Preliminary Steps in Radiography
Introduction to Radiography and Professional Practice
Definitions and Professional Roles:
- Radiologic Technology: A healthcare profession encompassing all diagnostic imaging technologists and radiation therapists.
- Radiographer: A radiologic technologist who administers ionizing radiation to perform radiographic procedures, producing diagnostic images at the request of a licensed medical practitioner.
- Required Knowledge & Skills: Effective clinical practice requires combining technical skills with extensive knowledge of physics, anatomy, physiology, and pathology.
Radiography Practice Standards:
- Formulated, maintained, and published by the American Society of Radiologic Technologists (ASRT).
- Define the scope of practice for radiography, detail educational and credentialing requirements, and set clear professional benchmarks.
- Section Components:
- Clinical Performance Standards: Define activities regarding patient care, operational procedures, and diagnostic evaluation.
- Quality Performance Standards: Define activities regarding equipment operation, safety, and quality assurance.
- Professional Performance Standards: Define expectations for professional conduct, ethics, and ongoing education.
- The current version is maintained on the ASRT website (
ASRT.org).
Ethics in Radiologic Technology:
- Definition: Ethics represents a health professional's moral responsibility and the science of appropriate conduct toward patients and healthcare team members.
- ARRT Standards of Ethics: Developed and maintained by the American Registry of Radiologic Technologists (ARRT) to translate professional values into concrete practice.
- Code of Ethics: Composed of 10 statements serving as an aspirational guide for professional behavior.
- Rules of Ethics: Composed of 22 mandatory, enforceable rules defining minimally acceptable professional conduct.
- CAMRT Code of Ethics: Developed by the Canadian Association of Medical Radiation Technologists (CAMRT) to delineate ethical expectations for practitioners in Canada (
CAMRT.ca).
Advanced Clinical Practice:
- Advanced practice roles enable experienced radiographers to act as a "radiologist extender" to perform or assist with:
- Patient care activities.
- Selected radiologic procedures.
- Initial image observations.
- Professional Titles:
- Radiologist Assistant (RA)
- Radiology Practitioner Assistant (RPA)
- Credentialing Requirements: Achieving RA or RPA status requires ARRT radiography certification, completion of advanced formal education, and supervised clinical experience under a radiologist preceptor, followed by passing an advanced-level certification examination.
Room Preparation and Infection Control
Care of the Examination Room:
- Radiographer Responsibility: Maintaining strict cleanliness of the examination room and equipment (X-ray tube housing, tabletop, and accessories) to minimize pathogen transmission and promote patient confidence.
- Room Preparation Protocols:
- Disinfect all contact surfaces prior to patient arrival.
- Ensure the room is clean, organized, and stocked with necessary positioning aids.
- Place a fresh pillowcase on positioning pillows.
Control of Pathogen Contamination:
- The Centers for Disease Control and Prevention (CDC) provides guidelines for infection control, with Standard Precautions for All Patient Care serving as the foundation.
- Core Standard Precautions:
- Hand Hygiene: Perform hand hygiene before and after each patient interaction.
- Personal Protective Equipment (PPE): Use PPE (gloves, gowns, masks, eye protection) whenever exposure to infectious material is anticipated.
- Respiratory Hygiene / Cough Etiquette: Implement measures to contain respiratory secretions.
- Patient Placement: Ensure appropriate isolation or bed assignment based on infection transmission risks.
- Equipment Disinfection: Properly clean and disinfect patient-care instruments, equipment, and environmental surfaces.
- Textiles & Laundry: Handle contaminated linens carefully to prevent pathogen dispersion.
- Safe Injection Practices: Follow strict aseptic technique; wear a surgical mask during lumbar punctures.
- Sharps Safety: Ensure safe handling and disposal of needles and sharp devices.
- Transmission-Based Precautions: Applied in addition to Standard Precautions for patients with known or suspected airborne, droplet, or contact transmissible infections.
Handwashing Practices:
- Handwashing is the single easiest and most convenient method to prevent the spread of microorganisms.
- Mandatory Handwashing Indications:
- Before and after caring for every individual patient.
- After examining patients with known communicable diseases.
- After coming into contact with blood, bodily fluids, secretions, or excretions.
- Prior to starting any invasive procedure.
- Prior to touching high-risk, immunocompromised, or vulnerable patients.

