minimum branch length of 10cm and accuracy of 0.1mm.
Anthropometric Box
Standard dimensions: +/- 40cm sides and height.
Facilitates measurement of heights, lengths, and breadths while seated.
Includes a cutout section for foot positioning.
Technical Error of Measurement (TEM)
Measurement variances are inevitable; reasons include:
Biological changes
Anthropometrist technique
Aim to minimize technical variability in measurements.
Key Concepts in Measurement Error
Precision:
Variability in repeated measurements on the same subject.
High precision equals low variability.
Competent anthropometrists strive for precision.
Reliability:
Alternative to precision; expressed as a correlation coefficient (Intraclass correlation coefficient, ICC).
Values range from 0 (no reliability) to 1 (perfect reliability).
Accuracy:
Correspondence of the measured value to the real value, determined by a skilled anthropometrist.
Validity:
The extent to which a measurement truly measures what it claims to.
Measures of Precision and Reliability
Technical Error of Measurement (TEM): Standard deviation of repeated independent measurements.
Comparison of TEM across different variables can be facilitated by converting to relative TEM (%TEM):
Formula: % TEM = (TEM / MEAN) x 100
Additional Notes on Measurements
TEM and ICC specifics:
Related to a particular variable, measuring technique, equipment, and tested population.
A minimum of 20 subjects and 2 trials recommended for accurate TEM and ICC determination.