spectroscopy V and VI

Spectroscopy High-Yield Notes (Units V and VI)

Built from the Spectroscopy Theory and Spectroscopy Applications decks. Items the slides flag as "know this" are marked [KNOW].

PART V. SPECTROSCOPY THEORY

V.A Electromagnetic radiation

Order of the spectrum [KNOW]

  • Lowest to highest energy: radio, microwave, IR, visible, UV, X-ray, gamma.

  • Visible runs from about 800 nm (red, low energy) to about 400 nm (blue/violet, high energy).

Light-matter interactions

Process

What happens

Where it shows up

Reflection

Light bounces off a surface

Mirrors and optics; attenuated total reflection (ATR) is an IR technique

Refraction

Light bends between two media

Prism monochromators

Diffraction

Light bends passing a small opening or fine grating

Grating monochromators; X-ray diffraction

Absorption

Molecule goes from ground to excited state

Core of absorption spectroscopy

Emission

Excited state relaxes, with or without a photon

Fluorescence, phosphorescence, atomic emission

Absorption energy

  • The photon energy must equal the gap: ΔE = hν = hc/λ.

  • Worked example, Hg lamp at 253.7 nm:

    • Step 1: E = (6.626e-34 J·s)(3e8 m/s) / (2.537e-7 m)

    • Step 2: top = 1.988e-25 J·m, so E = 7.84e-19 J per photon

    • Step 3: multiply by Avogadro's number to get 472 kJ per mole of photons

Quantum primer

  • Atoms have only electronic states. Molecules also vibrate and rotate, so they have many more transitions.

  • Light can move molecules between states, but not every transition is allowed.

Selection rules [KNOW]

  1. Photon energy must equal the transition gap: E = hν = ΔE.

  2. Spin multiplicity must not change: ΔS = 0.

    • S is the total spin quantum number (add up the spins, each ±½).

    • Multiplicity = 2S + 1.

    • S = 0 gives 2(0) + 1 = 1, a singlet.

    • S = 1 gives 2(1) + 1 = 3, a triplet.

  3. Symmetry rules.

    • UV-vis: σ to σ* and π to π* type transitions.

    • IR: there must be a change in dipole moment.

Absorption lines

  • Atoms give sharp line spectra in the gas phase. Example: Na 3s to 3p at 589.3 nm (about 203 kJ/mol).

  • Molecular gas-phase electronic spectra show resolved vibronic fine structure.

  • In solution the fine structure is lost and bands broaden, because of collisions and solvent perturbation of the quantum states.

Emission

  • Excited states have a finite lifetime (τ). The uncertainty principle means there is always an uncertainty in the excited-state energy, so every transition has a width (natural or uncertainty broadening).

  • Radiative relaxation emits a photon matching the energy gap.

  • Non-radiative relaxation emits no photon (collisional deactivation, heat released to the solvent).

  • Emission can be spontaneous or stimulated. Stimulated emission is the basis of lasers.

  • Atomic emission line sources: Hg lamp, Nd:YAG laser, hollow cathode lamps. Excitation is by electrical means.

Jablonski diagram vocabulary [KNOW]

  • IC (internal conversion): non-radiative, same multiplicity.

  • ISC (intersystem crossing): non-radiative, different multiplicity.

  • EC (external conversion): excitation lost to the solvent as heat.

  • Fluorescence: radiative, same multiplicity (fast).

  • Phosphorescence: radiative, different multiplicity (slow).

  • The Jablonski diagram has no x-axis.

V.B Generalized instrument design

Five components

  1. Radiation source (tunable or single wavelength, correct region of the spectrum).

  2. Wavelength selection (scanning or broadband; some techniques need more than one).

  3. Sample containment (static or dynamic).

  4. Detector (sensitive, responds in the wavelength range of interest).

  5. Signal processing and readout (Fourier transform, signal averaging).

Ideal source: high intensity, flat intensity versus wavelength, low variability.

Continuous sources (slides: "know examples, not details")

  • Many are black-body emitters. Intensity and wavelength depend on temperature, and λmax is proportional to 1/T.

Region

Source

Range or note

IR

Globar (SiC)

1000 to 2000 °C

IR

Nichrome wire

about 500 °C

Visible

Tungsten filament lamp

320 to 3500 nm

UV

Deuterium lamp

180 to 370 nm (summary table: 160 to 375)

UV

Xe arc lamp

250 to 600 nm

UV

High-pressure Hg arc lamp

254 to 580 nm, semi-continuous

Line sources

  • Low-pressure metal vapor lamps: Na (589.3 nm), Hg (254 nm).

