Comprehensive Notes on Scanning Electron Microscopy (SEM)

Scanning Electron Microscope (SEM)

Overview

  • The scanning electron microscope (SEM), also known as SCM, is a versatile instrument used in material and biological sciences.
  • It provides detailed topographic images of a sample surface and chemical contrast based on the atomic number of elements present.
  • The Hitachi SU N00 SEM features a secondary electron detector, backscattered electron detector, and a detector for low vacuum observation of charging samples.
  • It also has a STEM detector which functions similarly to a TEM but running at much lower accelerating voltage.
  • Resolution ranges from 500 nanometers to smaller than 1 nanometer, depending on the configuration.
  • It informs about the chemical composition of samples using backscatter electrons and electron dispersive spectroscopy (EDS).
  • Typically operates between 1 and 30 kilowatts.
  • The electron beam is focused into a small spot and scanned across the sample.
  • The image is built up pixel by pixel, line by line, using a raster scan.
  • Small angles of convergence (a few milliradians, approximately 0.05 degrees) give a large depth of field.
  • Data generated during the scan can be analyzed, allowing for spatially resolved chemical analysis.
  • The distance between the bottom of the objective lens and the sample is typically 5 to 10 millimeters.

Imaging Modes

  • Two main imaging modes: secondary electron imaging and backscattered electron imaging.
  • Secondary electrons produce topographic images that are visually appealing and easy to interpret.
  • Backscatter electrons are sensitive to the chemical composition of the sample.
  • The likelihood of a backscatter event is proportional to the atomic number of the sample.
  • A brighter signal indicates a higher atomic number.
  • For example, a darker area may be an aluminum alloy, while a lighter region is a copper alloy.
  • If the sample isn't smooth, the backscatter image will have shadows that highlight topography.
  • Large differences in composition can create compositional contrast in secondary electron images.

Key Definitions

  • Working Distance: The distance from the sample to the objective lens, usually measured in millimeters.
  • Objective Aperture: A strip of platinum with holes of different sizes (typically 10 to 100s of micrometers in diameter).
  • Probe Size: The diameter of the beam at the sample surface.
  • Interaction Volume: The volume of the sample that the beam interacts with.
  • Microscope resolution is defined by the volume of the sample that the signal is generated from

User-Controlled Parameters

  • Accelerating Voltage: Ranges from 30 kV down to 1 kV, with some instruments reaching 0.5 kV or lower.
  • Excitation of the Condenser System: Controls the intensity of the electron beam.
  • Objective Aperture: Used to define the convergence angle.
  • Working Distance: The distance between the objective lens and the sample.
  • Detectors: Choice depends on whether topography, chemistry, or both are of interest.

Accelerating Voltage Considerations

  • Higher voltage gives a shorter wavelength, which can lead to higher resolution.
  • Higher voltages can achieve smaller probe sizes.
  • At higher voltages, increased scattering over a larger volume of the sample can reduce resolution.
  • Voltage selection depends on sensitivity, elemental analysis needs, and excitation voltage of the sample.
  • Reducing voltage can minimize beam-induced damage on delicate samples.
  • For biology samples, 2 to 5 kV is common, while hard materials may require more than 15 kV.
  • For EDS, the voltage should be at least twice the excitation voltage of the element being analyzed.

Condenser System Excitation

  • Controls the electron beam as it's focused onto the sample using a series of lenses.
  • Controls the current density of the electron beam, also known as spot size or beam current.
  • Larger spot sizes equate to more electrons and a stronger signal but reduced resolution.
  • Smaller spot sizes provide high resolution but a weaker signal.

Objective Aperture

  • Changing the aperture size alters the angle of convergence.
  • Smaller apertures reduce the angle of convergence, improving the depth of field.
  • With a small enough angle, a significant range of the sample can appear in focus simultaneously.
  • Using smaller apertures can cause diffraction of the electron beam, spreading out the probe and reducing resolution.

Working Distance

  • Provides control over convergence angle and depth of focus.
  • Shorter working distances typically improve the signal for backscatter detectors.
  • Too short a working distance can reduce the amount of signal for secondary electronic detectors.
  • EDS detectors have a very specific working distance requirement.
  • Longer walking distances tend to reduce the resolution of the microscope.

Detectors

  • Fundamental detectors: secondary electron and backscattering electron detectors.
  • Secondary electrons are generated by interactions between the incident beam and weakly bound conduction band electrons.
  • They are emitted in random directions and have low energy (2 to 5 EV).
  • A positive voltage attracts the secondary electrons.
  • Collection depends on the topography of the sample; sharp edges facing the detector yield a brighter signal.
  • Backscattered electrons are primary beam electrons that have scattered enough to leave the sample again.
  • The probability of this happening is higher with higher atomic numbers.
  • For example, an aluminum-rich phase appears darker, while a nickel-rich phase is brighter.

