chapter 5 rad bio
radiation detection overview
radiation detection devices fall into 3 broad categories:
1. Equipment Survey Devices
These devices make sure x-ray machines are working correctly.
They check if the actual radiation output matches the settings on the control panel (mA and kVp).
They help confirm the machine is producing the correct amount of radiation.
This protects patients from getting too much radiation.
2. Environmental Survey Devices
These devices check for radiation in the environment.
They detect radioactive contamination.
They measure how much radiation is present in work or public areas.
They help keep workers and the public safe.
Example: Measuring radiation levels outside an x-ray room (like checking radiation through lead walls).
3. Personal Dosimetry
These devices measure how much radiation a worker is exposed to.
Used by healthcare workers and others who work around radiation.
Tracks occupational dose over time.
Ionization Chambers
An ionization chamber is a metal chamber filled with air or gas.
How It Works:
Radiation enters the chamber.
Radiation interacts with the air inside.
It knocks electrons out of their orbits (this is called ionization).
The free electrons create a small electrical current.
That current is measured and converted into a radiation value.
What Ionization Chambers Can Measure:
Exposure (C/kg)
Air kerma (Gy)
Equivalent dose (Sv)
Most Common Types of Ionization Chambers
1. Geiger-Müller (GM) Counter
Often called a Geiger counter.
Used in nuclear medicine departments.
Checks for radioactive contamination.
This is a type of environmental survey meter.
2. Pocket Ionization Chamber
Used for personal dosimetry.
Can be read immediately by looking at the dial.
Gives an instant reading of radiation exposure.
Proportional Counter
What it does
Used mainly in laboratories, NOT diagnostic imaging
Detects:
Alpha particles
Beta particles
Low-level radioactive contamination
Special capability
Can differentiate alpha from beta radiation
Limitations
Alpha travels only a short distance → probe must be very close to the source
Not useful in radiography or fluoroscopyScintillation Detectors
What They Do
Scintillation detectors create visible light when radiation hits them.
X-ray or gamma photons enter the detector.
The detector material produces flashes of light.
That light is then converted into an electrical signal.
The signal is measured and recorded.
Where They Are Used
Environmental radiation surveys
Used by:Nuclear medicine technologists
Medical physicists
Anyone checking an area for radiation
Indirect capture digital radiography (DR) receptors
These use a scintillator to first convert x-rays into light.
What They Measure
Scintillation detectors are often used to measure:
Radioactivity
Becquerel (Bq)
Counts per minute (CPM)
Counts per second (CPS)
They can also measure:
Photon energy
Measured in keV (kiloelectron volts)
Or MeV (megaelectron volts)
Gamma Spectroscopy (Photon Spectroscopy)
Gamma spectroscopy is the process of measuring the specific energy levels of photons in a beam.
It tells us how much energy each photon has.
Energy is usually measured in keV or MeV.
This helps identify different radioactive materials.
Simple Way to Remember It
Scintillator = Light first
Radiation → Light → Electrical signal → Measurement
scintillation detectors: has layers (1-3 component in order)
The scintillation crystal is the part of the detector that is most sensitive to radiation.
This means it is very good at absorbing incoming x‑ray photons.When x‑ray photons hit this crystal:
The crystal gives off a small flash of light — this is called scintillation.
The crystal is light‑emissive because it absorbs the x‑ray energy and converts it into visible light, which is then detected and turned into the final image.
the light photons then interact with the photo cathode layer and create electrons, this is when the radiation signal is first converted into an electrical signal by the electrons
the photomultiplier tube: signal in photocathode is to weak to measure, in this layer the electrons are multiplied to create a greater signal therefore creating a useful measurement
Semiconductor Detectors
Semiconductor detectors are used in many types of radiation detection devices.
Where They Are Used
Handheld detectors – Used for environmental radiation surveys. These can be made with semiconductor detectors.
X-ray testing detectors – Used to check if the x-ray tube is working properly. These also use semiconductors.
Automatic Exposure Control (AEC) cells – Built using the same semiconductor technology.
