L4

Fundamentals of Cardiac Electrophysiology

  • Membrane Potential Definition:

    • All living cells possess a membrane potential (MP\text{MP}), defined as the difference in electrical charge across the semipermeable cell membrane.

    • In resting cardiac muscle cells, there is an unequal distribution of ions resulting in a higher concentration of positive charges (+\text{+}) outside the cell membrane and a higher concentration of negative charges (-\text{-}) inside the intracellular space.

    • During changes in membrane potential, specialized ion channels open and close, allowing oppositely charged ions to traverse the membrane.

  • Action Potential (AP\text{AP}) and Contraction:

    • The propagation of an electrical action potential triggers excitation-contraction coupling, ultimately leading to mechanical muscle contraction.

    • The resting membrane potential (RMP\text{RMP}) of a standard ventricular cardiac muscle fiber is approximately −90 mV-90\,\text{mV}.

    • Depolarization occurs when the membrane potential rapidly shifts from −90 mV-90\,\text{mV} up to −0 mV-0\,\text{mV} (and slightly positive).

    • Repolarization is the physiological process that re-establishes the negative resting state back to −90 mV-90\,\text{mV}.

    • An electrocardiographic monitor plots these electrical voltage shifts along the vertical axis against elapsed time along the horizontal axis.

  • Cardiac Refractory Periods:

    • Absolute Refractory Period: The phase during which the cardiac cell is entirely incapable of generating another action potential, regardless of the strength of the incoming stimulus; no depolarization can occur.

    • Relative Refractory Period: The phase late in repolarization where some cardiac muscle fibers are fully repolarized while others remain depolarized; a sufficiently strong stimulus during this vulnerable window can initiate a premature or abnormal action potential.

Action Potential Phases in Ventricular Myocytes

  • Phase 0 (Rapid Depolarization):

    • Initiated when an incoming electrical impulse (originating from the Sinoatrial Node) reaches the cardiac muscle cell, making the membrane potential less negative.

    • Once the membrane potential is depolarized to its critical threshold potential (TP\text{TP}), voltage-gated fast Na+\text{Na}^+ channels open immediately.

    • A massive and rapid influx of extracellular sodium (Na+\text{Na}^+) ions enters the cell, producing the rapid positive upstroke (Phase 0) of the ventricular action potential.

  • Phase 1 (Early Partial Repolarization):

    • The fast Na+\text{Na}^+ channels rapidly close and inactivate, halting sodium entry.

    • Specialized transient outward potassium (K+\text{K}^+) channels open briefly, permitting an efflux of K+\text{K}^+ ions out of the cell.

    • This small outward positive ionic current causes the intracellular voltage to become slightly less positive, creating a minor downward notch in the action potential waveform.

  • Phase 2 (Plateau Phase):

    • Voltage-sensitive slow calcium channels (L-type Ca2+\text{Ca}^{2+} channels) open, allowing extracellular calcium (Ca2+\text{Ca}^{2+}) to slowly enter the cell.

    • Concurrently, potassium (K+\text{K}^+) continues to leak out of the cell through delayed rectifier channels.

    • The inward influx of positive Ca2+\text{Ca}^{2+} ions balances the outward efflux of positive K+\text{K}^+ ions, causing the trans-membrane potential to hover near 0 mV0\,\text{mV}.

    • Physiological Importance of the Plateau:

    • The plateau phase lasts between 0.2 sec0.2\,\text{sec} and 0.3 sec0.3\,\text{sec}, which is significantly longer than the depolarization phase.

    • By maintaining prolonged cellular depolarization, this plateau prevents sustained tetanic contractions and extends the refractory period, thereby optimizing the pumping effectiveness and ventricular filling of the heart.

  • Phase 3 (Final Rapid Repolarization):

    • The slow Ca2+\text{Ca}^{2+} channels fully close, terminating calcium influx.

    • The membrane becomes selectively and highly permeable to potassium (K+\text{K}^+), allowing continuous potassium efflux.

    • This rapid outward flow of positive charge drives the membrane potential back down to its resting baseline, restoring the resting membrane potential (RMP\text{RMP}) of −90 mV-90\,\text{mV}.

Electrophysiological Properties and Heart Cell Types

  • Core Electrophysiological Properties:

    • Automaticity: The intrinsic ability of specialized cardiac cells to initiate an electrical impulse spontaneously without external neural stimulation.

    • Rhythmicity: The capacity of cardiac cells to depolarize in a predictable, regular, and periodic rhythm.

    • Excitability: The ease or inclination with which cardiac cells respond to a stimulus by depolarizing.

    • Contractility: The mechanical property enabling muscle fibers to shorten and generate force in response to electrical stimulation.

  • Primary Types of Cardiac Cells:

    • Pacemaker Cells: Highly specialized cells characterized by innate automaticity that spontaneously fire electrical impulses to set the heart rate.

