Comprehensive Notes on Scanning Electron Microscopy (SEM)
Scanning Electron Microscope (SEM)
Overview
- The scanning electron microscope (SEM), also known as SCM, is a versatile instrument used in material and biological sciences.
- It provides detailed topographic images of a sample surface and chemical contrast based on the atomic number of elements present.
- The Hitachi SU N00 SEM features a secondary electron detector, backscattered electron detector, and a detector for low vacuum observation of charging samples.
- It also has a STEM detector which functions similarly to a TEM but running at much lower accelerating voltage.
- Resolution ranges from 500 nanometers to smaller than 1 nanometer, depending on the configuration.
- It informs about the chemical composition of samples using backscatter electrons and electron dispersive spectroscopy (EDS).
- Typically operates between 1 and 30 kilowatts.
- The electron beam is focused into a small spot and scanned across the sample.
- The image is built up pixel by pixel, line by line, using a raster scan.
- Small angles of convergence (a few milliradians, approximately 0.05 degrees) give a large depth of field.
- Data generated during the scan can be analyzed, allowing for spatially resolved chemical analysis.
- The distance between the bottom of the objective lens and the sample is typically 5 to 10 millimeters.
Imaging Modes
- Two main imaging modes: secondary electron imaging and backscattered electron imaging.
- Secondary electrons produce topographic images that are visually appealing and easy to interpret.
- Backscatter electrons are sensitive to the chemical composition of the sample.
- The likelihood of a backscatter event is proportional to the atomic number of the sample.
- A brighter signal indicates a higher atomic number.
- For example, a darker area may be an aluminum alloy, while a lighter region is a copper alloy.
- If the sample isn't smooth, the backscatter image will have shadows that highlight topography.
- Large differences in composition can create compositional contrast in secondary electron images.
Key Definitions
- Working Distance: The distance from the sample to the objective lens, usually measured in millimeters.
- Objective Aperture: A strip of platinum with holes of different sizes (typically 10 to 100s of micrometers in diameter).
- Probe Size: The diameter of the beam at the sample surface.
- Interaction Volume: The volume of the sample that the beam interacts with.
- Microscope resolution is defined by the volume of the sample that the signal is generated from
User-Controlled Parameters
- Accelerating Voltage: Ranges from 30 kV down to 1 kV, with some instruments reaching 0.5 kV or lower.
- Excitation of the Condenser System: Controls the intensity of the electron beam.
- Objective Aperture: Used to define the convergence angle.
- Working Distance: The distance between the objective lens and the sample.
- Detectors: Choice depends on whether topography, chemistry, or both are of interest.
Accelerating Voltage Considerations
- Higher voltage gives a shorter wavelength, which can lead to higher resolution.
- Higher voltages can achieve smaller probe sizes.
- At higher voltages, increased scattering over a larger volume of the sample can reduce resolution.
- Voltage selection depends on sensitivity, elemental analysis needs, and excitation voltage of the sample.
- Reducing voltage can minimize beam-induced damage on delicate samples.
- For biology samples, 2 to 5 kV is common, while hard materials may require more than 15 kV.
- For EDS, the voltage should be at least twice the excitation voltage of the element being analyzed.
Condenser System Excitation
- Controls the electron beam as it's focused onto the sample using a series of lenses.
- Controls the current density of the electron beam, also known as spot size or beam current.
- Larger spot sizes equate to more electrons and a stronger signal but reduced resolution.
- Smaller spot sizes provide high resolution but a weaker signal.
Objective Aperture
- Changing the aperture size alters the angle of convergence.
- Smaller apertures reduce the angle of convergence, improving the depth of field.
- With a small enough angle, a significant range of the sample can appear in focus simultaneously.
- Using smaller apertures can cause diffraction of the electron beam, spreading out the probe and reducing resolution.
Working Distance
- Provides control over convergence angle and depth of focus.
- Shorter working distances typically improve the signal for backscatter detectors.
- Too short a working distance can reduce the amount of signal for secondary electronic detectors.
- EDS detectors have a very specific working distance requirement.
- Longer walking distances tend to reduce the resolution of the microscope.
Detectors
- Fundamental detectors: secondary electron and backscattering electron detectors.
- Secondary electrons are generated by interactions between the incident beam and weakly bound conduction band electrons.
- They are emitted in random directions and have low energy (2 to 5 EV).
- A positive voltage attracts the secondary electrons.