Patient Care, Communication, and Motion Management
- Basic Patient Care & Communication:
- Patient Explanation: Technologists must provide coherent patients with a clear explanation of the procedure, detailing what to expect, potential discomfort, and instructions to ensure comfort and compliance.
- Clinical History Collection:
- Verifying the correct exam was ordered against clinical indications.
- Observing physical conditions or anatomical abnormalities to relay to the radiologist.

- Types and Control of Motion:
- Uncontrolled motion during exposure severely degrades recorded image detail and spatial resolution.

- Involuntary Motion:
- Causes: Heartbeat, chills, peristalsis, muscle tremors, spasms, and severe pain.
- Primary Control Mechanism: Utilizing the shortest possible exposure time ().
- Voluntary Motion:
- Causes: Nervousness, discomfort, excitability, mental illness, fear, age (pediatric or geriatric), and uncontrolled breathing.
- Control Mechanisms: Giving clear instructions, providing patient comfort, adjusting positioning support devices, and applying physical immobilization.
- Uncooperative Patients: For patients unable to cooperate (e.g., young children, confused elderly, or patients with severe mental illness), decreasing exposure time is the primary method of motion control.
- Equipment Motion: Unintended movement originating from the X-ray tube mount, table, or cassette tray.

Image Receptors and Accessory Equipment
- Image Receptor (IR) Definition & Types:
- An Image Receptor is defined as a device that receives the energy of the remnant X-ray beam and forms the diagnostic image of the body part.
- Four Primary Types:
- Solid-State Digital Detectors: Flat-panel digital radiography (DR) detectors utilizing amorphous silicon or amorphous selenium, available as fixed tabletop/bucky units, tethered panels, or wireless plates.
- Photostimulable Storage Phosphor Image Plate (PSP IP): Computed Radiography (CR) plates coated with photostimulable phosphor that store latent image energy until read by a laser scanner inside a CR reader.
- Fluoroscopic Image Receptor: Real-time imaging systems using either an image intensifier tube coupled to a television camera/monitor or a flat-panel digital fluoroscopy detector.
- Cassette with Film: Traditional screen-film cassettes requiring chemical film processing.




- Accessory Equipment:
- Positioning Aids: Radiolucent sponges, sandbags, and specialized leg holders (e.g., Ferlic leg holder) used to stabilize the body part.
- Grids: Precision devices composed of lead strips placed between the patient and the IR to absorb scattered radiation, enhancing contrast.
- Compensating Filters: Absorptive devices placed in the X-ray beam path to compensate for variations in tissue thickness and density, yielding uniform image brightness.
Radiographic Technical Factors and Exposure Technique Charts
Prime Technical Factors:
- Milliamperage (mA): Controls X-ray photon quantity and rate of exposure.
- Kilovolt Peak (kVp): Controls X-ray beam energy, quality, and penetrability.
- Exposure Time (seconds): Duration of beam activation; multiplied by mA to determine total milliampere-seconds ().
- Control Systems: Manual technical factor selection or Automatic Exposure Control (AEC) chambers.
Technique Charts:
- Standardized charts must be available in every examination room and attached to all mobile units.
- Foundation Parameters Required for Technique Chart Establishment:
- Milliampere-seconds ()
- Kilovolt peak ()
- Automatic Exposure Control () chamber selection and sensitivity
- Source-to-Image Receptor Distance ()
- Relative patient or part thickness ()
- Grid selection and ratio
- CR/DR exposure indicators
- IR or collimated field dimensions
- Electrical supply characteristics (generator phase and frequency)