    • Conducting Cells: Specialized pathways designed to conduct electrical impulses rapidly and efficiently throughout the cardiac chambers.

    • Myocardial (Contractile) Cells: Muscle cells containing contractile proteins that shorten and pump blood in response to electrical activation.

Classification of cardiac cell types based on specialized electrophysiological functions

The Cardiac Conduction System and Pacemaker Hierarchy

Anatomy of the specialized cardiac conduction system
  • Sinoatrial (SA) Node:

    • Located in the superior posterolateral wall of the right atrium (RA\text{RA}), near the entrance of the superior vena cava (SVC\text{SVC}).

    • Functions as the primary physiological pacemaker because it possesses the highest intrinsic degree of automaticity.

    • Self-Excitation Physiology:

    • The resting membrane potential (RMP\text{RMP}) of SA nodal fibers is approximately −55 mV-55\,\text{mV}.

    • The cell membranes of the SA node are naturally leaky to Na+\text{Na}^+ ions, causing a slow, steady inward leak of sodium (the "funny" current) that progressively elevates intracellular positivity.

    • The critical threshold potential (TP\text{TP}) for opening voltage-gated channels in the SA node is −40 mV-40\,\text{mV}.

    • The voltage differential between resting potential and threshold is calculated as:       −40 mV−(−55 mV)=15 mV-40\,\text{mV} - (-55\,\text{mV}) = 15\,\text{mV}

    • Because of this low differential (15 mV15\,\text{mV}), the SA node reaches threshold potential much earlier than ventricular or lower pacemaker cells.

    • The SA node lacks a prolonged Phase 2 plateau phase, allowing rapid cycling between depolarization and repolarization.

    • Overdrive Suppression: Every cardiac cell possesses latent pacemaker capabilities, but the SA node fires fastest, depolarizing downstream cells before they can reach threshold on their own, suppressing their latent automaticity.

    • Intrinsic firing rate: 60–100 BPM60\text{--}100\,\text{BPM}, modulated continuously by the autonomic nervous system.

  • Atrial Conduction Pathways:

    • From the SA node, impulses spread radially across the atria along internodal pathways:

    • Anterior Internodal Tract: Extends anteriorly, giving off the Bachmann bundle (interatrial pathway) which crosses the interatrial septum to rapidly depolarize the left atrium (LA\text{LA}).

    • Middle Internodal Tract: Descends through the atrial septum to the AV node.

    • Posterior Internodal Tract: Curves posteriorly toward the AV node.

    • Atrial depolarization initiates coordinated atrial contraction, filling the resting ventricles (the "atrial kick").

  • Atrioventricular (AV) Node:

    • Located in the lower right atrium near the interatrial septum, just above the tricuspid valve.

    • Known clinically as the "gatekeeper" of ventricular conduction.

    • Intrinsic secondary pacemaker firing rate: 40–60 BPM40\text{--}60\,\text{BPM}.

    • Conduction Velocity and Delay:

    • The impulse travels from the SA node across atrial tissue to the AV node in approximately 0.03 sec0.03\,\text{sec}.

    • Inside the AV node, the impulse is delayed for approximately 0.13 sec0.13\,\text{sec}.

    • The total elapsed time from the generation of an impulse at the SA node to its delivery into the ventricles is:       0.03 sec+0.13 sec=0.16 sec0.03\,\text{sec} + 0.13\,\text{sec} = 0.16\,\text{sec}

    • Protective Function of the Delay:

    • Ensures the atria fully contract and empty their volume into the ventricles before ventricular systole commences.

    • Shields the ventricles from excessively rapid, chaotic atrial rhythms (such as atrial flutter or atrial fibrillation) that could compromise cardiac output.

  • Bundle of His and Purkinje System:

    • Bundle of His (AV Bundle): The sole normal electrical connection bridging the insulating fibrous skeleton between the atria and the ventricles.

    • Divides into the Right Bundle Branch (RBB\text{RBB}) and Left Bundle Branch (LBB\text{LBB}) running down the interventricular septum.

    • The bundle branches divide extensively into arborized terminal Purkinje fibers, which rapidly distribute the electrical impulse directly to ventricular subendocardial myocytes.

    • Intrinsic tertiary pacemaker firing rate of Purkinje fibers: 15–40 BPM15\text{--}40\,\text{BPM}.

  • Pacemaker Failure and Stokes-Adams Syndrome:

    • Downward Displacement of the Pacemaker: If the SA node fails or Sinoatrial exit block occurs, the next fastest available pacemaker cell takes over pacing duties after a brief pause.

    • Delayed Ventricular Takeover: If conduction through the AV junction or higher bundles is acutely blocked, the lower-order Purkinje cells can pause for up to 30 seconds30\,\text{seconds} before establishing their slow escape rhythm (15–40 BPM15\text{--}40\,\text{BPM}).

    • During this prolonged 30 second30\,\text{second} pause, cardiac output plummets to zero.