  • Hollow cathode lamp (HCL): element specific.

  • Lasers: based on stimulated emission.

    • Problem: emission and absorption rates scale with the number of molecules in the excited and ground states.

    • Need a population inversion (more molecules in the excited state), which is impossible in a two-level system.

    • So lasers need three or four levels with suitable lifetimes.

    • Output is high intensity, monochromatic and coherent.

Monochromators (wavelength selectors)

Type

How it works

Key points

Filter

Absorbs unwanted light

Low-pass absorbs above the cutoff; high-pass absorbs below the cutoff; notch has both cutoffs. Broadband.

Prism

Refraction (dispersion) plus an exit slit

Materials: quartz for UV, glass for visible, NaCl or KBr for IR. Non-linear dispersion, better separation at short wavelengths, so a fixed slit gives different bandwidths.

Grating

Diffraction and interference off a finely spaced grating

Rotate the grating to choose the wavelength. Linear dispersion.

  • Bandwidth is the span of wavelengths leaving the monochromator.

  • Gratings beat prisms on: linear dispersion, larger useful range, compact size.

  • Gratings lose on: cost, more stray light, higher-order harmonics (overtones).

Detectors [KNOW the comparison]

Detector

Principle

Pros

Cons

Photovoltaic (350 to 750 nm)

Photon makes an electron-hole pair in a semiconductor, giving a current. Not powered.

No power supply, rugged

Limited range, low sensitivity

Phototube

Powered. Photon ejects electrons from a cathode, collected at the anode.

More sensitive, works in UV and vis

Dark current, costs more

PMT

Series of dynodes, each amplifying (about 100 V steps). Gain of 1e5 to 1e8 electrons per photon.

Very sensitive, versatile

Dark current, cost

Photodiode array

Array replaces the exit slit and measures all wavelengths at once

Simultaneous detection

More expensive and less sensitive than a PMT

Thermocouple (IR)

Two dissimilar metals, voltage at the junction depends on temperature

Detects IR heating

Must sense ΔT of 0.001 to 0.005 °C

Golay cell (IR)

Heating changes gas pressure, read optically

Detects IR heating

Same sensitivity demand

  • Geometry trap [KNOW]: with a single PMT the sample must go after the grating. With a photodiode array the sample goes between the source and the grating.

V.C Quantitation with absorption

Beer-Lambert law

  • Transmittance: T = I/I0 (range 0 to 1). Percent transmittance: %T = 100 × T.

  • Absorbance: A = -log(T) = -log(I/I0).

  • A = ε b c, where ε is molar absorptivity (L mol^-1 cm^-1), b is path length (cm), and c is concentration (mol/L).

Worked example: riboflavin

  • Given: ε = 3.24e4 L mol^-1 cm^-1, b = 1 cm, 30% decrease in radiant power.

  • Step 1: 30% decrease means P/P0 = 0.70.

  • Step 2: A = -log(0.70) = 0.155.

  • Step 3: c = A / (ε b)

  • Step 4: c = 0.155 / (3.24e4 × 1)

  • Step 5: c = 4.8e-6 M.

Quantitation methods

  • Calibration curve: interpolate between standards. Use internal standards. Only valid where Beer's law is linear.

  • Standard addition: add increasing amounts of standard to fixed volumes of sample.

    • Plot A versus the amount of standard added.

    • The x-intercept equals the negative of the initial concentration.

    • Used when the matrix could change the response.

Worked example: Cu in an alloy

  • Sample 0: 100 mg alloy in 100 mL, A0 = 0.30.

  • Sample 1: 100 mg alloy in 95 mL, plus 5 mL of 0.1 M Cu(NO3)2, so the volume is again 100 mL. A1 = 0.45.

  • Step 1: Added Cu = (5 mL × 0.1 M) / 100 mL = 0.005 M.

  • Step 2: A0 = k·c0 and A1 = k·(c0 + 0.005).

  • Step 3: A1 / A0 = 0.45 / 0.30 = 1.5.

  • Step 4: So c0 + 0.005 = 1.5 c0, which gives 0.005 = 0.5 c0, so c0 = 0.010 M.

  • Step 5: Moles Cu = 0.010 mol/L × 0.100 L = 1.0e-3 mol.

  • Step 6: Mass Cu = 1.0e-3 mol × 63.55 g/mol = 0.0636 g = 63.6 mg.

