describing pulsed waves chapter 4

Pulsed Ultrasound Imaging Systems

  • Definition: Pulsed ultrasound imaging systems produce short bursts or pulses of acoustic energy to create anatomical images.

  • Characteristics of Pulsed Sound:

    • A pulse consists of individual cycles but propagates as a single unit.

    • Pulsed sound has a beginning and an end, traveling together.

Key Components of Pulsed Ultrasound

Pulse Duration
  • Definition: The actual time from the start of a pulse to the end of that pulse.

    • Represents the time during which the transducer is actively transmitting (on time).

  • Units: Microseconds (µs) - millionths of a second.

  • Typical Values: Ranges from 0.3 to 2.0 µs.

  • Determined By: Sound source.

  • Adjustable: No; the pulse duration cannot be altered.

Properties of Pulse Duration
  • Mathematical Relationship:

    • Directly proportional to the number of cycles in the pulse

    • Directly proportional to the cycle period (the time duration of one cycle)

    • Inversely proportional to frequency

  • Effects on Imaging Quality: Longer pulse duration (many cycles with long periods) results in lesser clarity, while shorter durations produce clearer images.

Spatial Pulse Length
  • Definition: The distance of a pulse from beginning to end.

  • Units: Millimeters (mm).

  • Typical Values: Soft tissue ranges from 0.1 to 1.0 mm.

  • Determined By: Both the source and medium.

  • Adjustable: No; a transducer has a fixed spatial pulse length.

Properties of Spatial Pulse Length
  • Mathematical Relationship:

    • Directly proportional to the number of cycles in the pulse

    • Directly proportional to the wavelength

    • Inversely proportional to frequency

  • Imaging Quality: Shorter spatial pulse lengths are more desirable for accurate imaging, whereas longer lengths may compromise quality.

Pulse Repetition Period (PRP)
  • Definition: The time from the start of one pulse to the start of the next pulse.

  • Units: Milliseconds (ms) and microseconds (µs).

  • Typical Values: Ranges from 100 µs to 1 ms (1,000 µs).

  • Determined By: Sound source and imaging depth.

  • Adjustable: Yes; sonographers can adjust the scan's depth of view.

PRP and Depth Relationship
  • Inverse Relationship:

    • Shallow depths result in shorter PRP.

    • Deeper structures lead to longer PRP.

  • Components of PRP: Comprised of the transmit time (on time or pulse duration) and the receive time (off time).

Pulse Repetition Frequency (PRF)
  • Definition: The number of pulses an ultrasound system transmits into the body each second.

  • Units: Hertz (Hz) - pulses per second.

  • Typical Values: Ranges from 1,000 to 10,000 Hz.

  • Determined By: Sound source and maximum imaging depth of the system.

  • Adjustable: Yes; adjusting the depth of view alters the PRF.

  • Relationship to Depth:

    • Inversely related: When depth of view increases, PRF decreases.

    • When depth decreases, PRF increases.

  • PRP vs PRF: They share an inverse relationship, explained mathematically as:

    • Example Calculations:

    • PRP in milliseconds (ms) relates to kilohertz (kHz)

    • Duty Factor

  • Definition: The percentage fraction of time the system transmits a pulse.

  • Units: None (expressed as a percentage).

  • Typical Values: Ranges from 0.002 to 0.005 (or 0.2 to 0.5%), with 99.8% receiving in continuous wave sound, which equals 1 (100%).

  • Determined By: Sound source.

  • Adjustable: Yes; the duty factor can be changed by altering image depth.

Duty Factor Calculation:
Summary of Duty Factor Values
  • Max Value: 1 or 100% in continuous wave sound.

  • Min Value: 0%. Typical value for duty factor is around 0.2%, indicating the system is listening for 500 times longer than it's transmitting.

Relationship Summary

  • When adjusting imaging depth, several parameters change:

    • PRP, PRF, and duty factor change.

  • Single cycle can describe both pulsed and continuous waves.

  • Continuous Doppler: 100% duty factor;

  • Pulsed Doppler: Typically has around a 0.2% duty factor.

Imaging Examples

  • Liver imaging depth is typically around 16 cm.

  • Thyroid imaging depth is typically around 4 cm.