- Essential Projections Technical Reference Data (Thoracic Viscera):
- Chest: Lungs and Heart — PA Projection: Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
- Chest: Lungs and Heart — Lateral Projection: Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
- Chest: Lungs and Heart — PA Oblique Projection: Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
- Chest: Lungs and Heart — AP Projection (Nongrid, 40'' SID): Part thickness: ; ; ; Collimation: (); CR: , Dose: .
- Chest: Lungs and Heart — AP Projection (DR Nongrid, 40'' SID): Part thickness: ; ; ; Collimation: (); DR: , Dose: .
- Chest: Lungs and Heart — AP Projection (Bucky/Grid, 72'' SID): Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
- Pulmonary Apices — AP Axial Projection: Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
- Lungs and Pleurae — Lateral Decubitus Projection: Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
- Lungs and Pleurae — Dorsal / Ventral Decubitus Projection: Part thickness: ; ; ; Collimation: (); CR: , Dose: ; DR: , Dose: .
Radiation Protection, Geometry, and Beam Alignment
- Gonad Shielding Guidelines:
- AAPM position statements, endorsed by the ACR and other professional bodies, indicate routine gonadal shielding can negatively impact examination efficacy (e.g., obscuring anatomy or interfering with AEC).
- State regulations may still mandate gonadal shielding; technologists must remain aware of institutional and regulatory guidelines.

- Image Receptor Placement Configurations:
- Lengthwise Position: Long axis of the IR aligned parallel with the long axis of the body (most commonly used, e.g., AP Abdomen).
- Crosswise Position: Long axis of the IR aligned perpendicular to the long axis of the body (e.g., AP Pelvis).
- Diagonal Position: IR rotated diagonally to cover long bones (e.g., AP Leg including knee and ankle joints).
- Built-in Flat-Panel Detectors: Typically (), movable lengthwise along the bucky track.

Central Ray (CR) Alignment:
- The central or principal photon beam is termed the Central Ray (CR).
- The CR is centered to the anatomy of interest and the IR.
- Standard goal: CR perpendicular to the part and IR to prevent distortion.
- Reasons for Angling the CR:
- Avoid superimposition of structures.
- "Straighten out" curved structures.
- Align the beam through an angled joint space.
- Avoid distortion of an angled structure.
Source-to-Image Receptor Distance (SID):
- Distance from the anode focal spot inside the X-ray tube to the IR surface.
- Key factor affecting magnification, spatial resolution, and patient dose.
- Standard Operating SIDs:
- (): Traditional standard for general examinations.
- (): Modern standard adopted across many facilities to reduce magnification.
- (): Standard for chest and cervical spine imaging to minimize Object-to-Image Receptor Distance (OID) magnification.
- Specialized projections in Merrill's Atlas range from () to (); a minimum of () is recommended if unspecified.
Source-to-Skin Distance (SSD):
- Distance from the X-ray tube focal spot to the patient's skin surface.
- Influences patient entrance exposure dose.
- NCRP Guidelines: SSD shall not be less than () and should not be less than ().

- Collimation and Shuttering Rules:
- Restricting the radiation field size to include only the anatomy of interest.
- Two Primary Purposes:
- Minimizes patient radiation dose.
- Reduces scatter radiation, improving contrast resolution.
- Ethical Mandate: Collimating larger than the required field size violates the ARRT Code of Ethics and ASRT Practice Standards.

- Shuttering (Electronic Cropping):
- Software process adding a dark border around a collimated field in DR systems for viewing aesthetics.
- Prohibition against Misuse: Electronic shuttering must never be used to replace physical beam collimation. Post-exposure cropping increases patient dose and creates legal liability if cropped diagnostic information leads to a missed diagnosis.
Anatomic Markers, Image Display, and Evaluation
Anatomic Side Markers:
- Medicolegal requirements mandate physical lead Right () or Left () markers on all radiographs prior to exposure.
- Rules for Marker Placement:
- Must be visible on the image without obscuring anatomy.
- Placed within the collimation field border.
- Positioned outside of lead shielding.
- Post-acquisition digital annotation is discouraged due to the risk of lateral site identification errors.
Image Display Standards:
- Radiographs are viewed in standard anatomic position.

- Standard Display Exceptions: Images of extremities—specifically hands, fingers, wrists, feet, and toes—are displayed as seen from the perspective of the X-ray tube, with digits pointing upward toward the ceiling.