  • Step 7: %Cu = 63.6 mg / 100 mg = about 64% Cu.

Deviations from Beer's law

  • Expect a linear response. Negative deviations are usual, positive ones are possible. Best fix: dilute.

Why %T error matters. A is logarithmic but we measure %T, and the error in %T is constant. Same 1% T error, ε = 1000, b = 1 cm:

Reading

Absorbance

Resulting c

Effect of the 1% T step

%T 99 to 98

0.0044 to 0.0088

4.4e-6 to 8.8e-6 M

about +100%

%T 59 to 58

0.229 to 0.237

2.29e-4 to 2.37e-4 M

about +3%

%T 2 to 1

1.70 to 2.00

1.70e-3 to 2.00e-3 M

about +15%

  • Best results (error under 3.5%) come from staying between about 15% and 65% T, which is roughly A = 0.82 down to 0.18.

Chemical deviations

  • Beer's law holds only for dilute solutions. Molecules must not interact, because interactions alter electron clouds and energy levels.

  • This gives an upper practical limit near A = 0.82.

  • Too dilute, or ε too small, and the drop in intensity can't be measured well (A below 0.15). Then choose another detector.

Instrumental deviations

  • Bandwidth: ε changes with λ, so always measure at the maximum of the ε versus λ curve. Use a monochromator with small bandwidth.

  • Stray radiation also causes deviations.


PART VI. SPECTROSCOPY APPLICATIONS

VI.A Atomic absorption spectroscopy (AAS)

Atomic lines are very sharp (no vibrations or rotations)

  • Natural linewidth is about 1e-5 nm (uncertainty principle on the excited-state lifetime).

  • Doppler broadening: atoms move randomly, so their kinetic energy couples to ΔE. Blue shift is to shorter λ, red shift is to longer λ, and the peak broadens.

  • Pressure broadening: collisions perturb the energy levels.

  • Observed widths are still tiny: about 2 to 5 e-3 nm.

The catch: samples are rarely free gas-phase atoms. You must nebulize (turn liquid or solid into an aerosol) and atomize (free the atom from its partners). This needs a nebulizer and a heat source (flame or furnace).

Flame processes (example: Cr in an alloy)

  1. Dissolve in HNO3 to get Cr3+ and NO3-.

  2. Nebulize to get an aqueous aerosol of Cr(NO3)3.

  3. Desolvation gives solid Cr(NO3)3.

  4. Vaporization gives gas Cr(NO3)3.

  5. Atomization gives Cr (g). This is what you want.

  6. Unwanted side paths: excitation (Cr*), ionization (Cr+), oxidation (CrOx).

Flame goals: high atomization; low excitation, low ionization, low oxide formation. Flame temperature depends on the oxidant and fuel, the part of the flame, and the burner design.

Degeneracy (g) [KNOW]

  • Rotational states: g = 2J + 1.

  • Vibrational states: g = 1 always.

  • Electronic states: for isolated atoms the slide gives 2l + 1 (this is likely 2J + 1; check against your lecture notes). For molecules, σ and n have g = 1 and π has g = 2.

  • The instructor says g-values will be given on exams when they aren't 1.

  • Why it matters: excited-state population depends on temperature and on g, so higher T means more excitation (unwanted in AAS).

Flame choice (trends to know, not every number)

  • Air-acetylene is about 2300 °C. Nitrous oxide-acetylene is about 2800 to 2950 °C. Oxygen-acetylene is about 3100 °C. Air-methane is the coolest listed (about 1840 °C).

  • Atomization efficiency depends on both flame and element. Examples from the table: Al is under 0.00005 in air-C2H2 but 0.2 in N2O-C2H2; Cr is 0.065 versus 0.6; Sn is 0.04 versus 0.8. Hotter flames atomize refractory, oxide-forming elements better.

  • Ionization is worst for easily ionized elements (low IP): Cs (2.89 eV), Rb, K, Na, and Ba in air-C2H2. A hotter flame ionizes more.

Flame summary

Process

Wanted?

Favored by or worst for

Atomization

Yes

Volatile elements, low mass, pre- and post-transition metals; increases with T

Excitation

No

Groups 1 and 2; increases with T

Ionization

No

Groups 1 and 2, plus rows 4, 5 and 6; increases with T

Oxide formation

No

Early transition metals and main-group metals; worse in the oxidizing (upper) part of the flame

Instrument path (in order): hollow cathode lamp, chopper, nebulizer and flame, grating monochromator, PMT.

