describing pulsed waves chapter 4
Pulsed Ultrasound Imaging Systems
Definition: Pulsed ultrasound imaging systems produce short bursts or pulses of acoustic energy to create anatomical images.
Characteristics of Pulsed Sound:
A pulse consists of individual cycles but propagates as a single unit.
Pulsed sound has a beginning and an end, traveling together.
Key Components of Pulsed Ultrasound
Pulse Duration
Definition: The actual time from the start of a pulse to the end of that pulse.
Represents the time during which the transducer is actively transmitting (on time).
Units: Microseconds (µs) - millionths of a second.
Typical Values: Ranges from 0.3 to 2.0 µs.
Determined By: Sound source.
Adjustable: No; the pulse duration cannot be altered.
Properties of Pulse Duration
Mathematical Relationship:
Directly proportional to the number of cycles in the pulse
Directly proportional to the cycle period (the time duration of one cycle)
Inversely proportional to frequency
Effects on Imaging Quality: Longer pulse duration (many cycles with long periods) results in lesser clarity, while shorter durations produce clearer images.
Spatial Pulse Length
Definition: The distance of a pulse from beginning to end.
Units: Millimeters (mm).
Typical Values: Soft tissue ranges from 0.1 to 1.0 mm.
Determined By: Both the source and medium.
Adjustable: No; a transducer has a fixed spatial pulse length.
Properties of Spatial Pulse Length
Mathematical Relationship:
Directly proportional to the number of cycles in the pulse
Directly proportional to the wavelength
Inversely proportional to frequency
Imaging Quality: Shorter spatial pulse lengths are more desirable for accurate imaging, whereas longer lengths may compromise quality.
Pulse Repetition Period (PRP)
Definition: The time from the start of one pulse to the start of the next pulse.
Units: Milliseconds (ms) and microseconds (µs).
Typical Values: Ranges from 100 µs to 1 ms (1,000 µs).
Determined By: Sound source and imaging depth.
Adjustable: Yes; sonographers can adjust the scan's depth of view.
PRP and Depth Relationship
Inverse Relationship:
Shallow depths result in shorter PRP.
Deeper structures lead to longer PRP.
Components of PRP: Comprised of the transmit time (on time or pulse duration) and the receive time (off time).
Pulse Repetition Frequency (PRF)
Definition: The number of pulses an ultrasound system transmits into the body each second.
Units: Hertz (Hz) - pulses per second.
Typical Values: Ranges from 1,000 to 10,000 Hz.
Determined By: Sound source and maximum imaging depth of the system.
Adjustable: Yes; adjusting the depth of view alters the PRF.
Relationship to Depth:
Inversely related: When depth of view increases, PRF decreases.
When depth decreases, PRF increases.
PRP vs PRF: They share an inverse relationship, explained mathematically as:
Example Calculations:
PRP in milliseconds (ms) relates to kilohertz (kHz)
Duty Factor
Definition: The percentage fraction of time the system transmits a pulse.
Units: None (expressed as a percentage).
Typical Values: Ranges from 0.002 to 0.005 (or 0.2 to 0.5%), with 99.8% receiving in continuous wave sound, which equals 1 (100%).
Determined By: Sound source.
Adjustable: Yes; the duty factor can be changed by altering image depth.
Duty Factor Calculation:
Summary of Duty Factor Values
Max Value: 1 or 100% in continuous wave sound.
Min Value: 0%. Typical value for duty factor is around 0.2%, indicating the system is listening for 500 times longer than it's transmitting.
Relationship Summary
When adjusting imaging depth, several parameters change:
PRP, PRF, and duty factor change.
Single cycle can describe both pulsed and continuous waves.
Continuous Doppler: 100% duty factor;
Pulsed Doppler: Typically has around a 0.2% duty factor.
Imaging Examples
Liver imaging depth is typically around 16 cm.
Thyroid imaging depth is typically around 4 cm.