xray tube
Cathode assembly: Negative electrode in the x-ray tube a. Contains two filaments—small and large
b. Filaments are made of tungsten (because of its high melting point); a small amount of thorium is added to reduce vaporization and prolong tube life
c. Filaments are heated slightly when the x-ray machine is turned on; no electrons are ejected at this low level of heating
d. During x-ray exposure, one filament is heated to a level that causes electrons to be “boiled off” in preparation for x-ray production (which is known as thermionic emission)
e. Over time, filaments vaporize and coat the inner surface of the x-ray tube with tungsten, leading to tube failure
f. Cathode assembly also includes the focusing cup g. The focusing cup surrounds the filaments on three sides
h. The focusing cup has a negative charge applied, which tends to concentrate electrons boiling off the filaments into a narrower stream and repel them toward the anode
i. Electron concentration keeps the electrons aimed at a smaller area of the anode
j. Some x-ray tubes use the focusing cup as an electronic grid that can turn the tube current on and off rapidly, allowing for very short and precise exposure times, such as times needed for rapid serial expo sures (referred to as a grid-controlled tube)
Anode: Positive electrode in the x-ray tube
a. Consists of a metal target made of a tungsten– rhenium alloy (because of its high melting point and high atomic number) embedded in a disk (or base) of molybdenum with a motor to rotate the target
b. Must be able to tolerate extremely high levels of heat produced during x-ray production
Glass envelope with window
a. The cathode and target are inside the glass envelope; some models use a partial metal envelope
b. Glass envelope also contains a vacuum so that electrons from the filament do not collide with atoms
c. Tube window: thinner section of glass envelope that allows x-rays to escape
Tube housing: Encase the x-ray tube
a. Made of aluminum with lead lining
b. Supports and protects the tube, restricts leakage radiation during exposure, and provides electrical insulation
c. Anode rotates from 3300 to 10,000 rpm, depending on tube design
d. Rotation is achieved using of an induction motor located outside the x-ray tube, which turns a rotor located inside the x-ray tube; the target is attached at the end of the rotor
e. Rotation of the target allows for greater heat dissipation
f. Rotation of the target is stopped by a braking action provided by the induction motor
g. The x-ray machine should never be shut off immediately after an exposure; the machine should not be shut off until the target has stopped rotating
h. Without the braking action, the target may spin for 30 minutes, causing great strain on the bearings i. Electrons strike the target on the focal track (some times called the focal spot)
j. The focal track is beveled, producing the target angle
k. The target angle allows for a larger actual focal spot (area bombarded by electrons) and also produces a smaller effective focal spot (the area seen by the IR)
l. The larger the actual focal spot, the greater the heat capacity; the smaller the effective focal spot, the greater the radiographic image sharpness
m.This effect is called the line-focus principle
n. Target angle may be 7 to 20 degrees, depending on tube design
o. Exposure switch should be activated in one continuous motion, activating the rotor and then the exposure button
p. The equipment allows the rotor to come up to speed before making the exposure
q. The radiographer does not control this by acti vating the rotor and waiting to press the exposure button
r. Activating the rotor and allowing it to operate by itself results only in unnecessary heating of the filament and wear on the induction motor, shortening tube life
c. Also contains an oil in which the x-ray tube is immersed to assist with cooling and additional electrical insulation