  • HCL: cathode made of the element under study. Inert gas (Ne or Ar) is ionized, ions strike the cathode and ablate metal atoms, some come off excited (M*) and emit lines unique to M. Bandwidth is 0.002 to 0.005 nm. Gives individual lines.

  • Chopper: the flame also emits light (a DC signal). Modulating the HCL lets the detector separate lamp light from flame light. With lamp light, P = P(HCL) + P(flame); with the lamp blocked, P = P(flame). The difference isolates the lamp signal.

  • Nebulizer: aspirates sample using the oxidant flow, giving droplets of 10 to 40 µm. Uniform droplets give a more stable signal; baffles (spoilers) remove large droplets.

  • Monochromator: small-bandwidth grating. Biggest problems are flame light noise and absorption by impurities.

Atomizers: flame versus graphite furnace


Flame AAS (FAAS)

Graphite furnace AAS (GFAAS)

Setup

Laminar slot burner, 10 cm path, atoms in flame about 1 ms, continuous flow

No nebulizer; 0.5 to 100 µL loaded straight into furnace

Temperature program

Constant

Dry 100 to 250 °C; ash 500 to 1000 °C; atomize 1500 to 2000 °C

Signal

Steady

Transient (all sample used)

LOD

ppm to ppb

ppb to ppt (better)

Precision

about 1% relative

about 5% relative

Speed

Seconds

Minutes

Cost

Lower

Higher

Extras

Flame noise present

Small sample, high sensitivity, no flame noise

Interferences and fixes [KNOW]

Problem

Fix

Poorly volatile salts (for example phosphate with the analyte cation)

Releasing agent: a cation (Sr, La) that binds the bad anion and frees the analyte. Protecting agent: a volatile chelator (8-hydroxyquinoline) that binds the analyte cation and shields it.

Oxide formation

Add a metal (V, Ti) that preferentially reacts with oxygen

Ionization

Add an easily ionized metal (an alkali metal) that supplies electrons to convert M+ back to M

Spectral interference

Overlap from other atomic lines is rare. Molecular (especially oxide) impurities have broad bands that overlap. Fix: a second D2 lamp reads the background just above or below the atomic line, with a second chopper alternating the two lamps; subtract the background.

  • Commercial pre-mixed modifiers contain element-specific releasing and protecting agents.

AAS summary

  • More than 60 elements (limited by available HCLs).

  • Excellent for quantitation, poor for qualitative work (change the lamp for every element).

  • LOD: low ppm to ppb (FAAS), low ppb to ppt (GFAAS).

  • Beer's law linear over about 2 to 3 orders of magnitude.

  • Precision 1 to 2% relative (about 5% for GFAAS). Highly selective.

  • Cost: instrument $50 to 75K by one slide (the comparison table says about $30 to 50K); about $300 to 500 per lamp.

VI.B Atomic emission spectroscopy (AES)

  • The oldest spectroscopy (colored flames). Atoms are excited, then emit light, for example Na* to Na at 589.0 nm.

  • Sources: flame (rarely used now), stars (astronomy), plasma torch (inductively coupled or DC).

  • ICP: ionized argon (Ar+ and e-) at 4000 to 8000 °C. A spark starts it and radio-frequency radiation sustains it. Sample is nebulized into the torch, atomized and excited. The many electrons make the plasma reducing, so oxidation and ionization are suppressed.

  • Monochromator: very high-quality grating (0.01 nm resolution).

  • Big advantage: multi-element analysis in one run.

    • Scanning: exit slit, one wavelength at a time, computer-controlled, inefficient because most of the time is spent between lines. Uses a single PMT.

    • Simultaneous: no exit slit. Detect with a photodiode array, multiple PMTs on a Rowland circle, or a CCD camera with an Echelle grating.

AAS versus ICP-AES [KNOW]


AAS

ICP-AES

Qualitative

Poor (change lamps)

Excellent

Quantitative

One element at a time

Many elements at once

LOD

FAA ppm to ppb; GFAA ppb to ppt

ppm to ppt

Linearity

Relatively poor (1 to 3 orders)

Excellent (ppb to %)

Precision

FAAS 1 to 2%; GFAAS about 5%

1 to 2%

Technical demands

Low

High

Cost

about $30 to 50K

about $100 to 200K

VI.C Molecular absorption (UV-vis)

Regions: visible 400 to 800 nm; common UV 180 to 400 nm; vacuum UV 10 to 180 nm. O2 and N2 absorb near 180 nm, so work below that must be done in vacuum.