    • Stokes-Adams Syndrome: A clinical condition where a patient abruptly experiences cerebral ischemia, profound presyncope, or transient loss of consciousness (syncope) secondary to sudden complete heart block and delayed ventricular escape takeover. The episode can last from several seconds to a few minutes and typically necessitates immediate artificial pacemaker implantation.

  • Ectopic Foci and Ectopic Beats:

    • An ectopic focus is an irritable cardiac site outside the SA node that generates an electrical impulse prematurely.

    • Mechanisms of generation:

    • Secondary escape focus when the SA node fails to fire.

    • Irritation or injury (such as local tissue hypoxia) that causes an abnormal resting membrane potential, triggering spontaneous premature depolarizations.

    • Common clinical triggers: myocardial ischemia, atherosclerotic plaque irritation, nicotine toxicity, excessive caffeine consumption, and various pharmacological agents.

Electrocardiography Overview and Clinical Indications

  • Principles of the Electrocardiogram (ECG):

    • Depolarization and repolarization of millions of myocardial cells produce macroscopic electrical vector fields that propagate through conductive body tissues to the skin surface.

    • Surface electrodes record these tiny electrical potential differences via a specialized voltmeter (the ECG machine).

  • Diagnostic Indications for an ECG:

    • Evaluation of patients presenting with symptoms suggestive of underlying cardiovascular compromise:

    • Acute chest pain or substernal pressure.

    • Shortness of breath (SOB\text{SOB}).

    • Dyspnea accompanied by heart palpitations.

    • Unexplained systemic weakness or sudden lethargy.

    • Syncope or presyncope.

    • Detection of rhythm disturbances (arrhythmias and conduction blocks).

    • Localization and staging of myocardial ischemia, injury, or infarction.

    • Routine pre-operative screening in surgical candidates.

ECG Paper Architecture and Measurements

Standard ECG grid detailing time calibrations on the horizontal axis and voltage calibrations on the vertical axis
  • Grid Calibration Standards:

    • The standard paper feed speed of an ECG recorder is 25 mm/sec25\,\text{mm/sec}.

    • Horizontal Axis represents Time:

    • 1 small square=1 mm=0.04 sec1\,\text{small square} = 1\,\text{mm} = 0.04\,\text{sec} (40 msec40\,\text{msec}

    • 1 large square=5 small squares=5 mm=0.20 sec1\,\text{large square} = 5\,\text{small squares} = 5\,\text{mm} = 0.20\,\text{sec} (200 msec200\,\text{msec}

    • 5 large squares=25 mm=2.5 cm=1 inch=1.0 sec5\,\text{large squares} = 25\,\text{mm} = 2.5\,\text{cm} = 1\,\text{inch} = 1.0\,\text{sec}

    • Hash marks placed along the margin are spaced at intervals of 15 large squares15\,\text{large squares} (3.0 sec3.0\,\text{sec}) or 30 large squares30\,\text{large squares} (6.0 sec6.0\,\text{sec}).

    • Vertical Axis represents Voltage / Amplitude:

    • 1 small square=1 mm=0.1 mV1\,\text{small square} = 1\,\text{mm} = 0.1\,\text{mV}

    • 1 large square=5 small squares=5 mm=0.5 mV1\,\text{large square} = 5\,\text{small squares} = 5\,\text{mm} = 0.5\,\text{mV}

    • 2 large squares=10 small squares=10 mm=1.0 mV2\,\text{large squares} = 10\,\text{small squares} = 10\,\text{mm} = 1.0\,\text{mV} (standard calibration box height).

Deflections and Waveforms: P, QRS, T, and U Waves

  • Isoelectric Baseline:

    • The flat, neutral voltage line observed on an ECG tracing during periods when no net electrical current flows through the cardiac muscle.

    • Measured directly prior to the onset of the P wave or immediately following the termination of the T (or U) wave.

  • P Wave:

    • Represents atrial depolarization initiated by the SA node.

    • Appearance of a normal, uniform P wave indicates that the atria have received an electrical stimulus and will contract to fill the ventricles.

  • QRS Complex:

    • Represents rapid ventricular depolarization.

    • Precedes and correlates mechanically with ventricular systole (mechanical contraction).

    • Concurrently conceals atrial repolarization, which occurs at the same time but is masked by the larger ventricular electrical vector.

    • Diagnostic criteria prioritize QRS width over height when assessing intraventricular conduction delays.

    • Nomenclature of Individual Deflections:

    • Q wave: The initial downward (negative) deflection preceding any upward deflection.

    • R wave: The first positive (upward) deflection above the baseline.

    • S wave: The first negative (downward) deflection occurring after an R wave.

    • R′R' (R prime) wave: A second positive deflection occurring in the same QRS complex.

    • S′S' (S prime) wave: A second negative deflection occurring after an R′R' wave.

    • A complex entirely composed of a single negative deflection is termed a QS complex.