- Required Image Identification:
- Date of examination.
- Patient name or identification number.
- Anatomic side marker ( or ).
- Healthcare institution name.

- Image Evaluation Criteria:
- Technologists must evaluate every image for acceptability, safety compliance, and procedural goals.
- Criteria include verifying patient ID, marker placement, collimation, shielding, and absence of technical artifacts.
Radiographic Considerations for Obese Patients
Obesity Demographics and Definitions:
- Obesity prevalence has doubled in the last 15 years; of American adults are overweight, obese, or morbidly obese.
- Body Mass Index (BMI) Classifications:
- Obese:
- Morbidly Obese:
- Body weight and diameter determine whether radiographic procedures can be performed.
Equipment Considerations:
- Weight limits on radiographic tables; older tables held , while redesigned modern tables support up to .
- Diameter limits on fluoroscopy towers, CT scanners, and MRI units.
- Mobile DR detectors facilitate positioning by providing greater tube-to-table clearance.
Patient Transfer and Positioning:
- Utilization of larger wheelchairs, heavy-duty stretchers, team movement, proper body mechanics, transfer boards, air mattresses, and ceiling-mounted power lifts.


Communication Guidelines:
- Use empathetic communication, avoid mentioning weight, and clearly explain movement and positioning steps.
Landmark Location in Morbidly Obese Patients:
- Most surface bony landmarks are unpalpable due to adipose tissue.
- Jugular Notch Landmark Method: Locate the palpable jugular notch and measure inferiorly along the midline to locate the pubic symphysis plane based on height:
- Height under (): Pubic symphysis is () inferior to jugular notch.
- Height (): Pubic symphysis is () inferior to jugular notch.
- Height over (): Pubic symphysis is () inferior to jugular notch.

Questions & Discussion
Question 1: What is the easiest method to control the spread of pathogens?
- Options: A. Use of mask on patients | B. Radiographers wearing masks | C. Wearing gloves for all procedures | D. Hand hygiene
- Answer: D. Hand hygiene
- Explanation: Hand hygiene is the simplest and most effective method to prevent cross-contamination and pathogen transmission in medical environments.
Question 2: All of the following are examples of involuntary motion, except:
- Options: A. Peristalsis | B. Heartbeat | C. Breathing | D. Chills
- Answer: C. Breathing
- Explanation: Respiration is under voluntary patient control and can be stopped with clear breathing instructions. Peristalsis, heartbeat, and chills are involuntary movements.
Question 3: How do radiographers help reduce the risk of imaging involuntary motion?
- Options: A. Use immobilization | B. Appropriate patient instructions | C. Clear communications | D. Use the shortest possible exposure time
- Answer: D. Use the shortest possible exposure time
- Explanation: Involuntary motion cannot be controlled by verbal instructions or immobilization. Decreasing exposure time to the minimum achievable duration is required.
Question 4: All of the following are required identification on a radiographic image, except the:
- Options: A. Side marker | B. Patient name or ID number | C. Referring physician's name | D. Name of the facility
- Answer: C. Referring physician's name
- Explanation: Standard legal image identification requires examination date, patient name/ID number, side marker ( or ), and institution name. The referring physician's name is not required on the image.
Question 5: A technique chart must include all of the following, except:
- Options: A. Pathology type | B. | C. | D. Grid
- Answer: A. Pathology type
- Explanation: Exposure technique charts detail technical factors (, , , grid, ) for specific anatomical projections, not pathology classifications.
Question 6: Which of the following will affect whether a radiographic examination can be performed on an obese patient?
- Options: A. Height | B. Weight | C. BMI | D. Communication skills
- Answer: B. Weight
- Explanation: Patient weight and body diameter directly determine equipment compatibility due to table load limits and gantry/aperture clearance dimensions.
Question 7: Which guideline is used to locate the pubic symphysis on an obese patient who is approximately tall?
- Options: A. from the jugular notch | B. from the jugular notch | C. from the jugular notch | D. from the jugular notch
- Answer: C. from the jugular notch
- Explanation: For patients between and tall, the pubic symphysis plane lies inferior to the jugular notch.