Quality Control in Ultrasound Imaging

Ultrasound Quality Control: Test Objects and Phantoms

Introduction to Quality Control

  • Quality control is essential in imaging departments, especially those using ultrasound, to ensure equipment functions properly.
  • Testing is performed by service engineers, and some tests can be done by clinical sonographers.
  • Test phantoms or test objects are used to evaluate system performance.

Test Objects and Phantoms

  • Traditional Phantoms:
    • Fluid-filled with wires at known distances.
    • Used to check calibration by measuring distances between wires.
    • Used to assess resolution.
  • Modern Multipurpose Phantoms:
    • Hydrogel-based with a membrane for scanning.
    • Low attenuation coefficient (similar to water, e.g., 0.7 \, \text{dB/cm/MHz}) to mimic tissue properties.
    • Test resolution, calibration, and contrast resolution.
    • Contain structures that mimic cysts and echogenic masses to evaluate how well different structures are defined.
  • Speed of Sound:
    • Phantoms mimic soft tissue; the speed of sound within the phantom should match that of soft tissue.
    • Data sheets specify the speed of sound for the device (e.g., 1540 \, \text{m/s}).
  • Tissue Mimicking:
    • Phantoms are designed to mimic tissue structures, allowing for realistic system testing.

Types of Phantoms

  • General Ultrasound Training Phantoms:
    • Various phantoms are available for different scenarios, like breast phantoms.
  • Biopsy Phantoms:
    • Contain vessels for practicing needle tip location and insertion.
  • Pelvic Phantoms:
    • Useful for transvaginal ultrasound training, with simulated ovaries and a fluid-filled bladder to learn anatomical relationships.

Terminology

  • "Test objects" and "phantoms" can be used interchangeably.
  • Modern phantoms are multipurpose, testing various aspects of the ultrasound system, including calibration, resolution, and lesion detection.
  • Dedicated phantoms exist for specific tests, like slice thickness evaluation.

Resolution Types and Phantoms

  • Axial Resolution:
    • Phantoms help determine how close two structures can be, above and below each other, and still be resolved as separate entities.
  • Lateral Resolution:
    • Dependent on beam width (left to right on the image).
    • Phantoms contain a series of wires to assess how close structures can still be distinguished.
  • Slice Thickness:
    • Refers to the beam thickness in and out of the image plane.
    • Specialized phantoms measure slice thickness.

Uses of Test Objects

  • Compare different transducers.
  • Monitor a transducer's performance over time.
  • Assess registration: ensuring the location of a structure on the image matches its real-life location.

Registration

  • Correct placement of echoes is crucial.
  • Poor registration occurs when structures appear distorted on the image compared to their actual arrangement in the test object.

Main Bang Artifact

  • Some test objects have wires near the surface to test for the main bang artifact in the near field.
  • Different transducers show varying clarity in the near field.

Measuring Resolution

  • Axial Resolution:
    • Evaluated using closely spaced dots. If the dots are too close, they appear as one on the image.
  • Lateral Resolution:
    • Evaluated with closely spaced wires. The ability to distinguish the wires indicates lateral resolution.
    • Lateral resolution is influenced by beam width.
    • Lateral resolution can vary at different depths or focus settings.

Caliper Calibration

  • Test objects with known distances between wires are used to verify the accuracy of caliper measurements on the ultrasound system.
  • Measurements are taken in both vertical and horizontal planes.
  • A 2% error or less is generally considered acceptable, according to manufacturers.

Beam Profile

  • Test objects with small wires at increasing depths can visualize the beam profile.
  • The appearance of these wires indicates the beam's focus and width at different depths.
  • Traditional transmit focus methods may produce an hourglass-shaped beam profile, with the best lateral resolution at the focus.

Sensitivity and Lesion Detection

  • Phantoms often contain subtle lesions at various depths to test the system's sensitivity in detecting slight differences in echogenicity.
  • The ability to detect these lesions is assessed over time to monitor the transducer's performance.

Uniformity

  • The hydrogel within the phantom should have uniform density and acoustic properties.
  • After adjusting TGC settings, the image should appear uniform with consistent echogenicity and architecture throughout.
  • Patchiness suggests potential issues with the transducer or system at certain depths.

Dead Elements

  • Test phantoms can help detect dead elements in the transducer.
  • A clear vertical line on the image indicates a dead element, provided there is not too much compounding or post-processing.

Contrast Resolution

  • Specialized phantoms contain cones or simulated lesions with varying echogenicity.
  • Contrast resolution is determined by the smallest difference in echogenicity that is visible.
  • Harmonic imaging or different transducers can be compared to evaluate contrast resolution.

Slice Thickness Revisited

  • The ultrasound beam has a three-dimensional aspect including beam width and slice thickness (in and out of the image plane).
  • Echoes from regions outside the 2D scan plane reduce contrast resolution, so a thinner slice thickness improves contrast resolution.

Slice Thickness Phantom

  • Specialized phantoms contain a diagonal reflecting surface.
  • When scanned correctly, the image displays a band whose thickness indicates the relative slice thickness.

Other Quality Control Considerations

  • Printout image quality depends on the printer or camera (if applicable).
  • Monitor calibration is crucial for accurate image display.
  • Post-processing and offline measurement systems (e.g., PACS) must be calibrated correctly.