Fates of the excited state (M):* photochemistry (A + B), radiationless decay to heat, or fluorescence/phosphorescence. Lifetime ranges from about 1e-9 s to hours.

Orbitals

  • π and π* require multiple bonds. n requires lone pairs.

  • CH4: σ only. H2C=CH2: σ and π. NH3: σ and n. H2C=O: σ, π and n.

Allowed transitions [KNOW]

Transition

Typical λmax

Notes

σ to σ*

Very short (CH4 about 125 nm)

Vacuum UV, not analytically useful

n to σ*

about 150 to 250 nm

Saturated organics with lone pairs (H2O, ROH, R2O, RNH2). These are common solvents, which sets the solvent cutoff.

π to π*

about 200 to 400 nm

Compounds with multiple bonds; very intense, but often at too short a wavelength

n to π*

about 400 to 700 nm

Lone pair plus multiple bond; weaker than π to π* but more accessible

Color and complementary color: a compound absorbs one color and appears as its complement.

Absorbed (nm)

Color absorbed

Appears

400 to 430

violet

yellow-green

430 to 480

blue

yellow

480 to 500

blue-green

red

500 to 560

green

purple

560 to 580

yellow-green

violet

580 to 590

yellow

blue

590 to 630

orange

green-blue

630 to 750

red

blue-green

Inorganic color

  • Transition metals have unfilled d (or f) shells. In an octahedral field, dx2-y2 and dz2 (the eg set) point at the ligands and are destabilized; dxy, dyz and dxz (the t2g set) point between ligands and are stabilized. The t2g to eg transition falls in the visible.

  • The gap Δ depends on the ligand, the metal's row, and its oxidation state.

  • Weak-field ligand: CrCl6 3-, λmax = 736 nm. Strong-field ligand: Cr(CN)6 3-, λmax = 380 nm. Bigger Δ means higher energy and shorter wavelength.

  • Charge-transfer bands: an electron moves between ligand and metal, for example [Fe(III)(H2O)5SCN]2+ going to [Fe(II)(H2O)5SCN•]2+ (ligand to metal).

Applications

  • Qualitative: not very useful (broad bands).

  • Quantitative: very useful. ε is 10 to 1e5, giving a usable range of 1e-4 to 1e-7 M. Precision 1 to 3% relative. Selectivity is not great. Fast enough for kinetics, titrations and chromatography detection.

  • Considerations: solvent cutoff, Beer's law linearity, work at λmax.

  • Multicomponent analysis: absorbances are independent and additive, so

    • A at λ1 = ε(x,λ1) b c(x) + ε(y,λ1) b c(y)

    • A at λ2 = ε(x,λ2) b c(x) + ε(y,λ2) b c(y)

    • Measure at as many wavelengths as there are species and solve the simultaneous equations.

  • Isosbestic point: for two interconverting species, the wavelength where both have the same absorbance. Its presence suggests a clean two-species conversion.

  • Photometric titration: monitor A at a chosen wavelength. R, T or P (reactant, titrant, product) may absorb, which gives different curve shapes.

Instruments

  • Photometer: no wavelength dispersion (filters), very cheap.

  • Spectrophotometer: has wavelength selection.

    • Single beam: P0 measured once before the sample and reused. Simplest, but does not account for P0 drift.

    • Double beam: source split in two, P0 and P measured continuously (two detectors or a chopper). Compensates for source flicker and the source's power distribution. More expensive, and weaker signal from the extra optics.

    • Both are usually run in scan mode. A photodiode array measures all wavelengths at once; note the sample position relative to the monochromator.

VI.D Molecular emission (fluorescence and phosphorescence)

Jablonski sequence [KNOW]

  1. Absorption (about 1e-15 s, fastest), often into excited vibrational levels of an excited electronic state. If light exits at the same energy: Rayleigh scattering. If it exits to a different vibrational level of the ground state: Raman scattering.

  2. Intramolecular vibrational relaxation (about 1e-12 s) drops the molecule to v = 0 of the excited electronic state. All later processes start from v = 0.

  3. Internal conversion (same molecule) and external conversion (to another molecule or solvent): radiationless, same multiplicity, need vibrational overlap.

  4. Fluorescence: radiative, same multiplicity, short lived (ns to ms by the slide), broad band because many final vibrational levels are reachable. Fluorescence from S1 is preferred analytically because it is at a different wavelength than the excitation light.

  5. Intersystem crossing: radiationless, different multiplicity. Spin forbidden but spin-orbit allowed. Needs strong spin-orbit coupling, from a heavy atom (Z above about 30) on or near the molecule, including in the solvent.