- Collection depends on the topography of the sample; sharp edges facing the detector yield a brighter signal.
- Backscattered electrons are primary beam electrons that have scattered enough to leave the sample again.
- The probability of this happening is higher with higher atomic numbers.
- For example, an aluminum-rich phase appears darker, while a nickel-rich phase is brighter.
Interaction Volume
- Secondary electrons have low energy, so they can't travel far through the sample before being reabsorbed.
- The resolution of a secondary electron image is related to the probe size of the beam.
- Backscattered electrons have as much energy as the incident electrons, so they can escape from deeper within the sample.
- The interaction volume is a more realistic indicator of resolution when using backscattered electrons.
- Using a low voltage will reduce the interaction volume but increase the probe size and signal strength.
Energy Dispersive Spectroscopy (EDS)
- A qualitative and quantitative X-ray microanalytical technique that provides information about the chemical composition of a sample.
- Generally works for elements above atomic number 3, with specialized detectors that can detect lithium too.
- Incident X-rays cause an electron from the sample to be ejected. Subsequently is replaced by an electron from the higher energy and causes the emission of a photon from the sample.
- The spectra is generated with energy on the X-axis and frequency on the Y-axis.
- Characteristic X-rays appear as strong, well-defined peaks.
Impact of Accelerating Voltage on EDS Spectrum
- The incident electron needs to have enough energy to eject the core shell electron.
- The incident beam electrons must have at least as much energy as the X-ray being observed.
- Typically, double the energy of the peak being observed is used as a threshold.
- If the beam was 10 KV, the frequency of peaks would start to decrease, and at 5 KV, certain peaks might disappear entirely.
- The possible energy transitions should be considered and balance the voltage required to see the X-rays.
EDS Mapping
- The SCM generates a megapixel by pixel image, allowing it to do this with all signals.
- Every signal of the image has a spectra associated with it.
- We can create an image of elemental distribution in the sample.
Sample Preparation
- Samples are mounted on a sample stub (aluminum sample holder).
- Samples need to be dry due to the high vacuum in the SEM.
- They also need to be conductive to allow electrons to escape and prevent imaging problems.
- A flow chart summarizes the sample preparation process, ending in a dry and conductive sample.
- Fresh living tissue requires substantial processing, while a lump of metal may not need any prep.
- Tissue samples are typically fixed chemically, then dried using critical point drying with liquid carbon dioxide or chemical drying with HMDS.
- The sample is then coated with a thin layer of conducting material like carbon, gold, or platinum.
Specialized SCM Techniques
- Cryo-SEM: Snap freezing a sample and imaging it under cryogenic temperatures allows imaging without fixation and dehydration.
- This introduces its own complications, such as damage from ice crystals.
- Transfer chambers with spot coaters and articulated arms are used to fracture and coat the surface.
- Ultramicrotome in SEM: A miniaturized version of a device used in TEM is placed inside the SCM to cut slices less than 100 nanometers thick.
- Repeated cutting and imaging allows for building a 3D volume.
- Backscatter detectors and heavy metal staining are used to see the details.
- Focused Ion Beam (FIB): Uses a beam of ions to cut layers of the sample.
Vacuum Modes
- High Vacuum Mode: The normal mode of operation minimizes scattering of the electron beam.
- Low Vacuum Mode: Reduces the vacuum in the chamber, allowing electrons to dissipate.
- Since backscattering electrons and characteristic X-rays are generally higher energy than secondary electrons, their detection is not critically dependent on a high vacuum being maintained in the specimen chamber.
- Secondary Electron Detection in Low Vacuum: Secondary electrons travel through the cloud of low-pressure air, leading to photon emission.
- SCMs with VP mode are equipped with a detector to measure the intensity of this glow to create a topographic image.
- Environmental SCM: Pressure is high enough to sustain liquid water in the microscope, enabling imaging of fresh wet biological samples without special preparation.
Limitations of SEM
- Imaging wet samples requires special setups like low vacuum, environmental SCM, special detectors, or cryogenic conditions.
- SEM require samples to be dry
- Non-conductive samples require coating and preparation.
- SEM only produces grayscale images.
- SEM cannot accurately measure height like AFM does.
- Sub-surface imaging is limited due to the interaction volume.
- Imaging through fluid is limited due to interaction volume.
- Atomic imaging is not possible due to resolution and probe size limitations.
- Elemental analysis below the micrometer scale is difficult because the sharp probe not possible to make.
- Charged molecules can interfere with electron detection.