Advanced Biosensor and Microfluidic Technologies for Rapid Pathogen Diagnosis and Cancer Research

Rapid Pathogen Diagnosis

  • Current Limitations:

    • Pathogen identification currently takes up to 7272 hours, often involving PCR-based methods, leading to delays in patient treatment.

  • Novel Approach:

    • Significantly reduces diagnosis time to approximately 22 hours.

    • Utilizes biosensors capable of identifying pathogens even at very low concentrations.

  • Biosensor Mechanism:

    • Carbon Nanotube (CNT): Acts as a probe delivery system, facilitating the entry of probes into bacteria via electroporation.

    • RNA Probe: A single-stranded probe that is complementary to a specific bacterial gene RNA (e.g., 16S16S rRNA).

    • Fluorophore-Quencher System:

      • In the absence of target bacterial RNA, a shorter quencher strand is bound to the fluorophore on the probe. This binding causes fluorescence quenching (no signal) due to a relatively higher free energy of the quencher-probe complex.

      • When target bacterial RNA is present, the probe binds to the target RNA, which leads to the detachment of the fluorophore from the quencher, resulting in the release of a detectable fluorescent signal.

  • Specificity and Detection Capability:

    • Successfully identifies E. coli (indicated by a red signal) and P. aeruginosa (indicated by a green signal) simultaneously within the same experiment.

    • Demonstrated effective detection even with very low bacterial counts present in the sample.

    • Experimental videos confirm the system functions with viable pathogens, as bacteria are observed moving.

High-Throughput Antibiotic Resistance Screening

  • Microfluidic Platform:

    • Bacteria are introduced into individual microfluidic channels, with each channel containing a small, defined population (e.g., a "column" of bacteria).

  • Growth Monitoring:

    • The system tracks bacterial growth over time, observing an increase in length or count (e.g., from one entity to ten entities) in the absence of antibiotic treatment.

  • Resistance Assay Procedure:

    • Sample Preparation: Healthy human blood samples cultured with bacteria, as well as patient samples positive for E. coli, are loaded into the microchannels.

    • Incubation and Treatment: Samples are incubated for 3030 minutes, after which specific antibiotics are introduced.

    • Outcome Assessment:

      • If the bacteria are sensitive to the applied antibiotic, their growth ceases (e.g., only one original cell remains).

      • If the bacteria are resistant, their growth continues uninterrupted.

    • Time Efficiency: This method significantly reduces the time for antibiotic resistance determination to approximately 11 hour, a dramatic improvement over traditional multi-day methods.

    • Example: Sample number 66 was specifically identified as resistant to antibiotic C.

Pathogen Identification Based on Physical Properties

  • Core Principle: Utilizes a microfluidic device to separate and identify bacterial pathogens based on their distinct size and shape characteristics.

  • Separation Mechanism:

    • Bacteria are introduced into microfluidic channels, and varying pressures are applied.

    • Size-Based Trapping: Bacteria with larger sizes are easily trapped at lower fluid pressures. Conversely, smaller bacteria require higher pressures to be effectively trapped and separated.

  • Shape-Based Classification:

    • Coccus: Identifies spherical-shaped bacteria.

    • Bacillus: Identifies rod-shaped bacteria.

    • These morphological classifications are performed by observing bacterial behavior under different applied pressures on the microscope.

  • Clinical Application and Benefits:

    • Successfully applied to clinical samples. For instance, in a cohort of 2525 patients, 77 patients were found to have pathogens resistant to