Comprehensive Notes – Spirometry, Oxygen & CO₂ Monitoring, and Indirect Calorimetry
Application Scenarios – Spirometry & Lung-Volume Measurements
- Bellows spirometer reads lower exhaled volumes than hospital lab
- No leaks & patient unchanged → most likely error: failure to correct to BTPS (Body Temperature, Pressure, Saturated) conditions.
- Exhaled gas cools on the way to bellows ⇒ volume shrinks.
- Correcting to 37∘C restores true lung volume.
- Hand-held turbine (respirometer) used for FVC
- Patient blows an apparent 5L.
- Device accurate only when peak flow ≤ 300L⋅min−1.
- Higher flows add inertia → vanes overspin → falsely high volume.
- Post-op abdominal patient, shallow breaths on bed rest
- Start incentive spirometry to give visual feedback & encourage deep breaths.
- Adequate analgesia remains essential to prevent splinting.
Objective 1 – Oxygen Analysis
- Clinical need: know exact FiO2 delivered.
- General accuracy: ±2% of reading (e.g., 30 % ⇒ actual 28–32 %).
Polarographic Analyzer (Clark electrode)
- Electrochemical cell; measures partial pressure, converts to %O₂.
- O₂ diffuses across Teflon membrane → current proportional to P<em>O</em>2 → display %.
- Pressure-dependent: higher ambient/barometric or circuit pressure exaggerates reading.
- Calibrate at both 21% & 100% O₂; failure ⇒ replace electrode.
- Battery-powered; continuous or spot use.
Galvanic Cell Analyzer
- Also electrochemical & pressure-dependent.
- Generates its own current; no external power.
- Slower response but often longer cell life.
Electrical (paramagnetic/thermal) O₂ analyzers
- Common in PFT labs (e.g., helium dilution), less in bedside care.
Objective 2 – Pulse Oximetry
- Provides continuous or intermittent SpO₂; detects only functional hemoglobin (HbO₂ + Hb).
Physics
- Spectrophotometry: measures differential absorption at two λ (usually 660 nm red, 940 nm IR) by HbO₂ vs. Hb.
- Photoplethysmography: isolates pulsatile change in light absorption to distinguish arterial blood from static tissues.
- Accuracy ±2–4% for SpO₂ 70–100 % (good perfusion, no interference).
Probe Considerations
- Secure, limit ambient light; choose site (finger, earlobe, forehead) with adequate perfusion.
Interferences
- Intravascular dyes: methylene blue, indigo carmine ↓ SpO₂; indocyanine green minimal effect.
- Motion, low perfusion, vasoconstriction, dark skin, nail polish, ambient light all degrade accuracy.
- Dysfunctional Hb (COHb, MetHb) not differentiated → CO poisoning falsely high SpO₂.
SpO₂ vs. PaO₂ Relationship
- Follows O₂-Hb dissociation curve.
- PaO<em>2=60mmHg⇒SpO</em>2≈90%.
- Curve plateau: changes in PaO₂ >100 mmHg barely alter SpO₂.
Pulse-Ox vs. Co-oximetry (lab gold standard)
- Co-oximetry uses ≥4 wavelengths, measures SaO₂ + COHb + MetHb; accuracy ±1–2 %.
- Pulse-ox cheap, continuous, but blind to dyshemoglobins.
CPG Highlights
- Indications: monitor O₂Hb saturation; evaluate response to therapy; guide during bronchoscopy.
- Contra-indications: need for pH/PaCO₂/total Hb; abnormal Hb.
- Record: probe type/site, FiO₂, patient appearance, HR agreement, simultaneous ABGs.
Objective 3 – Capnometry & Capnography
- Measures P<em>CO</em>2 in exhaled gas; waveform = capnogram.
Technologies
- Colorimetric (disposable): pH-indicator paper turns yellow with >0.5% CO₂ → confirms tracheal intubation.
- Infrared Spectroscopy (most common)
- CO₂ absorbs IR at 4.26μm.
- Types: single-beam with rotating chopper cells; double-beam with reference gas.
- Interference: H<em>2O, N</em>2O absorb nearby wavelengths → drying agents & filters used.
- Configurations:
• Sidestream—small sample suctioned (≈0.5 L·min⁻¹); delay, clog risk.
• Mainstream—sensor inline at airway; zero delay but adds dead space & weight.
- Mass Spectrometry: ions separated by mass/charge; expensive; multi-patient ORs.
- Raman scattering (rare).
Capnogram Phases
- I : anatomic dead space (no CO₂).