Morphologic variants of the QRS complex
  • T Wave:

    • Represents ventricular repolarization, corresponding to ventricular relaxation (diastole).

    • Typically rounded and upright (positive) in leads where the QRS complex is predominantly positive.

    • Size Discrepancy: The T wave is broader and exhibits lower amplitude than the QRS complex because repolarization is a slower, less synchronous metabolic process than depolarization.

    • Pathological Variants:

    • Inverted T Waves: Indicate subendocardial or transmural myocardial ischemia.

    • Peaked (Tall/Tented) T Waves: Strongly indicate acute hyperkalemia.

Inverted T waves indicating myocardial ischemiaPeaked, tall T waves characteristic of hyperkalemia
  • U Wave:

    • A small, low-amplitude deflection following the T wave; often absent or inconspicuous in standard tracings.

    • Thought to represent late repolarization of the Purkinje fibers or mid-myocardial M-cells.

    • Clinical Significance:

    • Inverted U Waves: Strongly correlate with active myocardial ischemia, critical left anterior descending (LAD\text{LAD}) coronary artery disease (CAD\text{CAD}), or left ventricular (LV\text{LV}) volume/pressure overload.

    • Prominent U Waves (>2 mm> 2\,\text{mm} Amplitude): Associated with severe hypokalemia, hypercalcemia, thyrotoxicosis, or digitalis toxicity.

ECG Intervals and Segments

  • PR Interval:

    • Measured from the beginning of the P wave to the very beginning of the QRS complex (specifically the Q wave, or R wave if no Q is present).

    • Represents the total time required for an electrical impulse to leave the SA node, transit the atrial myocardium, undergo physiological delay inside the AV node, and reach the ventricles.

    • Normal Duration: 0.12–0.20 sec0.12\text{--}0.20\,\text{sec} (3–5 small squares3\text{--}5\,\text{small squares}).

    • Shortens physiologically with increased heart rates.

    • Pathology:

    • Prolonged PR interval (>0.20 sec> 0.20\,\text{sec}): Indicates first-degree atrioventricular heart block.

    • Shortened PR interval (<0.12 sec< 0.12\,\text{sec}): Suggests an ectopic junctional rhythm (retrograde atrial activation) or pre-excitation syndromes.

  • QRS Duration:

    • Measured from the onset of the Q wave (or R wave) to the J point where the S wave returns to the isoelectric baseline.

    • Represents the total duration of intraventricular conduction and ventricular depolarization.

    • Normal Duration: 0.08–0.10 sec0.08\text{--}0.10\,\text{sec} (must not exceed 2.5 small squares2.5\,\text{small squares} or 0.10 sec0.10\,\text{sec}).

    • A QRS duration exceeding 0.12 sec0.12\,\text{sec} indicates bundle branch block or ventricular ectopic conduction.

  • ST Segment:

    • Measured from the J point (the junction between the end of the QRS complex and the start of the ST segment) to the beginning of the T wave.

    • Represents the early, plateau phase of ventricular repolarization (Phase 2 of the cellular action potential).

    • Normally flat (isoelectric), merging into the initial upward limb of the T wave.

    • Normal Physiological Variations:

    • ST elevation: Normal if <2 mm< 2\,\text{mm}.

    • ST depression: Normal if <0.5 mm< 0.5\,\text{mm}.

Normal baseline, ST segment elevation, and ST segment depression
  • QT Interval:

    • Measured from the beginning of the QRS complex to the end of the T wave.

    • Represents the entire duration of ventricular electrical systole (combined ventricular depolarization and repolarization).

    • Normal Duration: Varies inversely with heart rate, typically <0.40 sec< 0.40\,\text{sec}.

    • Prolonged QT Interval and Arrhythmia Risk:

    • A prolonged QT interval indicates that ventricular repolarization is abnormally delayed.

    • Prolongs the relative refractory phase ("vulnerable period").

    • If a premature ventricular complex strikes during the peak or downslope of the T wave (the R-on-T phenomenon), it can trigger polymorphic ventricular tachycardia (Torsades de Pointes) or ventricular fibrillation.

Clinical Case Studies: Ischemia and Infarction

  • Case 1: Andy (Exercise Stress Testing and Myocardial Ischemia):

    • Patient Profile: Andy is a 56-year-old male56\text{-year-old male} undergoing an exercise treadmill stress test.

    • ECG Finding: The technician documents an ST-segment depression measuring 1 mm1\,\text{mm}.

    • Clinical Interpretation: ST depression exceeding 0.5–1.0 mm0.5\text{--}1.0\,\text{mm} reflects subendocardial myocardial ischemia triggered by inadequate coronary perfusion during increased cardiac workload.

  • Case 2: Bob (Acute Chest Pain and Myocardial Injury):

    • Patient Profile: Bob experiences acute, severe substernal chest pain while admitted to a medical ward.

    • ECG Finding: Emergency 12-lead ECG demonstrates marked ST-segment elevation exceeding normal baseline parameters.

