Comprehensive Notes – Spirometry, Oxygen & CO₂ Monitoring, and Indirect Calorimetry

Application Scenarios – Spirometry & Lung-Volume Measurements

  • Bellows spirometer reads lower exhaled volumes than hospital lab
    • No leaks & patient unchanged → most likely error: failure to correct to BTPS (Body Temperature, Pressure, Saturated) conditions.
    • Exhaled gas cools on the way to bellows ⇒ volume shrinks.
    • Correcting to 37C37\,^{\circ}\text C restores true lung volume.
  • Hand-held turbine (respirometer) used for FVC
    • Patient blows an apparent 5L5\,\text L.
    • Device accurate only when peak flow ≤ 300L⋅min1300\,\text{L·min}^{-1}.
    • Higher flows add inertia → vanes overspin → falsely high volume.
  • Post-op abdominal patient, shallow breaths on bed rest
    • Start incentive spirometry to give visual feedback & encourage deep breaths.
    • Adequate analgesia remains essential to prevent splinting.

Objective 1 – Oxygen Analysis

  • Clinical need: know exact FiO2FiO_2 delivered.
  • General accuracy: ±2%2\% of reading (e.g., 30 % ⇒ actual 28–32 %).
Polarographic Analyzer (Clark electrode)
  • Electrochemical cell; measures partial pressure, converts to %O₂.
  • O₂ diffuses across Teflon membrane → current proportional to P<em>O</em>2P<em>{O</em>2} → display %.
  • Pressure-dependent: higher ambient/barometric or circuit pressure exaggerates reading.
  • Calibrate at both 21%21\% & 100%100\% O₂; failure ⇒ replace electrode.
  • Battery-powered; continuous or spot use.
Galvanic Cell Analyzer
  • Also electrochemical & pressure-dependent.
  • Generates its own current; no external power.
  • Slower response but often longer cell life.
Electrical (paramagnetic/thermal) O₂ analyzers
  • Common in PFT labs (e.g., helium dilution), less in bedside care.

Objective 2 – Pulse Oximetry

  • Provides continuous or intermittent SpO₂; detects only functional hemoglobin (HbO₂ + Hb).
Physics
  • Spectrophotometry: measures differential absorption at two λ (usually 660 nm red, 940 nm IR) by HbO₂ vs. Hb.
  • Photoplethysmography: isolates pulsatile change in light absorption to distinguish arterial blood from static tissues.
  • Accuracy ±24%2–4\% for SpO₂ 70–100 % (good perfusion, no interference).
Probe Considerations
  • Secure, limit ambient light; choose site (finger, earlobe, forehead) with adequate perfusion.
Interferences
  • Intravascular dyes: methylene blue, indigo carmine ↓ SpO₂; indocyanine green minimal effect.
  • Motion, low perfusion, vasoconstriction, dark skin, nail polish, ambient light all degrade accuracy.
  • Dysfunctional Hb (COHb, MetHb) not differentiated → CO poisoning falsely high SpO₂.
SpO₂ vs. PaO₂ Relationship
  • Follows O₂-Hb dissociation curve.
    • PaO<em>2=60mmHgSpO</em>290%PaO<em>2 = 60\,\text{mmHg} \Rightarrow SpO</em>2 \approx 90\%.
    • Curve plateau: changes in PaO₂ >100 mmHg barely alter SpO₂.
Pulse-Ox vs. Co-oximetry (lab gold standard)
  • Co-oximetry uses ≥4 wavelengths, measures SaO₂ + COHb + MetHb; accuracy ±1–2 %.
  • Pulse-ox cheap, continuous, but blind to dyshemoglobins.
CPG Highlights
  • Indications: monitor O₂Hb saturation; evaluate response to therapy; guide during bronchoscopy.
  • Contra-indications: need for pH/PaCO₂/total Hb; abnormal Hb.
  • Record: probe type/site, FiO₂, patient appearance, HR agreement, simultaneous ABGs.

Objective 3 – Capnometry & Capnography

  • Measures P<em>CO</em>2P<em>{CO</em>2} in exhaled gas; waveform = capnogram.
Technologies
  1. Colorimetric (disposable): pH-indicator paper turns yellow with >0.5%0.5\% CO₂ → confirms tracheal intubation.
  2. Infrared Spectroscopy (most common)
    • CO₂ absorbs IR at 4.26μm4.26\,\mu m.
    • Types: single-beam with rotating chopper cells; double-beam with reference gas.
    • Interference: H<em>2OH<em>2O, N</em>2ON</em>2O absorb nearby wavelengths → drying agents & filters used.
    • Configurations:
      Sidestream—small sample suctioned (≈0.5 L·min⁻¹); delay, clog risk.
      Mainstream—sensor inline at airway; zero delay but adds dead space & weight.
  3. Mass Spectrometry: ions separated by mass/charge; expensive; multi-patient ORs.
  4. Raman scattering (rare).
Capnogram Phases
  • I : anatomic dead space (no CO₂).
  • II : rapid up-stroke (mix dead + alveolar).
  • III : alveolar plateau; slope indicates V/Q inhomogeneity.
  • IV : sharp drop on inspiration.
Abnormal Waveforms & Causes
  • Prolonged Phase III, no plateau → obstructive diseases (e.g., COPD).
  • Rebreathing, oscillations, curare cleft, hyper/hypoventilation, Cheyne-Stokes all have distinctive shapes.
Clinical Uses
  1. Confirm airway placement & monitor supraglottic devices.
  2. Evaluate CPR quality (EtCO₂ ∝ pulmonary perfusion).
  3. Optimize ventilation, detect circuit disconnections.
  4. Trend pulmonary disease severity & treatment response.
  5. Estimate metabolic rate (volumetric CO₂ elimination).