Interaction Volume

  • Secondary electrons have low energy, so they can't travel far through the sample before being reabsorbed.
  • The resolution of a secondary electron image is related to the probe size of the beam.
  • Backscattered electrons have as much energy as the incident electrons, so they can escape from deeper within the sample.
  • The interaction volume is a more realistic indicator of resolution when using backscattered electrons.
  • Using a low voltage will reduce the interaction volume but increase the probe size and signal strength.

Energy Dispersive Spectroscopy (EDS)

  • A qualitative and quantitative X-ray microanalytical technique that provides information about the chemical composition of a sample.
  • Generally works for elements above atomic number 3, with specialized detectors that can detect lithium too.
  • Incident X-rays cause an electron from the sample to be ejected. Subsequently is replaced by an electron from the higher energy and causes the emission of a photon from the sample.
  • The spectra is generated with energy on the X-axis and frequency on the Y-axis.
  • Characteristic X-rays appear as strong, well-defined peaks.

Impact of Accelerating Voltage on EDS Spectrum

  • The incident electron needs to have enough energy to eject the core shell electron.
  • The incident beam electrons must have at least as much energy as the X-ray being observed.
  • Typically, double the energy of the peak being observed is used as a threshold.
  • If the beam was 10 KV, the frequency of peaks would start to decrease, and at 5 KV, certain peaks might disappear entirely.
  • The possible energy transitions should be considered and balance the voltage required to see the X-rays.

EDS Mapping

  • The SCM generates a megapixel by pixel image, allowing it to do this with all signals.
  • Every signal of the image has a spectra associated with it.
  • We can create an image of elemental distribution in the sample.

Sample Preparation

  • Samples are mounted on a sample stub (aluminum sample holder).
  • Samples need to be dry due to the high vacuum in the SEM.
  • They also need to be conductive to allow electrons to escape and prevent imaging problems.
  • A flow chart summarizes the sample preparation process, ending in a dry and conductive sample.
  • Fresh living tissue requires substantial processing, while a lump of metal may not need any prep.
  • Tissue samples are typically fixed chemically, then dried using critical point drying with liquid carbon dioxide or chemical drying with HMDS.
  • The sample is then coated with a thin layer of conducting material like carbon, gold, or platinum.

Specialized SCM Techniques

  • Cryo-SEM: Snap freezing a sample and imaging it under cryogenic temperatures allows imaging without fixation and dehydration.
  • This introduces its own complications, such as damage from ice crystals.
  • Transfer chambers with spot coaters and articulated arms are used to fracture and coat the surface.
  • Ultramicrotome in SEM: A miniaturized version of a device used in TEM is placed inside the SCM to cut slices less than 100 nanometers thick.
  • Repeated cutting and imaging allows for building a 3D volume.
  • Backscatter detectors and heavy metal staining are used to see the details.
  • Focused Ion Beam (FIB): Uses a beam of ions to cut layers of the sample.

Vacuum Modes

  • High Vacuum Mode: The normal mode of operation minimizes scattering of the electron beam.
  • Low Vacuum Mode: Reduces the vacuum in the chamber, allowing electrons to dissipate.
  • Since backscattering electrons and characteristic X-rays are generally higher energy than secondary electrons, their detection is not critically dependent on a high vacuum being maintained in the specimen chamber.
  • Secondary Electron Detection in Low Vacuum: Secondary electrons travel through the cloud of low-pressure air, leading to photon emission.
  • SCMs with VP mode are equipped with a detector to measure the intensity of this glow to create a topographic image.
  • Environmental SCM: Pressure is high enough to sustain liquid water in the microscope, enabling imaging of fresh wet biological samples without special preparation.

Limitations of SEM

  • Imaging wet samples requires special setups like low vacuum, environmental SCM, special detectors, or cryogenic conditions.
  • SEM require samples to be dry
  • Non-conductive samples require coating and preparation.
  • SEM only produces grayscale images.
  • SEM cannot accurately measure height like AFM does.
  • Sub-surface imaging is limited due to the interaction volume.
  • Imaging through fluid is limited due to interaction volume.
  • Atomic imaging is not possible due to resolution and probe size limitations.
  • Elemental analysis below the micrometer scale is difficult because the sharp probe not possible to make.
  • Charged molecules can interfere with electron detection.