    • Clinical Interpretation: Acute ST-segment elevation reflects acute transmural myocardial tissue injury and impending infarction (ST-Elevation Myocardial Infarction, STEMI), mandating immediate reperfusion therapy.

Electrode Placement and Lead Systems

  • Electrode Application Principles:

    • Electrodes consist of disposable adhesive pads containing a conductive gel interface to optimize skin impedance.

    • Lead wires should always be snapped securely onto the electrode pads before adhering them to the patient to prevent uncomfortable mechanical pressure against the patient's skin.

  • 3-Electrode System (Standard Bipolar Limb Leads):

    • Uses 3 electrodes placed on specific body surfaces:

    • Right Arm (RA\text{RA}): White electrode.

    • Left Arm (LA\text{LA}): Black electrode.

    • Left Leg (LL\text{LL}): Red electrode.

    • In continuous bedside cardiac monitoring, placing electrodes on the torso equidistant from the heart minimizes muscular motion artifact.

Placement of the standard 3-electrode monitoring configuration
  • 5-Electrode System:

    • Provides 7 monitoring views (Leads I, II, III, aVRaV_R, aVLaV_L, aVFaV_F, and a selected precordial lead, usually V1V_1).

    • Electrode Positions and Color Coding:

    • RA (White): 2nd intercostal space (ICS\text{ICS}), right midclavicular line.

    • LA (Black): 2nd intercostal space (ICS\text{ICS}), left midclavicular line.

    • RL (Green): Lower right abdomen/torso (acts as electrical ground/reference).

    • LL (Red): Lower left abdomen/torso.

    • Chest / V (Brown): 4th intercostal space (ICS\text{ICS}) at the right sternal border (position of lead V1V_1).

    • Clinical Utility: The brown (V1V_1) lead provides an optimal anatomical view of the sternum and the right side of the heart, aiding in the differentiation of bundle branch blocks and ventricular arrhythmias.

Torso locations for the 5-electrode monitoring system

Einthoven’s Triangle and Vector Deflections

Einthoven's triangle showing the bipolar limb lead axes
  • Bipolar Limb Leads:

    • Introduced by Willem Einthoven; each lead measures the potential difference between two specific electrodes:

    • Lead I: Potential between RA\text{RA} (-) and LA\text{LA} (+); views the lateral heart wall.

    • Lead II: Potential between RA\text{RA} (-) and LL\text{LL} (+); views the inferior heart wall.

    • Lead III: Potential between LA\text{LA} (-) and LL\text{LL} (+); views the inferior heart wall.

  • Einthoven's Law:

    • The electrical voltage (amplitude) recorded in Lead II is mathematically equal to the sum of the voltages recorded in Lead I and Lead III:     Lead II=Lead I+Lead III\text{Lead II} = \text{Lead I} + \text{Lead III}

    • Example calculation: If Lead I measures 0.5 mV0.5\,\text{mV} and Lead III measures 0.8 mV0.8\,\text{mV}, then Lead II must equal:     Lead II=0.5 mV+0.8 mV=1.3 mV\text{Lead II} = 0.5\,\text{mV} + 0.8\,\text{mV} = 1.3\,\text{mV}

  • Principles of Deflection:

    • The net electrical depolarization wave propagates anatomically from the atria down toward the ventricles (base to apex).

    • Positive (Upward) Deflection: Generated on the ECG when an electrical wave of depolarization travels toward a positive (+\text{+}) electrode.

    • Negative (Downward) Deflection: Generated when a depolarization wave travels away from a positive electrode (or toward a negative electrode).

    • The "camera perspective" is always situated at the positive pole looking toward the heart.

Generation of an upward deflection as depolarization moves toward a positive electrodeGeometric viewing angles of the three bipolar limb leads
  • Unipolar Augmented Limb Leads (aVRaV_R, aVLaV_L, aVFaV_F):

    • Employ a single exploring positive electrode while the machine electronically computes the negative reference from the remaining electrodes.

    • Designated as "augmented" (aa) because the recorder boosts the voltage amplitude to match bipolar leads.

    • Axes and Orientations:

    • aVRaV_R: Exploring electrode on the right shoulder/arm (+); records an inverted tracing under normal conduction.

    • aVLaV_L: Exploring electrode on the left shoulder/arm (+); views the high lateral wall.

    • aVFaV_F: Exploring electrode on the left foot (+); views the inferior cardiac wall.

Unipolar augmented limb leads and their directional axes

Methods for Determining Heart Rate

  • Method 1: The Sequence / Rule of 300 (For Regular Rhythms):

    • At standard chart speed (25 mm/sec25\,\text{mm/sec}), 5 large squares=1.0 sec5\,\text{large squares} = 1.0\,\text{sec}.

    • Over a full minute, 5 large squares×60 sec=300 large squares5\,\text{large squares} \times 60\,\text{sec} = 300\,\text{large squares}.