Objective 4 – Transcutaneous P<em>O</em>2P<em>{O</em>2} & P<em>CO</em>2P<em>{CO</em>2}

  • Heated skin electrodes (42–45C45^{\circ}\text C) increase perfusion & melt lipids → better diffusion.
Electrode Types
  • PtcO₂: Clark polarographic with Teflon membrane.
  • PtcCO₂: Stowe-Severinghaus pH glass with bicarbonate buffer.
Accuracy & Influences
  • Best in neonates (thin skin, good perfusion).
  • Hypotension, vasopressors, edema, thick skin ↓ accuracy.
  • PtcCO₂ slightly overestimates PaCO₂ (local metabolism); PtcO₂ often underestimates PaO₂ (skin consumption).
Technical Points
  • Calibrate before each placement; change site q 4–6 h to prevent burns.
  • Heater power ↑ implies ↑ perfusion → usually good; ↓ may indicate poor blood flow or sensor fault.
  • Validate against ABG at initiation & periodically.
  • Hazards: skin injury, false results leading to wrong therapy.
Indications
  1. Trend adequacy of oxygenation/ventilation.
  2. Evaluate therapy; early hypoperfusion index (PtcO₂/FiO₂).
  3. Wound care, peripheral revascularization assessment.

Objective 5 – Calorimetry (Nutrition–Respiration Interface)

Importance
  • Malnutrition ⇒ diaphragm weakness, ↓ immune defense.
  • Overfeeding ⇒ ↑ VCO2VCO_2, harder to ventilate; excessive carbohydrate especially raises CO₂ burden.
Indirect Calorimetry
  • Measures VO<em>2VO<em>2 & VCO</em>2VCO</em>2 to compute energy expenditure.
  • Respiratory Quotient (RQ)RQ=VCO<em>2VO</em>2RQ = \frac{VCO<em>2}{VO</em>2}
    • Carbohydrate ≈ 1.0, Fat ≈ 0.7, Protein ≈ 0.8.
    • RQ > 1 ⇒ lipogenesis/overfeeding; RQ < 0.7 ⇒ underfeeding/ketosis.
Open-Circuit Method
  • Separate inspiratory/expiratory flows; FiO₂ stable (<0.60 ideal); leaks invalidate data.
Closed-Circuit Method
  • Patient rebreathes O₂-filled bag; CO₂ absorbed; O₂ use inferred from volume drop; bulky but no FiO₂ limit.
Advantages over Predictive Equations
  • Real-time, patient-specific; prevents over/underfeeding; works on ventilated pts.

Supplementary Clinical Scenarios

  • Blender set 80 %, analyzer reads 70 % → analyzer not calibrated at 21 % & 100 %.
  • Smoke-inhalation pt. with SpO₂ 100 % on 5 L NC
    • Do not wean O₂.
    • Pulse-ox misreads COHb as HbO₂ → false high.
    • Treat with 100 % high-flow O₂ to displace CO.
  • COPD pt. on ventilator shows slow-rising Phase III → high airway resistance causing slow alveolar emptying.
  • Transcutaneous monitor power ↑ 5 % → heater compensating for ↑ local perfusion; beneficial, expect improved accuracy.

Key Numerical & Formula Summary

  • Polarographic/galvanic O₂ analyzer accuracy: ±2%2\%.
  • Turbine respirometer accurate flows ≤300L⋅min1300\,\text{L·min}^{-1}.
  • Pulse-ox accuracy: ±24%2–4\% for SpO₂ 70–100 %.
  • Capnograph sidestream sample flow ≈500mL⋅min1500\,\text{mL·min}^{-1}.
  • Heater temp for transcutaneous electrodes: 4245C42–45^{\circ}\text C.
  • Cardiac index effects: >2.2L⋅min1m22.2\,\text{L·min}^{-1}\text m^{-2} ⇒ PtcO₂/PaO₂ ≈0.5; <1.51.5 ⇒ ≈0.1.

Practical, Ethical & Safety Notes

  • Always correlate non-invasive monitors with ABGs when clinical picture disagrees.
  • Misinterpretation (e.g., CO poisoning, poor calibration) may cause grievous harm—ensure calibration, understand device limitations.
  • Regularly document settings, sites, patient status to maintain continuity of care & meet infection-control standards (clean probes per manufacturer).