  6. Phosphorescence: radiative, different multiplicity, also spin forbidden, so it is slow (ms to hours).

  7. Competing losses: photodissociation (bond breaks after promotion into an antibonding orbital) and photochemistry (reactive center formed).

  • Each pathway has a rate constant: kIC, kEC, kISC, kF, kP, kd (dissociation), kr (reaction). The fraction that fluoresces depends on kF compared with the sum of all the competing rate constants.

Why fluorescence is more sensitive than absorbance [KNOW]

  • Absorbance: A = -log(P/P0) = ε b c. If you raise P0, P rises with it, so A does not change.

  • Fluorescence: F = k'(P0 - P). F depends on the number of excited states created, so more P0 gives more excited states and more F.

  • Detection is at a different wavelength from excitation, so background is near zero.

  • Result: sub-ppb LOD and linearity over 4 to 6 orders of magnitude.

Deviations from linearity at high concentration or high ε

  • Higher-order terms (Taylor series) curve the plot. Fix: dilute.

  • Self-quenching: collisional deactivation (external conversion) of excited molecules.

  • Self-absorption: emitted photons match an energy gap in other analyte molecules and get absorbed. Because emission is isotropic, the re-emission goes in random directions, so detected F power falls.

Instruments

  • Fluorimeter: filters, no scanning. Source, filter, sample, filter, detector (detector at right angles to the source beam).

  • Spectrofluorimeter: gratings; an excitation monochromator and an emission monochromator. Double-beam versions add a reference PMT and a sample PMT with a difference amplifier.

Fluorescence summary

  • Good selectivity: most things don't fluoresce (tag those that don't). Aromatics often fluoresce, π to π* is favored, and polycyclic aromatic hydrocarbons fluoresce readily.

  • Fluorescence is more common than phosphorescence; typically you see only one.

  • Moderately expensive: pulsed lamps or lasers, two good gratings, very sensitive PMTs.

VI.E Master comparison [KNOW]

Technique

Qualitative

Quantitative

Selectivity

Sensitivity

Beer's-law linearity

Flame AAS

No (needs lamp)

Excellent (1 to 2%)

Very, if interferences handled

LOD 0.3 ppb to 20 ppm

Relatively poor

GFAAS

No (needs lamp)

Very good (1 to 5%)

Very, if interferences handled

Better than flame

OK

AES

Excellent

Excellent (1 to 2%)

Very, if interferences handled

Sub ppb

Excellent

UV-vis absorption

Poor

Very good (1 to 3%)

Not great (broad bands)

1e-4 to 1e-7 M, depends on ε

OK

Fluorescence / phosphorescence

Poor

Excellent

Very

Sub ppb

Excellent (best)


RAPID-FIRE REVIEW

Equations

  • E = hν = hc/λ

  • A = -log(T) = ε b c

  • Spin multiplicity = 2S + 1

  • F = k'(P0 - P)

  • Atomic degeneracy: 2J + 1 (slide shows 2l + 1); σ and n: 1; π: 2; vibration: 1; rotation: 2J + 1

Likely exam traps

  • Single PMT: sample after the grating. Diode array: sample before.

  • Prisms: non-linear dispersion. Gratings: linear dispersion but more stray light and overtones.

  • Fluorescence keeps the same multiplicity; phosphorescence changes it, so it is slower.

  • In AAS you want atomization, not excitation, ionization or oxide formation.

  • Releasing agent binds the interfering anion; protecting agent binds the analyte.

  • Ionization fix: add an easily ionized alkali metal. Oxide fix: add V or Ti.

  • Plasma is reducing, which is why ICP suppresses oxides and ionization.

  • Quantify at λmax, and aim for A of about 0.18 to 0.82 (15 to 65% T).

  • Double beam fixes source drift and flicker; single beam does not.

  • Fluorescence beats absorbance in sensitivity because P0 helps F but cancels in A.

  • Standard addition: x-intercept is the negative of the starting concentration.

Slide inconsistencies to confirm with your instructor

  • AAS cost: $50 to 75K on the AAS summary slide versus about $30 to 50K in the AAS/ICP-AES table.

  • Deuterium lamp range: 180 to 370 nm (source slide) versus 160 to 375 (summary table).

  • Atomic electronic degeneracy printed as 2l + 1, likely meant 2J + 1.

  • Na D line quoted as 589.3 nm in one place and 589.0 nm in another.