    • When the ventricular rhythm is regular, identify an R wave that lands on a heavy grid line, count the number of large squares (xx) to the next consecutive R wave, and calculate:     Heart Rate (BPM)=300x\text{Heart Rate (BPM)} = \frac{300}{x}

    • Standard Memorized Sequence:

    • 1 large box=300 BPM1\,\text{large box} = 300\,\text{BPM}

    • 2 large boxes=150 BPM2\,\text{large boxes} = 150\,\text{BPM}

    • 3 large boxes=100 BPM3\,\text{large boxes} = 100\,\text{BPM}

    • 4 large boxes=75 BPM4\,\text{large boxes} = 75\,\text{BPM}

    • 5 large boxes=60 BPM5\,\text{large boxes} = 60\,\text{BPM}

    • 6 large boxes=50 BPM6\,\text{large boxes} = 50\,\text{BPM}

Sequence method for rapid heart rate estimation on ECG grid paper
  • Method 2: The 6-Second Method (For Irregular Rhythms):

    • Necessary when calculating heart rates in irregular rhythms, such as Second-Degree AV block (Mobitz Type I / Wenckebach) or atrial fibrillation.

    • Count the total number of complete QRS complexes present within a 6 second6\,\text{second} tracing interval (30 large squares30\,\text{large squares}, typically demarcated by top margin tick marks spaced 3 seconds3\,\text{seconds} apart).

    • Multiply that count by 1010 (since 6 sec×10=60 sec=1 minute6\,\text{sec} \times 10 = 60\,\text{sec} = 1\,\text{minute}):     Heart Rate (BPM)=Number of QRS complexes in 6 seconds×10\text{Heart Rate (BPM)} = \text{Number of QRS complexes in 6 seconds} \times 10

WEEK 4: LECTURE 4: SELF NOTE

September 28, 2026

10:08 AM

Membrane Potentials:

  • All cells have MP, a difference in electrical charge across a membrane

  • Cardiac muscle cells:     -   ^ (+) outside membrane

                                          - ^ (-) inside membrane

  • During MP, ions of opposite charge crosses the membrane

 

Action Potential (AP ultimately leads to muscle contraction!) :

  • Cardiac muscle fiber depolarizes when RMP (resting membrane potential) -90 mV to 0 mV

  • Repolarizes at -90 mV 

  • Monitor plots voltage against time

  • Refactory period:  - absolute: no AP can be reached, no depolarization

  • Relative period:  - some heart muscle's depolarized ( contracts) and repolarized (relaxes)

 

Phases of action potential:


   Phase 0 = Depolarization:

  • Signal comes from SA node and makes membrane less negative

  • Membrane reaches threshold potential (TP):   - fast Na+ channel opens: Na+ rapidly enters cell causing depolarization

                                                                                   - This generates upstroke 0f AP

Phase 1: Early repolarization:

  • Fast Na+ channel closes

  • K+ channel open briefly and K+ leaves the cell

  • Membrane potential becomes slightly less positive

 

Phase 2: Plateau phase:

  • Slow Ca2+ channel finally opens allowing Ca2+ into cell

  • K+ continues to leak out of cell

  • Ca2+ influx balances K+ efflux

  • Membrane potential hovers around 0mV creating plateau

Q: Why's plateau phase important?   A: lasts 0.2-0.3s (longer than depolarization). Responsible for enhancing pumping effectiveness of heart.

 

Phase 3: Repolarization:

  • Channels close

  • However, membrane only still permeable to K+ since it restores the RMP by repolarizing it.

 

Cardiac properties:

  • Automaticity = ability to depolarize spontaneously

  • Rhythmicity = ability to depolarize in regular rhythm

  • Excitability = inclination to depolarize spontaneously

  • Contractility = ability to shorten muscle fibers

 

Types of heart cells:

  • Pacemaker cells = specialized cells having ability to generate electrical activity spontaneously "automaticity"  

  • Conducting cells = cells that conduct electrical impulse throughout the heart

  • Myocardial cells = cells that contract in response to electrical stimuli and pump the blood

 

 Sinus Node = pacemaker of the heart:

  • Highest degree of automacticity (depolarizes on its own)

  • Located in RA

  • Impulse follows interatrial via (Bachmann)/internodal tracts

  • Causes cell depolarization, which leads to atrial contraction that fill the ventricles

 

Pacemaker:

Every cardiac cell can be a potential pacemaker, however no one normally initiate depolarization since the SA node depolarizes first before they can reach their threshold potentials

 

Sinus Node:

  • RMP of SA node fibers = -55 mV: - naturally leakier to Na+ ions:

                                                                                                                    - intracellular more +

                                                            - Threshold potential is -40 mV:

                                                                                                                     - RMP to TP difference of 15 mV [ -40 - (-55) = 15 mV ]

                                                                                                                                                                                                                         - Reaches TP before Ventricular CE (cell excitability)

  • SA node doesn't have plateau

 