- II : rapid up-stroke (mix dead + alveolar).
- III : alveolar plateau; slope indicates V/Q inhomogeneity.
- IV : sharp drop on inspiration.
- Prolonged Phase III, no plateau → obstructive diseases (e.g., COPD).
- Rebreathing, oscillations, curare cleft, hyper/hypoventilation, Cheyne-Stokes all have distinctive shapes.
Clinical Uses
- Confirm airway placement & monitor supraglottic devices.
- Evaluate CPR quality (EtCO₂ ∝ pulmonary perfusion).
- Optimize ventilation, detect circuit disconnections.
- Trend pulmonary disease severity & treatment response.
- Estimate metabolic rate (volumetric CO₂ elimination).
Objective 4 – Transcutaneous P<em>O</em>2 & P<em>CO</em>2
- Heated skin electrodes (42–45∘C) increase perfusion & melt lipids → better diffusion.
Electrode Types
- PtcO₂: Clark polarographic with Teflon membrane.
- PtcCO₂: Stowe-Severinghaus pH glass with bicarbonate buffer.
Accuracy & Influences
- Best in neonates (thin skin, good perfusion).
- Hypotension, vasopressors, edema, thick skin ↓ accuracy.
- PtcCO₂ slightly overestimates PaCO₂ (local metabolism); PtcO₂ often underestimates PaO₂ (skin consumption).
Technical Points
- Calibrate before each placement; change site q 4–6 h to prevent burns.
- Heater power ↑ implies ↑ perfusion → usually good; ↓ may indicate poor blood flow or sensor fault.
- Validate against ABG at initiation & periodically.
- Hazards: skin injury, false results leading to wrong therapy.
Indications
- Trend adequacy of oxygenation/ventilation.
- Evaluate therapy; early hypoperfusion index (PtcO₂/FiO₂).
- Wound care, peripheral revascularization assessment.
Objective 5 – Calorimetry (Nutrition–Respiration Interface)
Importance
- Malnutrition ⇒ diaphragm weakness, ↓ immune defense.
- Overfeeding ⇒ ↑ VCO2, harder to ventilate; excessive carbohydrate especially raises CO₂ burden.
Indirect Calorimetry
- Measures VO<em>2 & VCO</em>2 to compute energy expenditure.
- Respiratory Quotient (RQ)RQ=VO</em>2VCO<em>2
- Carbohydrate ≈ 1.0, Fat ≈ 0.7, Protein ≈ 0.8.
- RQ > 1 ⇒ lipogenesis/overfeeding; RQ < 0.7 ⇒ underfeeding/ketosis.
Open-Circuit Method
- Separate inspiratory/expiratory flows; FiO₂ stable (<0.60 ideal); leaks invalidate data.
Closed-Circuit Method
- Patient rebreathes O₂-filled bag; CO₂ absorbed; O₂ use inferred from volume drop; bulky but no FiO₂ limit.
Advantages over Predictive Equations
- Real-time, patient-specific; prevents over/underfeeding; works on ventilated pts.
Supplementary Clinical Scenarios
- Blender set 80 %, analyzer reads 70 % → analyzer not calibrated at 21 % & 100 %.
- Smoke-inhalation pt. with SpO₂ 100 % on 5 L NC
- Do not wean O₂.
- Pulse-ox misreads COHb as HbO₂ → false high.
- Treat with 100 % high-flow O₂ to displace CO.
- COPD pt. on ventilator shows slow-rising Phase III → high airway resistance causing slow alveolar emptying.
- Transcutaneous monitor power ↑ 5 % → heater compensating for ↑ local perfusion; beneficial, expect improved accuracy.
- Polarographic/galvanic O₂ analyzer accuracy: ±2%.
- Turbine respirometer accurate flows ≤300L⋅min−1.
- Pulse-ox accuracy: ±2–4% for SpO₂ 70–100 %.
- Capnograph sidestream sample flow ≈500mL⋅min−1.
- Heater temp for transcutaneous electrodes: 42–45∘C.
- Cardiac index effects: >2.2L⋅min−1m−2 ⇒ PtcO₂/PaO₂ ≈0.5; <1.5 ⇒ ≈0.1.
Practical, Ethical & Safety Notes
- Always correlate non-invasive monitors with ABGs when clinical picture disagrees.
- Misinterpretation (e.g., CO poisoning, poor calibration) may cause grievous harm—ensure calibration, understand device limitations.
- Regularly document settings, sites, patient status to maintain continuity of care & meet infection-control standards (clean probes per manufacturer).