Protein Analysis: Biuret, Dye Binding, and Immunoturbidimetry Notes
Learning Objectives for Protein Analysis
Upon completion of the Protein Analysis unit (Outcome 1, Module 3), students should be able to:
Describe the Biuret Method for total protein according to:
Principle.
Major reagents used.
Instrumentation required.
Specimen requirements.
Sensitivity.
Sources of error.
Describe the dye binding and the immunoturbidimetry methods for albumin according to:
Principle.
Major reagents used.
Instrumentation required.
Specimen requirements.
Sensitivity.
Sources of error.
Total Protein Measurement: The Biuret Method
The Biuret reaction is the standard method for determining total protein concentration.
Principle of the Reaction:
The reaction involves ions reacting with the nitrogen atoms present in the peptide bonds of proteins.
This binding reaction produces a complex with a distinct violet color.
The resulting intensity of the violet color is measured using spectrophotometry.
Instrumentation:
Absorbance for the Biuret reaction is measured at a wavelength of .
Components of the Biuret Reagent
The Biuret reagent consists of several specific chemical components, each serving a functional purpose in the assay:
(Copper Sulfate): Dissolved with ; provides the ions necessary for the complex formation.
(Sodium Hydroxide): Provides the alkaline environment required for the reaction.
Sodium Potassium Tartrate: Added to the reagent mixture.
Potassium Iodide: Added to the reagent mixture.
Biuret Method Performance: Sensitivity and Specificity
Sensitivity:
The method is sensitive enough to accurately measure protein concentrations found in plasma specimens.
It is not suitable for specimens containing low protein concentrations, such as Urine or Cerebrospinal Fluid (CSF).
Specificity:
The Biuret reagent specifically reacts with peptide bonds.
It does not react with dipeptides or free amino acids.
Analytical Interferences in the Biuret Method
Several substances can interfere with the accuracy of the total protein measurement, leading to false results:
Hemoglobin:
Hemoglobin naturally absorbs light at .
This leads to a false positive result (falsely elevated protein concentration).
Lipemia:
Lipemic samples scatter light rather than allowing it to pass through cleanly.
This light scattering results in a false positive result.
Specimen Requirements and Procedural Impacts for Total Protein
Sample Type:
Both Serum and Plasma may be utilized for testing.
Plasma concentrations are typically slightly higher than serum concentrations (noting the presence of fibrinogen).
Specimen Stability:
Specimens are stable for up to week when stored at Room Temperature (RT).
Specimens remain stable for up to month if refrigerated.
Collection Factors Influencing Results:
Posture:
Changes in patient posture will alter the protein concentration.
Recumbent patients (lying down) typically show results that are to lower than ambulatory patients (moving around).
Tourniquet Time:
Prolonged application of a tourniquet leads to hemoconcentration.
This causes a false increase in the measured protein concentration.
Albumin Analysis via Dye Binding Methods
Principle:
The method is based on the ability of albumin to bind to a wide variety of organic dyes.
Measurement is based on the shift in the absorption maximum of the dye when it becomes bound to albumin.
The color change can be measured even in the presence of unbound dye in the solution.
Reagents (The Dyes):
Bromocresol Green (BCG):
This is the most common dye used in clinical settings.
It offers better sensitivity and precision for the assay.
Bromocresol Purple (BCP):
This dye has half of the absorption maximum when compared to BCG.
It is considered more specific for albumin than BCG.
Reaction Conditions for Albumin Dye Binding
Reaction pH:
The reaction is maintained at a pH of .
Charge Interaction:
At a pH of , Albumin takes on a positive charge.
This positive charge allows it to bind to the negatively charged BCG dye.
Instrumentation:
Absorbance is measured at a wavelength of .
Relationship:
The reaction is proportional; the amount of color/absorbance is directly related to the concentration of albumin.
Errors and Interferences in Albumin Dye Binding
Manual Method Errors:
Time is a critical factor; if the reaction is allowed to proceed too long, other proteins in the sample begin to react with the dye.
Accuracy and specificity of the assay are improved by taking absorbance readings shortly after mixing the sample and reagent.
Specimen Choice:
Serum is the preferred specimen of choice.
Chemical Interferences:
Hemoglobin:
Interferes with the actual binding of Albumin to BCG dye.
This results in a false increase in measured albumin.
Hemolysis:
Historically unsure of the exact mechanism, but it may increase the speed at which other (non-albumin) proteins bind to the BCG dye.
This results in a false increase.
Heparin:
Identified as a source of interference.
Collection Factors:
Posture:
Recumbent concentrations are to lower compared to ambulatory concentrations.
Tourniquet Time:
Long tourniquet application causes hemoconcentration, resulting in a false increase.
Albumin Analysis via Immunoturbidimetry
Principle:
The method utilizes Sheep Antibodies—specifically Anti-Albumin (Human).
These antibodies react with the albumin in the sample to form large immune complexes.
These complexes cause light to scatter.
Proportionality and Detection:
The concentration of albumin is directly proportional to the amount of light scatter or absorbance measured.
The concentration of albumin is inversely proportional to the amount of light detected at the sensor (as scatter increases, less light reaches the detector).
Instrumentation and Sensitivity:
This method can detect concentrations as low as .
It is significantly more sensitive than most dye binding methods.
Specimen Choice:
Serum is the preferred specimen of choice.
Sources of Error:
Lipemia: Lipids in the sample will increase light scatter, leading to erroneous results.
General Scatter: Anything that increases light scatter within the specimen will act as a source of error.