________ the SA node sets the pace and supresses all other cardiac cell pacemaker tendencies ________

 

Sinus Node Self - excitation:

  • RMP increases as Na+ leaks into cell:

                                                                    - when threshold of -40 mV reached, leads to depolarization of SA node:

                                                                                                                                                                                                         - impulse that spreads throughout all heart muscles

  • Repolarizations leads back to RMP and cycles again

  • 60-100 BPM

  • Strongly influenced by the autonomic nervous system


 

Ectopic beats/focus:

  • Impulse originating outside SA node

  • Occurs: - If SA node fails, for next most excitable myocardial cell excites and depolarizes spontaneously

               - if another area of heart is irritated (hypoxic), it can generate a abnormal RMP and then initiate a premature beat

  • Causes:

                           - ischemia

                           - irritation caused by plaque build up

                           - toxicity  from nicotine

                            - caffeine

                            - drugs

 

Downward displacement of pacemaker:

  • All cells in the cardiac conduction system can be potential pacemakers

  • Fastest cell = winner takes all, aka SA node

Q: what happens if SA node stops working?    A: lower order cells begin to fire after a few second pause, AKA: downward displacement of the pacemaker

 

Atrioventricular (AV) Node:

  • 2nd place!

  • "gatekeeper"

  • Fires at rate of 40-60 BPM

  • Takes signal from SA node and sends it to rest of the heart

 

Atrioventricular conduction:

  • AV node located in RA (like SA node)

  • SA node travels across atrial muscle in 0.03s to AV node, but impulse is delayed another 0.13s

  • Total of 0.16s from when SA node generates impulse until ventricles receive it.

Q: why can't SA node immediately send signal to ventricles?   A: Ventricles would contract before atria have time to fill

 

AV node:

  • Protects ventricles from excessively rapid atrial rates that ventricles can't tolerate

Q: in cases where AV junction's damaged (EX: MI) what do you suspect happens to the electrical impulse passing into the ventricles?

A: It's excessively delayed resulting in slower ventricular rates or heart blocks (HR will also decrease

 

Bundle of His, AKA AV bundle:


 - only electrical conductive connection between the atria and the ventricle

-Conducts signal to left and right bundle branches

-Bundle branches divide to form Purkinje fibers, when these Purkinje fibers are stimulated it leads to ventricular contraction

 

Purkinje Fibers:

  • Lower order pacemakers, not as reliable

  • If impulse from above is blocked, purkinje cells pause for 30s before establishing rate of 15-40 bpm

  • During the pause, there's no cardiac output due to delayed ventricular takeover. This is stokes-Adams syndrome, where one suddenly feels faint and passes out and can last few seconds or minute.

Q: treatment?   A: Pacemaker

 

ECG: millions of AP conducted through myocardial tissue during depolarization creating electrical currents that can be seen on specialized ECG paper:

  • Consists of electrodes placed on skin and attached to voltmeter (aka ECG) that detects small voltage changes as heart depolarizes and repolarizes

  • Electrodes aka leads: - chest

                                        - limbs


 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ECG Allows us to:          - detect abnormal rhythm

                                            - conduction problems

                                            - location of injury

ECG Indications:

  • Used to evaluate patient's with S&S of cardiac disease

  • Used for pts c/o:   - chest pain

                                 - SOB

                                 - Dyspnea with heart palpitations

                                 - weakness

                                 - lethargy

                                  - syncope

  • Detect abnormalities that have or are occurring

  • Preoperatively

 

ECG Components:

  • Y axis = voltage

  • X axis = time

  • Intervals and segments = spaces between waves/complexes

 

 

 

P wave:

  • Atrial depolarization

  • Once P wave appears , atria contracts

Q: what can we assume if we can see a P wave?       A: That the signal comes from above the ventricles!

 

QRS COMPLEX:

  • Ventricular depolarizations In Q,R,S. Based on leads recorded, you may not see Q and S.

  • Atria repolarization wave hidden by QRS

  • Electrical activity associated with systole (contraction of ventricle)


  • Consists of several distinct waves, each wave has a letter.

  • QRS configuration can vary from one lead to next and from patient to patient

  • If first deflection of QRS is downward (in lead II) = it's labelled Q

  • The initial upward deflection = R wave

  • Following an R, the first negative deflection = S

  • If QRS has second positive deflection = R'

  • If there's 2nd S = Labelled S'

T wave:

  • Ventricular repolarization:   - Ventricular relaxation (diastole)

                                                             - normally upright and rounded

                                                             - can appear peaked (=hyperkalemia)

                                                             - can appear inverted (=ischemia)

Q: why is the T wave smaller than QRS?    A: because repolarization takes longer and it's spread out



U wave:

  • Small deflection

  • Often not seen

  • Thought to represent the final phase of ventricular repolarization

  • Clinical significance not fully understood


  • Prominent U wave (>2mm) may indicate:  - hypokalemia

                                                                            - hypercalcemia

                                                                            - thyrotoxicosis

                                                                             - digitalis toxicity

  • Inverted U wave may indicate:  - Ischemia

                                                          - LAD CAD

                                                          - LV overload

Intervals and Segments:

  • PR interval:    - Sinus node impulse reaches ventricle

                         - measured from beginning of P to beginning of Q

                         - 0.12 - 0.20s: 3-5 small squares

                         - Often smaller in fast HR

                                   - Prolonged PR interval = heart block.