Direct capture digital receptors – These use semiconductor materials to directly detect x-rays.
Main Components and How They Work
1. Semiconductor Crystal
The most important part is the semiconductor crystal.
It converts x-ray photons into electrons.
When photons hit the crystal, they knock electrons loose.
It is called a semiconductor because it only conducts electricity when it is stimulated by radiation (photons).
Because of this, it is sometimes called a photoconductor.
Important Difference from Scintillators:
There is no photocathode in a semiconductor detector.
The electrons are already created inside the crystal itself.
In scintillators, light is produced first and then converted into electrons.
In semiconductors, x-rays are converted directly into electrons.
2. Amplifier (Multiplier)
The electrical signal produced is very weak at first.
The signal goes to an amplifier (also called a multiplier).
The amplifier increases the strength of the electrical signal.
This creates a stronger, usable signal.
3. Signal Conversion
Finally, the electrical signal is converted into a useful measurement.
Semiconductors are similar to scintillators because both are used to measure radioactivity.
Common measurements include:
Becquerel (Bq)
Counts per minute (CPM)
Counts per second (CPS)
Photon energy in keV or MeV
Measuring photon energy like this is called gamma spectroscopy.
Radiation Dose Limits
Occupational (Radiation Workers)
These limits protect workers from radiation effects.
Whole Body
50 mSv per year
Purpose: Minimize stochastic effects (like cancer risk).
Eyes
150 mSv per year
Purpose: Prevent radiation-induced cataracts.
The limit is higher because cataracts occur at higher doses.
Skin (Hands & Extremities)
500 mSv per year
Purpose: Prevent deterministic effects (like skin burns).
Skin can tolerate higher doses than internal organs.
Public Dose Limits
The public is limited to 10% of occupational limits.
Whole body: 5 mSv per year
Eyes: 15 mSv per year
Skin & extremities: 50 mSv per year
Occupational Dosimetry
Occupational dosimetry means monitoring radiation dose received by workers.
There are two main types:
Thermoluminescent Dosimeter (TLD)
Optically Stimulated Luminescent Dosimeter (OSL or OSLS)
TLD (Thermoluminescent Dosimeter)
What It’s Made Of
Contains a lithium fluoride crystal.
The crystal absorbs radiation similarly to how the human body absorbs radiation.
How It Works
Radiation hits the crystal.
Electrons inside the crystal become excited.
The energy is stored in the crystal for months.
When it’s time to read the badge, it is placed in a TLD reader.
The reader heats the crystal.
Heat causes the crystal to release light.
The amount of light released is measured.
The light is proportional to the radiation dose received.
Disadvantages of TLD
Can fade in heat (sunlight or hot areas may release stored energy).
Can only be read once.
If there is a reading error, the information is lost.
OSL (Optically Stimulated Luminescent Dosimeter)
What It’s Made Of
Contains an aluminum oxide crystal.
OSL has three filters over the detector:
Aluminum (least absorption)
Tin
Copper (most absorption)
Why these matter:
Radiation passing through each filter is attenuated differently based on its energy.
High-energy radiation → similar readings under all filters
Low-energy radiation → highest reading under the aluminum filter
Sensitivity of OSL
Detects as low as 10 µSv (1 mrem)
X‑ray & gamma range: 5 keV – 40 MeV
Measures up to:
10 Sv (1000 rem) for photons
10 Sv (1000 rem) for beta radiation
Limited for neutrons:
Up to 200 µSv (fast neutrons)
Up to 250 µSv (thermal neutrons)
How It Works
Radiation hits the crystal.
Electrons become excited and store energy.
Energy stays stored until reading time.
The dosimeter is placed in an OSL reader.
It is exposed to laser light (not heat).
The crystal releases stored energy as visible light.
The light is measured to calculate the radiation dose.
Advantages of OSL
More sensitive (more precise readings). as low as 10sv for xray and gamma ray photons in the energy range 5 KeV and 10 MeV
not affected by heat, moisture, pressure
Can be read multiple times.
More accurate overall.
Disadvantage
More expensive than TLD.
Some facilities use TLD instead because it costs less.