                                  - <0.12 seconds = junctional

                                  changes should have you thinking there's issue at the AV junction


 

  • QRS:      -  0.08-0.10s (doesn't exceed 2 1/2 squares)

               - width represents ventricular depolarization

               - measured from baseline to baseline (J point)

 

  • ST Segment:   - J point to beginning of T wave

                          - Early phase of ventricular repolarization

                          - begins to curve slightly upward at beginning of T wave

                                     - length varies with HR

                                    - normally flat

                                     - slight variations are normal:   - Elevated = < 2 mm

                                                                                             -  Depressed = <0.5 mm


 


 

Q: Andy's 56 yr old male completing stress exercise test when technician notes ST depression of 1 mm, what may be cause?

A: myocardial ischemia

 

Q: you receive a call to wards for a pt experiencing acute chest pain. ECG reveals following, what's cause?


A: myocardial tissue injury

 

QT interval:

  • Beginning of QRS to end of T

  • Start of ventricular depolarization to end of ventricular repolarization

  • < 0.40 seconds (varies with HR), some muscle fibers are repolarized in this phase


Q: What happens If QT interval is prolonged?

A: - prolonged QT means ventricles take longer than normal to recover (repolarize) after each heartbeat

      - this creates longer vulnerable period

      - if extra beat occurs during this vulnerable period (R-on-T phenomenon), it can trigger a life threating ventricular arrhythmia.

 

Isoelectric Line:

  • Flat line just before P wave or right after T or U wave


 

ECG Graph Paper:

  • Paper permitting time measurement along x axis and voltage along y axis

  • Small and large squares made of light and dark lines

  • Small squares = 1 mm apart, represent 0.04s

  • Heavy lines 5 mm apart, represent 0.20s

Q: how long is 5 large squares?

A:  - 25 mm (2.5cm or 1 inch)

      - 1 second


  • 0.2 equals 1 large square or 5 small squares

  • 1 tiny square = 0.04s

  • Vertically each small square rep 0.1 mV

  • Large square rep 0.5 mV

  • Short vertical markers placed every 15 large squares (or 3s) apart to allow time intervals to be easily analyzed


 

Electrodes:

 

Electrical activity:

  • Attach electrodes to adhesive pads with conductive gel on body surface

  • Lead wires permit connection between ECG unit and electrodes

  • Lead wires should be attached to electrodes before placed on skin, to avoid unnecessary pressure to skin surface

  • Lead wires often marked to ensure proper placement of pt's body


 

Placement - 3 Lead:

  • Right arm (RA)

  • Left arm (LA)

  • Left leg (LL)

  • These display 3 bipolar leads: I, II, III

  • For best results, place electrodes on chest wall equidistant from heart (rather than specific limbs)


 

Einthoven's Triangle:

  • Named after Willem Einthoven

  • Introduced 3 bipolar limb leads

  • 3 leads = 3 views

  • Einthoven discovered that the voltage (amplitude) of ORS complex recorded in lead II is always equal to sum of voltages of amplitudes in leads I + III (Einthoven's Law)

EX: Lead I = 0.5 mV, Lead III = 0.8 mV. Lead II should equal 0.5 + 0.8 = 1.3 mV


 

Deflections:      A heart depolarizes from atria to ventricle…

  • Thus, electrical current depolarizes in same direction

  • If current flows toward a positive lead = upward deflection. If current flows away from positive lead = downward deflection.


 

3 electrode system (limb leads):

  • The ECG machine "designates each electrode as either (+) electrode or (-) electrode depending on "lead" setting

  • Camera's always at positive electrode


 

Assessing HR:

  • If HR regular:


 


 

  • If HR rhythm is irregular, count number of QRS complexes in 6 second interval.

  • 6 seconds = 30 large boxes

  • Then, multiple this number by 10. (because 6x10=60 seconds…aka 1 minute)


 

5 Electrode System:

  • Provides 2 additional views compared to 3 lead

  • Brown lead:   - gives view of right side of heart

                          - placed on right chest, 4th ICS to right of sternum

                          - V1 provides image of sternum + partial R side of heart

 

  • Green lead: - right side across red lead

we can place limb leads on torso to reduce artifact


 

Unipolar Limb Leads:

  • Single electrode of positive polarity (heart's considered the electrode)

  • Placed on same points as bipolar limb leads

  • Provide larger amplitude for QRS than bipolar leads, AKA: augmented limb leads